Magazine et Edition

Volume 29 (2023): Edition 4 (December 2023)

Volume 29 (2023): Edition 3 (September 2023)

Volume 29 (2023): Edition 2 (June 2023)

Volume 29 (2023): Edition 1 (March 2023)

Volume 28 (2022): Edition 4 (December 2022)

Volume 28 (2022): Edition 3 (September 2022)

Volume 28 (2022): Edition 2 (June 2022)

Volume 28 (2022): Edition 1 (March 2022)

Volume 27 (2021): Edition 4 (December 2021)

Volume 27 (2021): Edition 3 (September 2021)

Volume 27 (2021): Edition 2 (June 2021)

Volume 27 (2021): Edition 1 (March 2021)

Volume 26 (2020): Edition 4 (December 2020)

Volume 26 (2020): Edition 3 (September 2020)

Volume 26 (2020): Edition 2 (June 2020)

Volume 26 (2020): Edition 1 (March 2020)

Volume 25 (2019): Edition 4 (December 2019)

Volume 25 (2019): Edition 3 (September 2019)

Volume 25 (2019): Edition 2 (June 2019)

Volume 25 (2019): Edition 1 (March 2019)

Volume 24 (2018): Edition 4 (December 2018)

Volume 24 (2018): Edition 3 (September 2018)

Volume 24 (2018): Edition 2 (June 2018)

Volume 24 (2018): Edition 1 (March 2018)

Volume 23 (2017): Edition 4 (December 2017)

Volume 23 (2017): Edition 3 (September 2017)

Volume 23 (2017): Edition 2 (June 2017)

Volume 23 (2017): Edition 1 (March 2017)

Volume 22 (2016): Edition 4 (December 2016)

Volume 22 (2016): Edition 3 (September 2016)

Volume 22 (2016): Edition 2 (June 2016)

Volume 22 (2016): Edition 1 (March 2016)

Volume 21 (2015): Edition 1 (December 2015)

Volume 20 (2014): Edition 1 (March 2014)

Volume 19 (2014): Edition 2 (March 2014)

Volume 19 (2013): Edition 1 (March 2013)

Volume 18 (2012): Edition 2 (June 2012)

Volume 18 (2012): Edition 1 (March 2012)

Volume 17 (2011): Edition 4 (January 2011)

Volume 17 (2011): Edition 3 (January 2011)

Volume 17 (2011): Edition 2 (January 2011)

Volume 17 (2011): Edition 1 (January 2011)

Volume 16 (2010): Edition 2 (January 2010)

Volume 16 (2010): Edition 1 (January 2010)

Volume 15 (2009): Edition 4 (January 2009)

Volume 15 (2009): Edition 3 (January 2009)

Volume 15 (2009): Edition 2 (January 2009)

Volume 15 (2009): Edition 1 (January 2009)

Volume 14 (2008): Edition 4 (January 2008)

Volume 14 (2008): Edition 3 (January 2008)

Volume 14 (2008): Edition 2 (January 2008)

Volume 14 (2008): Edition 1 (January 2008)

Volume 13 (2007): Edition 4 (January 2007)

Volume 13 (2007): Edition 3 (January 2007)

Volume 13 (2007): Edition 2 (January 2007)

Volume 13 (2007): Edition 1 (January 2007)

Détails du magazine
Format
Magazine
eISSN
1898-0309
Première publication
30 Dec 2008
Période de publication
4 fois par an
Langues
Anglais

Chercher

Volume 29 (2023): Edition 2 (June 2023)

Détails du magazine
Format
Magazine
eISSN
1898-0309
Première publication
30 Dec 2008
Période de publication
4 fois par an
Langues
Anglais

Chercher

0 Articles
Accès libre

Objective evaluation method using multiple image analyses for panoramic radiography improvement

Publié en ligne: 04 May 2023
Pages: 85 - 91

Résumé

Abstract Introduction

In the standardization of panoramic radiography quality, the education and training of beginners on panoramic radiographic imaging are important. We evaluated the relationship between positioning error factors and multiple image analysis results for reproducible panoramic radiography.

