The influence of method precision on assessing radiological hazards for patients using radium water in balneotherapy


Zitieren

Fig. 1.

γ-ray spectrogram of radium water.
γ-ray spectrogram of radium water.

Humans hazard and its uncertainty

Measurement method Energy-efficiency calibration Calculation method Radioactivity (Bq/L) Effective dose EH (nSv) Uncertainty (%)
γ-Spectrometry with HPGe detector medium efficiency detector Calibration source Genie2000 56.4 2.44 2.2
Evaluation of 186.2 keV FEAP 54.5 2.35 5.0
Evaluation of 214Pb and 214Bi FEAPs 51.1 2.21 5.0
LabSOCS Genie2000 53.2 2.30 2.6
Evaluation of 186.2 keV FEAP 51.4 2.22 4.0
Evaluation of 214Pb and 214Bi FEAPs 51.0 2.20 4.0
γ-Spectrometry with HPGe detector high-efficiency detector Calibration source Genie2000 68.0 2.94 4.3
Evaluation of 186.2 keV peak 71.2 3.08 4.2
Evaluation of 214Pb and 214Bi FEAPs 59.0 2.55 2.8
LabSOCS Genie2000 65.8 2.84 3.4
Evaluation of 186.2 keV peak 68.9 2.98 4.2
Evaluation of 214Pb and 214Bi FEAPs 56.8 2.45 2.6
Liquid scintillation 57.9 2.50 10.4
Emanation method 51.1 2.21 16.3