Otwarty dostęp

Intra-rater and inter-rater reliability of the process of obtaining cross-sectional area and echo intensity measurements of muscles from ultrasound images


Zacytuj

Fig. 1.

Ultrasound image measurements analyzed for reliability
Ultrasound image measurements analyzed for reliability

Intra-rater reliability statistics ultrasound derived measurements of the leg extensor muscles

Scale Depth Rectus femoris Vastus lateralis
Variable (pixels/cm) (cm) CSA (cm2) El (AU) CSA (cm2) El (AU)
RATER 1 2 1 2 1 2 1 2 1 2 1 2
Average 41.5 41.4 0.82 0.77 6.29 6.21 10.04 10.02 21.78 21.66 12.04 11.91
CV% 2.51 1.89 10.97 10.12 11.66 6.72 6.11 9.93 6.01 5.54 6.95 5.61
P value 0.45 0.42 0.86 0.52 0.47 0.25 0.40 0.36 0.13 0.07 0.32 0.22
ICC 3,1 0.85 0.88 0.96 0.98 0.74 0.67 0.36 0.32 0.85 0.88 0.55 0.47
SEM 1.69 1.60 0.05 0.04 0.68 0.77 1.00 1.05 1.44 1.38 0.71 0.72
SEM (%) 4.07 3.86 6.10 5.19 10.81 12.40 9.96 10.48 6.61 6.37 5.90 6.05
MD 4.68 4.43 0.14 0.11 1.88 2.15 2.79 2.92 4.00 3.83 1.97 2.00

Inter-rater reliability statistics ultrasound derived measurements of the leg extensor muscles

Scale Depth Rectus Femoris Vastus Lateralis
Variable (pixels/cm) (cm) CSA (cm2) El (AU) CSA (cm2) El (AU)
Average 41.54 0.79 6.30 10.00 21.72 11.98
cv % 2.01 7.59 7.82 3.91 3.36 2.95
P value 0.68 0.4 0.12 0.21 0.98 0.45
ICC 3,1 0.87 0.97 0.87 0.77 0.92 0.91
SEM 1.62 0.04 0.47 0.56 1.11 0.28
SEM (%) 3.90 5.06 7.46 5.60 5.11 2.34
MD 4.49 0.13 1.30 1.56 3.06 0.78
eISSN:
2451-070X
Język:
Angielski
Częstotliwość wydawania:
4 razy w roku
Dziedziny czasopisma:
Medicine, Basic Medical Science, other