An analytical “quality by design” approach in RP-HPLC method development and validation for reliable and rapid estimation of irinotecan in an injectable formulation
Publicado en línea: 20 jul 2020
Páginas: 57 - 79
Aceptado: 02 abr 2020
DOI: https://doi.org/10.2478/acph-2021-0008
Palabras clave
© 2021 Navya Ajitkumar Bhaskaran et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
The objective of the present study was to develop a robust, simple, economical and sensitive HPLC-UV method using the “quality-by-design” approach for the estimation of irinotecan (IRI) in marketed formulations. RP-HPLC method was developed by applying Box-Behnken design with Hyper-Clone (Phenomenex®) C18 column (250 × 4.6 mm id, particle size 5 µm, ODS 130 Å) as a stationary phase. Acetonitrile and 20 mmol L−1 potassium phosphate buffer (pH 2.5) containing 0.1 % triethylamine in a ratio of 45:55 % (