Introduction

The aim of radiotherapy is to deliver the prescribed dose to the target volume according to the treatment plan. So, effective and safe radiotherapy should be as accurate as possible in every stage of the treatment preparation and delivery process. The calculation of the number of MUs is an important step in the chain of activities leading to the final stage of computerized treatment plan preparation. According to the international recommendations, at least one independent verification of the calculation of the number of MUs is required for each and every patient treated with external radiotherapy. There are several guidelines for MU calculation verification [1]. The simplest and the fastest method of MU verification is by using a hand-held simple PC software [2]. In Poland, the quality assurance (QA) of radiotherapy is regulated by the Regulation of the Polish Minister of Health [3]. The simplest method, a simple hand-held PC software, has usually several disadvantages, such as not taking into account patient geometry, tissue heterogeneities, the field-shaping device (MLC, blocks aperture), etc. More advanced solutions available commercially are independent of the MU verification calculation software, such as Rad-Calc [4], DIAMOND [5] or Mobius3D [6], where patient geometry and beam shape are considered. The dose delivered to a patient can also be verified by different methods of in vivo dosimetry [7]. However, the most advanced method of dose calculation verification is calculating the dose distribution using a second independent treatment planning system (TPS) as proposed by Report of AAPM Task Group 114 [8]. In our center, two TPSs are available, Eclipse (version 10.0, Varian) and Oncentra MasterPlan (version 3.3, Elekta). In our procedure, the 3D conformal treatment plans were prepared with the Eclipse system and verified with the OMP.

The main goal of this study is to present the comparison results of dose distribution performed with Eclipse and OMP. The doses calculated for PTV with these two systems are presented. We have also presented the results of dosimetric verification of dose distribution calculations carried out with both systems in the inhomogeneous phantom. The problems in the interpretation of results were also discussed.

Materials and methods
Patient data

The analysis and assessment of differences in dose calculations obtained with two treatment planning systems (TPSs) were performed for patients treated with 6 MV and 15 MV X-rays, respectively, in our radiotherapy center. The results of calculations for randomly selected patients were analysed in this study. Patients were irradiated at the following treatment sites: head and neck, brain, breast, oesophagus, lung, gynecology, prostate, rectum, soft tissues (sarcomas), and others. Further, a three-dimensional conformal radiotherapy (3D-CRT) plan was prepared for each patient. All treatment plans were delivered on a Varian CL 600 C/D (X6 MV) or CL 2300 C/D (X6 MV/X15 MV) accelerators. The differences were analysed for 6 MV and 15 MV energies separately. There were 152 plans for 6 MV beams and 128 plans for 15 MV beams (see Table 1).

The number of analysed plans according to calculation models

EnergyNumber of treatment plans
X6 MV152
X15 MV128

Computed tomography (CT) was carried out for each patient according to the internal protocols used in our hospital. Subsequently, GTV/CTV/PTV (gross tumour volume/clinical target volume/planning target volume) and organs at risk (OARs) were delineated on 3 mm-thick transversal slices for patients with brain or head and neck cancer and 5 mm-thick transversal slices in case of other localizations. All structures were delineated according to ICRU recommendation [9, 10]. The following cases were analysed:

head and neck: two perpendicular or oblique wedged fields,

brain: non-standard geometry; two to four beams,

breast: two opposed wedged beams with the medial field border aligned to avoid irradiation of heart and ipsilateral lung,

oesophagus: three wedged fields,

lung: three to five fields, some of them with wedges,

gynecology: box technique or three fields (open anteroposterior (AP) and two lateral wedged fields),

prostate, rectum: three wedged fields (open AP and two lateral wedged fields), and

soft tissues: the non-standard geometry of beams. All treatment beams were set up isocentrically.

Dose calculation algorithms and TPSs

For each patient, a 3D-CRT treatment plan was prepared in the Eclipse TPS. The analytical anisotropic algorithm (AAA) [11] with 2.5 mm grid size was used for the calculation of dose distribution. Each treatment plan was prepared according to ICRU Reports no. 50 and 62 and the internal protocols were used for OARs. Further, the dose distribution in primary TPS (Eclipse) was always normalized to the mean dose in the PTV. Subsequently, the DICOM RT data were transferred to OMP, and the dose distribution was calculated. In the OMP, the collapse cone convolution (CCC) algorithm [12] with 2 mm cubic grid was used for the second dose distribution calculation. Also, the same number of MUs was utilized for each beam.

Commissioning of Eclipse and OMP

Both TPSs were carefully checked prior to the first clinical use. The input data used for modelling of Varian accelerators beams, the percent depth doses, profiles and output factors were calculated using water phantom for both TPSs and compared with the input data. The gamma method with gamma 2 mm, 2% was used to compare percent depth doses and profiles. The gamma values larger than 1 were obtained in the build-up region and for the smallest beam sizes, i.e. 2 cm × 2 cm and 3 cm × 3 cm only. The actual CT conversion HU-electron density curve was entered into both systems.

