Improved estimation of left ventricular volume from electric field modeling

Published Online: 27 Dec 2021
Volume & Issue: Volume 12 (2021) - Issue 1 (January 2021)
Page range: 125 - 134
Received: 16 Jun 2021
Journal Details
License
Format
Journal
eISSN
1891-5469
First Published
01 Jan 2010
Publication timeframe
1 time per year
Languages
English
Introduction

The use of left ventricular assist devices (LVADs) in the treatment of advanced heart failure or cardiogenic shock has proven to improve patient outcomes significantly [1, 2, 3]. The advanced monitoring of pressures and flows in the circulation would allow to adapt LVAD therapy to individual needs and reduce adverse events [4, 5]. In the past, various control strategies of an LVAD that adapt to the loading conditions of the heart measured by different sensor modalities were studied by several groups [6, 7, 8, 9, 10, 11] without being clinical routine. In general, reliable monitoring of hemodynamics has been shown to reduce rehospitalization [12]. In particular, left ventricular volume (LVV) is one potential input variable for physiological VAD control [13, 14]. Bioimpedance measurements have been identified as a potential tool for the assessment of LVV [15]. As shown previously by [14, 16], an implanted LVAD can even provide a suitable platform for sensor integration.

Techniques for intracardiac bioimpedance measurement have been developed over the last 40 years. The invasive nature of the LVAD can be utilized to place the measurement setup, including the electrodes directly into the heart, without the need for additional surgical access to the patient. The established methods of Wei et al. [17], and Baan et al. [18, 19] sought to derive an analytical estimate of the left ventricular radius from the electric field distribution spanned inside the ventricle. LVV can then be calculated from this radius using ventricular volume models. The methods of Wei and Baan are regarded as the gold standard in intracardiac bioimpedance measurement, in research, and commercially available pV-loop catheters but are not clinical routine.

Both methods are based on a multitude of simplifying assumptions regarding the electric field distribution inside the heart caused by bioimpedance measurements. Baan et al. [18] assume a linear conductance-volume relationship. This results in a false assumption of the dependency between impedance and cylindrical volume segments. Furthermore, the methods by Baan and Wei neglect the boundary to the conductive myocardium and background tissues. Lastly, only the electric field distribution of the excitation field is modeled, and the measurement field is omitted, which contradicts the reciprocity condition [20]. As a result, both methods overestimate LVV [18, 21, 22]. In addition, Kass et al. [23] found that changes in conductance are very small over one heart cycle in patients with dilated hearts and reduced ejection fraction, which is not represented by these methods. The experimental results from Wu et al. [22] support this conclusion.

Wei et al. also provide an empirical solution to determine LVV. However, this approach necessitates, in addition to a given blood conductivity, patient-specific geometric calibration using echocardiographic volumetry. Further, parallel muscle conductance significantly influences absolute LVV measurements [21]. As the refraction of the electric field at the muscle boundary is not sufficiently considered by Baan and Wei, its proportion needs to be elaborately determined and then subtracted from the measurement signal [24, 25, 26, 19].

This study aims to establish new relationships between the geometric properties of the left ventricle and the distribution of the electric field generated by intracardiac bioimpedance measurements, simplifying and improving the assessment of LVV.

Analytical calculation methods

This section presents the analytical calculation methods (ACM) developed to improve absolute LVV estimation. In this study, we focus on tetrapolar measurement configurations. Larson et al. [27] found that tetrapolar configurations show better sensitivity distribution in the blood cavity compared to the commonly used multipolar configurations.

Wei et al. [17] postulated a nonlinear conductance-volume relationship designed and validated for infinite and homogeneous blood media.

The analytical solution of this relationship was derived by modeling the electric field of two point charges, as depicted in Fig. 1, From electric field theory, the electric field distribution ~Esup of a superposition of two point charges in cylindrical coordinates (, , z) is given by

Figure 1

Electric field distribution of to spheres with opposite charges Q (red) and -Q (blue). The measurement electrodes (green) are distanced by L. Background tissue (σba = 0) is assumed.

Esup =Q4πϵ((zd2)ez+ρeρ(zd2)2+ρ23(z+d2)ez+ρeρ(z+d2)2+ρ2), $$\begin{align}& {{{\vec{E}}}_{\text{sup }}}= \\ & \frac{Q}{4\pi \epsilon }\cdot \left( \frac{\left( z-\frac{d}{2} \right)\cdot {{{\vec{e}}}_{z}}+\rho \cdot {{{\vec{e}}}_{\rho }}}{{{\sqrt{{{\left( z-\frac{d}{2} \right)}^{2}}+{{\rho }^{2}}}}^{3}}}-\frac{\left( z+\frac{d}{2} \right)\cdot {{{\vec{e}}}_{z}}+\rho \cdot {{{\vec{e}}}_{\rho }}}{\sqrt{{{\left( z+\frac{d}{2} \right)}^{2}}+{{\rho }^{2}}}} \right), \\ \end{align}$$

assuming that the two charges ±Q $\pm Q$are located at z=d2 $z=\frac{d}{2}$and Z=d2. $Z=-\frac{d}{2}.$The distance d between the charges represents the distance between the injection electrodes in a tetrapolar setup. The current IWei that is induced by this dipole arrangement is determined integrating the current density J=σblEsup  $\vec{J}={{\sigma }_{\text{bl}}}\cdot {{\vec{E}}_{\text{sup }}}$in the z-plane at z = 0 [28],

/Wei=ϕ=02πρ=r0RσblEsup ,z(z=0)dA=Q4πϵ4πσbld(1d2+4R21d2+4r02), $$\begin{align}& {{/}_{\text{Wei}}}=\int_{\phi =0}^{2\pi }{\int_{\rho ={{r}_{0}}}^{R}{{{\sigma }_{\text{bl}}}}}\cdot {{{\vec{E}}}_{\text{sup },\text{z}}}(z=0)\,\text{d}\vec{A} \\ & =\frac{Q}{4\pi \epsilon }\cdot 4\pi {{\sigma }_{\text{bl}}}\cdot d\cdot \left( \frac{1}{\sqrt{{{d}^{2}}+4{{R}^{2}}}}-\frac{1}{\sqrt{{{d}^{2}}+4r_{0}^{2}}} \right), \\ \end{align}$$

with dA = dd in cylindrical coordinates. R marks the radius of the blood cavity, r0 describes the diameter of the injection electrodes, and σbl the conductivity of blood.

The voltage drop VWei is described by the integration of E~sup between the two measurement electrodes along the z-axis [28]. Assuming a distance L between the measurement electrodes, it follows,

VWei=L2L2Esup dz=4Q4πϵ2Ld2L2. $${{V}_{\text{Wei}}}=-\int_{-\frac{L}{2}}^{\frac{L}{2}}{{{E}_{\text{sup }}}}dz=-4\cdot \frac{Q}{4\pi \epsilon }\cdot \frac{2L}{{{d}^{2}}-{{L}^{2}}}.$$

Finally, the conductance GWei of the left ventricle is calculated using Ohm’s law,

GWei =IWei VWei =πσbld2L24L1d2+4r021d2+4R2, $$\begin{align}& {{G}_{\text{Wei }}}=\frac{{{I}_{\text{Wei }}}}{{{V}_{\text{Wei }}}} \\ & \,\,\,\,\,\,\,\,\,\,\,\,=\frac{\pi {{\sigma }_{\text{bl}}}\left( {{d}^{2}}-{{L}^{2}} \right)}{4L}\cdot \left( \frac{1}{\sqrt{{{d}^{2}}+4r_{0}^{2}}}-\frac{1}{\sqrt{{{d}^{2}}+4{{R}^{2}}}} \right), \\ \end{align}$$

which is similar to the result from [17]. Applying Eq. 4, R can be derived based on the known geometric properties of the measurements setup and the measured GWei, Wei et al. then applied a cylindrical volume model (LV V Wei = R2 L) to estimate LVV [17].

