Probing deep lung regions using a new 6-electrode tetrapolar impedance method
, and
Jan 08, 2023
About this article
Published Online: Jan 08, 2023
Page range: 116 - 124
Received: Dec 01, 2022
DOI: https://doi.org/10.2478/joeb-2022-0016
Keywords
© 2022 Mahjabin Mobarak, Muhammad Abdul Kadir and K Siddique-e Rabbani, published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Fig.1
![Electrode connections for transfer impedance measurement using the 6-electrode TPIM technique (left, 1a) and that for a 4 electrode TPIM (right, 1b) used for comparison. The sensitive regions are indicated in dark red, shades representing qualitative magnitudes (not to scale). Reproduced from [9] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_001.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T230908Z&X-Amz-Expires=3600&X-Amz-Signature=58118ae4cbd0393a690c7282db3182b48e940c25f07d2bbc34ae25ecc4ba6f7f&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig.2

Fig.3
![Near to real life 3D model of the thorax. The positioning of electrodes for the measurements undertaken are also shown. Reproduced from [17] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_003.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T230908Z&X-Amz-Expires=3600&X-Amz-Signature=2316a8420138f4d4e9019cc4108e660ec3d8382786555951f77db17f0b417f26&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig.4
![Some details of a cross section of the modelled thorax. Reproduced from [17] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_004.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T230908Z&X-Amz-Expires=3600&X-Amz-Signature=c7c9a31747ba225e2dfc389d0c2bca68a899f027e3d311de49b377ba1cc25162&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig.5

Fig.6

Fig.7

Fig.8

Fig.9
![6-electrode TPIM covering both lung regions. Reproduced from [9] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_009.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T230908Z&X-Amz-Expires=3600&X-Amz-Signature=0b7762e2c6d943ad3e17f27b988ad71d2168edcee66c6f56edefab502d9cc624&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Simulation results for potential electrode separation of 7 cm
6-TPIM | 4-TPIM, Front | 4-TPIM, back | |||||||
---|---|---|---|---|---|---|---|---|---|
Inspired | Expired | % change* | Inspired | Expired | % change* | Inspired | Expired | % change* | |
Total Imp, Ω | 10.6 | 7.8 | 34.7 | 10.4 | 8.0 | 30.4 | 10.6 | 7.7 | 37.9 |
% CLTI(1) | 24.2 | 35.6 | -32.1 | 20.5 | 31.2 | -34.1 | 28.3 | 40.4 | -29.8 |
% CLTI(2) | 25.0 | 38.0 | -34.1 | 22.2 | 35.3 | -37.1 | 28.2 | 41.2 | -31.5 |
% OCLTI | 3.6 | 6.8 | 8.2 | 13.4 | -0.3 | 2.1 |
Simulation results for potential electrode separation of 10 cm
6-TPIM | 4-TPIM, Front | 4-TPIM, back | |||||||
---|---|---|---|---|---|---|---|---|---|
Inspired | Expired | % change* | Inspired | Expired | % change* | Inspired | Expired | % change* | |
Total Imp, Ω | 13.9 | 10.6 | 31.4 | 13.6 | 10.3 | 32.2 | 14.0 | 9.9 | 42.1 |
% CLTI(1) | 26.3 | 36.4 | -27.6 | 21.9 | 33.4 | -34.4 | 29.8 | 42.8 | -30.5 |
% CLTI(2) | 27.2 | 38.0 | -28.6 | 23.5 | 36.9 | -36.3 | 29.8 | 43.5 | -31.4 |
% OCLTI | 3.2 | 4.6 | 7.4 | 10.6 | 0.2 | 1.6 |
Ratio of values for 10 cm Electrode Separation to those for 7 cm ES
6-TPIM | 4-TPIM Front | 4-TPIM back | |
---|---|---|---|
% CLTI(1) Inspiration | 1.09 | 1.07 | 1.05 |
% CLTI(1) Expiration | 1.02 | 1.07 | 1.06 |
%OCLTI Inspiration | 0.91 | 0.90 | |
%OCLTI Expiration | 0.68 | 0.79 | 0.75 |