Data analysis

Data were analysed for 6 MV plans and for 15 MV plans separately. For each plan, the dose disparity between Eclipse and OMP calculations were expressed in terms of the difference between mean doses in the PTV and the doses at the isocenter point. The dose difference was calculated using the formula (1): Δk=(xExOMP-1)100%{\Delta _k} = \left({{{{x^{\rm{E}}}} \over {{x^{{\rm{OMP}}}}}} - 1} \right) \cdot 100\% where: k – plan’s number, xE, xOMP – dose statistics computed with Eclipse and OMP for plan number k, respectively.

Further, the mean of differences (2) and the standard deviation (STD) of these differences (3) were calculated, for each group of plans, using the following formulas: Δ¯=1kn=1kΔk\overline \Delta = {1 \over k}\sum\limits_{n = 1}^k {{\Delta _k}}SΔ¯=1(k-1)kn=1k(Δ¯-Δk)2{S_{\overline \Delta}} = \sqrt {{1 \over {(k - 1) \cdot k}}\sum\limits_{n = 1}^k {{{(\overline \Delta - {\Delta _k})}^2}}}

Note that the Wilcoxon matched-pairs signed rank test was used for statistical analyses of differences. The differences were considered as statistically significant if p <0.001 (two-tailed). All statistical analyses were performed using SPSS software (version 19.0, IBM Corporation).

Dose calculation with and without inhomogeneity correction

Additional calculations without inhomogeneity correction were performed for 24 plans in which the dose distribution was highly influenced by inhomogeneities; further, large dose differences between Eclipse and OMP calculations were obtained. The disparity of the mean dose in the PTV and dose at the isocenter point were computed using formulas (1), (2), and (3).

Dosimetric measurements

The calculations of both TPSs were compared with measurements which were performed using the inhomogeneous phantom [13]. The phantom is presented in Fig. 1. The phantom was the solid polystyrene phantom of 15 cm width, 15 cm length and 15 cm height. The physical density was 1.04 g/cm3, and the relative electron density to water was 1.00. One of the polystyrene slabs allows the positioning of the ionization chamber in the phantom. Phantom also consists of two exchangeable inhomogeneity slabs to mimic the bone (15 cm × 15 cm × 2 cm) and lung (15 cm × 15 cm × 5 cm). The measurements were performed with the calibrated cylindrical ionization chamber TM 31010 (PTW) with a sensitive volume of 0.125 cm3, connected to Unidos (PTW) electrometer. An ionization chamber was placed at a physical depth of 10 cm.

Fig. 1

Polystyrene/lung solid phantom.

Three phantom configurations (polystyrene, polystyrene with lung-equivalent material, and polystyrene with bone-equivalent material) were CT scanned. All acquired images were exported to Eclipse TPS. During CT imaging, the ionization chamber slot was filled with solid water. Also, for each phantom configuration, six various geometries of treatment plans were prepared (Table 2) for both 6 MV and 15 MV photon beams. DICOM RT data for each plan was transferred to OMP TPS where dose distribution was calculated.

Geometries of treatment plans used for measurements

No.Number of fieldsGantry directionsSize of field (cm × cm)WedgeShape of field
Polystyrene

113 × 3Nonesquare
218 × 8Nonesquare
330°, 90°, 270°8 × 8, 8 × 8, 8 × 8Nonesquare
430°, 70°, 290°8 × 8, 8 × 8, 8 × 8Nonecircular
530°, 70°, 290°8 × 8, 8 × 8, 8 × 8none, 45 L, 45 Rcircular
630°, 70°, 290°8 × 8, 8 × 8, 8 × 8none, 45 OUT, 45 INcircular

Polystyrene with bone equivalent material

713 × 3Nonesquare
818 × 8Nonesquare
930°, 90°, 270°8 × 8, 10 × 8, 10 × 8Nonesquare
1030°, 45°, 315°8 × 8, 9.5 × 8, 9.5 × 8Nonecircular
1130°, 45°, 315°8 × 8, 9 × 8, 9 × 8none, 45 L, 45 Rcircular
1230°, 45°, 315°8 × 8, 8 × 9, 8 × 9none, 45 OUT, 45 INcircular

Polystyrene with lung equivalent material

1313 × 3Nonesquare
1418 × 8Nonesquare
1530°, 90°, 270°8 × 8, 9.5 × 8, 9.5 × 8Nonesquare
1630°, 50°, 310°8 × 8, 9 × 8, 9 × 8Nonecircular
1730°, 60°, 300°8 × 8, 9 × 8, 9 × 8none, 45 L, 45 Rcircular
1830°, 50°, 310°8 × 8, 8 × 9, 8 × 9none, 45 OUT, 45 INcircular

After the dose calculations in Eclipse and OMP TPSs, irradiations of the solid phantom in three configurations were performed with Varian CL 2300 C/D. The numbers of MUs were calculated to deliver 2 Gy at the isocenter point. The actual output factor was taken into account.