Building on this ACM to estimate R from G, we repeal some of its simplifications and propose two new ACMs for improved estimation of the ventricular radius from bioimpedance measurements.

Plate ACM

The electric field refracts at the boundary of two materials (1 and 2) with conductivities σ1 and σ2, respectively. For the normal and tangential components of the current density fields J1 and J2 applies [28]

J1n=J2nandJ1tσ2=J2tσ1. $${{J}_{1n}}={{J}_{2n}}\ \ \ \text{and}\ \ \ {{J}_{1t}}\cdot {{\sigma }_{2}}={{J}_{2t}}\cdot {{\sigma }_{1}}.$$

In order to consider these dependencies in an analytical description of the electric field distribution, the incident angles at each point of the boundary have to be known. However, the initial fields E1 and E2 are unknown and thus an analytical solution is not derivable.

Assuming a high conductivity of material 1 and a low conductivity of material 2 (σ1>σ2), $\left( {{\sigma }_{1}}>{{\sigma }_{2}} \right),$the current density is refracted in tangential direction for material 1 according to Eq. 5, As a result, the field is distorted and compressed into the high conducting material. Based on this observation, we assume that in intracardiac bioimpedance measurements the field is compressed into the blood-filled cavity, as blood is more conductive than muscle tissue (σbl = 0.7 S/m > σm = 0.17 S/m at 20 kHz, see [29]).

Thus, we propose to model the refraction of the electric field by superposing the electric field of one charged sphere QatZ=d2 $Q\,\text{at}\,Z=\frac{d}{2}$with the electric field of one charged plate Qatz=0(Epla ). $-{{Q}^{\prime }}\,\text{at}\,z=0\left( {{{\vec{E}}}_{\text{pla }}} \right).$This arrangement is depicted in Fig. 2, The size of the plate is equal to the cross-sectional area of the blood cavity with radius R, Due to its equally distributed charge and its bounded size, the electric field is compressed into the blood filled volume.

Figure 2

Electric field distribution by applying the voltages -V0 (red) on a sphere and V0=+V0 $V_{0}^{\prime }=+{{V}_{0}}$on a spare charge plate (blue). The measurement electrodes (green) are distanced by L. Muscle tissue (σm) with a thickness of b and background tissue (σba) are modeled.

The plate carries the same charge as the sphere, which has a voltage of V0 and the radius r0,

Q4πϵ=!V0r0. $$\frac{-{{Q}^{\prime }}}{4\pi \epsilon }\overset{!}{\mathop{=}}\,{{V}_{0}}\cdot {{r}_{0}}.$$

Here, ⌈Q0 is equally distributed on the plate. E~pla is then defined by [28]

Epla=V0r02R2(1zz2+R2)ez, $${{\vec{E}}_{\text{pla}}}={{V}_{0}}\cdot {{r}_{0}}\cdot \frac{2}{{{R}^{2}}}\left( 1-\frac{z}{\sqrt{{{z}^{2}}+{{R}^{2}}}} \right){{\vec{e}}_{z}},$$

It yields the complete field E~sup,pla by a summation of Eq. 7 with the electric field of a sphere at Z=d2, $Z=\frac{d}{2},$

Esup,pla =V0r0.((zd2)ez(ρr0)eρ(zd2)2+(ρr0)222R2(1zz2+R2)ez) $$\begin{align}& {{{\vec{E}}}_{\text{sup,pla }}}=-{{V}_{0}}{{r}_{0}}. \\ & \,\,\,\left( \frac{\left( z-\frac{d}{2} \right)\cdot {{{\vec{e}}}_{z}}\left( \rho -{{r}_{0}} \right){{{\vec{e}}}_{\rho }}}{{{\sqrt{{{\left( z-\frac{d}{2} \right)}^{2}}+{{\left( \rho -{{r}_{0}} \right)}^{2}}}}^{2}}}-\frac{2}{{{R}^{2}}}\left( 1-\frac{z}{\sqrt{{{z}^{2}}+{{R}^{2}}}} \right){{{\vec{e}}}_{z}} \right) \\ \end{align}$$

The current Isup,pla excited by this field is the integration of the current density Jsup,pla =σbl Esup,pla  ${{\vec{J}}_{\text{sup,pla }}}={{\sigma }_{\text{bl }}}\cdot {{\vec{E}}_{\text{sup,pla }}}$in the z-plane

(z=0) analogous to Eq. 2,Isup,pla =8r0V0πσbl(4r024r0R+d2d4r028r0R+4R2+d22+r02R2) $$\begin{align}& \text{(}z\,\text{=}\,\text{0) analogous to Eq}\text{. 2,} \\ & {{I}_{\text{sup,pla }}}= \\ & \,\,-8{{r}_{0}}{{V}_{0}}\pi {{\sigma }_{\text{bl}}}\left( \frac{4r_{0}^{2}-4{{r}_{0}}R+{{d}^{2}}}{d\sqrt{4r_{0}^{2}-8{{r}_{0}}R+4{{R}^{2}}+{{d}^{2}}}}-2+\frac{r_{0}^{2}}{{{R}^{2}}} \right) \\ \end{align}$$

The voltage drop Vsup,pla at the location of the measurement electrodes is derived by integrating the electric field along the z-axis considering the symmetry of the setup to the z = 0 plane,

Vsup, pla =20L2Esup,pla dz=8r0V0.( 2(Ld)LR2+2R2(Rr0)2+(L2)2 +2d2R2(Rr0) ). $$\begin{align}& {{V}_{\text{sup, pla }}}=-2\cdot \int_{0}^{\frac{L}{2}}{{{E}_{\text{sup,pla }}}}\text{dz}=-8{{r}_{0}}{{V}_{0}}. \\ & \,\,\,\left( \frac{2}{(L-d)}-\frac{L}{{{R}^{2}}}+\frac{2}{{{R}^{2}}}\sqrt{{{\left( R-{{r}_{0}} \right)}^{2}}+{{\left( \frac{L}{2} \right)}^{2}}} \right. \\ & \,\,\,\left. +\frac{2}{d}-\frac{2}{{{R}^{2}}}\cdot \left( R-{{r}_{0}} \right) \right). \\ \end{align}$$

This yields the following ACM of the conductance Gsup,pla of the substitute plate configuration,

Gsup,pla =πσbl4r024r0R+d2d4r028r0R+4R2+d22+r02R22(Ld)LR2+2d2R2(Rr0)+2R2(Rr0)2+(L2)2 $$\begin{align}& {{G}_{\text{sup,pla }}}=\pi {{\sigma }_{\text{bl}}} \\ & \frac{\frac{4r_{0}^{2}-4{{r}_{0}}R+{{d}^{2}}}{d\sqrt{4r_{0}^{2}-8{{r}_{0}}R+4{{R}^{2}}+{{d}^{2}}}}-2+\frac{r_{0}^{2}}{{{R}^{2}}}}{\frac{2}{(L-d)}-\frac{L}{{{R}^{2}}}+\frac{2}{d}-\frac{2}{{{R}^{2}}}\cdot \left( R-{{r}_{0}} \right)+\frac{2}{{{R}^{2}}}\sqrt{{{\left( R-{{r}_{0}} \right)}^{2}}+{{\left( \frac{L}{2} \right)}^{2}}}} \\ \end{align}$$

