Quick and adequate intravenous fluid resuscitation is crucial in the management of critically ill patients, however excessive fluid administration has been shown to contribute to mortality(1,2).
It has been established that clinical examination alone is unreliable; therefore, more objective means of intravascular volume assessment have arisen(3–5). Ultrasound measurement of the inferior vena cava (IVC) diameter and its respiratory variability have been proposed as a simple, non-invasive tool to estimate the fluid volume status and predict fluid responsiveness(6–8). This vital clinical information may determine the choice of critical treatment, and have a decisive impact on patients’ outcome.
In many studies, IVC collapsibility index (IVC-CI) over 40% was acknowledged as the cut-off value to differentiate “fluid responders (FR)” and “non- responders (FNR)” in spontaneously breathing patients(9,10).
IVC diameter assessment has been implemented into various simplified protocols used during evaluation of patients in a critical condition(11), and is thus often performed by clinicians in life-threatening scenarios. There is, however, an inconsistency in reporting, describing and determination of the recommended acquisition technique, including the methodology of performing IVC measurement(12). Although both cardiac and convex transducers are used for IVC diameter assessment in clinical practice, we are not aware of any study comparing the consistency and accuracy of measurements performed with both probes.
The primary objective of the present study was to compare the consistency in IVC diameter measurements and the dynamic IVC-derived collapsibility index, obtained with convex and cardiac transducers. A secondary aim was to assess the agreement of patient allocation to one of the two groups: “fluid responder” or “fluid non-responder”, based on IVC-CI calculation with two different probes.
A prospective observational study was performed in compliance with Helsinki declaration. Written informed consent was obtained from all patients. Institutional Bioethics Committee of the John Paul II Hospital in Kraków, Poland approved the study protocol (Ref. No.: DW-0700-017/14).
The methodology was consistent with international guidelines for observational studies(13). The study was conducted in February 2015 in the Emergency Department (ED) of a tertiary care centre. The inclusion criteria were: consecutive patients aged >18 years old, spontaneously breathing, admitted to the cardiac ED with chest pain. Exclusion criteria were: atrial fibrillation, dyspnoea, inability to lie down in a supine position, and difficulty to obtain interpretable ultrasound images from the subcostal acoustic window.
All bedside ultrasound examinations were performed by two certified sonographers, both with at least 5-year experience in echocardiography and emergency ultrasound. Each patient was examined twice. The first examination was conducted with a cardiac and the second one with a convex probe. All examinations were performed with a portable ultrasound system equipped with a 1–5 MHz transthoracic phased-array (cardiac) and a 3.5–5 MHz curvilinear (convex) transducers (CX 50 Philips, Eindhoven, Netherlands).
All IVC views were obtained in a supine position. The inferior vena cava was visualized longitudinally in a sub-xiphoid view. The infrahepatic segment of the IVC was imaged as it entered the right atrium. At that stage, the images were stored as 5-second digital loops on the machine’s hard drive.
During the second stage of the study, all digital loops (half performed with a cardiac, and half with a convex probe) stored in the ultrasound machine memory were reviewed by 20 clinicians with experience in focused cardiac ultrasound (at least 200 POCUS examinations’ experience). All participants were blinded to the study allocation, and not aware they assessed twice IVC diameter of the same patient examined with two different transducers. To minimize the possible inconsistencies in the measurement technique, all participating physicians underwent a 30-minutes didactic course focused on relevant sonographic details of the study. Maximal and minimal IVC diameters (IVC max and IVC min, retrospectively) were measured in two dimensional (2D) mode, distally to the hepatic vein-IVC junction, over a single respiratory cycle, by tracking the distance between anterior and posterior walls perpendicular to the long axis of the vessel.
In the final stage of the study, reports and measurements performed by the participating clinicians were analysed by an independent researcher. Comparisons of the measurements were conducted for each pair of transducers (cardiac vs convex) for each patient. IVC-CI was calculated for each of the digital loops. The IVC collapsibility index (IVC-CI) was defined as: IVC-CI = (IVC max – IVC min) / IVC max and expressed as percentage. Based on IVC-CI calculation, patients were categorized to “fluid responder” (if IVC-CI ≥ 40%) or “fluid non-responder” (if IVC-CI <40%) group.