Material and methods

Using a panoramic radiography system and a dental phantom, reference images were acquired on the Frankfurt plane along the horizontal direction, midsagittal plane along the left–right direction, and for the canine on the forward–backward plane. Images with positioning errors were acquired with 1–5 mm shifts along the forward– backward direction and 2–10° rotations along the horizontal (chin tipped high/low) and vertical (left–right side tilt) directions on the Frankfurt plane. The cross-correlation coefficient and angle difference of the occlusion congruent plane profile between the reference and positioning error images, peak signal-to-noise ratio (PSNR), and deformation vector value by deformable image registration were compared and evaluated.

Results

The cross-correlation coefficients of the occlusal plane profiles showed the greatest change in the chin tipped high images and became negatively correlated from 6° image rotation (r = −0.29). The angle difference tended to shift substantially with increasing positioning error, with an angle difference of 8.9° for the 10° chin tipped low image. The PSNR was above 30 dB only for images with a 1-mm backward shift. The positioning error owing to the vertical rotation was the largest for the deformation vector value.

Conclusions

Multiple image analyses allow to determine factors contributing to positioning errors in panoramic radiography and may enable error correction. This study based on phantom imaging can support the education of beginners regarding panoramic radiography.

Mots clés

  • panoramic radiography
  • quantitative evaluation
  • deformable image registration
  • peak signal-to-noise ratio
Accès libre

High resolution 2D plastic scintillator detectors for radiotherapy departments

Publié en ligne: 05 Jun 2023
Pages: 92 - 103

Résumé

Abstract Introduction

Plastic scintillation detectors (PSD) have been developed for over four decades and are widely used in a variety of fields, but one can find relatively few reports of their clinical use compared to other dosimetric solutions.

Material and methods

The inexpensive detector setup made of a Saint-Gobain BC-400 plastic scintillator and commercially available on the market CMOS-based DSLR Pentax camera was investigated. Build PSD detectors were irradiated with 6, 10 and 15 MV flattening filtered (FF) and 6 and 10 MV flattening filter free (FFF) photon beams using a clinical linear accelerator. Data were processed using Matlab software to remove background and artefacts. A comparison of the spatial resolution parameters to the Gafchromic EBT3 films was performed.

Results

Average dose difference between TPS and PSD was 1.1%. The measured spatial resolution was 0.29 mm, and it differed from the film by 1.1%. MTF50 for PSD was 0.57 mm higher than the Gafchromic film. Signal to dose fit function with an R-square equal to 0.999 was established. The standard deviation of mean pixels value for a series of measurements was below 0.1%, for variable dose rate dependence was below 0.6% and for different energies 1.1%.

It was demonstrated that such a setup allows a satisfactory signal-to-dose dependence and provides high spatial resolution at an affordable price compared to a 2D ion chamber or a diode detector array. Moreover, PSDs are reusable and provide a simple readout compared to Gafchromic films commonly used in radiotherapy departments.

Conclusions

Variable parameters of the camera allow to select signal values at the optimal level. The system presented excellent signal stability, high image resolution and a simple signal-to-dose relationship which encourages further work to investigate PSDs for use in radiation therapy departments.

Mots clés

  • PSD
  • dosimetry
  • radiotherapy
  • plastic scintillation detectors
  • Gafchromic films
Accès libre

Comparison of noise-power spectrum and modulation-transfer function for CT images reconstructed with iterative and deep learning image reconstructions: An initial experience study

Publié en ligne: 05 Jun 2023
Pages: 104 - 112

Résumé

Abstract Introduction

Deep learning image reconstruction (DLIR) is a very recent image reconstruction method that is already available for commercial use. We evaluated the quality of DLIR images and compared it to the quality of images from the latest adaptive statistical iterative reconstruction (ASIR-V) algorithm in terms of noise-power spectrum (NPS) and modulation-transfer function (MTF).

