Adaptation of ENDS Device Measurements into Puffing Topography Variables and Clinical Implications

, , , , , ,  und   
16. Mai 2025

Zitieren
COVER HERUNTERLADEN

Figure 1.

Diagram of the flowrate versus pressure-drop measurement apparatus.
Diagram of the flowrate versus pressure-drop measurement apparatus.

Figure 2.

Three puffing regimens using a square wave (A), sine wave (B) and gap wave (C). All puff profiles were 55 mL in volume with the square and sine wave using that volume over 3 seconds while the gap profile was over 2.5 seconds.
Three puffing regimens using a square wave (A), sine wave (B) and gap wave (C). All puff profiles were 55 mL in volume with the square and sine wave using that volume over 3 seconds while the gap profile was over 2.5 seconds.

Figure 3.

Mean (SD) pressure-drop measurement of the 30 JUUL2® devices. Standard deviation values for some data points are within the data point. The dark vertical rectangle is the estimated pressure-drop required for device activation.
Mean (SD) pressure-drop measurement of the 30 JUUL2® devices. Standard deviation values for some data points are within the data point. The dark vertical rectangle is the estimated pressure-drop required for device activation.

Figure 4.

Correlation of laboratory measured device mass loss versus estimated aerosol mass (A) and aerosol collected mass versus estimated aerosol mass (B) from the JUUL2® device for square wave (circles), sine wave (squares), and gap wave (triangles) puffing profile for Virginia Tobacco pods.
Correlation of laboratory measured device mass loss versus estimated aerosol mass (A) and aerosol collected mass versus estimated aerosol mass (B) from the JUUL2® device for square wave (circles), sine wave (squares), and gap wave (triangles) puffing profile for Virginia Tobacco pods.

Figure 5.

Correlation of laboratory measured device mass loss versus estimated aerosol mass (A) and aerosol collected mass versus estimated aerosol mass (B) from the JUUL2® device for square wave (circles), sine wave (squares), and gap wave (triangles) puffing profile for Polar Menthol pods.
Correlation of laboratory measured device mass loss versus estimated aerosol mass (A) and aerosol collected mass versus estimated aerosol mass (B) from the JUUL2® device for square wave (circles), sine wave (squares), and gap wave (triangles) puffing profile for Polar Menthol pods.

Percent of device actuation occurrences as a function of flowrate and pressure-drop measurements_

Flowrate setting (cm3/s) Pressure-drop (Pascal) Percent of missing device activations
Mean SD N
5 47 7.3 2 100
7.5 92 12.8 2 100
10 151 14.9 2 18.3
12.5 227 29.5 2 5.0
16.7 380 36.3 2 3.3
20.8 539 51.9 2 5.0
25.0 727 48.7 2 0

Summary of duration of device activation for three puffing regimens_

Device number Mean percent of puff durationa
Virginia Tobacco Polar Menthol
Square Sine Gap Square Sine Gap
1 95.0 76.1 91.3 95.1 76.1 92.0
2 95.4 77.2 90.0 95.8 73.5 90.1
3 95.9 77.2 90.9 95.5 76.1 90.8
4 95.5 74.5 92.3 96.5 77.4 93.3
5 96.1 76.6 90.4 96.8 75.9 92.7
6 91.3 74.6 88.3 95.4 77.4 91.9
7 95.9 73.8 86.9 96.6 76.6 91.1
8 95.5 78.9 89.1 95.8 76.6 90.2
9 97.1 77.4 92.5 96.2 76.9 92.8
10 97.0 76.2 89.7 95.7 77.1 92.8
11 93.8 76.3 90.4 96.8 78.1 93.3
12 93.0 76.8 89.1 95.5 78.2 92.1
13 92.4 75.4 88.2 96.8 78.2 93.1
14 94.6 74.6 91.2 95.8 77.1 92.1
15 93.8 77.6 90.3 95.4 75.5 91.5
Mean 94.8 76.2 90.0 96.0 76.7 92.0
SD 1.6 1.4 1.5 0.6 1.2 1.0

Summary of average measured device mass loss, aerosol collected mass, and estimated aerosol mass from the JUUL2® system_

Puffing regimen 50 Puff collection mean mass ± standard deviation (mg) (N = 15)
Virginia Tobacco Polar Menthol
Device mass loss Aerosol collected mass Estimated aerosol mass Device/pod mass loss Aerosol collected mass Estimated aerosol mass
Square 466.6 ± 22.7 474.2 ± 22.1 408.0 ± 20.7 526.8 ± 30.1 532.1 ± 30.3 402.1 ± 35.0
Sine 362.8 ± 15.4 371.3 ± 15.9 308.6 ± 13.2 441.3 ± 19.1 448.5 ± 19.4 349.4 ± 9.7
Gap 330.6 ± 25.1 335.8 ± 25.4 304.0 ± 9.6 399.2 ± 24.0 404.0 ± 24.4 338.3 ± 6.9