Electrical impedance spectroscopy is a widely used tool for characterization the structure of tissues and cell cultures [1]. In the cases where the properties of objects are changing in time (e.g., heart muscle) or the objects are moving as cells in a microfluidic channel, the coverage of the frequency range of interest within a short timeframe demands to satisfy the criteria of the linear time-invariant (LTI) system. If the properties of a sample under test (SUT) are changing significantly during a measurement timeframe, the corresponding spectra will be distorted. Recently, a novel method was proposed for the measurement of periodically time-varying impedances using multisine excitation [2]. However, the most of impedances do not change periodically. For example, the impedance of cells moving fast between the electrodes of high throughput microfluidic system or the impedance of bacteria attaching to the electrodes of a biosensor, also the impedance of food products changing during their processing. That is, the short timeframe of every single non-repeatable measurement becomes valuable.
As a result, the limited amount of energy of excitation signal spreads between multiple signal components with different frequencies during a short timeframe. Therefore, the power of corresponding individual components, equal to the square of their root-mean-square (RMS) values, decreases. The task to use the limited energy resources of signals most effectively and flexibly becomes necessary.
Concurrently, such an important criterion of the efficiency of impedance measurements – the signal-to-noise ratio (SNR) of measured signal – is proportional to the power of every frequency components in it, decreases as well. In electrical bioimpedance (EBI) measurements, unfortunately, the SNR cannot be improved by increasing the overall amplitude of the summary excitation signal, since it is limited to low values because of both two issues: a) satisfying the criteria of linearity of the LTI systems and b) fulfilling the security needs for living tissues [3], [4].
Even for non-biologic measurements, the allowable input signal range and power supply voltage of electronic components, both limit the allowed amplitude of signals. Moreover, due to the high crest-factor (CF) of noisy signals, the probability of saturation of measurement channels increases. That leads to nonlinearities, which involve corresponding uncertainties into measurement results.
As discussed below, there are still several possibilities to improve the SNR in EBI measurements by optimizing the properties of excitation signals and specifying some parameters of the measurement system.
The typical structure of EBI spectroscopy system is given in Fig 1. An excitation signal source (generator of the excitation waveform) has inputs for setting the required timing, frequency range, and amplitude. The signal from this generator provides an input excitation for the complex impedance to be measured
Next block, performs both, signal conditioning (amplification, filtering, normalizing) and analog-to-digital (A/D) conversion. A part of this block (front end) is often located near the SUT for obtaining better performance.
Discrete or fast Fourier transform (DFT or FFT) of the response signal from the SUT provides the amplitude |
More exactly, to measure an unknown impedance, we can choose one of the two approaches. Either we apply a known voltage excitation across the SUT and measure the current response flowing through, or we inject a known current excitation into that object and measure the voltage response across it. We can also measure both, current and voltage, simultaneously. The last method is preferable if the excitation source is not stable enough in a required frequency range or the parasitic impedances cause too large changes on the known parameters of the excitation signal [5].
Typically, the signal-to-noise ratio (SNR) of the measured signal is considered as the main efficiency criterion of the measurement system. The SNR of the EBI spectroscopy system depends on several factors, including the frequency bandwidth and measurement time. These parameters may be as well considered as efficiency criterions.
In several cases, e.g., for implantable and wearable measurement devices, the energy consumption, and physical dimensions are important. Also the reduced complexity of the measurement system is required frequently.
In turn, the parameters as frequency bandwidth, measurement time, and SNR of measurements are mutually related.
To characterize the properties of biological objects, the α-and β-dispersion frequency ranges are to be covered, usually. In the single-cell cytometry, capacitance and conductivity of the cell membrane and conductivity of the cytoplasm influence mainly the β-dispersion area that exhibits a characteristic frequency between several kHz up to tens of MHz [5]. In the case of living tissues, the spectrum of bio-impedance shifts toward lower frequencies and the spectra of the β-dispersion range cover typically 3-4 decades within the kHz to MHz band [1], [6]. However, even if only the β dispersion area is of interest, some part of lower frequency area must be included since the α-dispersion changes the plateau, from which the β-dispersion begins.
Presentation of signals can be converted mutually between time and frequency domains by the aid of mathematical operators known as transforms. The use of Discrete Fourier Transform (DFT) or more efficient Fast Fourier Transform (FFT) for the conversion from time to frequency domain is essential in fast impedance spectroscopy. The inverse transforms (DFT-1 and FFT-1) apply to opposite conversions. Though the signals are classified as periodic and aperiodic ones, however, the aperiodic signals can be addressed also as periodic having infinitely long period in the time domain. In the frequency domain, such approach produces infinitely small frequency steps leading to a continuous spectrum. In signal processing practice, a signal component with the longest but finite period
The smallest positive value of
Due to the discrete nature of signal processing, one more condition must be fulfilled for the synchronous DFT: the measurement interval
The spectral leakage errors can be reduced by employing special window functions in the time domain. However, this increases the complexity of signal processing and produces additional error components. Hence, if feasible, the synchronous measurement is always preferred.
