Comparison of validity, repeatability and reproducibility of the Peer Assessment Rating (PAR) between digital and conventional study models


Zacytuj

Introduction

The validity, reliability and inter-method agreement of Peer Assessment Scores (PAR) from acrylic models and their digital analogues were assessed.

Method

Ten models of different occlusions were digitised, using a 3 Shape R700 laser scanner (Copenhagen, Denmark). Each set of models was conventionally and digitally PAR-scored twice in random order by 10 examiners. The minimum time between repeat measurements was two weeks. The repeatability was assessed by applying Carstensen’s analysis. Inter-method agreement (IEMA) was assessed by Carstensen’s limit of agreement (LOA).

Results

Intra-examiner repeatability (IER) for the unweighted and weighted data was slightly better for the conventional rather than the digital models. There was a slightly higher negative bias of -1.62 for the weighted PAR data for the digital models. IEMA for the overall weighted data ranged from −8.70 – 5.45 (95% Confidence Interval, CI). Intra-class Correlation Coefficients (ICC) for the weighted data for conventional, individual and average scenarios were 0.955 (0.906 – 0.986 CI), 0.998 (0.995 – 0.999 CI). ICC for the weighted digital data, individual and average scenarios were 0.99 (0.97 – 1.00) and 1.00. The percentage reduction required to achieve an optimal occlusion increased by 0.4% for the digital scoring of the weighted data.

Conclusion

Digital PAR scores obtained from scanned plastic models were valid and reliable and, in this context, the digital semi-automated method can be used interchangeably with the conventional method of PAR scoring.

eISSN:
2207-7480
Język:
Angielski
Częstotliwość wydawania:
Volume Open
Dziedziny czasopisma:
Medicine, Basic Medical Science, other