The measured doses were compared with the doses calculated in Eclipse and OMP for each of the following, namely, the phantom configuration, energy, and geometry of the treatment plan. The comparison was performed using the equation: ΔMEASUREMENTTPS=(xTPSxMEASUREMENT-1)100%\Delta _{{\rm{MEASUREMENT}}}^{{\rm{TPS}}} = \left({{{{x^{{\rm{TPS}}}}} \over {{x^{{\rm{MEASUREMENT}}}}}} - 1} \right) \cdot 100\% where: xTPS – dose computed by treatment planning system, xMEASUREMENT – measured dose.

Results

Tables 3 and 4 present the differences between the mean dose to the PTV and dose to isocenter calculated with Eclipse and OMP for 6 MV and 15 MV beams. The results are given separately in terms of the mean value ± standard deviations of differences for each type of energy. The distributions of these differences are presented in Figs. 2 and 3.

Differences between Eclipse and OMP calculations of mean doses to the PTV

EnergyCasesAverage (%) ± STD (%)Range (%)p-value (p <0.001)
X6 MV1521.4 ± 1.0−3.4, 6.00.0
X15 MV1282.5 ± 0.6−0.2, 4.50.0

Differences between Eclipse and OMP calculations of dose to isocenter

EnergyCasesAverage (%) ± STD (%)Range (%)p-value (p <0.001)
X6 MV1521.3 ± 1.9−2.7, 10.60.0
X15 MV1282.1 ± 1.0−0.7, 5.60.0

Fig. 2

Frequency histograms of differences between Eclipse and OMP calculations of mean doses to the PTV for 6 MV (a) and 15 MV (b).

Fig. 3

Frequency histograms of differences between Eclipse and OMP calculations of doses to isocenter for 6 MV (a) and 15 MV beams (b).

Table 5 presents the differences between mean doses and doses to isocenter calculated in both systems for 24 plans in which inhomogeneities highly influenced on dose distribution. The mean doses and the doses to isocenter were calculated for both with and without inhomogeneity correction. If inhomogeneity correction was applied, the mean values of differences exceeded 1% and 2% for 6 MV and 15 MV beams, respectively. However, if inhomogeneity correction was not applied for 6 MV beams, the mean value differences between the mean doses to the PTV, and doses to isocenter, was smaller than 0.5%. Also, for 15 MV beams, these differences were also smaller with inhomogeneity correction.

Statistical summary of differences between Eclipse and OMP calculation with and without inhomogeneity correction

EnergyPlansAverage (%) ± STD (%)Range (%)p-value (p <0.001)
Mean dose with inhomogeneity correction

X6 MV241.3 ± 0.50.1, 2.10.000
X15 MV242.6 ± 0.61.6, 4.50.000

Dose at isocenter with inhomogeneity correction

X6 MV241.1 ± 1.5−1.0, 6.80.000
X15 MV242.2 ± 1.10.4, 5.20.000

Mean dose without inhomogeneity correction

X6 MV240.4 ± 0.4−0.8, 1.00.000
X15 MV241.0 ± 0.30.5, 1.60.000

Dose in isocenter without inhomogeneity correction

X6 MV240.2 ± 0.7−1.0, 2.40.110
X15 MV240.5 ± 0.9−2.1, 2.80.001

Table 6 presents the comparison of doses measured using the phantom and calculated in both TPSs. For Eclipse system, the differences measured with homogenous polystyrene phantom and with polystyrene phantom under lung and bone inhomogeneity are found to be 2%. The largest differences were obtained for bone. Much smaller differences were obtained for the OMP system. For the OMP, the discrepancies between measurements and calculations were smaller than 1%, except in three cases for X6 MV and three cases for X15 MV.

