The influence of method precision on assessing radiological hazards for patients using radium water in balneotherapy
25 cze 2024
O artykule
Kategoria artykułu: Original Paper
Data publikacji: 25 cze 2024
Zakres stron: 87 - 91
Otrzymano: 03 paź 2023
Przyjęty: 22 kwi 2024
DOI: https://doi.org/10.2478/nuka-2024-0012
Słowa kluczowe
© 2024 Joanna Lemańska et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Fig. 1.

Humans hazard and its uncertainty
Measurement method | Energy-efficiency calibration | Calculation method | Radioactivity (Bq/L) | Effective dose EH (nSv) | Uncertainty (%) |
---|---|---|---|---|---|
γ-Spectrometry with HPGe detector medium efficiency detector | Calibration source | Genie2000 | 56.4 | 2.44 | 2.2 |
Evaluation of 186.2 keV FEAP | 54.5 | 2.35 | 5.0 | ||
Evaluation of 214Pb and 214Bi FEAPs | 51.1 | 2.21 | 5.0 | ||
LabSOCS | Genie2000 | 53.2 | 2.30 | 2.6 | |
Evaluation of 186.2 keV FEAP | 51.4 | 2.22 | 4.0 | ||
Evaluation of 214Pb and 214Bi FEAPs | 51.0 | 2.20 | 4.0 | ||
γ-Spectrometry with HPGe detector high-efficiency detector | Calibration source | Genie2000 | 68.0 | 2.94 | 4.3 |
Evaluation of 186.2 keV peak | 71.2 | 3.08 | 4.2 | ||
Evaluation of 214Pb and 214Bi FEAPs | 59.0 | 2.55 | 2.8 | ||
LabSOCS | Genie2000 | 65.8 | 2.84 | 3.4 | |
Evaluation of 186.2 keV peak | 68.9 | 2.98 | 4.2 | ||
Evaluation of 214Pb and 214Bi FEAPs | 56.8 | 2.45 | 2.6 | ||
Liquid scintillation | – | 57.9 | 2.50 | 10.4 | |
Emanation method | – | 51.1 | 2.21 | 16.3 |