Head and neck (H&N) cancer is an excellent indication for radiation therapy (RT) which is an organ preserving method maintaining the quality of life of patients. Brachytherapy (BT) with or without external beam RT (EBRT) can play an essential role in the treatment of certain tumor localizations in the H&N area except cases with bone invasion or proximity of large vessels, delivering an ablative dose to the target volume while sparing the critical and normal tissues - due to the rapid dose fall-off - which is not safely feasible with EBRT alone.1, 2, 3 Image guided high-dose-rate BT (IGBT) with computed tomography (CT) and/or magnetic resonance imaging (MRI) has been implemented improving the efficacy of BT.4 IGBT decreases irradiated doses to critical structures without compromising target coverage.5 However, IGBT is an invasive procedure requiring special skills and interdisciplinary co-operation with a H&N surgeon, as well as patience on the part of patients who have to endure e the insertion of catheters.6, 7 Recently, intensity modulated RT (IMRT) has been introduced in clinical practice and achieved higher dose conformity with better organs at risk (OARs) sparing compared to 3-dimensional conformal radiation therapy.8 Volumetric modulated arc therapy (VMAT), an improved technique of IMRT offers reduced irradiation time compared to IMRT.9 Nowadays IGBT meets the challenge of high precision EBRT such as VMAT, which is a non-invasive modality and does not require special skills to deliver a high conformal dose to the target volume while saving critical normal structures. To our knowledge no detailed dosimetric comparisons of IGBT with VMAT have so far been reported in the H&N region, such publications are available only for breast and cervical cancer.10, 11
Head and neck (H&N) cancer is an excellent indication for radiation therapy (RT) which is an organ preserving method maintaining the quality of life of patients. Brachytherapy (BT) with or without external beam RT (EBRT) can play an essential role in the treatment of certain tumor localizations in the H&N area except cases with bone invasion or proximity of large vessels, delivering an ablative dose to the target volume while sparing the critical and normal tissues - due to the rapid dose fall-off - which is not safely feasible with EBRT alone.1, 2, 3 Image guided high-dose-rate BT (IGBT) with computed tomography (CT) and/or magnetic resonance imaging (MRI) has been implemented improving the efficacy of BT.4 IGBT decreases irradiated doses to critical structures without compromising target coverage.5 However, IGBT is an invasive procedure requiring special skills and interdisciplinary co-operation with a H&N surgeon, as well as patience on the part of patients who have to endure e the insertion of catheters.6, 7 Recently, intensity modulated RT (IMRT) has been introduced in clinical practice and achieved higher dose conformity with better organs at risk (OARs) sparing compared to 3-dimensional conformal radiation therapy.8 Volumetric modulated arc therapy (VMAT), an improved technique of IMRT offers reduced irradiation time compared to IMRT.9 Nowadays IGBT meets the challenge of high precision EBRT such as VMAT, which is a non-invasive modality and does not require special skills to deliver a high conformal dose to the target volume while saving critical normal structures. To our knowledge no detailed dosimetric comparisons of IGBT with VMAT have so far been reported in the H&N region, such publications are available only for breast and cervical cancer.10, 11
The purpose of this study is to present a dosimetric analysis regarding planning target volume (PTV) and OARs with the comparison of IGBT and VMAT for localized H&N cancer using identical CT data and contours.
Thirty-eight consecutive patients with T1-4 mobile tongue (n = 17), floor of mouth (n = 9) and base of tongue (n = 12) cancer treated between January 2013 and March 2017 at the National Institute of Oncology, Budapest, Hungary were selected for this study (Table 1). Primary lesions or tumor bed were treated with CT image based IGBT alone (n = 22) or after EBRT as a boost (n = 16). All T3-4 cases (n = 12) were base of tongue cancer treated exclusively with radiotherapy (EBRT + BT boost). The T4 tumors invaded the deep muscles of the tongue, without invasion into the mandible or other regions. Tumor excision was carried out in 18 patients. Ipsi- and contralateral submandibular glands (15 and 1) (iSMGs, cSMGs) were removed during surgery.
