Adaptation of ENDS Device Measurements into Puffing Topography Variables and Clinical Implications
Online veröffentlicht: 16. Mai 2025
Seitenbereich: 40 - 50
Eingereicht: 01. Okt. 2024
Akzeptiert: 12. März 2025
DOI: https://doi.org/10.2478/cttr-2025-0005
Schlüsselwörter
© 2025 Michael J. Oldham et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.
Puffing topography attempts to quantify the intensity of tobacco product use in a common set of terminology and variables (e.g., number of puffs, puff duration, puff volume, interpuff interval, etc.). However, with each new type of inhaled tobacco product, the importance of certain variables changes and some of the terminology needs to be redefined based upon the specific type of product (e.g., cigarettes, heat-not-burn, and electronic nicotine delivery systems (ENDS), etc.). As noted in a recent review of puffing topography (1), the variability in ENDS products preclude use of some of the puff topography variables that have been historically used for cigarettes. For example, because cigarettes are relatively homogenous the number of cigarettes smoked/day provide a relatively good estimate of product use intensity (1, 2). This doesn’t necessarily apply to ENDS since depending on the e-liquid storage capacity of the ENDS device and the per-puff aerosol mass generated, a single pod or tank full of e-liquid can last throughout a single day or several days.
A recent review of puffing topography provided the advantages and disadvantages of two common methods that have been used to measure puffing topography variables, video recording and use of specialized add-on devices (Clinical Research Support System – CReSS handheld devices and SODIM Smoking Puff Analyzer Mobile Device – SPA-M from Körber Technologies GmbH, Germany, wireless portable use monitor – wPUMTM from the Rochester Institute of Technology; note these devices do not interface with the ENDS product electronics) (1). A similar specialized add-on device for open tank systems provided results similar to a CReSS device in a comparative clinical trial (3, 4). All of these specialized add-on devices use one or more pressure sensors to detect a puff, meaning that within each specialized add-on device there is a minimum pressure difference (pressure drop) required to trigger puff recording. This pressure drop threshold also means that each specialized add-on device will not recognize some part of each human puff that is below the minimum pressure difference, resulting in underestimation of puff duration and volume (5). These specialized add-on devices may also change the users’ behavior. In recent work (6), it was found that the measured topography variables may be influenced by use of one of these specialized add-on devices, CReSS pocket, which echoes concerns regarding use of these specialized add-on devices that date from the 1980s (7). Two recently published studies reported topography variables being directly measured by the ENDS device (8, 9). These two studies used the Joyetech® eVic™ Supreme ENDS device but did not provide any data on the minimum pressure drop required to trigger puff recording nor the accuracy of the measured number of puffs or puff duration reported. F
It is important to realize that every one of the abovementioned measurement techniques is measuring a unique aspect of human puff topography and comparison among these different measurement techniques are likely to result in small differences in the measured human topography variables. For example, use of different criteria when evaluating the video recording for the beginning and ending of puffing (e.g., duration of the light-emitting diode if present on the device and visible, when the device is inserted and removed from the mouth, change in cheek shape, abdominal volume change, etc.) will create variability in puff duration measurements. Although, this variability can be reduced by using the same criteria with two independent evaluators, it will still be different from puff duration as measured by specialized add-on devices (e.g., CReSS pocket, SPA-M, and wPUMTM) or ENDS devices that use a pressure drop threshold to measure puff duration. By design, a pressure drop threshold omits a certain part of a human puff profile (i.e., beginning and ending parts that are below the pressure drop threshold), but has to be balanced against the incidence of phantom or false puffs due to normal daily events that exceed the pressure drop threshold (e.g., pod insertion and removal, rapid movement of the device when no puff is taken, being dropped, or tossed onto a counter/surface, or ambient pressure changes due to riding in an elevator, etc.). Incorporation of a pressure sensor and photometer into an existing ENDS device (12), would only measure the duration of aerosol production of the ENDS device, once the e-liquid is heated sufficiently to produce an aerosol and the photometer’s threshold value is exceeded. Similar to a pressure drop threshold, photometers also have a threshold and therefore will omit a small duration of aerosol production at the beginning and end of a puff. B
In addition to the Joyetech eVic Supreme, the PUB device, and the modified Aspire Nautilus Prime X, some of the next generation ENDS products also are capable of providing measurements of selected device parameters. The selected device parameters include device activation duration (duration of battery current to heater), counting number of device activations and an estimate of aerosol mass produced by each puff (or device activation). It is measurement of these selected device parameters, their translation into standardized puffing topography parameters (i.e., puff duration and number of puffs) and the estimated aerosol mass that is the focus of this laboratory study for the JUUL2® System.
