Intra-rater and inter-rater reliability of the process of obtaining cross-sectional area and echo intensity measurements of muscles from ultrasound images


Zitieren

Fig. 1.

Ultrasound image measurements analyzed for reliability
Ultrasound image measurements analyzed for reliability

Intra-rater reliability statistics ultrasound derived measurements of the leg extensor muscles

Scale Depth Rectus femoris Vastus lateralis
Variable (pixels/cm) (cm) CSA (cm2) El (AU) CSA (cm2) El (AU)
RATER 1 2 1 2 1 2 1 2 1 2 1 2
Average 41.5 41.4 0.82 0.77 6.29 6.21 10.04 10.02 21.78 21.66 12.04 11.91
CV% 2.51 1.89 10.97 10.12 11.66 6.72 6.11 9.93 6.01 5.54 6.95 5.61
P value 0.45 0.42 0.86 0.52 0.47 0.25 0.40 0.36 0.13 0.07 0.32 0.22
ICC 3,1 0.85 0.88 0.96 0.98 0.74 0.67 0.36 0.32 0.85 0.88 0.55 0.47
SEM 1.69 1.60 0.05 0.04 0.68 0.77 1.00 1.05 1.44 1.38 0.71 0.72
SEM (%) 4.07 3.86 6.10 5.19 10.81 12.40 9.96 10.48 6.61 6.37 5.90 6.05
MD 4.68 4.43 0.14 0.11 1.88 2.15 2.79 2.92 4.00 3.83 1.97 2.00

Inter-rater reliability statistics ultrasound derived measurements of the leg extensor muscles

Scale Depth Rectus Femoris Vastus Lateralis
Variable (pixels/cm) (cm) CSA (cm2) El (AU) CSA (cm2) El (AU)
Average 41.54 0.79 6.30 10.00 21.72 11.98
cv % 2.01 7.59 7.82 3.91 3.36 2.95
P value 0.68 0.4 0.12 0.21 0.98 0.45
ICC 3,1 0.87 0.97 0.87 0.77 0.92 0.91
SEM 1.62 0.04 0.47 0.56 1.11 0.28
SEM (%) 3.90 5.06 7.46 5.60 5.11 2.34
MD 4.49 0.13 1.30 1.56 3.06 0.78