The United Nations Office on Drugs and Crime (UNODC) World Drug Report for 2017 estimated that around 250 million people used an illicit drug at least once in 2015 (1). Based on the European Monitoring Centre for Drugs and Drug Addiction (EMCDDA) report for 2014 and 2017, illicit drug use in Serbia is at similar levels as in other countries. The report indicated drug consumption as a major social and serious public health problem, especially in the younger population (2, 3). This is in accordance with the results of a school survey which found that 15.1 % of Serbian secondary school students (15–16-yearolds) had tried psychoactive controlled substances at least once (4).

In response to widespread illicit drug consumption, on-site drug testing has become widely utilised because of its convenience and cost-efficiency. Urine drug screening (UDSTs) with immunochromatographic test strips is a rapid, non-invasive, inexpensive, and easy-to-handle on-site method that has found wide application in many fields, such as anti-doping control in sports, drug abuse rehabilitation monitoring, parental control, and forensic and clinical toxicology screening (5, 6, 7, 8, 9, 10, 11). UDSTs are particularly useful at the workplace for pre-employment screening or random on-the-job drug abuse testing (11, 12). UDSTs can be used by experienced laboratory personnel but are also available in commercial kits for general public and are usually sold in pharmacies with manufacturer instructions.

UDST devices are based on competition between a drug and a drug-dye conjugate for binding an anti-drug antibody immobilised on nitrocellulose membrane. If the urine sample is drug-free, the drug-dye conjugate and the immobilised anti-drug antibody create a coloured line indicating a negative test. However, if urine contains a drug above the cut-off level, the drug inhibits the drug-dye conjugate to bind to the antibody, and the result is positive (no line appears on the strip).

Their many advantages notwithstanding, UDSTs suffer of some shortcomings as well, including low accuracy and specificity. They can produce false positive results due to cross-reactivity with other non-targeted drugs of similar chemical structure or certain food ingredients. They can also produce false negative results if samples have been tampered with (11, 12, 13, 14).

According to literature, drug users tamper with samples to obtain negative results in three ways. The first is by consuming great amounts of fluid and/or taking diuretics to dilute urine to the point where a drug is no longer detectable (in vivo adulteration). The second is by replacing their own urine with a drug-free sample (synthetic or obtained from another person). However, when testing comes unannounced, drug consumers resort to simple addition of urine adulterating chemicals, in hope of obtaining a false negative drug test result (in vitro and in situ adulteration). There are many chemical substances that can produce this effect, including acids, alkalis, surfactants, and oxidising agents, and some of them have been designed to beat drug tests available on the market (5, 7, 11, 15, 16, 17, 18, 19).

For drug testing at the workplace there are two recommended ways to prevent adulteration or substitution of urine samples. If urine is collected in privacy [without (in)direct supervision by another person], this should take place in a toilet with water dyed in blue or turned off to prevent dilution (18). Alternatively, institutions and employers without this option resort to direct supervision (19). Samples can also be checked for temperature and integrity immediately after collection. In some cases, however, compliance with these recommendations is low, and urine collection goes unmonitored, which may lead to cheating at urine drug tests.

Because of the rise of narcotic consumption and drug abuse in early adolescence in Serbia, more and more parents are screen-testing their children with UDSTs at home (2, 3). Current commercially available immunochromatographic UDSTs can detect metabolites of the most common recreational drugs available on the Serbian black market such as cannabis, heroin, cocaine, amphetamine, or 3,4-methylenedioxymethamphetamine (MDMA) (20). At home, however, urine samples can easily be diluted or adulterated, so if there is any doubt, laboratory testing is the best choice (21, 22).

Households keep many readily-available chemicals that can adulterate urine samples, but little is known about how exactly they interfere with UDST results. Therefore, the aim of this study was to investigate the potential of the most popular urine adulterants (based on an Internet search) to turn UDST test results from positive to false negative for the most commonly abused psychoactive substances in the country/region. We also wanted to see whether pH and urine biochemical parameter test strips could help to check whether a sample has been changed by the tested adulterants.

