Validation of transient time method to calibrate oil flow meters in closed conduits using 123I as the radiotracer
Kategoria artykułu: Original Paper
Data publikacji: 08 cze 2021
Zakres stron: 69 - 74
Otrzymano: 20 mar 2020
Przyjęty: 19 lis 2020
DOI: https://doi.org/10.2478/nuka-2021-0009
Słowa kluczowe
© 2021 Eduardo R. Gonçalves et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
To assure the credibility of commodity transfer operations in the oil and gas industry, calibrated flow meters are used to quantify the movement of fluids in the pipelines. The purpose of this paper is to validate the transient time method (TTM) to calibrate oil flow meters installed in restricted areas, using the 123I-labelled oil as a radiotracer. Traditionally, as proposed by the standard ISO 2975-7:1977 [