Gsup,pla is a function of parameters of the measurement setup (r0, d, L), the geometry of the ventricle (R), and

the blood conductivity (σbl). With an increased distance of the blood-muscle boundary R to the z-axis, the electric field simplifies to the field of two spherical charges as proposed by Wei et al. [17]. This effect is not considered in the plate ACM (cf. Eq. 11), as Q' is equally distributed on the plate. Consequently, the plate configuration should only be applied for small R (R < L). Therefore, the Wei ACM (cf. Eq. 4) is adapted to consider the conductance of muscle and background tissue. This was achieved by extending its integral limits and using the conductivity of muscle/background to calculate the resulting electrical current in Eq. 2, The plate ACM is then used to estimate the conductance of the blood cavity Gsup,pla, Assuming that Gsup,pla can be modeled parallel to the conductance of the muscle Gm and background Gba, the complete compartment Gplate is described by their summation,

Gplate =Gsup, pla +Gm+Gba . $${{G}_{\text{plate }}}={{G}_{\text{sup, pla }}}+{{G}_{\text{m}}}+{{G}_{\text{ba }}}.$$
Lead ACM

The second ACM to describe the relationship between the conductance measured and the radius of the ventricle is based on lead field theory. The concept of lead field theory was introduced by Helmholtz et al. [20]. Its fundamental idea is the reciprocity of the measurement and injection field: in tetrapolar measurement configurations, it is irrelevant which pair of electrodes is used for current injection or voltage measurement. The impedance measured is identical for both setups. As a result, the macroscopic impedance was defined in the lead field theory of Malmivuo and Plonsey [30]. Accordingly, the impedance of a tetrapolar arrangement is calculated by the volume integral of the product of the normalized current densities JI and JM of the injection and the measurement field.

Within the ventricle, we model the electric field distributions resulting from either an injection at the inner electrodes, or an injection at the outer electrodes, using the electric field distribution Esup of two spheres (cf. Eq. 1), interchanging the values of d and L, From ~ JI;M = σ ~E, the injection current density JI and the measurement current density JM are then derived.

Figure 3 c) schematically shows the superposition of the injection (red) and the measurement field (green). Unfortunately, it is not possible to solve the volume integral over the product of JI and JM analytically. Consequently, we propose a simplified solution to account for the influence of the measurement field.

Figure 3

Left: electric field distribution of the injection at the outer electrodes with distance d. Middle: electric field distribution of the injection at the inner electrodes with distance L. Right: electric field distribution considering both the injection and the measurement field according to lead field theory.

From reciprocity follows that the conductance measured is independent of interchanges between the injection and measurement electrodes. Hence,

Gd=!GL $${{G}_{\text{d}}}\overset{!}{\mathop{=}}\,{{G}_{\text{L}}}$$

needs to be valid. The indices of G indicate the distance between the electrodes used for injection (outside: d, inside: L), whereas the other electrode pair is used for voltage measurement. Considering the configurations shown in Fig. 3 a) and b), the differences in potential Vd and VL of both injection locations modeled by two spherical charges are

Vd=L2L2Esph,ddz=4V0r02LL2d2,VL=d2d2Esph,Ldz=4V0r02LL2d2, $$\begin{align}& {{V}_{d}}=-\int_{-\frac{L}{2}}^{\frac{L}{2}}{{{E}_{\text{sph},\text{d}}}}\text{d}z=-4{{V}_{0}}{{r}_{0}}\cdot \frac{2L}{{{L}^{2}}-{{d}^{2}}}, \\ & {{V}_{L}}=-\int_{-\frac{d}{2}}^{\frac{d}{2}}{{{E}_{\text{sph},\text{L}}}}\text{d}z=-4{{V}_{0}}{{r}_{0}}\cdot \frac{2L}{{{L}^{2}}-{{d}^{2}}}, \\ \end{align}$$

and thus Vd = VL = Vall in correspondence to lead field theory.

The currents Id and IL excited by both fields,

Id=ρ=r0Rϕ=02πσblEsph,dρdρdϕ=4πσblr0V0(4r0(r0R)+d2d4(r0R)2+d21), $$\begin{align}& {{I}_{\text{d}}}=\int_{\rho ={{r}_{0}}}^{R}{\int_{\phi =0}^{2\pi }{{{\sigma }_{\text{bl}}}}}\cdot {{E}_{\text{sph},\text{d}}}\cdot \rho \,\text{d}\rho \,\text{d}\phi \\ & \,\,\,\,\,\,=-4\pi {{\sigma }_{bl}}{{r}_{0}}{{V}_{0}}\cdot \left( \frac{4{{r}_{0}}\left( {{r}_{0}}-R \right)+{{d}^{2}}}{d\sqrt{4{{\left( {{r}_{0}}-R \right)}^{2}}+{{d}^{2}}}}-1 \right), \\ \end{align}$$ IL=ρ=r0Rϕ=02πσblEsph,Lρdρdϕ=4πσb1r0V0(4r0(r0R)+L2L4(r0R)2+L21), $$\begin{align}& {{I}_{\text{L}}}=\int_{\rho ={{r}_{0}}}^{R}{\int_{\phi =0}^{2\pi }{{{\sigma }_{\text{bl}}}}}\cdot {{E}_{\text{sph},\text{L}}}\cdot \rho \,\text{d}\rho \,\text{d}\phi \\ & \,\,\,\,\,=-4\pi {{\sigma }_{b1}}{{r}_{0}}{{V}_{0}}\cdot \left( \frac{4{{r}_{0}}\left( {{r}_{0}}-R \right)+{{L}^{2}}}{L\sqrt{4{{\left( {{r}_{0}}-R \right)}^{2}}+{{L}^{2}}}}-1 \right), \\ \end{align}$$

however, are not equal and the reciprocity condition of Eq. 13 is not satisfied. Consequently, the assumption that the electric field of the tetrapolar configuration is described by two ideal spheres is not valid. As a result, the description of the currents needs to be adjusted. According to Eq. 13 and Eq. 14, a straightforward solution to fulfill the reciprocity theorem is given by

IdVallxd+ILVallxL=0, $$-\frac{{{I}_{\text{d}}}}{{{V}_{\text{all}}}}\cdot {{x}_{\text{d}}}+\frac{{{I}_{\text{L}}}}{{{V}_{\text{all}}}}\cdot {{x}_{\text{L}}}=0,$$

with xd and xL being unknown factors that model the influence of the measurement field on the injection field. This simple solution implies a multiplicative rather than an additive adaptation of the fields. Hence, if no current is injected (Id = IL = 0), no offset current remains. However, Eq. 17 is under-determined. Therefore, the following limitations for xd and xL are defined:

xd and xL are unit-less,

the contribution of both currents Id and IL is balanced, as the current densities in lead field theory also contribute equally [30],

Xd=XL1. $$\bullet \,\,-{{X}_{\text{d}}}=X_{\text{L}}^{-1}.$$

These limitations lead to the unique solutions

Xd=ILIdandXL=IdIL. $${{X}_{\text{d}}}=\frac{\sqrt{{{I}_{\text{L}}}}}{\sqrt{{{I}_{\text{d}}}}}\ \ \ \text{and}\ \ \ {{X}_{\text{L}}}=\frac{\sqrt{{{I}_{\text{d}}}}}{\sqrt{{{I}_{\text{L}}}}}.$$