Materials and methods
Internet search for household adulterants

We ran an inclusive Internet literature search with Google, Google Scholar, and PubMed search engines using the following key words/strings “urine adulteration”, “tampering methods” and “false negative immunoassay” in English and Serbian. Information about urine tampering was also collected from various on-line drug forums. Internet search revealed a range of potential adulterants kept in the household, and we selected the most common acids, alkalis, surfactants, oxidising agents, additives, as well as one drug and one supplement to test their potential for changing positive strip test results into false negative (Table 1).

Adulterants considered for testing and the quantities used for those that passed urine appearance test

Household adulterant Form Quantity used
Acids
        WC Sanitar (9 % HCl) 0.3 mL
        Vinegar (white distilled vinegar; 9 % acetic acid) liquid 0.2 mL
        Lemon juice 0.4 mL
        Citric acid solid 0.1 g

Alkalis
        Cevtok (30 % NaOH) liquid 0.1 mL

Oxidizing agents liquid
        Varikina (bleach; 5 % NaOCl) 0.3 mL

Surfactants
        Asepsol (5 % quaternary ammonium compounds, such as benzalkonium chloride) 0.2 mL
        Visine® eye drops (active substance: tetrahydrozoline-hydrochloride 0.05 % w/v; non-medical ingredients: benzalkonium-chloride, sodium chloride, boric acid, sodium borate, disodium EDTA) liquid 0.4 mL
        Liquid soap*
        Household cleaning products Frosh, Aro and Skala*

Miscellaneous substances
        Primosept GA (8 % glutaraldehyde) liquid 0.4 mL
        Zn salts (tablets, supplements)* solid
        Nitrite and nitrate salts (meat product additives)* solid

* household chemicals that changed urine appearance and were not tested further for adulteration

Urine sampling

Urine samples used in this study made part of those submitted to our toxicological laboratory for drug screening and were confirmed positive for one of the following substances: amphetamine, MDMA, tetrahydrocannabinol, heroin, cocaine, diazepam, or alprazolam (Table 2). The last two were selected as the most commonly abused benzodiazepines. Drug-free urine samples were prepared from pooled urine collected from three healthy donors. They were used as control in urine sample integrity tests.

Urine drug (metabolite) concentrations and UDST results before adulteration with household chemicals

Drug (metabolite) Urine concentration (mg/L) UDST result
MDMA 0.83 +

Amphetamine 2.28 +

Cocaine (benzoylecgonine) 5.69 +

Diazepam 0.077 +

Alprazolam 0.065 +

Tetrahydrocannabinol (carboxy-tetrahydrocannabinol) 112.2 +

Heroine (6-acetylmorphine) 0.23 +

+” – test positive result

For analysis we used the Waters Alliance® liquid chromatograph with a Waters Micromass® ZQ™ mass spectrometer (Waters Corporation, Milford, MA, USA). All analytical methods have been validated for everyday use at laboratory accreditation. Diazepam and alprazolam were determined without previous hydrolysis and their concentrations are low, but the strip test was positive due higher concentrations of their metabolites (23).

Immunochromatographic UDSTs

We purchased the test strips that were available on the Serbian market at the time of the study from a wholesaler who supplies laboratories and pharmacies. For the detection of amphetamine, MDMA, cocaine, and benzodiazepines we used One Step Rapid® (Ameritek®, Everett, WA, USA), while for tetrahydrocannabinol, morphine, heroin, and codeine we used AbuGnost® (Biognost®, Zagreb, Croatia). All UDST kits were used according to manufacturer instructions. Two drops of urine were pipetted into the sample well of the test strip. The results were interpreted after 3–5 min. These tests are qualitative, so a positive result means that the drug which is being screened for is most likely present in urine at a concentration above the cut-off level. Cut-off levels for MDMA, amphetamine, cocaine, benzodiazepine, tetrahydrocannabinol, and opiates (morphine, heroine, codeine) are 500 ng/mL, 1000 ng/mL, 300 ng/mL, 300 ng/mL, 50 ng/mL, and 300 ng/mL, respectively. Manufacturers also warn about limited reliability of UDSTs if urine is adulterated. The test is invalid if there is no line in the control region, which indicates malfunctioning test strips and/or urine adulteration.