The measurement interval,
Although only one period of a periodical signal fulfills the requirement of synchronous DFT (
However, this improvement comes at the cost of increased total measurement time (observation time)
The minimal measurement interval
One possibility is to decrease the observation time
However, multifrequency signals have a drawback: their crest factor (CF) - the ratio of signal’s peak value in respect to its RMS level – is often higher in comparison with a single frequency signal (sine wave). Moreover, even if the CF of the multifrequency signal is similar or even below the CF of a single sine wave
The highest frequency
To improve the SNR, higher RMS magnitudes of the frequency components of interest (FCI) are desirable. For a graspable comparison of different waveforms, some variables must be normalized as constants. It is reasonable to normalize the amplitude to the unit value ± 1 and the duration to 1 second. Normalizing the duration of the signal against 1 second also provides a normalized presentation of the spectra since the fundamental frequency is reciprocal to the duration
In general, RMS magnitudes of the frequency components (bins) of excitation signals are not equal. The shape of a magnitude spectrum may also be decaying, rising or oscillating around some mean value. The mean value of the RMS magnitudes of a number of frequency bins that fall in the half-power limit range is used here as a figure of merit for this general case.
According to Parseval theorem, the total energy in the frequency domain must be the same as calculated in the time domain. In the case of using DFT, this relation is following:
where
The only signal with a single line spectrum is the sine wave. Sinusoidal signal
where
The infinitely short δ-pulse can be considered as an ideal excitation signal for the fast impedance spectroscopy since it covers all the frequencies in the shortest time frame. However, in practice, there is a crucial drawback – because the excitation pulses are very short ones with limited amplitudes, the energy of pulse spread over a number of frequency components, will be very low. The reason is that according to the left side of (3), the energy content of the signal is proportional to the surface area of the waveform, and in the frequency domain, this energy is divided among spectral components. In the case of very short pulses, the distribution of frequencies is dense, almost all the higher harmonics of the first frequency
Other types of short pulses, e.g., the chirps as short-time frequency sweeps, could also be used for EBI measurements [8]. Chirps with a single cycle or even less can be generated, too. The RMS values of the first six components of the exponentially modulated short chirp [9] (Fig. 3) are higher in comparison with the pulses shown in Fig.2, but still low since the surface area of the waveform (hatched), being proportional to the energy content of the signal, is small.
The further symmetrical extension of the excitation signal shown in Fig. 2, runs out into a rectangular waveform in Fig. 4. In this case, the energy content of the signal is maximal. However, the spectrum contains only odd harmonics (1st, 3rd, 5th, 7th), which are declining fast with frequency by 1/
Another type of signal that covers a large number of harmonic frequencies with almost equal amplitudes is a step wave (Fig. 5). This type of excitation is often used in a time domain based impedance measurement [5], [10]. The total energy content of the signal starting from zero is near maximal, but there are also serious drawbacks. At first, most of the energy of the signal is concentrated in a DC component that is mostly undesired in EBI measurements. If the rise time of the step waveform tends to zero, the number of harmonics tends to infinity, but their magnitudes reach to zero. In practice, the rise time of the step waveform is larger than a zero moment, and the amplitudes of higher harmonics are decaying.
A step waveform with the rise time of 1/500 of the period
The energy content of the shifted waveform is proportional to a cross-hatched area, and its spectral content is similar to very short pulses.
The DC component can be canceled also by adding a second part of opposite polarity. However, the waveform resumes in near rectangular signal with a well known decaying shape of the magnitude spectrum (Fig. 4). It must be emphasized here, that DFT interprets the signal as periodic, and according to (1) all the changes of the waveform must be included in a period.
The situation described here also illustrates the drawback of time-domain impedance measurement method with a short pulse or step waveforms. In the case of unbalanced step waveforms, most of the excitation signal energy is useless for spectroscopy (concentrated in a DC component). Faster measurement cycle comes at the cost of lower SNR since the energy of harmonic spectral components is created only during the front part of the step or duration of the balanced pulses. Another reason for the low level of spectral components is that the energy spreads over many harmonic components. Averaging improves the SNR, however, this contradicts with the idea of fast measurements. Moreover, the use of longer excitation with a higher level of desired spectral components is more efficient under similar conditions since only uncorrelated noises can be effectively suppressed by averaging.