All testing was performed at the Juul Labs Regulatory Chemistry Laboratory in Durham, North Carolina, which is an ISO 17025 accredited facility. The JUUL2® System consists of a JUUL2® device and a single use JUUL2® pod (1.2 mL volume). A total of 30 commercial JUUL2® devices obtained from finished goods inventory were used in this laboratory study. For the first part of the laboratory study an empty pod was used in each of the 30 JUUL2® devices. For the second part of the laboratory study, 15 of the 30 JUUL2® devices were used with pods filled with either Virgina Tobacco (18 mg/mL nicotine) and the other 15 JUUL2® devices were used with pods filled with Polar Menthol (18 mg/mL nicotine) e-liquids. Both the Virginia Tobacco and Polar Menthol e-liquids contain United States Pharmacopeia compliant propylene glycol, glycerin, and nicotine along with proprietary flavors. A single lot of Virgina Tobacco (18 mg/mL nicotine) and Polar Menthol (18 mg/mL nicotine) e-liquid-filled pods from finished goods inventory were used and all e-liquid filled pods used were approximately 16 months post manufacture and were within their shelf-life period of 18 months. For both parts of the laboratory study all devices were fully charged prior to use and for the second part of the study the JUUL2® devices with pods were oriented 45 degrees to gravity for testing. The study was conducted in a controlled temperature 22 ± 2 °C and humidity 60 ± 5% relative humidity laboratory environment.
In the first part of the laboratory study, different flowrates were set using a mass-flow controller (Allicat, MC-20SLPM-D-DB9M/5M, 5IN, Tucson, AZ, USA) that was connected to a digital manometer (Bonoutil® 522, Hangzhou, Zhejiang, China), which was protected from any aerosol generation by a Cambridge filter pad (Performance Systematix Inc., acquired by Selig Group, Grand Rapids, MI, USA) ( Figure 1). The inherent pressure-drop of this test system (without the JUUL2® system attached) was determined at each flowrate and was subtracted from the pressure-drop value measured when a JUUL2® device with empty pod were attached. Each of the JUUL2® devices with an empty pod were attached to the Cambridge filter pad holder using a labyrinth seal and the mass-flow controller was set to one of the following flowrates: 0.3, 0.45, 0.6, 0.75, 1.0, 1.25, and 1.5 standard L/min (note these standard flowrates translate to 5, 7.5, 10.0, 12.5, 16.7, 20.8, and 25.0 standard cm3/s). At each flowrate an empty pod was inserted into the JUUL2® device, the mass flow controller was used to set the desired flowrate, a reading from the digital manometer was recorded, flow was stopped using the mass flow controller and the pod was removed. Two replicate measurements per flowrate for each JUUL2® device were performed with the first replicate starting at 0.3 L/min and finishing at 1.5 L/min and the second replicate starting at 1.5 L/min and finishing at 0.3 L/min. Because the JUUL2® device activation stops after six seconds the average of the first six seconds of the digital manometer readings was used to correlate to the maximum draw strength reported in the JUUL2® device logs to determine whether the flowrate triggered device activation. The number of pressure versus flow replicate measurements that didn’t trigger device activation were also recorded.