Methods

We scanned a Revolution QA phantom (GE Healthcare, USA) and a 20 cm water phantom (GE Healthcare, USA) with our 512 multi-slice computed tomography (CT) scanner. Images of the tungsten wire within the Revolution QA phantom were reconstructed with a 50 mm field of view (FOV). The images were reconstructed with various ASIR-V strengths (i.e. 0, 10, 20, 30, 40, 50, 60, 70, 80, 90, 100%) and DLIRs (i.e. low, medium, and high) to assess the MTF. The images from the 20 cm water phantom were reconstructed with the same configuration to assess the NPS.

Results

The MTF was similar for both reconstruction algorithms of DLIR and ASiR-V. The peak frequency (fp) of the DLIR low was comparable to that from ASIR-V at 50, 60, 70%; the DLIR medium was comparable to ASIR-V at 80%; and the DLIR high was comparable to ASIR-V at 100%. The average frequency (fA) of the DLIR low was comparable to that from ASIR-V at 40%; the DLIR medium was comparable to ASIR-V at 50%; and the DLIR high was comparable to ASIR-V at 70%. Both the DLIR and ASIR-V were able to reduce noise, but they had a different texture.

Conclusions

The noise in the DLIR images was more homogenous at high and low frequencies, while in the ASIR-V images, the noise was more concentrated at high frequencies. The MTF was similar for both reconstruction algorithms. The DLIR method showed a better noise reduction than the ASIR-V reconstruction.

Mots clés

  • ASIR-V
  • DLIR
  • noise power spectrum
  • modulation transfer function
Accès libre

Osseointegration properties of domestic bioactive calcium phosphate ceramics doped with silicon

Publié en ligne: 23 Jun 2023
Pages: 113 - 129

Résumé

Abstract Introduction

The relevance of this study lies in the fact, that today the search for biocompatible materials for the management of bone defects is of importance. Such materials could become an alternative to transplants. For the replacement of bone defects, two-phasic bioactive ceramics of hydroxyapatite and β-tricalcium phosphate is a very attractive biomaterial due to its excellent biocompatibility and osteoconductivity, but the results of its use are quite controversial due to insufficient bioactivity. The purpose of this work is to investigate the osseointegration properties of two-phase bioactive ceramics doped with silicon (HTdSi), both as a single component and a component in combination with platelet-rich fibrin, as well as in comparison with the well-known imported analogue – BIO, which consists of β-tricalcium phosphate, also as an independent component and a component in combination with platelet-rich fibrin. In the experiment, the rabbits of the New Zealand white breed at the age of 3 months and with an average weight of 2.5 kg were used. The terms of implantation are 30, 60, 90 and 180 days. The advantages of the domestic bio-composite are substantiated on the basis of clinical, radiological and histological studies.

Material and methods

In the experiment, the rabbits of the New Zealand white breed at the age of 3 months and with an average weight of 2.5 kg were used. The terms of implantation are 30, 60, 90 and 180 days.

Results

The osteointegration properties of two-phase bioactive ceramics doped with silicon (HTdSi), both as a single component and in combination with platelet fibrin, were investigated, as well as in comparison with the known imported analog - BIO, which contains β-tricalcium phosphate, both as a single component and in combination with platelet fibrin.

Conclusions

The advantages of domestic biocomposite are substantiated on the basis of clinical, radiological and histological studies.

Mots clés

  • bioactive calcium-phosphate ceramics
  • osseointegration
  • osteoinduction
  • clinical and radiological studies
  • histological studies
Accès libre

Effect of the Small Field of View and Imaging Parameters to Image Quality and Dose Calculation in Adaptive Radiotherapy

Publié en ligne: 23 Jun 2023
Pages: 130 - 142

Résumé

Abstract Background

The use of cone beam computed tomography (CBCT) for dose calculation in adaptive radiotherapy has been investigated in many studies. Proper acquisition and reconstruction of preset parameters could improve the accuracy of dose calculation based on CBCT images. This study evaluated the impact of the modified image acquisition and preset reconstruction parameter available in X-Ray Volumetric Imaging (XVI) to improve CBCT image quality and dose calculation accuracy.