Dosimetry of PTV
IGBT | VMAT | ||
---|---|---|---|
Parameters | n=38 | n=38 | |
mean S.D | mean S.D | ||
V95 (%) | 92.1 3.0 | 98.4 0.9 | <0.05 |
V98 (%) | 90.2 3.2 | 90.4 3.7 | >0.05 |
V100 (%) | 89.0 3.4 | 76.7 8.9 | <0.05 |
D90 (%) | 98.6 4.7 | 98.2 0.8 | <0.05 |
D100 (%) | 58.6 9.0 | 87.0 3.2 | <0.05 |
Abbreviations: S.D., standard deviation; PTV, planning target volume; IGBT, image guided high-dose-rate brachytherapy; VMAT, volumetric modulated arc therapy; n, number of patients for analysis; VX, relative volume of the PTV receiving at least X% of the PD; DX, minimum relative dose of the PD delivered to X% of the PTV.
The process of BT planning was described in details in our previous publication.12 Under general anesthesia in the operating theatre plastic catheters (median 7, range 3–12) were implanted into the target volume located in 17 patients on the left, in 18 on the right side, and in 3 in the central region of the tongue/ floor of mouth/base of tongue. After implantation, all patients underwent CT imaging with 3 mm slice thickness including the primary tumor or tumor bed, the spinal cord, parotid glands (PAGs) on both sides and SMGs. The images were transferred to the Oncentra Brachy v4.3 (Elekta, Brachytherapy, Veenendaal, The Netherlands) treatment planning system. Gross tumor volume (GTV), clinical target volume (CTV) and PTV were contoured on CT images as follows. In non-surgical cases (definitive) palpation, visual inspection and MR images without CT-MRI fusion were used to determine GTV in the patients. CTV was defined as GTV + 5mm limited to mandible. In postoperative cases CTV was directly defined similarly to definitive cases using the preoperative MRI. For all cases PTV was the same as CTV without using any margin. Based on CT image sets, the mandible, the spinal cord, PAGs on both sides and SMGs - as OARs - were delineated. Mandible was not part of the PTV. After catheter reconstruction, treatment plans were made with geometrical optimization, complemented with graphical optimization by adjusting the isodose line in order to appropriately cover the PTV by the prescribed dose (PD) and keep the doses to OARs as low as possible.13 Our aim was to obtain less than 0.40 for the dose non-uniformity ratio (DNR), the ratio of volumes receiving 1.5 times the PD and those receiving the PD, (V150/V100). At our institute the fractionation schedule was 15 x 3 Gy (45 Gy) for IGBT alone and 7 x 3 Gy (21 Gy) for IGBT after 50 Gy EBRT.
For VMAT plans the same CT images and structure set were used as for IGBT. The contours applied for IGBT plans were transferred from the Oncentra Brachy to the Eclipse v11 (Varian Medical Systems, Palo Alto, CA) treatment planning system using DICOM RT protocol. Due to this process, all contours of PTV and OARs were identical in both planning systems. In case of IGBT alone patients VMAT was set up for a total dose of 70 Gy (35 x 2 Gy), and for EBRT+IGBT patients a total dose of 20 Gy (10 x 2 Gy) was planned as a boost. VMAT plans were created with 6 MV photon energy beams using double partial arcs from gantry angle 110° to 250° to ensure the homogeneity of PTV dose coverage. As regards the dose constraints for OARs and target volume coverage, VMAT plans were optimized using the Varian RapidArc progressive resolution optimization algorithm (PRO) and AAA dose calculation algorithm. The objective for PTV was to deliver a dose of 95–107% of the PD. The dose constraints for PTV70Gy were as follows: V98% > 90%, V95% > 95%, V50% = 100%, V2% < 107%, for the spinal cord V2% < 45 Gy, for parotid glands Dmean < 24 Gy, for mandible V0.03cc < 70 Gy.