Diagram of the flowrate
In the second part of the laboratory study the 30 JUUL2® devices were randomly divided into two equal groups. One group of JUUL2® devices was used with Virginia Tobacco (18 mg/mL nicotine) pods and the second group of JUUL2® devices was used with Polar Menthol (18 mg/mL nicotine) pods. For each group three puffing regimens were used:

The gap was created using a three-way solenoid valve (Parker/71315, New Britain, CT, USA) placed between the piston and JUUL2® devices with one way vented to atmosphere that was controlled with a pressure switch (SMC/ZSE30A-NO1-P, Noblesville, IN, USA) and a calibrated digital timer (Eaton Moeller/Easy 621-DC-TC, Cologne, Germany). Each group of JUUL2® devices were puffed in groups of five devices simultaneously using a 20-port SM450 RH linear smoking machine (Cerulean, Milton Keynes, UK). The first five ports were programmed for the square wave puff profile (14), ports 6–10 were programmed for the sine wave puff profile and ports 11–15 were programmed for the gap wave puffing profile. Prior to puffing JUUL2® devices, the puffing profile on each port was verified using a calibrated puff profiler (Cerulean VFA450RH, Milton Keynes, UK) designed for use with the SM450 RH linear puffing machine (Cerulean, Milton Keynes, UK). Each group of JUUL2® devices were rotated among these 15 ports. At each rotation, the JUUL2® devices were fully charged, and a new pod was inserted (Virginia Tobacco 18 mg/mL nicotine or Polar Menthol 18 mg/mL nicotine). For each puff profile, a total of 50 puffs from each JUUL2® device/pod combination were performed in 10 puffs/block with each block separated by a 2-min rest period. JUUL2® devices with pods were weighed at the beginning and after conclusion of the 50 puffs (called device mass loss), as was the filter collecting all of the aerosol mass generated from each JUUL2® device/pod combination (called aerosol collected mass). These gravi-metric estimates of aerosol mass produced were compared to aerosol mass estimate from the JUUL2® device (i.e., the sum of estimated aerosol mass from the JUUL2® device for all device activations) provided in the JUUL2® device logs. This estimate mass is based upon a thermal model where e-liquid parameters are determined experimentally (e.g., latent heat of vaporization of the liquid, specific heat of the liquid in use, heat losses to the environment, etc.), which enables an estimation of the power required to vaporize an amount of e-liquid and therefore how much e-liquid was vaporized.
To assess if there were differences in device activation duration between the formulations (Viginia Tobacco
The mean flowrate

Percent of device actuation occurrences as a function of flowrate and pressure-drop measurements.
Flowrate setting (cm3/s) | Pressure-drop (Pascal) | Percent of missing device activations | ||
---|---|---|---|---|
Mean | SD | N | ||
5 | 47 | 7.3 | 2 | 100 |
7.5 | 92 | 12.8 | 2 | 100 |
10 | 151 | 14.9 | 2 | 18.3 |
12.5 | 227 | 29.5 | 2 | 5.0 |
16.7 | 380 | 36.3 | 2 | 3.3 |
20.8 | 539 | 51.9 | 2 | 5.0 |
25.0 | 727 | 48.7 | 2 | 0 |
N = number of replicate measurements; SD = standard deviation
For the square wave puffing regimen, the duration of JUUL2® device’s activation averaged 95% (Table 2) of the 3 second puff duration with a small but statistically significant difference between the Virginia Tobacco (94.8%) and Polar Menthol (96.0%) formulations (p = 0.0227). Statistical analysis indicated testing order did influence (p < 0.00001) percentage of the device activation duration for both Virginia Tobacco and Polar Menthol pods using the square wave puffing regimen. For the sine wave puffing regimen, the duration of the JUUL2® device’s activation averaged 76.4% of the 3-s puff showing that more of the sine wave compared to the square wave was below the 95–150 Pa pressure-drop required for device activation (Table 2). No statistically significant differences were noted between percent of the 3-s puff that device activation occurred for Virginia Tobacco (76.4%) and Polar Menthol (76.7%) formulations (p = 0.3084) and there was no order of testing influence on the percent of the 3-s puff that device activation constituted. For the gap wave puffing regimen, the duration of the JUUL2® device’s activation averaged 91% of the 2.5-s puff duration (Table 2). There was a small but statistically significant difference in the percent of the 2.5 second puff device activation between pods with Virginia Tobacco (90.0%)
Summary of duration of device activation for three puffing regimens.
Device number | Mean percent of puff durationa | |||||
---|---|---|---|---|---|---|
Virginia Tobacco | Polar Menthol | |||||
Square | Sine | Gap | Square | Sine | Gap | |
1 | 95.0 | 76.1 | 91.3 | 95.1 | 76.1 | 92.0 |
2 | 95.4 | 77.2 | 90.0 | 95.8 | 73.5 | 90.1 |
3 | 95.9 | 77.2 | 90.9 | 95.5 | 76.1 | 90.8 |
4 | 95.5 | 74.5 | 92.3 | 96.5 | 77.4 | 93.3 |
5 | 96.1 | 76.6 | 90.4 | 96.8 | 75.9 | 92.7 |
6 | 91.3 | 74.6 | 88.3 | 95.4 | 77.4 | 91.9 |
7 | 95.9 | 73.8 | 86.9 | 96.6 | 76.6 | 91.1 |
8 | 95.5 | 78.9 | 89.1 | 95.8 | 76.6 | 90.2 |
9 | 97.1 | 77.4 | 92.5 | 96.2 | 76.9 | 92.8 |
10 | 97.0 | 76.2 | 89.7 | 95.7 | 77.1 | 92.8 |
11 | 93.8 | 76.3 | 90.4 | 96.8 | 78.1 | 93.3 |
12 | 93.0 | 76.8 | 89.1 | 95.5 | 78.2 | 92.1 |
13 | 92.4 | 75.4 | 88.2 | 96.8 | 78.2 | 93.1 |
14 | 94.6 | 74.6 | 91.2 | 95.8 | 77.1 | 92.1 |
15 | 93.8 | 77.6 | 90.3 | 95.4 | 75.5 | 91.5 |
Mean | 94.8 | 76.2 | 90.0 | 96.0 | 76.7 | 92.0 |
SD | 1.6 | 1.4 | 1.5 | 0.6 | 1.2 | 1.0 |
SD = Standard Deviation
Square and sine wave puff duration = 3 seconds and for the gap wave puff duration = 2.5 seconds. Each value is the mean of 50 puffs.