Materials and methods

Calibration curves were generated by scanning the CIRS phantom using CBCT XVI Elekta 5.0.4 and Computed Tomography (CT) Simulator Somatom, which served as CT image reference. Rando and Catphan 503 phantoms were scanned with various acquisition and reconstruction parameters for dose accuracy and image quality tests. The image quality test is uniformity, low contrast visibility, spatial resolution, and geometrical scale test for each image by following the XVI image quality test module.

Results

Acquisition and reconstruction parameters have an impact on the Hounsfield Unit (HU) value that is used as the HU-Relative Electron Density (RED) calibration curve. The dose difference for all the calibration curves was within 1% and passed the gamma passing rate. Images acquired using 120 kVp, F1 (with Bowtie Filter), and 50 mA (F1-120-50-10) scored the highest Gamma Index (GI) of 98.5%. For the image quality test, it scored 1.20% on the uniformity test, 2.14% on the low contrast visibility test, and 11 lp/cm on the spatial resolution test. However, F1-120-50-10 reconstructed with different reconstructions scored 3.83% and 4 lp/cm in contrast and spatial resolution test, respectively.

Conclusion

CBCT reconstruction parameters work as a scatter correction. It could improve the dose accuracy and image quality. Nevertheless, without adequate CBCT acquisition protocols, it would produce an image with high uncertainty and cannot be fixed with reconstruction protocols. The F1-120-50-10 protocols generate the highest dose accuracy and image quality.

Mots clés

  • radiotherapy
  • adaptive radiotherapy
  • calibration curve
  • cone beam computed tomography
0 Articles
Accès libre

Objective evaluation method using multiple image analyses for panoramic radiography improvement

Publié en ligne: 04 May 2023
Pages: 85 - 91

Résumé

Abstract Introduction

In the standardization of panoramic radiography quality, the education and training of beginners on panoramic radiographic imaging are important. We evaluated the relationship between positioning error factors and multiple image analysis results for reproducible panoramic radiography.

Material and methods

Using a panoramic radiography system and a dental phantom, reference images were acquired on the Frankfurt plane along the horizontal direction, midsagittal plane along the left–right direction, and for the canine on the forward–backward plane. Images with positioning errors were acquired with 1–5 mm shifts along the forward– backward direction and 2–10° rotations along the horizontal (chin tipped high/low) and vertical (left–right side tilt) directions on the Frankfurt plane. The cross-correlation coefficient and angle difference of the occlusion congruent plane profile between the reference and positioning error images, peak signal-to-noise ratio (PSNR), and deformation vector value by deformable image registration were compared and evaluated.

Results

The cross-correlation coefficients of the occlusal plane profiles showed the greatest change in the chin tipped high images and became negatively correlated from 6° image rotation (r = −0.29). The angle difference tended to shift substantially with increasing positioning error, with an angle difference of 8.9° for the 10° chin tipped low image. The PSNR was above 30 dB only for images with a 1-mm backward shift. The positioning error owing to the vertical rotation was the largest for the deformation vector value.

Conclusions

Multiple image analyses allow to determine factors contributing to positioning errors in panoramic radiography and may enable error correction. This study based on phantom imaging can support the education of beginners regarding panoramic radiography.

Mots clés

  • panoramic radiography
  • quantitative evaluation
  • deformable image registration
  • peak signal-to-noise ratio
Accès libre

High resolution 2D plastic scintillator detectors for radiotherapy departments

Publié en ligne: 05 Jun 2023
Pages: 92 - 103

Résumé

Abstract Introduction

Plastic scintillation detectors (PSD) have been developed for over four decades and are widely used in a variety of fields, but one can find relatively few reports of their clinical use compared to other dosimetric solutions.