Chirps are considered here as signals of many cycles of sinusoidal oscillation with a changing „instantaneous“ frequency. A sine-wave chirp signal with an instantaneous phase θ(
with the „instantaneous“ frequency
In practice, however, generating of discrete signals is more convenient. The waveforms of the sinusoidal and discrete, so-called signum-chirp
The dependency of the mean RMS magnitudes on a number of frequency bins
Further analysis shows some deviation (ripple) of the RMS magnitudes
and is presented in Fig. 7 and Fig. 8.
The advantage of signum chirp (in comparison with a sinusoidal one) is higher RMS value of its spectral components, but the ripple is also considerably higher in the cases, when the number of desired harmonic components,
Another disadvantage of signum chirp is that some part of the energy, typically 15-20 % of total energy, always spreads into higher harmonics above k.
Another type of wideband excitation waveform with a dense distribution of frequencies is a maximum length sequence (MLBS). MLBS is a type of pseudorandom binary sequence, which can be easily generated by an
The shape of the magnitude spectrum becomes flattered as
where B ≈ 0.93 and
It may be concluded that the magnitudes of the spectral components of MLBS and signum chirp signals depend almost similarly on a number of frequency components
The excitation signals described above (except sinewave) have a dense distribution of frequencies. Spreading of the signal energy in such a way reflects in a lower power of separate frequency components. If the frequency response of SUT is smooth, as in the case of EBI measurements, the concentration of the signal energy into a limited number of components is more adequate than the dense distribution. A more detailed illustration will be given in next sections.
The accuracy of fitting the equivalent circuit model with the impedance measurement results depends on the number of frequency components. In general, the use of more frequencies improves the fitting accuracy [13]. However, the fitting accuracy also depends on the signal to noise ratio (SNR) of measurements. As already discussed in previous sections, the magnitudes of spectral components of the excitation signal are decreasing, when their number increases. It follows that the number of used spectral components has two contradictory effects on the accuracy of estimation of parameters of the equivalent circuit model. Usually, less than ten components are required in EBI spectroscopy.
The SNR is expressed through the ratio of signal and noise power
If the noise power density is not uniform in the frequency range of interest, then the better overall SNR is obtained when the shape of the excitation spectrum follows the power density spectrum of noise. Moreover, in-vivo electrical bioimpedance (EBI) measurements must comply with the safety regulations. The safe applied current for the BF and CF type devices is limited according to the IEC60601 standard.
In EBI measurements, the required dynamic range of the impedance magnitude may reach three decades and more [1], [5]. Therefore adaption of the magnitudes of spectral components of the excitation provides better SNR. From the excitation signals described above, only the sine wave and sinusoidal multicycle chirp allow free adjustment of the magnitudes of separate frequency components.
Binary multifrequency waveforms in Fig. 10a and Fig. 11a with two discrete values (±1, if normalized) look similar to the maximum length sequence (MLBS) shown in Fig. 9a. However, a sequence of pulses in the period
According to the Fourier principle, periodic signals can be represented by a set of discrete frequencies. A spectrum of the periodic BMS waveform (a pattern of pulse sequence) may be composed as the sum of spectra of its
Magnitude spectrum
As a result, the shapes of the BMS spectra differ principally from the envelope of sinc-function being characteristic to MLBS. The envelope of the spectrum lines of synthesized BMS may have as decaying, as well as flat (Fig. 10b) or even rising shape (Fig.11b).
The CF of the BMS waveforms is the smallest of possible (CF = 1), and the total energy of BMS excitation is two times higher compared to the sine or multi sine wave with the same amplitude. Considering the simplicity in generating binary signals, BMS also meets the criteria of reduced energy consumption and lower complexity. The disadvantage of the BMS is that some part of the energy always spreads into higher harmonics outside the desired spectral range. Therefore, characterization of the BMS only by the smallest CF = 1 is an overestimation of its merits.
Multisine excitation retains the advantages of the sine wave while reducing the measurement time. The specific advantage of the multisine signal is that only the desired components appear in the spectra. Adding up several (
Let us indicate here that the CF of the sum of same level consequent frequency components with optimal phases can be below the corresponding value of a single sine wave
Let us take an example. In the case of 36 added components CF = 1.358. If the total amplitude of the sum of signals with equal amplitudes is set to 1 V, then according to (10), RMS values of all components are 122.8 mV that is 5.76 times less than RMS value of the single sine wave with the amplitude of 1 V.
In the case of very sparse distribution of frequencies, the optimization of initial phases of multisine components is ineffective, and the CF tends to its worst possible case
Remark: even though some part of the BMS energy transfers to useless higher harmonics, the RMS levels of the components of well optimized BMS are higher than of the optimized multisine. A benefit of using BMS increases with the sparsity of frequency distributions. Fig. 13 illustrates RMS magnitudes of both excitation waveforms for the different frequency distributions.
