Adaptation of ENDS Device Measurements into Puffing Topography Variables and Clinical Implications
16. Mai 2025
Über diesen Artikel
Online veröffentlicht: 16. Mai 2025
Seitenbereich: 40 - 50
Eingereicht: 01. Okt. 2024
Akzeptiert: 12. März 2025
DOI: https://doi.org/10.2478/cttr-2025-0005
Schlüsselwörter
© 2025 Michael J. Oldham et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.
Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Percent of device actuation occurrences as a function of flowrate and pressure-drop measurements_
Flowrate setting (cm3/s) | Pressure-drop (Pascal) | Percent of missing device activations | ||
---|---|---|---|---|
Mean | SD | N | ||
5 | 47 | 7.3 | 2 | 100 |
7.5 | 92 | 12.8 | 2 | 100 |
10 | 151 | 14.9 | 2 | 18.3 |
12.5 | 227 | 29.5 | 2 | 5.0 |
16.7 | 380 | 36.3 | 2 | 3.3 |
20.8 | 539 | 51.9 | 2 | 5.0 |
25.0 | 727 | 48.7 | 2 | 0 |
Summary of duration of device activation for three puffing regimens_
Device number | Mean percent of puff duration |
|||||
---|---|---|---|---|---|---|
Virginia Tobacco | Polar Menthol | |||||
Square | Sine | Gap | Square | Sine | Gap | |
1 | 95.0 | 76.1 | 91.3 | 95.1 | 76.1 | 92.0 |
2 | 95.4 | 77.2 | 90.0 | 95.8 | 73.5 | 90.1 |
3 | 95.9 | 77.2 | 90.9 | 95.5 | 76.1 | 90.8 |
4 | 95.5 | 74.5 | 92.3 | 96.5 | 77.4 | 93.3 |
5 | 96.1 | 76.6 | 90.4 | 96.8 | 75.9 | 92.7 |
6 | 91.3 | 74.6 | 88.3 | 95.4 | 77.4 | 91.9 |
7 | 95.9 | 73.8 | 86.9 | 96.6 | 76.6 | 91.1 |
8 | 95.5 | 78.9 | 89.1 | 95.8 | 76.6 | 90.2 |
9 | 97.1 | 77.4 | 92.5 | 96.2 | 76.9 | 92.8 |
10 | 97.0 | 76.2 | 89.7 | 95.7 | 77.1 | 92.8 |
11 | 93.8 | 76.3 | 90.4 | 96.8 | 78.1 | 93.3 |
12 | 93.0 | 76.8 | 89.1 | 95.5 | 78.2 | 92.1 |
13 | 92.4 | 75.4 | 88.2 | 96.8 | 78.2 | 93.1 |
14 | 94.6 | 74.6 | 91.2 | 95.8 | 77.1 | 92.1 |
15 | 93.8 | 77.6 | 90.3 | 95.4 | 75.5 | 91.5 |
Mean | 94.8 | 76.2 | 90.0 | 96.0 | 76.7 | 92.0 |
SD | 1.6 | 1.4 | 1.5 | 0.6 | 1.2 | 1.0 |
Summary of average measured device mass loss, aerosol collected mass, and estimated aerosol mass from the JUUL2® system_
Puffing regimen | 50 Puff collection mean mass ± standard deviation (mg) (N = 15) | |||||
---|---|---|---|---|---|---|
Virginia Tobacco | Polar Menthol | |||||
Device mass loss | Aerosol collected mass | Estimated aerosol mass | Device/pod mass loss | Aerosol collected mass | Estimated aerosol mass | |
Square | 466.6 ± 22.7 | 474.2 ± 22.1 | 408.0 ± 20.7 | 526.8 ± 30.1 | 532.1 ± 30.3 | 402.1 ± 35.0 |
Sine | 362.8 ± 15.4 | 371.3 ± 15.9 | 308.6 ± 13.2 | 441.3 ± 19.1 | 448.5 ± 19.4 | 349.4 ± 9.7 |
Gap | 330.6 ± 25.1 | 335.8 ± 25.4 | 304.0 ± 9.6 | 399.2 ± 24.0 | 404.0 ± 24.4 | 338.3 ± 6.9 |