From this directly follows that the modeled current Ilead is

Ilead =ILId $${{I}_{\text{lead }}}=\sqrt{{{I}_{\text{L}}}\cdot {{I}_{\text{d}}}}$$

This results in the ACM of the conductance Glead,bl of the blood cavity,

Glead,bl =Ilead Vall =πσbl.(4r0(r0R)+d2d4(r0R)2+d21)(4r0(r0R)+L2L4(r0R)2+L21)2LL2d2. $$\begin{align}& {{G}_{\text{lead,bl }}}=\frac{{{I}_{\text{lead }}}}{{{V}_{\text{all }}}}=-\pi {{\sigma }_{\text{b}l}}. \\ & \frac{\sqrt{\left( \frac{4{{r}_{0}}\left( {{r}_{0}}-R \right)+{{d}^{2}}}{d\sqrt{4{{\left( {{r}_{0}}-R \right)}^{2}}+{{d}^{2}}}}-1 \right)\left( \frac{4{{r}_{0}}\left( {{r}_{0}}-R \right)+{{L}^{2}}}{L\sqrt{4{{\left( {{r}_{0}}-R \right)}^{2}}+{{L}^{2}}}}-1 \right)}}{\frac{2L}{{{L}^{2}}-{{d}^{2}}}}. \\ \end{align}$$

The complete conductance Glead further depends on the properties of muscle and background tissue. We assume that the lead calculation is valid for all R, so that the total conductance is calculated by

Glead =Glead,bl +Glead,m +Glead,ba . $${{G}_{\text{lead }}}={{G}_{\text{lead,bl }}}+{{G}_{\text{lead,m }}}+{{G}_{\text{lead,ba }}}.$$

Glead,m and Glead,ba are obtained by changing the conductivities and the integral limits for current estimation (cf. Eq.16 and 15) according to the muscle and background dimensions.

The relationship between R and G is given for the plate and the lead ACM in Eqs. 1112 and Eqs. 2021, respectively, where G is directly depended on R, However, the analytical conversion to R from G is impossible. Hence, the dependence of R from G is described by numerically derived characteristic lookup tables. These tables are based on the model descriptions of the R-dependent conductances, including the geometry of the tetrapolar setup, the conductivity of the blood, and the background conductivity. Dependent on the selected electrode configuration, a specific radius R can be targeted within the ventricle. Hence, from the estimation of R, a volume model as proposed byWyatt et al. [31] can then be applied for the estimation of LVV. Assuming that R is measured in the center of the ventricle, the volume would only be dependent on the height LVh of the ventricle:

LVV=56πR2LVh. $$LVV=\frac{5}{6}\cdot \pi \cdot {{R}^{2}}\cdot L{{V}_{h}}.$$
Validation methods

This section introduces the models that are applied to validate the ACMs. One highly controllable setup with glass containers and a setup with two in-vitro silicone phantoms are presented. Tetrapolar bioimpedance measurements are performed with a device described in Korn et al.[32].

Glass containers

Initial testing of the ACMs was performed in cylindrical glass containers. Their cross-sectional areas correspond to a perfect circle and thus fit the assumptions of radius estimation. Five glass containers with radii between 13.3mm and 50.4mm were selected as shown in Fig. 4 to represent a variety of physiological and pathological heart radii [33].

Figure 4

Left: five cylindrical glass containers representing simplistic volumes to test the measurement system and calculation methods. 1: radius 13.3 mm, 2: radius 17.5 mm, 3: radius 32.8 mm, 4: radius 38.6 mm, 5: radius 50.4 mm. Right: close-up on the measurement catheter and a possible cascaded measurement configuration.

All containers were filled with saline solution to conduct impedance measurements. Two distinct concentrations of saline solutions were prepared to model high conductive and low conductive materials. The solutions’ conductivity was measured with the conductivity meter HI 8733 (HANNA Instruments, Woonsocket, Rhode Island, USA) as σ1 = 1.68 S/m and σ2 = 0.62 S/m.

In-vitro phantoms

The glass containers do not accurately replicate ventricle anatomy and omit background material. Thus, the ACMs were tested in in-vitro phantoms to demonstrate proof-of-concept. This reduces animal trials and is in accordance with the 3R strategy [34].

The in-vitro phantoms were cast from silicone (A00, Silikonfabrik.de, Ahrensburg, Germany) and model the anatomy of the left ventricle derived from a left ventricle CT image series from the embodi3D database [35]. The silicone phantoms were either of pure (insulating) silicone or are enriched with carbon additives to replicate the conductive behavior of cardiac tissue [32]. The conductivity of the conductive phantom is approximately σm = 0.3 S/m, but not exactly determinable due to inhomogeneity in mixing (similar to real muscle tissue). Both phantoms are shown in Fig. 5 a) and were placed inside a tank. A syringe was used to displace water from the tank, which is then inversely displaced inside the phantom.

Figure 5

a) Left ventricular in-vitro phantoms made from silicone (left: carbon, right: pure), b) C-arm with manual test bench to measure phantom dimension, c) Cross-sectional areas obtained from CT measurements at the center electrode pair, adapted from [32].

For each phantom, a displacement of about 10 mL was realized in five steps, which resulted in a cumulative reduction of saline solution in the phantom of approximately 50 mL. The saline solution exhibited a conductivity of σbl = 0.7–0.8 S/m. The phantom was placed inside a water tank filled with a conductivity of σba = 0.2 S/m to replicate background tissue. The volumes of the phantoms were obtained from three-dimensional (3D) computer tomographic (CT) measurements from the silicone phantom as shown in Fig. 5 b). The equivalent circular radius was then estimated of the cross-sectional area between the center electrode pair for each volume step (cf.

Fig. 5 c)). Table 1 gives the volumes, the resulting areas, and the equivalent radius of all six volumes. Information on the position and deformation of the measurement catheter is gained from additional CT scans.

Ethical approval

The conducted research is not related to either human or animal use.

Results

All radii estimated by the different ACMs are compared to the reference radius by providing the bias (%) as the mean of the difference between the measured and calculated radii, and the limits of 95% agreement (LOA (%)) as 1.96 of the standard deviations of these differences in relation to the mean radius measured.

Estimation of glass container radius

The conductance measured inside each glass container is applied to the classical Wei ACM [17] (cf. Eq. 4), plate ACM (cf. Eq. 11), and lead ACM (cf. Eq. 20) to estimate the glass container radius. These estimations are depicted in Fig. 6 and compared to the actual radii of the glass cylinders. For all algorithms, the specific conductivity of the saline solution is given.

Figure 6

Estimated radii using the classical Wei, the plate, and the lead ACM compared to the known radii of the cylinders (Data). Left: conductivity of saline solution σ1 = 1.69 S/m, right: conductivity of saline solution σ2 = 0.62 S/m.

For small radii (< 20 mm), the lead and the plate ACM estimate radii that are close to the actual radii of the containers within both saline solutions. For all radii, the lead ACM outperforms Wei’s ACM, exhibiting a bias between -1.66% and -2.48% to the reference (bias Wei: 41.35% – 42.43%). The limits of agreement (LOA) are less than 16.33% (LOA Wei: 47.22% – 48.05%). The plate ACM underestimates radius (σ1. -23.92% ± 40.01%, σ2. -24.39% ± 39.90%). This underestimation is significant for larger radii (container 3–5), but negligible for radii below 20 mm.