Based on dose-volume-histograms (DVHs), treatment plans of both techniques were dosimetrically analyzed. For evaluating conformity of dose distribution to PTV, the relative volume of the PTV receiving at least 95, 98 and 100% of the PD (V95, V98 and V100) and the minimum relative dose of the PD delivered to 90 and 100% of the PTV (D90 and D100) were calculated. The minimum relative doses of the PD delivered to the most exposed 0.1, 1 and 2 cm3 of the organs ( D0.1cm3, D1cm3 and D2 cm3) were quantified for the assessment of doses to OARs. The salivary glands were classified as ipsilateral or contralateral based on the tumor location, and were dosimetrically evaluated. For the salivary glands, additional parameters such as the mean relative dose of the PD delivered to the organs (Dmean), D10, D30, D50, V10, V30 and V50 were also quantified. The dosimetric data for the parotid and submandibular glands were not evaluated for all patients, because in some cases the glands were excluded from the field-of-view of the CT scans, the volume was too small (for D1 and D2cm3 calculation), or the target volume (tumor or tumor bed) was centrally located (Tables 4 and 5). The spinal cord was not detected in one case because it was excluded from the field-of-view of the CT scans (Table 3).
Dosimetry of mandible and spinal cord
IGBT | VMAT | ||||
---|---|---|---|---|---|
Parameters | |||||
mean | S. D | mean | S.D | ||
Mandible, n=38 | |||||
D0.1cm3 (%) | 77.0 | 17.2 | 85.4 | 7.9 | <0.05 |
D1cm3 (%) | 56.9 | 13.4 | 74.5 | 9.6 | <0.05 |
D2cm3 (%) | 48.4 | 12.2 | 68.4 | 9.5 | <0.05 |
Spinal cord, n=37 in one patient the spinal cord was not detected because it was excluded from the field-of-view of the CT scans. | |||||
D0.1cm3 (%) | 9.7 | 2.6 | 12.3 | 4.1 | <0.05 |
D1cm3 (%) | 6.8 | 2.0 | 10.8 | 3.7 | <0.05 |
D2cm3 (%) | 5.9 | 1.9 | 10.0 | 3.6 | <0.05 |
Abbreviations: S.D., standard deviation; IGBT, image guided high-dose-rate brachytherapy; VMAT, volumetric modulated arc therapy; n, number of patients for analysis; DXcm3, minimum relative dose of the PD delivered to most exposed X cm3 of the organs.
Dosimetry of parotid gland
IGBT | VMAT | ||||
---|---|---|---|---|---|
Parameters | |||||
mean | S.D | mean | S. D | ||
ipsilateral side (n=34 in one patient the parotid gland was not detected because it was excluded from the field-of-view of the CT scans, and in three patients parotid glands were not evaluated because of central implantation. | |||||
Dmean (%) | 4.6 | 1.4 | 4.6 | 2.6 | >0.05 |
D0.1cm3 (%) | 11.2 | 2.8 | 18.0 | 7.1 | <0.05 |
D1cm3 (%) | 8.1 | 2.1 | 12.9 | 6.3 | <0.05 |
D2cm3 (%) | 7.0 | 2.0 | 10.5 | 5.7 | <0.05 |
D10 (%) | 7.3 | 2.1 | 11.2 | 5.6 | <0.05 |
D30 (%) | 5.5 | 1.6 | 5.5 | 4.0 | >0.05 |
D50 (%) | 4.5 | 1.3 | 3.0 | 3.1 | <0.05 |
V10 (%) | 3.1 | 4.0 | 18.0 | 16.9 | <0.05 |
V30 (%) | 0.0 | 0.0 | 0.7 | 3.9 | N.A |
V50 (%) | 0.0 | 0.0 | 0.1 | 0.5 | N.A |
contralateral side (n=35 three patients were not evaluated because of central implantation. | |||||
Dmean (%) | 3.0 | 0.9 | 3.9 | 2.1 | >0.05 |
D0.1cm3 (%) | 8.9 | 2.1 | 15.3 | 5.9 | <0.05 |
D1cm3 (%) | 5.9 | 1.6 | 10.8 | 5.9 | <0.05 |
D2cm3 (%) | 4.9 | 1.5 | 9.1 | 5.4 | <0.05 |
D10 (%) | 5.4 | 1.3 | 9.9 | 4.4 | <0.05 |
D30 (%) | 4.0 | 1.0 | 4.9 | 3.7 | >0.05 |
D50 (%) | 3.2 | 0.8 | 2.2 | 2.3 | <0.05 |
V10 (%) | 0.5 | 0.3 | 14.2 | 13.2 | <0.05 |
V30 (%) | 0.0 | 0.0 | 0.8 | 4.7 | N.A |
V50 (%) | 0.0 | 0.0 | 0.2 | 1.3 | N.A |
