Otwarty dostęp

Photon beam commissioning of an Elekta Synergy linear accelerator

, ,  oraz   
21 gru 2017

Zacytuj
Pobierz okładkę

The aim of this study is to present the results of commissioning of Elekta Synergy linear accelerator (linac). The acceptance test and commissioning were performed for three photon beams energies 4 MV, 6 MV and 15 MV and for the multileaf collimator (MLC). The percent depth doses (PDDs), in-plane and cross-plane beam profiles, head scatter factors (Sc), relative photon output factors (Scp), universal wedge transmission factor and MLC transmission factors were measured. The size of gantry, collimator, and couch isocenter were also measured.

Język:
Angielski
Częstotliwość wydawania:
4 razy w roku
Dziedziny czasopisma:
Medycyna, Inżynieria biomedyczna, Fizyka, Fizyka techniczna i stosowana, Fizyka medyczna