The discrepancy between Eclipse and OMP calculations and measurements for various beams geometry

No.X6 MVX15 MV

ΔMEASUREMENTECLIPSE(%)\Delta _{{\rm{MEASUREMENT}}}^{{\rm{ECLIPSE}}}(\%)ΔMEASUREMENTOMP(%)\Delta _{{\rm{MEASUREMENT}}}^{{\rm{OMP}}}(\%)ΔMEASUREMENTECLIPSE(%)\Delta _{{\rm{MEASUREMENT}}}^{{\rm{ECLIPSE}}}(\%)ΔMEASUREMENTOMP(%)\Delta _{{\rm{MEASUREMENT}}}^{{\rm{OMP}}}(\%)
Polystyrene phantom

12.20.32.20.8
22.2−0.52.8−0.4
32.00.23.00.4
42.40.42.80.1
52.1−0.12.1−0.3
62.1−0.12.3−0.4

Polystyrene with bone equivalent material

72.51.42.51.4
82.30.53.60.8
92.60.62.6−0.1
102.80.84.01.3
112.80.83.51.0
122.40.63.40.8

Polystyrene with lung equivalent material

132.91.31.50.7
143.3−0.33.1−0.4
151.7−0.22.3−0.2
162.91.42.90.8
171.80.62.20.3
181.60.22.30.2
Discussion

The independent verification of the MU calculation for 3D-CRT plan is recommended or even required by law. In our center, verification is performed by comparing the dose distribution calculations carried out with Eclipse and with OMP. Further, the comparison results are presented for 6 MV and 15 MV beams, separately. Also, the mean doses to the PTV and the doses to isocenter are compared.

On the average, the mean doses to the PTV calculated with Eclipse were larger than those calculated with OMP of 1.4% and 2.5% for 6 MV and 15 MV beams, respectively. Similar results were obtained for the doses to isocenter. A very similar result was obtained for the differences of dose to isocenter (Fig. 3), but the dispersion of differences is larger for the dose to isocenter than for the mean dose i.e., 1.9% vs. 1.0% for 6 MV and 1.0% vs. 0.6% for 15 MV. All the differences were statistically significant. The variations obtained between both systems were smaller than 3.5% for almost all plans in the mean doses for 6 MV. For 15 MV beams, these discrepancies were larger than 3.5% in 30 out of 128 plans. As for the doses to isocenter, for 6 MV and 15 MV beams, those discrepancies which amounted to larger than 3.5% were observed for 25 plans and 11 plans, respectively. Before getting the results published in this work, the 3.5% discrepancy has been treated as the first action level. According to our rules, when the discrepancy amounts to larger than 3.5%, the case should be analysed carefully. Next, since this large discrepancy could not be explained, the procedure used for intensity modulated radiation therapy (IMRT) plans was followed. The verification plan was prepared and measured with Octavius dosimetry system (PTW). The results of measurements did not all exceed the action level.

After getting the results obtained in this work, the first action level was changed to the range of (−1.5%, 5.5%). Our decision is based on the comparison of calculations and measurements in the inhomogeneous phantom. The calculation doses performed with Eclipse were systematically larger than the measurements of about 2–3% for both energies, while for OMP, these differences were much smaller. These results were in a sense consistent with the differences of calculations obtained for patients with both systems. The dose calculated in the OMP was almost always smaller on the average of 1.5% than in Eclipse. Also, we want to stress that, for both systems, comparisons of calculations and measurements carried out in a water phantom for many different sizes of beams and geometries provided an excellent result. The conclusion has been drawn that for Eclipse, the discrepancy between calculations and measurements has been the result of inhomogeneity correction. This conclusion is substantiated by using the results of the comparison obtained for 24 plans for which we performed the comparison again with and without inhomogeneity correction. When the inhomogeneity correction is switched off, the discrepancy between both systems is decreased by about 1% on an average. There might be at least two reasons for these discrepancies which were obtained for inhomogeneity. The first source of wrong calculations might be the erroneous HU-electron density curve used for performing calculations. The HU-electron density curve was measured, and we did not find any differences between the curve entered into the systems and the measured curve. The second reason might be the unique algorithms implemented for inhomogeneity calculations in the TPSs. The user has no influence on the algorithm used in the system.

According to recommendations, calculations should be checked for the number of MUs. The easiest way is to calculate the dose distribution or the number of MUs with another software. It is worth noting that this procedure may be unreliable and may lead to wrong decisions. Both programs may calculate the dose distribution with an error. Both systems may deliver the number of MUs as being larger or smaller than the correct ones. Subsequently, the user may receive the false positive results when comparing the results obtained with both systems. If one system calculates a larger number of MUs while the second system calculates a smaller number, the user may obtain a false negative result. Therefore, prior to using any program, it should always be validated. The validation should be carried out by the comparison of measurements and calculations. It is relatively easy to make measurements in the homogeneous phantom. It is much more difficult to perform measurements in the inhomogeneous phantom. Nevertheless, such an effort should always be made. The results of our measurements have revealed that the calculations carried out in the OMP are slightly more trustworthy. Consequently, we have decided to change the action level region.

Conclusions

The MU calculations have been verified with two different 3D TPSs, but the calculations carried out with these systems always require to be validated. The definition of the action level should account for the results of verification of the accuracy of calculation of each TPS. A verification procedure should also be performed in the inhomogeneous phantom.