Abbreviations: S.D., standard deviation; IGBT, image guided high-dose-rate brachytherapy; VMAT, volumetric modulated arc therapy; n, number of patients for analysis; Dmean, mean relative dose of the PD delivered to the organs; DXcm3, minimum relative dose of the PD delivered to most exposed X cm3 of the organs; DX, minimum relative dose of the PD delivered to X% of the organs; VX, relative volume of the organs receiving at least X% of the PD; N.A., not available.
Dosimetry of submandibular gland
IGBT | VMAT | ||||
---|---|---|---|---|---|
Parameters | |||||
mean | S.D | mean | S. D | ||
ipsilateral side | |||||
Dmean (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 16.4 | 10.7 | 21.9 | 19.9 | >0.05 |
D0.1cm3 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 27.4 | 15.6 | 39.9 | 23.3 | >0.05 |
D1cm3 (%), n=18 two and three patients, respectively, were additionally excluded from the analysis of D1cm3 and D2cm3 because the volume was too small. | 19.4 | 11.2 | 27.2 | 17.6 | >0.05 |
D2cm3 (%), n=17 two and three patients, respectively, were additionally excluded from the analysis of D1cm3 and D2cm3 because the volume was too small. | 16.1 | 9.4 | 20.8 | 14.7 | >0.05 |
D10 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 22.6 | 13.9 | 33.2 | 23.0 | >0.05 |
D30 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 18.3 | 11.7 | 25.8 | 21.9 | >0.05 |
D50 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 15.7 | 10.4 | 21.0 | 20.9 | >0.05 |
V10 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 61.3 | 38.6 | 63.9 | 33.6 | >0.05 |
V30 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 14.5 | 28.0 | 30.1 | 38.7 | >0.05 |
V50 (%), n=20 in fifteen patients the submandibular glands were resected by operation and three patients were not evaluated because of central implantation. | 1.4 | 3.7 | 14.1 | 27.4 | >0.05 |
contralateral side | |||||
Dmean (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 8.2 | 4.8 | 16.9 | 10.8 | <0.05 |
D0.1cm3 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 13.4 | 5.3 | 29.7 | 12.0 | <0.05 |
D1cm3 (%), n=32 two and two patients, respectively, were additionally excluded from the analysis of D1cm3 and D2cm3 because the volume was too small. | 9.4 | 3.2 | 21.7 | 8.6 | <0.05 |
D2cm3 (%), n=32 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 8.1 | 2.9 | 18.3 | 8.3 | <0.05 |
D10 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 11.1 | 5.6 | 25.3 | 12.2 | <0.05 |
D30 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 9.2 | 5.1 | 20.4 | 11.4 | <0.05 |
D50 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 8.0 | 4.7 | 16.7 | 11.4 | <0.05 |
V10 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 26.0 | 30.2 | 69.9 | 27.5 | <0.05 |
V30 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 1.2 | 7.0 | 14.1 | 25.1 | <0.05 |
V50 (%), n=34 in one patient the submandibular gland was resected by operation and three patients were not evaluated because of central implantation. | 0.0 | 0.0 | 2.7 | 12.1 | N.A |
Abbreviations: S.D., standard deviation; IGBT, image guided high-dose-rate brachytherapy; VMAT, volumetric modulated arc therapy; n, number of the patients for analysis; Dmean, mean relative dose of the PD delivered to the organs; DXcm3, minimum relative dose of the PD delivered to most exposed X cm3 of the organs; DX, minimum relative dose of the PD delivered to X% of the organs; VX, relative volume of the organs receiving at least X% of the PD, N.A., not available.
