Bandwidth and Common Mode Optimization for Current and Voltage Sources in Bioimpedance Spectroscopy
, , and
Dec 27, 2021
About this article
Published Online: Dec 27, 2021
Page range: 135 - 146
Received: Jul 05, 2021
DOI: https://doi.org/10.2478/joeb-2021-0016
Keywords
© 2021 Tobias Menden, Jascha Matuszczyk, Steffen Leonhardt and Marian Walter, published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Figure 1
![Circuitry of a symmetrical Enhanced Howland Current Source, adapted from [8]. The typical EHCS consists of R1 to R4 and Rx, The symmetric signal is generated by the negative output of OP1 over Rt, R5 mimics the current flowing through R1 and R3, The capacitor Cc increases the bandwidth of the current source.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e8e215d2f6c89dbc94e/j_joeb-2021-0016_fig_001.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250913%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250913T091158Z&X-Amz-Expires=3600&X-Amz-Signature=8503fe16359efb018eee014b201b17ec713c4179c4ae366f16377cbc189bf6ec&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Figure 2

Figure 3

Figure 4

Figure 5

Figure 6

Figure 7

Figure 8

Figure 9

Figure 10

Figure 11

Figure 12

Figure 13

Figure 14

Figure 15

The mean CMR without, with static and active compensation methods_ The variances σ of the VS (Cv = 82 pF) and current source (Cc = 2_6 pF) are given for the frequency range 1 kHz to 1 MHz_
Compensation | [dB] |
[dB] |
|||
---|---|---|---|---|---|
Off | -29.76 | 0 | -30.71 | 0 | |
Static | -79.04 | 0.004 | -130.08 | 9.03 | |
Active | -39.43 | 0 | -38.65 | 0.04 | |
Active & Static | -128.93 | 0.5 | -139.76 | 10.91 |
SNR and THD of the hardware realization_
EHCS | VS | ||
---|---|---|---|
SNR [dB] | 10 | 47,53 dB | 46,22 dB |
100 | 47,14 dB | 46,25 dB | |
THD [dBc] | 10 | -42.31 | -44.15 |
100 | -35.17 | -35.74 |