Material and methods

The inexpensive detector setup made of a Saint-Gobain BC-400 plastic scintillator and commercially available on the market CMOS-based DSLR Pentax camera was investigated. Build PSD detectors were irradiated with 6, 10 and 15 MV flattening filtered (FF) and 6 and 10 MV flattening filter free (FFF) photon beams using a clinical linear accelerator. Data were processed using Matlab software to remove background and artefacts. A comparison of the spatial resolution parameters to the Gafchromic EBT3 films was performed.

Results

Average dose difference between TPS and PSD was 1.1%. The measured spatial resolution was 0.29 mm, and it differed from the film by 1.1%. MTF50 for PSD was 0.57 mm higher than the Gafchromic film. Signal to dose fit function with an R-square equal to 0.999 was established. The standard deviation of mean pixels value for a series of measurements was below 0.1%, for variable dose rate dependence was below 0.6% and for different energies 1.1%.

It was demonstrated that such a setup allows a satisfactory signal-to-dose dependence and provides high spatial resolution at an affordable price compared to a 2D ion chamber or a diode detector array. Moreover, PSDs are reusable and provide a simple readout compared to Gafchromic films commonly used in radiotherapy departments.

Conclusions

Variable parameters of the camera allow to select signal values at the optimal level. The system presented excellent signal stability, high image resolution and a simple signal-to-dose relationship which encourages further work to investigate PSDs for use in radiation therapy departments.

Mots clés

  • PSD
  • dosimetry
  • radiotherapy
  • plastic scintillation detectors
  • Gafchromic films
Accès libre

Comparison of noise-power spectrum and modulation-transfer function for CT images reconstructed with iterative and deep learning image reconstructions: An initial experience study

Publié en ligne: 05 Jun 2023
Pages: 104 - 112

Résumé

Abstract Introduction

Deep learning image reconstruction (DLIR) is a very recent image reconstruction method that is already available for commercial use. We evaluated the quality of DLIR images and compared it to the quality of images from the latest adaptive statistical iterative reconstruction (ASIR-V) algorithm in terms of noise-power spectrum (NPS) and modulation-transfer function (MTF).

Methods

We scanned a Revolution QA phantom (GE Healthcare, USA) and a 20 cm water phantom (GE Healthcare, USA) with our 512 multi-slice computed tomography (CT) scanner. Images of the tungsten wire within the Revolution QA phantom were reconstructed with a 50 mm field of view (FOV). The images were reconstructed with various ASIR-V strengths (i.e. 0, 10, 20, 30, 40, 50, 60, 70, 80, 90, 100%) and DLIRs (i.e. low, medium, and high) to assess the MTF. The images from the 20 cm water phantom were reconstructed with the same configuration to assess the NPS.

Results

The MTF was similar for both reconstruction algorithms of DLIR and ASiR-V. The peak frequency (fp) of the DLIR low was comparable to that from ASIR-V at 50, 60, 70%; the DLIR medium was comparable to ASIR-V at 80%; and the DLIR high was comparable to ASIR-V at 100%. The average frequency (fA) of the DLIR low was comparable to that from ASIR-V at 40%; the DLIR medium was comparable to ASIR-V at 50%; and the DLIR high was comparable to ASIR-V at 70%. Both the DLIR and ASIR-V were able to reduce noise, but they had a different texture.

Conclusions

The noise in the DLIR images was more homogenous at high and low frequencies, while in the ASIR-V images, the noise was more concentrated at high frequencies. The MTF was similar for both reconstruction algorithms. The DLIR method showed a better noise reduction than the ASIR-V reconstruction.