In the square wave puffing regimen testing, approximately 1% of the device activations were anomalous in that they were split into two activations when only a single puff was taken on the SM450 RH linear smoking machine port. The percent of device activations split into two activations increased to approximately 4% for the sine wave puffing regimen testing and decreased to approximately 1.8% for the gap wave regimen. No meaningful differences were found between the Virginia Tobacco and Polar Menthol pods. These anomalies didn’t affect device activation duration measurements as the device activation duration of both activations were manually combined into a single duration. The 50-puff mean device mass loss and aerosol collected mass measurements were different between the Virginia Tobacco and Polar Menthol filled pods, such that mass loss was greater for the Polar menthol pods. However, there was little difference in the estimated aerosol mass from the JUUL2® System for the two formulations (Table 3).
Summary of average measured device mass loss, aerosol collected mass, and estimated aerosol mass from the JUUL2® system.
Puffing regimen | 50 Puff collection mean mass ± standard deviation (mg) (N = 15) | |||||
---|---|---|---|---|---|---|
Virginia Tobacco | Polar Menthol | |||||
Device mass loss | Aerosol collected mass | Estimated aerosol mass | Device/pod mass loss | Aerosol collected mass | Estimated aerosol mass | |
Square | 466.6 ± 22.7 | 474.2 ± 22.1 | 408.0 ± 20.7 | 526.8 ± 30.1 | 532.1 ± 30.3 | 402.1 ± 35.0 |
Sine | 362.8 ± 15.4 | 371.3 ± 15.9 | 308.6 ± 13.2 | 441.3 ± 19.1 | 448.5 ± 19.4 | 349.4 ± 9.7 |
Gap | 330.6 ± 25.1 | 335.8 ± 25.4 | 304.0 ± 9.6 | 399.2 ± 24.0 | 404.0 ± 24.4 | 338.3 ± 6.9 |
Association of the measured device mass loss and aerosol collected mass for pods with Virginia Tobacco 18 mg/mL e-liquid to the JUUL2® device estimated aerosol mass was different for each puffing regimen (

Correlation of laboratory measured device mass loss

Correlation of laboratory measured device mass loss
Based upon results of this laboratory study, it is possible to translate selected device parameters provided by the JUUL2® System into clinically relevant standardized puffing topography parameters (i.e., puff duration and number of puffs), as well as, estimated aerosol mass, with some caveats. For example, duration of device activation provided by the JUUL2® System is sensitive enough to be used as a surrogate for puff duration with the following caveats: it requires a minimum pressure drop of 95–150 Pa and it has a maximum activation duration of 6 seconds meaning aerosol is no longer produced after 6 seconds because battery current to the heater is turned off. The minimum pressure drop means that the part of the human puff that is below this pressure (95–150 Pa) will not be included (beginning and end of puff) in the estimate of puff duration represented by device activation duration. Our laboratory study demonstrated this is a function of puff shape when puff volume and duration are held constant, with a sine wave only representing on average 76.4% and a square wave puff of equal volume representing on average 95% of the 3 second puff duration.