The stair-step multifrequency waveforms (MFS) [16] and tri-level (ternary) waveforms (MFM) [17] are recently proposed for the broadband bioimpedance spectroscopy. Although these waveforms can be used as multifrequency excitation, there are also some limitations and disadvantages.
Calculations with (3) show that the total energy content of described 6-th order MFS and MFM waveforms are 61% and 83 %, respectively, in comparison with two level BMS. Taking into account that 83.6 % of the energy of the MFM signal falls into the desired frequency components, the overall energetic efficiency is 69.3 %. The typical energetic efficiency of the BMS signal depends on the distribution of frequencies and is in the range from 65 to 85 %.
Limitations of the described MFM waveforms are that only binary distribution of frequencies and decaying shapes of magnitude spectra are produced. MFS waveform provides a flat magnitude spectrum, but its total energy content is significantly lower in comparison with MFM and BMS. Furthermore, generation of MFM and MFS waveforms requires more resources in comparison with BMS.
The conducted research is not related to either human or animal use.
There are several criteria for rating the efficiency of EBI spectroscopy system as described in the second section. The criteria are related and are often contradictory, e. g., as the SNR and speed of measurements. A choice of proper excitation signal has an important role in acceptable satisfying of different requirements.
Even though the short signals cover wide bandwidth, they do not suit well for EBI spectroscopy. Since the amplitude of signals is strictly limited, the energy content of short pulses (proportional to the waveform area) remains low. In other terms - the crest factor CF of this type of signals is high (too high). The second disadvantage is that the energy of short pulses is spread widely over too many frequency components. The third drawback is that the amplitudes of individual spectral components are not freely adjustable.
Disadvantages of the step waveform have common roots with short pulses. If the step starts from zero, most of the energy is concentrated into a DC component, and the magnitudes of other spectral components are low. Though the shifted initial level cancels the DC component, the remaining waveform becomes similar to a short pulse.
Binary waveforms (MLS, signum chirp, and BMS) have the lowest crest factor (CF = 1) that guarantees highest energy content of the excitation signal in a measurement interval in comparison with other waveforms. The simplicity in generating of binary signals also meets the criteria of reduced energy consumption and lower complexity.
A common disadvantage of all the binary excitations is that some part of the energy always spreads onto higher harmonics of the desired frequency components. This part is mirrored at lower frequencies and distorts the spectra.
Another important aspect is that the overall SNR of impedance measurements also depends on the CF of the response signal. The use of binary waveforms may significantly increase it in comparison with the sinusoidal waveforms [18]. Limiting the bandwidth of binary excitation improves the situation. However, simple filtering also produces distortions and increases the CF of the excitation signal. Another solution is increasing the sampling rate, but this raises both the complexity and energy consumption.
Additional disadvantages of the MLS and signum chirp excitations are that the energy spreads onto too many frequency components, and the amplitudes of spectral components are not freely adjustable. BMS does not have these drawbacks. However, the adjustment range of magnitudes is limited, especially if the rising shape of the spectrum is desired.
Tri-level and stair-step multifrequency waveforms do not have clear advantages - their CF is significantly higher in comparison with BMS, but generation requires more resources.
The advantage of the multisine waveform is that only the desired components appear in the spectra, and the amplitudes of spectral components are freely adjustable. The disadvantages of multisine (in comparison with the BMS) are higher CF, especially in the case of sparse frequency distributions, and increased complexity of generation. However, in the cases where the higher accuracy of measurements is required, it is a good choice. Besides, the multisine waveforms can be used also for the composition of near-binary waveforms [18]. The CF of such signals is near 1.2 even in case of sparse frequency distributions and the distortions caused by unwanted higher frequency components of binary waveforms are eliminated.
To decrease the dynamic range of amplitudes, the use of several time-separated excitation signal sections with different frequency subranges may be an optimal solution.
In electrical bioimpedance (EBI) measurements, several limitations set specific demands on the properties of the excitation signals. Signal-to-noise ratio (SNR), measurement time, and the complexity of realization were considered as the main criteria of efficiency.
Multicomponent signals, both multisine and binary (BMS), have a clear advantage over the spread spectrum ones (short pulses, step waveform, binary chirps, pseudorandom sequences). The final choice is as follows – the multisine wave excitation gives the highest accuracy, but the multi-frequency binary excitation enables the lowest complexity.
The accuracy of fitting of the equivalent circuit impedance model with the impedance measurement results depends on the number of used frequency components. An optimal number of frequencies of the multifrequency excitation is a topic for further research.