Probing deep lung regions using a new 6-electrode tetrapolar impedance method
, e
08 gen 2023
INFORMAZIONI SU QUESTO ARTICOLO
Pubblicato online: 08 gen 2023
Pagine: 116 - 124
Ricevuto: 01 dic 2022
DOI: https://doi.org/10.2478/joeb-2022-0016
Parole chiave
© 2022 Mahjabin Mobarak, Muhammad Abdul Kadir and K Siddique-e Rabbani, published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Fig.1
![Electrode connections for transfer impedance measurement using the 6-electrode TPIM technique (left, 1a) and that for a 4 electrode TPIM (right, 1b) used for comparison. The sensitive regions are indicated in dark red, shades representing qualitative magnitudes (not to scale). Reproduced from [9] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_001.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T215851Z&X-Amz-Expires=3600&X-Amz-Signature=073aa95532faa16d8cd6cb42053e9b702e1353ab7ee3e7266ef0025f61cebbb3&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig.2

Fig.3
![Near to real life 3D model of the thorax. The positioning of electrodes for the measurements undertaken are also shown. Reproduced from [17] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_003.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T215851Z&X-Amz-Expires=3600&X-Amz-Signature=8ddd3d27993b94b7007e99a73e89cedc721a453fb11f892367255241594c5383&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig.4
![Some details of a cross section of the modelled thorax. Reproduced from [17] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_004.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T215851Z&X-Amz-Expires=3600&X-Amz-Signature=d4a3249255ef069739acec7d166d723d6694df26e26904a59b575e736b6b2eaf&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig.5

Fig.6

Fig.7

Fig.8

Fig.9
![6-electrode TPIM covering both lung regions. Reproduced from [9] with permission.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721eaa215d2f6c89dbc97b/j_joeb-2022-0016_fig_009.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250917%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250917T215851Z&X-Amz-Expires=3600&X-Amz-Signature=595e19406f79f888e1f18c44711834ad06a81680d614cb9dcc5fc263a337895d&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Simulation results for potential electrode separation of 7 cm
6-TPIM | 4-TPIM, Front | 4-TPIM, back | |||||||
---|---|---|---|---|---|---|---|---|---|
Inspired | Expired | % change* | Inspired | Expired | % change* | Inspired | Expired | % change* | |
Total Imp, Ω | 10.6 | 7.8 | 34.7 | 10.4 | 8.0 | 30.4 | 10.6 | 7.7 | 37.9 |
% CLTI(1) | 24.2 | 35.6 | -32.1 | 20.5 | 31.2 | -34.1 | 28.3 | 40.4 | -29.8 |
% CLTI(2) | 25.0 | 38.0 | -34.1 | 22.2 | 35.3 | -37.1 | 28.2 | 41.2 | -31.5 |
% OCLTI | 3.6 | 6.8 | 8.2 | 13.4 | -0.3 | 2.1 |
Simulation results for potential electrode separation of 10 cm
6-TPIM | 4-TPIM, Front | 4-TPIM, back | |||||||
---|---|---|---|---|---|---|---|---|---|
Inspired | Expired | % change* | Inspired | Expired | % change* | Inspired | Expired | % change* | |
Total Imp, Ω | 13.9 | 10.6 | 31.4 | 13.6 | 10.3 | 32.2 | 14.0 | 9.9 | 42.1 |
% CLTI(1) | 26.3 | 36.4 | -27.6 | 21.9 | 33.4 | -34.4 | 29.8 | 42.8 | -30.5 |
% CLTI(2) | 27.2 | 38.0 | -28.6 | 23.5 | 36.9 | -36.3 | 29.8 | 43.5 | -31.4 |
% OCLTI | 3.2 | 4.6 | 7.4 | 10.6 | 0.2 | 1.6 |
Ratio of values for 10 cm Electrode Separation to those for 7 cm ES
6-TPIM | 4-TPIM Front | 4-TPIM back | |
---|---|---|---|
% CLTI(1) Inspiration | 1.09 | 1.07 | 1.05 |
% CLTI(1) Expiration | 1.02 | 1.07 | 1.06 |
%OCLTI Inspiration | 0.91 | 0.90 | |
%OCLTI Expiration | 0.68 | 0.79 | 0.75 |