Improved estimation of left ventricular volume from electric field modeling
, , e
27 dic 2021
INFORMAZIONI SU QUESTO ARTICOLO
Pubblicato online: 27 dic 2021
Pagine: 125 - 134
Ricevuto: 16 giu 2021
DOI: https://doi.org/10.2478/joeb-2021-0015
Parole chiave
© 2021 Leonie Korn, Stephan Dahlmanns, Steffen Leonhardt and Marian Walter, published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Figure 1

Figure 2

Figure 3

Figure 4

Figure 5
![a) Left ventricular in-vitro phantoms made from silicone (left: carbon, right: pure), b) C-arm with manual test bench to measure phantom dimension, c) Cross-sectional areas obtained from CT measurements at the center electrode pair, adapted from [32].](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e8e215d2f6c89dbc94e/j_joeb-2021-0015_fig_005.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250913%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250913T005113Z&X-Amz-Expires=3600&X-Amz-Signature=590c8fb7bacc8d7eca8d64b8c68ece03cdf8bf3c8e7f8a431d8d6bcd442848f4&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Figure 6

Figure 7

Measured volumes inside the silicone ventricle from CT measurements_ Cross-sectional areas obtained between the center electrode pair_ Estimated equivalent center radius_
Step | Volume (mL) | Area (mm2) | Radius (mm) |
---|---|---|---|
1 | 117.90 | 1248.84 | 19.94 |
2 | 107.12 | 1022.55 | 18.04 |
3 | 97.96 | 855.50 | 16.50 |
4 | 87.20 | 683.15 | 14.75 |
5 | 76.84 | 506.83 | 12.70 |
6 | 69.46 | 401.12 | 11.30 |