The experiment shows that both lead and plate ACM can be used to determine the radius within cylindrical glass containers without further calibration of the algorithms to patient-individual geometric properties of the heart. The lead ACM covers a wide range of radii (13.34mm – 50.4 mm) with an accuracy above 95%, thus including radii of severely dilated ventricles that may appear during

heart failure. Compared to the Wei ACM, the biases are reduced by a factor of ten if the lead ACM is applied. In addition, the plate ACM fits well for radii that are encountered in physiological ranges [33].

In-vitro phantom measurements

In this section, the ACMs are applied to estimate the radii of two in-vitro phantoms for six different filling volumes (cf. Tab. 1). Figure 7 shows the estimated radii based on the conductances measured inside the phantoms compared to the center radii of the phantoms determined from 3D CT scans. For all algorithms, we applied measurements of the conductivity of the saline solution (σbl = 0.7 S/m), either an insulating (σm = 106 S/m) or conductive myocardium (σm = 0.3 S/m), and background conductivity (σba = 0.2 S/m). These data are applied to the calculation methods to estimate phantom radii.

Figure 7

Estimated radii by the ACMs (Wei, Plate, Lead) obtained from the conductances measured inside the in-vitro phantoms compared to reference radii (Data). Left: insulating phantom of pure silicone, right: silicone phantom enriched with carbon.

Measured volumes inside the silicone ventricle from CT measurements. Cross-sectional areas obtained between the center electrode pair. Estimated equivalent center radius.

StepVolume (mL)Area (mm2)Radius (mm)
1117.901248.8419.94
2107.121022.5518.04
397.96855.5016.50
487.20683.1514.75
576.84506.8312.70
669.46401.1211.30

Both the lead and the plate ACM better estimate the radii of both phantoms and all volumes than the Wei ACM, achieving the best results for measurements inside the carbon phantom. This phantom best replicates natural ventricles as it models both the anatomy and the conductivity of the ventricle and background tissue. Here, the radii estimated by the lead ACM deviate by less than 17.1% ± 26.72% (plate: 13.58% ± 26.03%) from the reference values. The application of the Wei ACM leads to a deviation of 141.51% ± 13.75%. For the silicone phantom, the plate and lead ACM biases are increased (plate: 34.14%, lead: -20.41%), the LOAs remain in a range of approximately 26%. The Wei ACM, however, exhibits a smaller difference in the silicone phantom (75.95% ± 12.93%) compared to the carbon phantom, as the silicone phantom better matches the assumptions of Wei et al. [17]. All considered ACM show a dynamic range that is different from the dynamics of the reference radius. The ACMs estimate a lower radius sensitivity on volume than the reference.

Discussion

The application of the newly proposed ACMs to estimate radius, and hence LVV (cf. 22), from bioimpedance measurements in glass cylinders and in in-vitro ventricle phantoms demonstrates that the plate and the lead ACM outperform the established method of Wei et al. [17]. The measurements inside the glass containers show the possibility of estimating their radii without calibration to patient-individual geometric properties of the heart with an accuracy of more than 95% using the lead ACM. The plate method fits well for radii smaller than 20 mm, but underestimates larger radii. The assumption that the charge Q0 is uniformly distributed on the plate simplifies this underestimation of larger radii. Consequently, the plate configuration should only be used for smaller radii and is not applicable for highly dilated hearts.

The validation using a carbon-silicone phantom illustrates the importance of considering heart and background conductivity in the ACMs, as the Wei ACM, which assumes insulating muscle, overestimates phantom radius by more than 100%. The plate and lead ACM better estimate radius with biases below 20%.

However, the dynamic changes in radius due to different volumes are not correctly replicated by the ACMs. These discrepancies may be caused by the ACMs or by a non-ideal behavior of the reference radius. CT measurements were not simultaneously conducted with bioimpedance measurements, and catheter position and volume changes may not have been exactly replicated. Therefore, a major challenge of in-vitro validation of the ACMs lies in the precise placement of the electrode catheter. A preliminary study showed that small displacements of the catheter in the radial direction do not significantly affect the results. However, a change of position in the longitudinal direction severely impacts radius estimation. Varying the longitudinal position of the catheter in a phantom with initial volume (step 1) by 1 cm, the impedance measured varies by 8Ωto 38Ω depending on the insertion depth (total insertion depth of 10 cm). Therefore, a fixed longitudinal position should be accounted for, e.g., by screw-in assemblies as used for implantation of pacemaker electrodes or when mounted onto the cannula of the LVAD.

In future work, the proper placement of electrodes in the left ventricle needs to be investigated. Therefore, the carrier for the electrode placement is critical. For example, a catheter-based LVAD has the potential to serve as a platform for the electrodes. If a setup proposed by Cysyk et al. [14] is used, the geometric factors of the electrodes must be adjusted in the proposed ACMS. Another option for electrode placement may be a catheter attached to the cannula that extends from the apex into the left ventricle.

The deformation of the phantom due to volume changes is most likely different from the deformation of a ventricle in vivo. In contrast to the native heart, the walls of a silicone phantom are passively controlled. Therefore, thinner wall sections contribute more to the contraction of the in-vitro model. This unlikely deformation is confirmed by the CT scans of different volume steps (cf. Fig. 5, c)) and might explain the difference of the conductance measured compared to the equivalent circular reference (cf. Fig. 7). Therefore, another approximation of the actual center radius might show fewer differences.

In the future, the deformation of the ventricle should be more realistically represented to validate the accuracy of the proposed ACMs further. Additionally, the integration of electrodes on catheter-based and apex-based LVADS and their influence on the performance of the algorithms needs to be evaluated and tested in in-vivo studies.

Conclusion

In this study, two new ACMs are proposed that overcome some of the limitations of intracardiac bioimpedance measurements accepted by established methods [17, 18]. The plate and lead ACM account for the background properties of the blood cavity. Additionally, the plate ACM is a workaround to model the refraction of the electric field at the blood-muscle boundary, whereas the lead model considers the contribution of the measurement (lead) field.

By applying these new methods, accuracy in the assessment of radii without calibration to geometric properties is increased from 60% for the established ACM of Wei et al. [17] to more than 95% in glass container experiments. Furthermore, the ventricular radius was estimated with an accuracy above 80% in in-vitro ventricle phantoms. Hence, we suggest using the lead ACM in future measurements of bioimpedance inside the left ventricle, as it covers a wide range of radii. If these estimates are applied to a volume model [31], the proposed ACMs can provide absolute volume assessment necessitating calibration of the longitudinal length only. These improvements in intracardiac bioimpedance measurement may describe a major step forward in LVAD therapy, as clinically established volume measurements from bioimpedance need further calibration to patient-individual stroke volume.

In-vitro silicone models are potent tools for rapidly testing highly invasive measurement techniques before animal experiments are required. They are durable, and the controlled environment allows for a high degree of reproducibility. Future work should focus on accurate reproduction of deformation along the cardiac cycle and measurements in in-vivo trials to further validate the proposed methods.

Figure 1

Electric field distribution of to spheres with opposite charges Q (red) and -Q (blue). The measurement electrodes (green) are distanced by L. Background tissue (σba = 0) is assumed.
Electric field distribution of to spheres with opposite charges Q (red) and -Q (blue). The measurement electrodes (green) are distanced by L. Background tissue (σba = 0) is assumed.

