Indicator | Low S/CO | High S/CO | Negative S/CO |
---|---|---|---|
Mean | 3.734 | 15.495 | 0.461 |
SD | 0.063 | 0.268 | 0.014 |
CV | 1.68% | 1.73% | 3.03% |
Indicator | Low S/CO | High S/CO | Negative S/CO |
---|---|---|---|
Mean | 4.018 | 16.284 | 0.474 |
SD | 0.065 | 0.365 | 0.016 |
CV | 1.63% | 2.24% | 3.38% |
Year | CV% |
---|---|
2020 | 3.61% |
2021 | 3.36% |
2022 | 3.67% |
Mean | 3.55% |
LICA500 in our hospital | LICA500 in Taizhou People’s Hospital | Agreement | |
---|---|---|---|
Positive | Negative | ||
Positive | 10 | 0 | 100% |
Negative | 0 | 10 | 100% |
COI | Photostimulated chemiluminescence assay positive | WB positive | WB negative |
---|---|---|---|
1.0–10.0 | 26 | 1 | 25 |
10.0–20.0 | 2 | 0 | 2 |
20.0–30.0 | 2 | 0 | 2 |
30.0–40.0 | 2 | 2 | 0 |
> 40.0 | 47 | 47 | 0 |
Total | 79 | 50 | 29 |