Electrical Impedance tomography – recent applications and developments
, , , , and
Oct 06, 2021
About this article
Article Category: Review article
Published Online: Oct 06, 2021
Page range: 50 - 62
Received: Mar 08, 2021
DOI: https://doi.org/10.2478/joeb-2021-0007
Keywords
© 2021 Sofiene Mansouri, Yousef Alharbi, Fatma Haddad, Souhir Chabcoub, Anwar Alshrouf, and Amr A. Abd-Elghany, published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Fig.1

Fig. 2

Fig. 3
![Chest image reconstruction by EIT [21].](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e8e215d2f6c89dbc94e/j_joeb-2021-0007_fig_003.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250911%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250911T055323Z&X-Amz-Expires=3600&X-Amz-Signature=a79276929b061df2243b06b27d1fc2e9f391c02f6bef63b3328a443173ec3c07&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig. 4

Fig. 5
![Comparison of the phantom reconstruction obtained with the linear inverse solver (left), and the proposed method using ANN and PSO (right). Both targets are indicated by red circles [54].](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e8e215d2f6c89dbc94e/j_joeb-2021-0007_fig_005.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250911%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250911T055323Z&X-Amz-Expires=3600&X-Amz-Signature=340fd84d3ff716c50ca3e957b839bde992081b363f1b5d1f8385a7f2535fa044&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig. 6
![Series of dynamic images showing air filling during inspiration by the PulmoVista500 system [81].](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e8e215d2f6c89dbc94e/j_joeb-2021-0007_fig_006.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250911%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250911T055323Z&X-Amz-Expires=3600&X-Amz-Signature=de72b313f96e625e155ddbabc7535bfebb2a564eda46ec0645590b18204cc92a&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Fig. 7
![Cardiac monitoring by CardioInspect system [88].](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e8e215d2f6c89dbc94e/j_joeb-2021-0007_fig_007.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250911%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250911T055323Z&X-Amz-Expires=3600&X-Amz-Signature=abbe80a7e64c952452ab98915ad8c2ff54563681ae7d7d8b6eaab08ac30326ec&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Comparison between different imaging modalities_
Medical imaging modality | CT | US | MRI | EIT |
---|---|---|---|---|
X-rays | High frequency sound | Radio waves | Impedance | |
Ionizing radiation | Non-Ionizing radiation | Non-Ionizing radiation | Non-Ionizing radiation | |
High | Low | High | Low | |
50-200 μm | 50-500 μm | 25-100 μm | Low | |
<20 min | < 30 min | <40 min | <10 min | |
Moderate | Low | Very High | Low | |
Non portable | Portable | Non portable | Portable | |
Bone and tumor imaging, anatomic imaging | Visualize muscles, tendon and internal organs | Morphological and functional imaging | Rapid tomographic imaging, low cost, noninvasive | |
High cost Ionizing radiation | Operator dependency | Noisy, cost, low sensitivity | Not mature yet |
Image reconstruction algorithms_
Reconstruction algorithm | Description |
---|---|
Linear approach | - Solves the forward problem |
Simple stage reconstruction | - Solves the forward problem |
Sheffield Back-projection | - Solves the forward problem |
Newton-Raphson | - Solves the forward problem based on EMFm |
Optimization by particle swarms | - Solves the inverse problem |
Electrical conductivity for Human tissues_
Tissue | Conductivity ( |
---|---|
Cerebrospinal fluid | 1450 - 1800 |
Blood | 500 - 650 |
Scalp | 300 - 400 |
Brain | 300 - 420 |
Muscle | 200 - 400 |
Fat | 50 |
Bone | 6 |