Mots clés

  • ASIR-V
  • DLIR
  • noise power spectrum
  • modulation transfer function
Accès libre

Osseointegration properties of domestic bioactive calcium phosphate ceramics doped with silicon

Publié en ligne: 23 Jun 2023
Pages: 113 - 129

Résumé

Abstract Introduction

The relevance of this study lies in the fact, that today the search for biocompatible materials for the management of bone defects is of importance. Such materials could become an alternative to transplants. For the replacement of bone defects, two-phasic bioactive ceramics of hydroxyapatite and β-tricalcium phosphate is a very attractive biomaterial due to its excellent biocompatibility and osteoconductivity, but the results of its use are quite controversial due to insufficient bioactivity. The purpose of this work is to investigate the osseointegration properties of two-phase bioactive ceramics doped with silicon (HTdSi), both as a single component and a component in combination with platelet-rich fibrin, as well as in comparison with the well-known imported analogue – BIO, which consists of β-tricalcium phosphate, also as an independent component and a component in combination with platelet-rich fibrin. In the experiment, the rabbits of the New Zealand white breed at the age of 3 months and with an average weight of 2.5 kg were used. The terms of implantation are 30, 60, 90 and 180 days. The advantages of the domestic bio-composite are substantiated on the basis of clinical, radiological and histological studies.

Material and methods

In the experiment, the rabbits of the New Zealand white breed at the age of 3 months and with an average weight of 2.5 kg were used. The terms of implantation are 30, 60, 90 and 180 days.

Results

The osteointegration properties of two-phase bioactive ceramics doped with silicon (HTdSi), both as a single component and in combination with platelet fibrin, were investigated, as well as in comparison with the known imported analog - BIO, which contains β-tricalcium phosphate, both as a single component and in combination with platelet fibrin.

Conclusions

The advantages of domestic biocomposite are substantiated on the basis of clinical, radiological and histological studies.

Mots clés

  • bioactive calcium-phosphate ceramics
  • osseointegration
  • osteoinduction
  • clinical and radiological studies
  • histological studies
Accès libre

Effect of the Small Field of View and Imaging Parameters to Image Quality and Dose Calculation in Adaptive Radiotherapy

Publié en ligne: 23 Jun 2023
Pages: 130 - 142

Résumé

Abstract Background

The use of cone beam computed tomography (CBCT) for dose calculation in adaptive radiotherapy has been investigated in many studies. Proper acquisition and reconstruction of preset parameters could improve the accuracy of dose calculation based on CBCT images. This study evaluated the impact of the modified image acquisition and preset reconstruction parameter available in X-Ray Volumetric Imaging (XVI) to improve CBCT image quality and dose calculation accuracy.

Materials and methods

Calibration curves were generated by scanning the CIRS phantom using CBCT XVI Elekta 5.0.4 and Computed Tomography (CT) Simulator Somatom, which served as CT image reference. Rando and Catphan 503 phantoms were scanned with various acquisition and reconstruction parameters for dose accuracy and image quality tests. The image quality test is uniformity, low contrast visibility, spatial resolution, and geometrical scale test for each image by following the XVI image quality test module.

Results

Acquisition and reconstruction parameters have an impact on the Hounsfield Unit (HU) value that is used as the HU-Relative Electron Density (RED) calibration curve. The dose difference for all the calibration curves was within 1% and passed the gamma passing rate. Images acquired using 120 kVp, F1 (with Bowtie Filter), and 50 mA (F1-120-50-10) scored the highest Gamma Index (GI) of 98.5%. For the image quality test, it scored 1.20% on the uniformity test, 2.14% on the low contrast visibility test, and 11 lp/cm on the spatial resolution test. However, F1-120-50-10 reconstructed with different reconstructions scored 3.83% and 4 lp/cm in contrast and spatial resolution test, respectively.

Conclusion

CBCT reconstruction parameters work as a scatter correction. It could improve the dose accuracy and image quality. Nevertheless, without adequate CBCT acquisition protocols, it would produce an image with high uncertainty and cannot be fixed with reconstruction protocols. The F1-120-50-10 protocols generate the highest dose accuracy and image quality.

Mots clés

  • radiotherapy
  • adaptive radiotherapy
  • calibration curve
  • cone beam computed tomography