Comparing fracture openings in mortar using different imaging techniques
, , , , , oraz
30 mar 2024
O artykule
Kategoria artykułu: Original Study
Data publikacji: 30 mar 2024
Zakres stron: 77 - 90
Otrzymano: 15 gru 2023
Przyjęty: 23 lut 2024
DOI: https://doi.org/10.2478/sgem-2024-0004
Słowa kluczowe
© 2024 Jonathan Marliot et al., published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Figure 1:

Figure 2:

Figure 3:

Figure 4:

Figure 5:

Figure 6:

Figure 7:

Figure 8:

Figure 9:

Figure 10:

Summary of average fracture opening obtained using X-ray computed tomography (XRCT), 14C-PMMA autoradiographs (14C-PMMA) and heaviside-digital volumetric correlation (H-DVC) methods_ The numbers of data used to calculate the average values are indicated in brackets_
XRCT | 21.2 ± 9.1 [585] | 32.1 ± 13.1 [1549] | 120.1 ± 68.3 [3685] |
H-DVC | 18.6 ± 8.1 [741] | - | - |
14C-PMMA | 18.6 ± 5.7 [3414] | 26.1 ± 8 [3331] | 180.2 ± 72.4 [6051] |
Comparison of fracture densities calculated using the XRCT and 14C-PMMA methods_ These results were obtained from the same data used to calculate the mean aperture values presented in Table 2, except for the 14C-PMMA method for sample #16 where all the data were considered_
#5 | 3.3 ± 0.3 × 10−2 | 3.7 ± 0.4 × 10−2 |
#39 | 3.8 ± 0.4 × 10−2 | 3.5 ± 0.4 × 10−2 |
#16 | 9.1 ± 0.9 × 10−2 | 7.6 ± 0.8 × 10−2 |
Process duration and constraints for each method of aperture and density analysis_
OM | A few hours | Manual analysis of few points / Difficulties of fracture observation |
SEM | 1 day | Manual analysis of few points / long analysis time |
XRCT | 1 day | Fracture detection depends on image resolution |
H-DVC | A few weeks / months | Fracture detection depends on image resolution / long process and analysis time, two image acquisitions |
14C-PMMA | A few months | Use of radioactive tracer / long process time / semi-destructive method |
Summary of average fracture opening obtained using microscopy methods: optical microscopy (OM) and scanning electron microscopy (SEM)_ The number of observation points is indicated in brackets_
OM | 15.6 ± 5.5 [12] | 29.3 ± 14.6 [80] | 131.8 ± 85.7 [33] |
SEM | 15.8 ± 6.1 [59] | 25 ± 14.2 [121] | 129.5 ± 136.8 [137] |