Figure 2

Electric field distribution by applying the voltages -V0 (red) on a sphere and 
V0′=+V0
$V_{0}^{\prime }=+{{V}_{0}}$on a spare charge plate (blue). The measurement electrodes (green) are distanced by L. Muscle tissue (σm) with a thickness of b and background tissue (σba) are modeled.
Electric field distribution by applying the voltages -V0 (red) on a sphere and V0′=+V0 $V_{0}^{\prime }=+{{V}_{0}}$on a spare charge plate (blue). The measurement electrodes (green) are distanced by L. Muscle tissue (σm) with a thickness of b and background tissue (σba) are modeled.

Figure 3

Left: electric field distribution of the injection at the outer electrodes with distance d. Middle: electric field distribution of the injection at the inner electrodes with distance L. Right: electric field distribution considering both the injection and the measurement field according to lead field theory.
Left: electric field distribution of the injection at the outer electrodes with distance d. Middle: electric field distribution of the injection at the inner electrodes with distance L. Right: electric field distribution considering both the injection and the measurement field according to lead field theory.

Figure 4

Left: five cylindrical glass containers representing simplistic volumes to test the measurement system and calculation methods. 1: radius 13.3 mm, 2: radius 17.5 mm, 3: radius 32.8 mm, 4: radius 38.6 mm, 5: radius 50.4 mm. Right: close-up on the measurement catheter and a possible cascaded measurement configuration.
Left: five cylindrical glass containers representing simplistic volumes to test the measurement system and calculation methods. 1: radius 13.3 mm, 2: radius 17.5 mm, 3: radius 32.8 mm, 4: radius 38.6 mm, 5: radius 50.4 mm. Right: close-up on the measurement catheter and a possible cascaded measurement configuration.

Figure 5

a) Left ventricular in-vitro phantoms made from silicone (left: carbon, right: pure), b) C-arm with manual test bench to measure phantom dimension, c) Cross-sectional areas obtained from CT measurements at the center electrode pair, adapted from [32].
a) Left ventricular in-vitro phantoms made from silicone (left: carbon, right: pure), b) C-arm with manual test bench to measure phantom dimension, c) Cross-sectional areas obtained from CT measurements at the center electrode pair, adapted from [32].

Figure 6

Estimated radii using the classical Wei, the plate, and the lead ACM compared to the known radii of the cylinders (Data). Left: conductivity of saline solution σ1 = 1.69 S/m, right: conductivity of saline solution σ2 = 0.62 S/m.
Estimated radii using the classical Wei, the plate, and the lead ACM compared to the known radii of the cylinders (Data). Left: conductivity of saline solution σ1 = 1.69 S/m, right: conductivity of saline solution σ2 = 0.62 S/m.

Figure 7

Estimated radii by the ACMs (Wei, Plate, Lead) obtained from the conductances measured inside the in-vitro phantoms compared to reference radii (Data). Left: insulating phantom of pure silicone, right: silicone phantom enriched with carbon.
Estimated radii by the ACMs (Wei, Plate, Lead) obtained from the conductances measured inside the in-vitro phantoms compared to reference radii (Data). Left: insulating phantom of pure silicone, right: silicone phantom enriched with carbon.

Measured volumes inside the silicone ventricle from CT measurements. Cross-sectional areas obtained between the center electrode pair. Estimated equivalent center radius.

Step Volume (mL) Area (mm2) Radius (mm)
1 117.90 1248.84 19.94
2 107.12 1022.55 18.04
3 97.96 855.50 16.50
4 87.20 683.15 14.75
5 76.84 506.83 12.70
6 69.46 401.12 11.30

Sieweke JT, Berliner D, Tongers J, Napp LC, Flierl U, Zauner F, Bauersachs J, and Schäfer A. Mortality in patients with cardiogenic shock treated with the Impella CP microaxial pump for isolated left ventricular failure. European Heart Journal: Acute Cardiovascular Care 2018, 9.138–48, htt:.//dx,doi,org/10,1177/2048872618757393Sieweke JT Berliner D Tongers J Napp LC Flierl U Zauner F Bauersachs J and Schäfer A Mortality in patients with cardiogenic shock treated with the Impella CP microaxial pump for isolated left ventricular failure European Heart Journal: Acute Cardiovascular Care 2018 9 13848 10.1177/204887261875739329405734Open DOISearch in Google Scholar

Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado III RM, Long JW, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. New England Journal of Medicine 2009, 361.2241–51, 10.1056/NEJMoa0909938Slaughter MS Rogers JG Milano CA Russell SD Conte JV Feldman D Sun B Tatooles AJ Delgado III RM Long JW et al Advanced heart failure treated with continuous-flow left ventricular assist device New England Journal of Medicine 2009 361 224151 10.1056/NEJMoa090993819920051Open DOISearch in Google Scholar

Wilson SR, Mudge Jr GH, Stewart GC, and Givertz MM. Evaluation for a ventricular assist device: selecting the appropriate candidate. Circulation 2009, 119.2225–32, 10.1161/CIRCULATIONAHA.109.850610Wilson SR Mudge Jr GH Stewart GC and Givertz MM Evaluation for a ventricular assist device: selecting the appropriate candidate Circulation 2009 119 222532 10.1161/CIRCULATIONAHA.109.85061019398678Open DOISearch in Google Scholar

Chong MA, Wang Y, Berbenetz NM, and McConachie I. Does goal-directed haemodynamic and fluid therapy improve peri-operative outcomes?: a systematic review and meta-analysis. European Journal of Anaesthesiology (EJA) 2018, 35.469–83, 10, 1097/EJA.0000000000000778Chong MA Wang Y Berbenetz NM and McConachie I Does goal-directed haemodynamic and fluid therapy improve peri-operative outcomes?: a systematic review and meta-analysis European Journal of Anaesthesiology (EJA) 2018 35 46983 10.1097/EJA.000000000000077829369117Open DOISearch in Google Scholar

Stephens RS and Whitman GJR. Postoperative critical care of the adult cardiac surgical patient. Part I: routine postoperative care. Critical Care Medicine 2015, 43.1477–97, 10.1097/CCM.0000000000001059Stephens RS and Whitman GJR Postoperative critical care of the adult cardiac surgical patient Part I: routine postoperative care. Critical Care Medicine 2015 43 147797 10.1097/CCM.000000000000105925962078Open DOISearch in Google Scholar

Rüschen D, Opitz S, Platen P von, Korn L, Leon-hardt S, and Walter M. Robust physiological control of rotary blood pumps for heart failure therapy. at-Automatisierungstechnik 2018, 66.767–79, 10.1515/auto-2018-0014Rüschen D Opitz S Platen P von Korn L Leon-hardt S and Walter M Robust physiological control of rotary blood pumps for heart failure therapy at-Automatisierungstechnik 2018 66 76779 10.1515/auto-2018-0014Open DOISearch in Google Scholar

Wu EL, Stevens MC, Nestler F, Pauls JP, Bradley AP, Tansley G, Fraser JF, and Gregory SD. A Starling-like total work controller for rotary blood pumps: An in vitro evaluation. Artificial organs 2020, 44.40–53, 10.1111/aor.13570Wu EL Stevens MC Nestler F Pauls JP Bradley AP Tansley G Fraser JF and Gregory SD A Starling-like total work controller for rotary blood pumps: An in vitro evaluation Artificial organs 2020 44 4053 10.1111/aor.1357031520408Open DOISearch in Google Scholar