Adulteration and testing

Adulterants were added to the drug-positive urine samples (1 mL) in different amounts simulating real-life scenarios in which people do not have the equipment at hand to precisely measure the volume, yet try not to change urine’s natural appearance to avoid being caught at cheating. We started with the lowest volume of 0.1 mL, as we expected it would be practically impossible to add lower quantities than this in real-life conditions. The highest volume of adulterant added to urine was 40 % vol/vol, because larger amounts of liquids changed urine colour (it became too light not to notice) and urine would foam or smell of chemicals. For solid adulterants, the upper limit was determined by their solubility in urine. We tested only the samples that appeared unchanged. If the UDST result was unclear, that is, if the line that appeared was barely visible, the amount of adulterant was increased until a distinct result was obtained. The final quantities of adulterants that did not change the appearance of urine are given in Table 1. All tests were performed in duplicate.

Other adulterants were excluded from further study because they changed the appearance of urine even at the lowest quantities. These were liquid soap and household cleaning products Frosh, Aro, and Skala, and Zn salts, which come only in an unpractical tablet form, and changed urine appearance even when pulverised. Nitrite and nitrate salts were not tested because we could not find any product based on either in retail stores.

Test strips for urine sample integrity check and adulterant detection

Urine samples with and without adulterants and drug-free control urine were tested for indicators of adulteration with the URIT 11G urine reagent strips (Medical Electronic Co. Ltd., Guangxi, China) following manufacturer instructions. Considering that the URIT 11G strips have a narrower pH range (pH 5–9), we also used pH test strips, whose range is pH 0–14 (Macherey-Nagel GmbH & Co. KG, Düren, Germany).

Results

Urine drug concentrations and UDST results obtained before adulteration are presented in Table 2. Table 3 shows how the tested adulterants affected UDST results. Vinegar (9 % acetic acid) turned nearly all UDST results false negative, with the exception of cocaine. Lemon juice followed suit in tampering, save for cocaine and opiates. Other adulterants were less efficient. Many of the adulterants invalidated the UDSTs, led by Cevtok (30 % NaOH), which invalidated all of them.

UDST results obtained after adulteration of drug-positive urine samples with household chemicals

Adulterating agent Urine positive for
Opiates Cannabinoids Amphetamine MDMA Cocaine Alprazolam Diazepam
WC Sanitar Invalid - Invalid Invalid + Invalid Invalid

Vinegar - - - - + - -

Lemon juice + - - - + - -

Citric acid + - Invalid + + Invalid Invalid

Cevtok Invalid Invalid Invalid Invalid Invalid Invalid Invalid

Varikina - - - + + + +

Asepsol + - Invalid Invalid Invalid Invalid Invalid

Primosept GA + + Invalid Invalid Invalid Invalid Invalid

Visine + + + + + + +

+” – test remained positive; “-” – test became negative; “Invalid” – omitting of control line or both lines on test strip

Control drug-free and drug-positive urine samples tested with URIT 11G urine reagent strips and pH strips before adulteration showed similar findings suggesting maintained sample integrity, as expected (Table 4). After adulteration, however, URIT 11G urine reagent strips and pH strips revealed manipulation with some urine samples. WC Sanitar (9 % HCl) raised ascorbic acid levels compared to control and rendered the sample invalid due to compromised blood in urine values, plus it did not pass the pH strip test (see bolded values in Table 4). Vinegar (9 % acetic acid), lemon juice, and citric acid did change parameter levels of the URIT 11G test but not enough to reveal tampering. The pH test strip, however, did reveal tampering with lemon juice and citric acid. Cevtok (30 % NaOH) rendered URIT 11G test strips invalid on nearly all parameters and failed the pH strip test. Varikina (bleach; 5 % NaOCl), in contrast, passed undetected, even though it did test positive for nitrite presence and raised urine blood cells count. Asepsol invalidated the sample over protein values. Primosept GA (8 % glutaraldehyde) also rendered the sample invalid over high ketone values. Visine® eye drops, in turn, passed undetected.

Discussion

To the best of our knowledge, until this study of ours, no one has reported research data about household chemicals interfering with urine drug test results. Many of the household chemicals used as adulterants in this study are available in other countries in the region and beyond, and our findings could be useful anywhere.

Urine collection for laboratory drug testing is still not completely regulated in Serbia, and no one knows how many urine samples have been collected without supervision. Even though there are no test strips to test urine for specific adulterants, URIT 11G and pH test did detect adulteration by five household chemicals. Both have their limitations, though. The pH test strips are limited to changes in pH, whereas the URIT 11G test strips have a narrow pH range.

