The Influence of Measurement Methodology on the Accuracy of Electrical Waveform Distortion Analysis
and
Apr 04, 2018
About this article
Published Online: Apr 04, 2018
Page range: 72 - 78
Received: Sep 30, 2017
Accepted: Mar 20, 2018
DOI: https://doi.org/10.1515/msr-2018-0011
Keywords
© 2018 Jacek Bartman et al., published by De Gruyter Open
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
The present paper covers a review of documents that specify measurement methods of voltage waveform distortion. It also presents measurement stages of waveform components that are uncommon in the classic fundamentals of electrotechnics and signal theory, including the creation process of groups and subgroups of harmonics and interharmonics. Moreover, the paper discusses selected distortion factors of periodic waveforms and presents analyses that compare the values of these distortion indices. The measurements were carried out in the cycle per cycle mode and the measurement methodology that was used complies with the IEC 61000-4-7 norm. The studies showed significant discrepancies between the values of analyzed parameters.