Colacino FM, Moscato F, Piedimonte F, Arabia M, and Danieli GA. Left ventricle load impedance control by apical VAD can help heart recovery and patient perfusion: a numerical study. Asaio Journal 2007, 53.263–77, 10.1097/MAT.0b013e31805b7e39Colacino FM Moscato F Piedimonte F Arabia M and Danieli GA Left ventricle load impedance control by apical VAD can help heart recovery and patient perfusion: a numerical study Asaio Journal 2007 53 26377 10.1097/MAT.0b013e31805b7e3917515714Open DOISearch in Google Scholar

Petrou A, Lee J, Dual S, Ochsner G, Meboldt M, and Schmid Daners M. Standardized comparison of selected physiological controllers for rotary blood pumps: in vitro study. Artificial organs 2018, 42.29–42, 10.1111/aor.12999Petrou A Lee J Dual S Ochsner G Meboldt M and Schmid Daners M Standardized comparison of selected physiological controllers for rotary blood pumps: in vitro study Artificial organs 2018 42 2942 10.1111/aor.1299929094351Open DOISearch in Google Scholar

Arndt A, Nüsser P, and Lampe B. Fully autonomous preload-sensitive control of implantable rotary blood pumps. Artificial organs 2010, 34.726–35, 10.1111/j.1525-1594.2010.01092.xArndt A Nüsser P and Lampe B Fully autonomous preload-sensitive control of implantable rotary blood pumps Artificial organs 2010 34 72635 10.1111/j.1525-1594.2010.01092.x20883392Open DOISearch in Google Scholar

AlOmari AH, Savkin AV, Stevens M, Mason DG, Timms DL, Salamonsen RF, and Lovell NH. Developments in control systems for rotary left ventricular assist devices for heart failure patients: a review. Physiological Measurement 2012, 34.R1, 10.1088/0967-3334/34/1/R1AlOmari AH Savkin AV Stevens M Mason DG Timms DL Salamonsen RF and Lovell NH Developments in control systems for rotary left ventricular assist devices for heart failure patients: a review Physiological Measurement 2012 34 R1 10.1088/0967-3334/34/1/R1Open DOISearch in Google Scholar

Abraham WT, Adamson PB, Bourge RC, Aaron MF, Costanzo MR, Stevenson LW, Strickland W, Neelagaru S, Raval N, Krueger S, et al. Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. The Lancet 2011, 377.658–66, 10.1016/S0140-6736(11)60101-3Abraham WT Adamson PB Bourge RC Aaron MF Costanzo MR Stevenson LW Strickland W Neelagaru S Raval N Krueger S et al Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial The Lancet 2011 377 65866 10.1016/S0140-6736(11)60101-3Open DOISearch in Google Scholar

Ochsner G, Wilhelm MJ, Amacher R, Petrou A, Cesarovic N, Staufert S, Röhrnbauer B, Maisano F, Hierold C, Meboldt M, et al. In vivo evaluation of physiologic control algorithms for left ventricular assist devices based on left ventricular volume or pressure. ASAIO Journal 2017, 63.568–77, 10.1097/MAT.0000000000000533Ochsner G Wilhelm MJ Amacher R Petrou A Cesarovic N Staufert S Röhrnbauer B Maisano F Hierold C Meboldt M et al In vivo evaluation of physiologic control algorithms for left ventricular assist devices based on left ventricular volume or pressure ASAIO Journal 2017 63 56877 10.1097/MAT.000000000000053328857904Open DOISearch in Google Scholar

Cysyk J, Newswanger R, Popjes E, Pae W, Jhun CS, Izer J, Weiss W, and Rosenberg G. Cannula tip with integrated volume sensor for rotary blood pump control: early-stage development. ASAIO Journal 2019, 65.318–23, http .//dx , doi , org/10 , 1097/MAT , 0000000000000818Cysyk J Newswanger R Popjes E Pae W Jhun CS Izer J Weiss W and Rosenberg G Cannula tip with integrated volume sensor for rotary blood pump control: early-stage development ASAIO Journal 2019 65 31823 10,1097/MAT,0000000000000818Open DOISearch in Google Scholar

Schmid Daners M and Dual SA. Pathophysiological Determinants Relevant in Blood Pump Control. Mechanical Support for Heart Failure, Springer, 2020 .253–77, 10.1007/978-3-030-47809-4_18Schmid Daners M and Dual SA Pathophysiological Determinants Relevant in Blood Pump Control Mechanical Support for Heart Failure, Springer 2020 25377 10.1007/978-3-030-47809-4_18Open DOISearch in Google Scholar

Dual SA, Zimmermann JM, Neuenschwander J, Cohrs NH, Solowjowa N, Stark WJ, Meboldt M, and Schmid Daners M. Ultrasonic sensor concept to fit a ventricular assist device cannula evaluated using geometrically accurate heart phantoms. Artificial Organs 2019, 43.467–77, 10.1111/aor.13379Dual SA Zimmermann JM Neuenschwander J Cohrs NH Solowjowa N Stark WJ Meboldt M and Schmid Daners M Ultrasonic sensor concept to fit a ventricular assist device cannula evaluated using geometrically accurate heart phantoms Artificial Organs 2019 43 46777 10.1111/aor.1337930357874Open DOISearch in Google Scholar

Wei CL, Valvano JW, Feldman MD, and Pearce JA. Nonlinear conductance-volume relationship for murine conductance catheter measurement system. IEEE Transactions on Biomedical Engineering 2005, 52.1654–61, 10.1109/TBME.2005.856029Wei CL Valvano JW Feldman MD and Pearce JA Nonlinear conductance-volume relationship for murine conductance catheter measurement system IEEE Transactions on Biomedical Engineering 2005 52 165461 10.1109/TBME.2005.85602916235651Open DOISearch in Google Scholar

Baan J, Jong TTA, Kerkhof PL, Moene RJ, Van Dijk AD, Der Velde ET van, and Koops J. Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter. Cardiovascular Research 1981, 15.328–34, 10.1093/cvr/15.6.328Baan J Jong TTA Kerkhof PL Moene RJ Van Dijk AD Der Velde ET van and Koops J Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter Cardiovascular Research 1981 15 32834 10.1093/cvr/15.6.3287296590Open DOISearch in Google Scholar

Baan J, Van Der Velde ET, De Bruin HG, Smeenk GJ, Koops J, Van Dijk AD, Temmerman D, Senden J, and Buis B. Continuous measurement of left ventricular volume in animals and humans by conductance catheter. Circulation 1984, 70.812–23, 10.1161/01.CIR.70.5.812Baan J Van Der Velde ET De Bruin HG Smeenk GJ Koops J Van Dijk AD Temmerman D Senden J and Buis B Continuous measurement of left ventricular volume in animals and humans by conductance catheter Circulation 1984 70 81223 10.1161/01.CIR.70.5.812Open DOISearch in Google Scholar

Helmholtz Hv. Ueber einige Gesetze der Vertheilung elektrischer Ströme in körperlichen Leitern, mit Anwendung auf die thierisch-elektrischen Versuche. Annalen der Physik und Chemie 1853, 89.353–77, http://dx.doi.org/10.1002/andp.18531650603Helmholtz Hv Ueber einige Gesetze der Vertheilung elektrischer Ströme in körperlichen Leitern, mit Anwendung auf die thierisch-elektrischen Versuche Annalen der Physik und Chemie 1853 89 35377 10.1002/andp.18531650603Open DOISearch in Google Scholar

