Signal processing precision of Coriolis mass flowmeters affects the measurement accuracy directly. To improve the measurement accuracy of Coriolis mass flowmeters, a phase and frequency matching-based signal processing method for Coriolis mass flowmeters is proposed. Estimated phase difference is obtained by means of frequency estimation, 90° phase shift, generating reference signals and cross-correlation. Simulated results demonstrate that the proposed method has better phase difference estimation and anti-interference performance than the Hilbert transform method, cross-correlation method, data extension-based correlation method, and quadrature delay estimator. Measurement results of Coriolis mass flowmeters verify the effectiveness and superiority of the proposed method in practice.