Design, Realization and Experiments with a new RF Head Probe Coil for Human Vocal Tract Imaging in an NMR device
, , oraz
28 maj 2012
O artykule
Data publikacji: 28 maj 2012
Zakres stron: 98 - 103
DOI: https://doi.org/10.2478/v10048-012-0017-6
Słowa kluczowe
This content is open access.
Magnetic resonance imaging (MRI) is nowadays widely used in medicine for diagnostic imaging and in research studies. The modeling of the human vocal tract acoustics has recently attracted considerable interest. This paper describes the design, realization and first MR scan experiments with a new head probe coil for vocal tract imaging in the open-air MRI equipment working in a weak magnetic field up to 0.2 T. The paper also describes an experimental setting for sound recording during the MR imaging.