Phase and Frequency Matching-based Signal Processing Method for Coriolis Mass Flowmeters

INFORMAZIONI SU QUESTO ARTICOLO

Cita

Signal processing precision of Coriolis mass flowmeters affects the measurement accuracy directly. To improve the measurement accuracy of Coriolis mass flowmeters, a phase and frequency matching-based signal processing method for Coriolis mass flowmeters is proposed. Estimated phase difference is obtained by means of frequency estimation, 90° phase shift, generating reference signals and cross-correlation. Simulated results demonstrate that the proposed method has better phase difference estimation and anti-interference performance than the Hilbert transform method, cross-correlation method, data extension-based correlation method, and quadrature delay estimator. Measurement results of Coriolis mass flowmeters verify the effectiveness and superiority of the proposed method in practice.