The mean value with standard deviation was used to describe dosimetric parameters for both techniques, and the data were compared using the non-parametric Mann-Whitney U test with GraphPad Prism version 5.01 for Windows (GraphPad Software, San Diego, CA). The level of statistical significance was set at
The dosimetric parameters of PTV are presented in Table 2. Dose distribution of BT and VMAT are presented in Figure 1. The dose conformity was excellent with both techniques. V95 was significantly better with VMAT (98.4%
As for OARs, the mandible and the spinal cord were better protected by IGBT (Table 3). For doses to small volumes (D0.1cm3, D1cm3 and D2cm3) significantly lower values were obtained with IGBT (
Doses to the salivary glands were generally lower with IGBT (Tables 4 and 5). Six parameters of ipsilateral PAGs (iPAGs) were significantly lower with IGBT than with VMAT (
The comparison of new technologies in the H&N region is a very interesting area of research.7 Sresty
Owosho
For tonsillar cancer patients Stieler
Dose to PAGs is important in terms of xerostomia, because they produce up to 70% of the total stimulated saliva.20, 21, 22, 23 Owosho
A major source of concern with VMAT and IMRT is the higher low dose radiation to surrounding normal tissue, which potentially increases the risk of secondary malignancy.17 As for our results, V10 with VMAT was appreciably large especially for PAG. Immobilization for precise irradiation and the possibility of tumor repopulation during the long treatment time can also be a problem with VMAT. On the other hand, technique sensitivity is one of the drawbacks of IGBT. However, some measures have been introduced for easy implantation, such as the use of a vinyl template or ultrasound guided technique.5
One of the limitations of our study is that the Task Group (TG)-43 formalism, implemented in our planning system for dose calculation, has not taken into consideration tissue heterogeneities.28 However, Peppa
Though this article deals with the dosimetric and not the therapeutic comparison of the above mentioned two techniques, we would like to emphasize that the fractionation schedule, which we used in our BT treatment, was based on the European and American recommendations, where the suggested boost dose is 21-30/3 Gy with HDR after 45–50 Gy EBRT and the definitive dose is 45–60/3 Gy.2, 30 Using alfa/beta = 10 Gy and without taking into consideration the time factor, the calculated biological effective dose (BED) for BT alone (15 x 3 Gy) was 58.5 Gy, for EBRT (25 x 2 Gy) and BT boost (7 x 3 Gy) 87.3 Gy (60 Gy + 27.3 Gy), for EBRT alone (35 x 2 Gy) 84 Gy and for EBRT (25 x 2 Gy) and EBRT boost (10 x 2 Gy) 84 Gy (60 Gy + 24 Gy). We applied lower BT dose for BT alone because of the lack of long-term experiences with HDR BT.
This is the first study with direct dosimetric comparison between IGBT and VMAT for H&N cancer applying various parameters. Both techniques provided excellent target coverage, but IGBT was found superior in protecting OARs. Adverse events, such as xerostomia and osteoradionecrosis, derived from irradiation of OARs could be serious problems in the H&N cancer radiotherapy. Therefore, it is clinically important to keep the dose for OARs as low as possible. In this respect the results confirm the important role of interstitial RT in the era of new external beam RT techniques. To translate the results of these dosimetric findings into clinical practice, more patients and long term follow-up with prospective collection of toxicities are necessary.