However, puff volume and duration are just as important, and the three variables are all inter-related in human topography. Using the pressure drop of 95–150 Pa and a 3 s-sine wave puff profile, the effect of different puff volumes ranging from 25 mL to 100 mL resulted in predicted device activation durations (puff duration) from 1.4–1.8 s for the 25 mL-puff volume to 2.8 s for the 100 mL-puff volume (
The number of device activations can be used as a surrogate for number of puffs with the following caveat: because of the sensitivity of the JUUL2® System, sometimes a single puff is split into two puffs with the frequency seeming to increase as the puff profile becomes more complex (e.g., more similar to human puffing) meaning that puffs with short durations (e.g., # 0.5 s) should be combined with the appropriate parent puff (11). The challenge of false or phantom puffs and splitting puffs have been previously reported for CReSSmicro™ and SPA-M add-on devices (5, 15). The differences between the device mass loss and aerosol collect mass measurements are likely due to the hygroscopicity of the aerosol collected on the filter (e.g., propylene glycol and glycerin at the laboratory relative humidity) (16, 17). Finally, the estimated aerosol mass provided by the JUUL2® System correlated with the measured device mass loss and aerosol-collected mass better when data from all puffing regimens was combined
To test translation of these selected JUUL2® device parameters into standardized puffing topography parameters, a simple algorithm was constructed and the JUUL2® device logs from this laboratory study were evaluated using the simple algorithm. The algorithm was designed to combine device activations (puffs) if they were less than 0.5 seconds in duration or within 1.5 seconds of each other (end of one to beginning of another) and if the subsequent device activation (total puff duration) did not exceed 4.3 seconds. If the combined device activations exceeded 4.3 seconds, the device activations (puffs) were kept separate. In processing the device logs for the square wave puffing profile using this algorithm, the percent of puff duration for the 15 devices with Virginia Tobacco and Polar Menthol formulation were 96.1% and 96.1%, respectively. When device logs were processed for the sine wave puffing profile, the percent of puff duration for the 15 devices with Virginia Tobacco and Polar Menthol formulation were 76.9% and 77.3%, respectively. For both the square wave and sine wave puff profiles the simple algorithm resulted in slight improvements in percent of puff duration. More importantly, the number of device activations exactly matched the number of puffs from the SM450 puffing machine, with no anomalous device activations noted for the square and sine wave puffing profiles. This suggests that algorithm-processed device logs can be used as an approximation to traditional topography measures, subject to the caveats noted above. The gap wave regimen device logs were not tested since there was an intentional 0.5-s gap between the two portions of the gap wave regimen.
The flowrate between 7.5 and 10 cm3/s resulting in the pressure drop required to activate the JUUL2® device of 95 and 150 Pa (0.96–1.53 cm water) is approximately half of the flowrate reported for the CReSSmicro™ add-on device of 20 cm3/s (5). It is similar to the lowest flowrate (10 cm3/s) used to evaluate the CReSS Pocket and SPA-M add-on devices (15) and calibrate the second generation wPUMTM add-on device (19). The flowrate between 7.5 and 10 cm3/s resulting in the pressure drop threshold required for JUUL2® device activation is also similar to the minimum flow rate (8.33 cm3/s) required to activate puffing for the Vuse Solo ENDS product (20) used in the PUB add-on device (11). Using a 3-s square wave puffing profile, the JUUL2® device activation duration was 95% or only 150 milliseconds different from the actual puff duration. In comparison the CReSS Pocket and SPA-M add-on devices with three different e-cigarettes were generally ± 10% of the actual puff duration with the CReSS Pocket add-on device in some cases being up to 18% shorter (15), while the puff duration accuracy for the wPUMTM add-on device is ± 100 microseconds (19). Using a 55 mL-puff volume in a sine wave profile the CReSS Micro add-on device measured on average only 66.5% of the puff duration compared to an average 76.4% for the JUUL2® System without the simple algorithm and 77.1% using the simple algorithm. Using a bell-shaped puffing profile the CReSS Pocket add-on device puff duration was up to 22% less than the actual puff duration while the SPA-M add-on device was within ± 10% of the actual puff duration (15).
Several publications have indicated challenges with topography data, noting that it should be reviewed for accuracy prior to analysis (5, 11, 15). O
Our laboratory study has several limitations including that only three puffing regimens were used and only one puff volume was used (55 mL) with each puffing regimen. The use of only three puffing regimens will not capture the variability of human puffing topography and may not be representative of human ENDS puffing topography. Although the square wave has been reported as the most representative of ENDS topography in a natural setting (22), there is not universal agreement on what puff profile is the most representative of human topography. P
In conclusion, our laboratory study demonstrated that selected JUUL2® device measured parameters can be translated into classical puffing topography variables (i.e., number of puffs and puff duration), and can provide an estimated aerosol mass generated by the JUUL2® System.