Salo RW, Wallner TG, and Pederson BD. Measurement of Ventricular Volume by Intracardiac Impedance: Theoretical and Empirical Approaches. IEEE Transactions on Biomedical Engineering 1986, 2.189–95, 10.1109/TBME.1986.325890Salo RW Wallner TG and Pederson BD Measurement of Ventricular Volume by Intracardiac Impedance: Theoretical and Empirical Approaches IEEE Transactions on Biomedical Engineering 1986 2 18995 10.1109/TBME.1986.325890Open DOISearch in Google Scholar

Wu CC, Skalak TC, Schwenk TR, Mahler CM, Anne A, Finnerty PW, Haber HL, Weikle RM, and Feldman MD. Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electric field homogeneity and parallel conductance. IEEE Transactions on Biomedical Engineering 1997, 44.266–77, 10.1109/10.563296Wu CC Skalak TC Schwenk TR Mahler CM Anne A Finnerty PW Haber HL Weikle RM and Feldman MD Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electric field homogeneity and parallel conductance IEEE Transactions on Biomedical Engineering 1997 44 26677 10.1109/10.563296Open DOISearch in Google Scholar

Kass DA, Midei M, Graves W, Brinker JA, and Maughan WL. Use of a conductance (volume) catheter and transient inferior vena caval occlusion for rapid determination of pressure-volume relationships in man. Catheterization and Cardiovascular Diagnosis 1988, 15.192–202, 10.1002/ccd.1810150314Kass DA Midei M Graves W Brinker JA and Maughan WL Use of a conductance (volume) catheter and transient inferior vena caval occlusion for rapid determination of pressure-volume relationships in man Catheterization and Cardiovascular Diagnosis 1988 15 192202 10.1002/ccd.1810150314Open DOISearch in Google Scholar

White PA, Brookes CI, Ravn HB, Stenbøg EE, Christensen TD, Chaturvedi RR, Sorensen K, Hjortdal VE, and Redington AN. The effect of changing excitation frequency on parallel conductance in different sized hearts. Cardiovascular Research 1998, 38.668–75, 10.1016/S0008-6363(98)00052-2White PA Brookes CI Ravn HB Stenbøg EE Christensen TD Chaturvedi RR Sorensen K Hjortdal VE and Redington AN The effect of changing excitation frequency on parallel conductance in different sized hearts Cardiovascular Research 1998 38 66875 10.1016/S0008-6363(98)00052-2Open DOISearch in Google Scholar

Georgakopoulos D and Kass DA. Estimation of parallel conductance by dual-frequency conductance catheter in mice. American Journal of Physiology-Heart and Circulatory Physiology 2000, 279.H443–H450, 10.1152/ajpheart.2000.279.1.H443Georgakopoulos D and Kass DA Estimation of parallel conductance by dual-frequency conductance catheter in mice American Journal of Physiology-Heart and Circulatory Physiology 2000 279 H443–H450 10.1152/ajpheart.2000.279.1.H443Open DOISearch in Google Scholar

Gawne TJ, Gray KS, and Goldstein RE. Estimating left ventricular offset volume using dual-frequency conductance catheters. Journal of Applied Physiology 1987, 63.872–6, 10.1152/jappl.1987.63.2.872Gawne TJ Gray KS and Goldstein RE Estimating left ventricular offset volume using dual-frequency conductance catheters Journal of Applied Physiology 1987 63 8726 10.1152/jappl.1987.63.2.872Open DOISearch in Google Scholar

Larson ER, Feldman MD, Valvano JW, and Pearce JA. Analysis of the spatial sensitivity of conductance/admittance catheter ventricular volume estimation. IEEE Transactions on Biomedical Engineering 2013, 60.2316–24, 10.1109/TBME.2013.2256134Larson ER Feldman MD Valvano JW and Pearce JA Analysis of the spatial sensitivity of conductance/admittance catheter ventricular volume estimation IEEE Transactions on Biomedical Engineering 2013 60 231624 10.1109/TBME.2013.2256134Open DOISearch in Google Scholar

Cheng DK et al. Fundamentals of engineering electro-magnetics. 1993Cheng DK et al Fundamentals of engineering electro-magnetics 1993Search in Google Scholar

Gabriel C, Gabriel S, and Corthout yE. The dielectric properties of biological tissues: I. Literature survey. Physics in Medicine & Biology 1996, 41.2231, 10.1088/0031-9155/41/11/001Gabriel C Gabriel S and Corthout yE The dielectric properties of biological tissues: I Literature survey. Physics in Medicine & Biology 1996 41 2231 10.1088/0031-9155/41/11/001Open DOISearch in Google Scholar

Malmivuo J and Plonsey R. Bioelectromagnetism: principles and applications of bioelectric and biomagnetic fields. Oxford University Press, New York, 1995 .482Malmivuo J and Plonsey R. Bioelectromagnetism: principles and applications of bioelectric and biomagnetic fields Oxford University Press New York 199548210.1093/acprof:oso/9780195058239.001.0001Search in Google Scholar

Wyatt HL, Meerbaum S, Heng MK, Gueret P, and Corday E. Cross-sectional echocardiography III. Analysis of mathematic models for quantifying volume of symmetric and asymmetric left ventricles. American Heart Journal 1980, 100.821–8, 10.1016/0002-8703(80)90062-9Wyatt HL Meerbaum S Heng MK Gueret P and Corday E Cross-sectional echocardiography III Analysis of mathematic models for quantifying volume of symmetric and asymmetric left ventricles. American Heart Journal 1980 100 8218 10.1016/0002-8703(80)90062-9Open DOISearch in Google Scholar

Korn L, Lyra S, Rüschen D, Telyshev D, Leonhardt S, and Walter M. In silico and in vitro conductivity models of the left heart ventricle. Journal of Electrical Bioimpedance 2020, 11.62–71, 10.2478/joeb-2020-0010Korn L Lyra S Rüschen D Telyshev D Leonhardt S and Walter M. In silico and in vitro conductivity models of the left heart ventricle Journal of Electrical Bioimpedance 2020 11 6271 10.2478/joeb-2020-0010753110433584905Open DOISearch in Google Scholar

Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, Flachskampf FA, Foster E, Goldstein SA, Kuznetsova T, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. European Heart Journal-Cardiovascular Imaging 2015, 16.233–71, http://dx.doi.org/10.1093/ehjci/jev014Lang RM Badano LP Mor-Avi V Afilalo J Armstrong A Ernande L Flachskampf FA Foster E Goldstein SA Kuznetsova T et al Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging European Heart Journal-Cardiovascular Imaging 2015 16 23371 10.1093/ehjci/jev01425712077Open DOISearch in Google Scholar

Russell WMS, Burch RL, et al. The principles of humane experimental technique. London: Methuen & Co. Ltd., 1959Russell WMS Burch RL et al The principles of humane experimental technique London: Methuen & Co. Ltd 1959Search in Google Scholar

Emboi3D. Heart and pulmonary artery tree from CT angiogram. 2017 Nov. Available from: https://www.embodi3d.com/files/file/59-heart-and-pulmonary-artery-tree-from-ct-angiogram/?%5C_fromLogin=Emboi3D Heart and pulmonary artery tree from CT angiogram 2017 Nov. Available from https://www.embodi3d,com/files/file/59-heart-and-pulmonary-artery-tree-from-ct-angiogram/?%5C_fromLogin=Search in Google Scholar