Accès libre

Diagnostic Efficacy of Photostimulated Chemiluminescence Assay for Detecting Anti-HIV Antibodies: A Retrospective Study

À propos de cet article

Citez

Fig. 1.

ROC curve of photostimulated chemiluminescence assay.
ROC curve of photostimulated chemiluminescence assay.

Coefficient of variation of the intra test for photostimulated chemiluminescence assay.

Indicator Low S/CO High S/CO Negative S/CO
Mean 3.734 15.495 0.461
SD 0.063 0.268 0.014
CV 1.68% 1.73% 3.03%

Coefficient of variation of the inter test for photostimulated chemiluminescence assay.

Indicator Low S/CO High S/CO Negative S/CO
Mean 4.018 16.284 0.474
SD 0.065 0.365 0.016
CV 1.63% 2.24% 3.38%

CV values for three years of in-house quality control.

Year CV%
2020 3.61%
2021 3.36%
2022 3.67%
Mean 3.55%

Agreement of inter-test measure photostimulated chemiluminescence assay.

LICA500 in our hospital LICA500 in Taizhou People’s Hospital Agreement
Positive Negative
Positive 10 0 100%
Negative 0 10 100%

Distribution ranges of COI values and WB test results in photostimulated chemiluminescence assay.

COI Photostimulated chemiluminescence assay positive WB positive WB negative
  1.0–10.0 26 1 25
10.0–20.0   2   0   2
20.0–30.0   2   0   2
30.0–40.0   2   2   0
> 40.0 47 47   0
Total 79 50 29