The study sample size resulted from logistic reasons, primarily the number of clinicians with experience in POCUS, which determined the number of patients enrolled in the study. As modelling of balanced systems is more efficient and tolerant of deviations from the assumptions of testing, we assumed the same number of physicians-evaluators and patients enrolled in the study (
Categorical data are reported as number and percentages. Proportions were compared by Pearson’s chi2 test.
All tests were two-sided. A
Twenty patients (13 male, aged 45–74 years old, median 59 years old) were included in the study. In all studied patients (n = 20) the ultrasound examination rendered an interpretable view of IVC, and enabled measurements of its diameters with cardiac and convex transducers.
Thus, two digital loops of each of 20 patients were stored on the machine’s hard drive. The whole analysed set comprised 800 elements (400 pairs). The participating clinicians represented the five following groups: 9 cardiology consultants, 2 internal medicine consultants, 2 anaesthesiology consultants, 5 cardiology residents and 2 internal medicine residents (with a minimum of four years of training completed). No differences were found between the two levels of the first main factor-type of transducer: in IVC min (
The results of inferior vena cava diameters and ANOVA analysis of the assessed factors
Parameter | Factor: transducer | Factor: clinician * | Interaction: transducer * clinician | |||
---|---|---|---|---|---|---|
Cardiac transducer | Convex transducer | Relative Delta [%] | ||||
IVCmin (mm) | 12.6 ± 7.25 | 12.1 ± 7.37 | 4.1 | 0.4127 | 0.0355 | 0.6557 |
IVCmax (mm) | 17.9 ± 6.30 | 17.3 ± 6.55 | 3.5 | 0.1785 | 0.0272 | 0.6946 |
IVC delta (mm) | 5.3 ± 2.85 | 5.2 ± 3.11 | 2.7 | 0.6411 | 0.0262 | 0.4222 |
IVC CI (%) | 35.4 ± 24.5 | 35.3 ± 23.79 | 0.3 | 0.9746 | 0.0069 | 0.7062 |
IVCmin – minimal inferior vena cava diameter; IVCmax – maximal inferior vena cava diameter; IVC delta – maximal-minimal vena cava diameter ; IVC-CI – inferior vena cava collapsibility index
We also did not notice a significant difference in the decision of allocation to one of the two groups: “fluid responder” [(cardiac (
Our study shows a good agreement between IVC diameter measurements and IVC collapsibility index when imaging was performed with convex and cardiac ultrasound probes. A novel finding in this study is the observation that allocation to one of the groups: “fluid responders” or “fluid non-responder” was not probe-dependent. Thus, both transducers can be used interchangeably to predict fluid responsiveness. This is important information in POC practice, since both transducers are used commonly during IVC assessment.
Although current American and European guidelines do not recommend any particular probe for IVC measurements, it has not been entirely clear whether the results obtained with these two transducers are equal, or at least similar(14). Although we did not observe it in our study, there are potentially few reasons for considerable discrepancies between measurements performed with these two probes. Both transducers have different characteristics and physical properties, such as operating at different centre frequencies (with bandwidth for cardiac 1–5 Mz vs convex 3.5–5 Mz), have different dimensions, footprints and shapes, and provide different image formats(15). All mentioned technical differences might translate into different imaging angles and planes when using convex or cardiac probes. Differences in imaging angulations, in turn, may cause significant differences in IVC diameters measured with these two transducers. Although in numerous studies various aspects of IVC measurements were assessed(6,15,16), we did not find any study comparing the consistency of measurements performed with cardiac and convex probes.
Many studies have shown that the potential source of discrepancies in measurement of IVC might be the image acquisition modality, methodology and the patient’s position during the assessment(17,18). Finnerty
There are some limitations of our study. The generalizability of our findings to other sonographers is limited by the relatively few numbers of examinations that obtained images for this study. All ultrasound examinations were performed by fairly experienced sonographers. It cannot be ruled out that results obtained by less competent physicians, especially in a life-threatening scenario, would have been different.
Cardiac and convex transducers can be used interchangeably for the estimation of dimensions of IVC and its respiratory variability.