In terms of tampering, our study shows that the UDST for cannabinoids is susceptible to urine adulterants, as it yielded six false negative results. This is in line with earlier reports of immunoassay tests for cannabinoid detection (radioimmunoassay – RIA, enzyme immunoassay – EIA, and fluorescence polarisation immunoassay – FPIA) being the most susceptible to adulteration (24, 25). The UDST for amphetamine was also affected by adulterants (five invalid tests and three false negatives). In contrast, the UDST for cocaine was resistant to most adulterants (no false negative results), which has also been reported by Schwarzhoff and Cody (24). However, Asepsol, Primosept GA, and Cevtok did invalidate the test.

The most effective adulterants were acids and alkalis. They probably act through changes in pH and enable drug-antibody binding (26). Vinegar was the most successful in producing false negative results, save for the cocaine test, and did not affect the URIT 11G test results enough to be detected. As it were, its finding of elevated ascorbic acid could be interpreted as a result of high vitamin C supplement intake. Its masking properties have already been reported for immunoassays such as FPIA and EIA for some drugs (11, 24, 25, 26, 27). In contrast, the least efficient adulterant in our study were the Visine® eye drops, as they did not change any of the original UDST findings, not even for cannabinoids, even though other reports say that it strongly interfered with immunoassays (FPIA, EIA, RIA) for the presence of cannabinoids (11, 25, 26, 27). To sum up, most of the investigated adulterants have the ability to change positive UDST strip results into false negative or render them invalid. In that respect vinegar and Asepsol stand out, as they can remain undetected even by experienced laboratory personnel.

Effects of urine adulteration on parameters investiated by URIT 11G urine reagent strips and pH test strips

URIT 11G urine reagent strips pH strips test

Ascorbic acid (mmol/L) Blood (cells/μL) Specific gravity Glucose (mmol/L) Protein g/L Bilirubin (μmol/L) Urobilinogen Nitrite (μmol/L) Ketone (mmol/L) Leukocytes (cells/μL) pH pH
Test strip sensitivity 0.6-0.85 5-15 / 2.2-2.8 0.1-0.3 8.6-17 17-33 18-33 0.5-1 15-0 / /

Test strip range 0-5.6 0-200 1.005-1.030 0-55 0-3 0-100 Normal-131 +/- 0-8 0-500 5-9 0-14

Control urine* 0 0 1.020 0-2.8 0 0-8.6 Normal - - - 6 5

Control urine +

added adulterant**

WC Sanitar 2.8 Invalid 1.030 0 0 0 Normal - - - 5 1

Vinegar 2.8 0 1.030 0-2.8 0 0 Normal - - - 5 4

Lemon juice 2.8 25 1.030 5.5 0.15 8.6 Normal - - - 5 3

Citric acid 2.8 10 1.030 0-2.8 0 0 Normal - - - 5 1

Cevtok Invalid 0 Invalid 0 Invalid 8.6 Invalid - - 70 >9 14

Varikina 0 80 1.015 0-2.8 0 0 Normal + - - 6 5

Asepsol 0 0 1.010 0 Invalid 0 Normal - - - 6 5

Primo sept GA 0 0 1.030 0 0.15 0 Normal - >8.0 - 5 5

Visine 0 0 1.025 0 0 0 Normal - - - 5 5

"-" no presence detected; Invalid - malfunction of test; Normal - within the range of normal physiological parameters; bold results indicate invalid test, pathophysiological conditions or urine manipulation; *similar results were obtained in all tested drag positive urine samples before adulteration; "similar results were obtained in all tested drag positive urine samples after adulteration

In order to detect urine adulteration where test strips for urine adulterants are not available, urine samples should be checked not only with biochemical test strips such as URIT 11G and pH test strips but also visually and for warmth (if checking immediately after taking the sample), to see if they have been tampered with. Our study calls for more efficient screening methods for detecting sample manipulation. But maybe even more important is to develop drug screening methods, immunochromatographic or other, that will be more or even completely resistant to adulterants. In the meanwhile, urine should be collected by competent institutions in so-called “dry toilets” to ensure the integrity of the specimen and still protect the privacy of the person tested.