Long-term outcome of multilayer flow modulator in aortic aneurysms


Zitieren

FIGURE 1.

Volume rendering of implanted multilayer flow modulators (MFMs) (A) and fluoroscopic image of implanted MFM (B) in thoracoabdominal aorta.
Volume rendering of implanted multilayer flow modulators (MFMs) (A) and fluoroscopic image of implanted MFM (B) in thoracoabdominal aorta.

Time and cause of reinterventions after multilayer flow modulator (MFM) implantation

Patient number Time after MFM implantation (months) Cause of reintervention; reintervention undertaken
1. 25 Endoleak type I and enlargement of the aneurysm; implantation of another MFM
2. 3 Insufficient proximal seal and collateral flow, resulting in enlargement of the aneurysm; implantation of another MFM
3. 1 Insufficient proximal seal; implantation of another MFM
4. 8 Enlargement of aneurysm and partial stenosis of subclavian artery; implantation of another MFM into previous MFM
6. 13 Stenosis of brachiocephalic truncus; implantation of a stent, which resulted in a stroke
9. 9 Migration of the MFM; implantation of another MFM
10. 74 Displacement of MFM and stent graft; implantation of another stent graft

Age of patients at procedure, year of procedure, and location of multilayer flow modulator (MFM)

Patient Age Procedure year Location of MFM
1. 48 2012 Thoracic aorta
2. 79 2012 Thoracic aorta
3. 80 2012 Thoracic aorta
4. 73 2013 Aortic arch
5. 65 2013 Thoracic aorta
6. 67 2016 Aortic arch
7. 68 2016 Thoracic aorta
8. 63 2016 Abdominal aorta
9. 71 2016 Thoracic aorta
10. 80 2017 Abdominal aorta
11. 67 2017 Aortic arch
12. 65 2017 Aortic arch
13. 76 2018 Thoracoabdominal aorta
14. 75 2018 Thoracoabdominal aorta
15. 81 2018 Abdominal aorta
16. 66 2018 Abdominal aorta
17. 66 2019 Thoracoabdominal aorta

Aneurysm size before and after multilayer flow modulator (MFM) implantation

Patient number Aneurysm volume (ml) Aneurysm diameter (mm) Difference between the last follow-up and before the MFM implantation Duration of follow-up (months)

Before MFM implantation At last follow-up Before MFM implantation At last follow-up Volume (ml [%]) Diameter (mm [%])
1. 345 854 65 107 509 [147] 42 [64] 76
2. 888 920 122 130 32 [4] 8 [7] 9
3. 309 943 70 110 634 [205] 40 [57] 50
4. 255 355 57 93 100 [39] 36 [63] 69
5. 75 62 26 27 −13 [−17] 1 [4] 14
6. 197 330 44 76 133 [68] 32 [73] 49
7. 343 549 65 85 206 [60] 20 [31] 28
8. 311 266 48 68 −45 [−14] 20 [42] 3
9. 575 667 76 81 92 [16] 5 [7] 14
10. 539 530 97 90 −9 [−2] −7 [−7] 40
11. 32 62 34 50 30 [94] 16 [47] 48
12. 411 411 54 55 0 [0] 1 [2] 33
13. 493 409 53 57 −84 [−17] 4 [8] 24
14. 262 332 60 75 70 [27] 15 [25] 23
15. 307 223 59 81 −84 [−27] 22 [37] 18
16. 130 113 58 58 −17 [−13] 0 [0] 19
17. 250 250 54 60 0 [0] 6 [11] 7