Acceso abierto

Diagnostic Efficacy of Photostimulated Chemiluminescence Assay for Detecting Anti-HIV Antibodies: A Retrospective Study


Cite

Fig. 1.

ROC curve of photostimulated chemiluminescence assay.
ROC curve of photostimulated chemiluminescence assay.

Coefficient of variation of the intra test for photostimulated chemiluminescence assay.

Indicator Low S/CO High S/CO Negative S/CO
Mean 3.734 15.495 0.461
SD 0.063 0.268 0.014
CV 1.68% 1.73% 3.03%

Coefficient of variation of the inter test for photostimulated chemiluminescence assay.

Indicator Low S/CO High S/CO Negative S/CO
Mean 4.018 16.284 0.474
SD 0.065 0.365 0.016
CV 1.63% 2.24% 3.38%

CV values for three years of in-house quality control.

Year CV%
2020 3.61%
2021 3.36%
2022 3.67%
Mean 3.55%

Agreement of inter-test measure photostimulated chemiluminescence assay.

LICA500 in our hospital LICA500 in Taizhou People’s Hospital Agreement
Positive Negative
Positive 10 0 100%
Negative 0 10 100%

Distribution ranges of COI values and WB test results in photostimulated chemiluminescence assay.

COI Photostimulated chemiluminescence assay positive WB positive WB negative
  1.0–10.0 26 1 25
10.0–20.0   2   0   2
20.0–30.0   2   0   2
30.0–40.0   2   2   0
> 40.0 47 47   0
Total 79 50 29
eISSN:
2544-4646
Idioma:
Inglés
Calendario de la edición:
4 veces al año
Temas de la revista:
Life Sciences, Microbiology and Virology