Stereotactic body radiation therapy (SBRT) given in 1–5 fractions is an effective treatment for spinal metastases.1 Spine SBRT involves tight planning margins and steep dose gradients to the surrounding organs at risk (OAR). Spinal cord, which is a serial organ, is the most important dose-limiting structure in spine SBRT planning. The risk of radiation myelopathy can be kept to ≤ 1% with meticulous radiotherapy planning and delivery.2 Multiple studies have demonstrated the safety and feasibility of using stereotactic radiotherapy for spinal metastases.3, 4 In many published de novo and adjuvant studies, spine SBRT has led to one year local control rates of 80–90%.5, 6 Given its safety and efficacy, use of spine SBRT in the United States has increased from 2% to 20% over the last decade.7
Spine SBRT is often delivered with dynamic arc, static intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) on linear accelerator (Linac), non-isocentric robotic delivery (CyberKnife) and Tomotherapy using computed tomography (CT)-based platforms.8, 9 At the time of planning, spinal cord volume is usually defined on diagnostic magnetic resonance (MR) images fused with planning kilovoltage computed tomography (kVCT). This approach can result in fusion errors on the order of 2 mm.10 Despite near-rigid full body immobilization and on-board CT-based imaging, inter-fraction and intra-fraction motion of the spinal cord necessitates a planning risk volume (PRV) margin of 1–2 mm to ensure safe treatment delivery.11
In a study that evaluated the effect of setup errors on dose distribution for spine SBRT12, investigators used cone beam computed tomography (CBCT) scans to assess the actual dose to the spinal cord PRV generated by expanding spinal cord by 2 mm. The difference in minimum dose to the upper 10% of the PRV (
Magnetic resonance-guided radiotherapy (MRgRT) delivery systems have now entered clinical practice at several major treatment centers. One such system is the ViewRay MRIdian LinacTM (ViewRay, Inc., Oakwood Village, Ohio, USA), which combines a 0.345T field strength split-bore magnet MRI with a 28 cm gap that contains 6 MV flattening filter free linear accelerator (Linac).13 The imaging field of view is 50 cm wide with 70 cm diameter of bore body coil, with a capability to acquire scans as fast as 17 sec or 25 sec using the true FISP (TRUFI) imaging sequence. The TRUFI imaging sequence on the ViewRay MRIdian LinacTM platform enables real-time visualization of the spinal cord and surrounding cerebrospinal fluid, thereby making MRIdian an optimal modality for image guided radiotherapy.14 This novel Linac system is equipped with a slightly de-focused double-stack multi-leaf collimator (MLC). This system is designed so that the beams have sharp penumbra with minimal leakage through the leaves. MLCs are designed to project field sizes from 0.2 x 0.4 cm2 up to 27.4 x 24.1 cm2. The MRIdian system uses step-and-shoot intensity modulated radiation therapy (IMRT) technique to deliver dose that is calculated with a Monte Carlo algorithm.
Previous studies have compared dosimetric data for normal tissues and target for different treatment planning stations and delivery techniques.9,15 In this study, we report on the quality of ViewRay MRIdian Linac treatment plans compared to TrueBeamTM STx (Varian Medical Systems, Palo Alto, CA) volumetric modulated arc therapy (VMAT) plans and clinical implementation of MRgRT for spine SBRT along with benefits of using MRgRT for spine SBRT.
Patients previously treated with vertebral body metastases between 2015 and 2018 were included in this retrospective study. This study was approved by the local institutional review board. For simulation, patients were immobilized in a BodyFIX bluebag (Elekta, Stockholm, Sweden) with vacuum wrap.16 They were scanned in supine position with arms elevated above the head. All scans were acquired on a Siemens SOMATOM Definition Edge scanner, with a slice thickness of 1 to 2 mm.
Planning CT and diagnostic MR scans were exported to MIM Maestro (MIM Software, Cleveland, OH, USA) for segmentation of target and OARs. Rigid registration was performed between CT simulation scan and the diagnostic MRI scan (MRI Dx) using the MIM optimization algorithm. Segmentation was done by radiation oncologists with expertise in spine SBRT. Radiographically visible tumor was contoured as gross tumor volume (GTV). Clinical target volume (CTV) was contoured using the international consensus guidelines.17 A geometric margin of 3 mm excluding the spinal cord was used to generate planning target volume (PTV), and a 2 mm margin was used for spinal cord planning risk volume (PRV). Similar principles were used to contour cauda equina.
Common practice for spine SBRT at our institution involves a prescription dose of 27 Gy to PTV in 3 fractions with a simultaneous integrated boost (SIB) of 30 Gy to GTV, which is based on published prospective studies.3,5 For dosimetric comparison, the same set of contours was used to generate both MRIdian and VMAT plans. Treatment planning objectives were: at least 95% of the target volume receives the prescribed dose; hotspots were limited to 110% within 1 cm of the target volume and 105% outside. For spinal cord and spinal cord PVR maximum dose was constrained to 18 Gy and 20 Gy respectively. Mean dose to kidneys was restricted to less than or equal to 10 Gy. For lungs, volume of lungs receiving 5 Gy, 12.5 Gy, 20 Gy and 12 Gy was restricted to 50%, 15%, 10% and 1000 cc respectively.
CT scans and contours were exported to Pinnacle treatment planning system (TPS) to generate VMAT plans for TrueBeam TM STx. VMAT plans were generated using three co-planar 6 MV arcs with gantry angles varying from 178° to 182° (CCW), 183° to 178° (CW), 178° to 182° (CCW) and collimator set at 330°, 25° and 320°. TrueBeamTM STx is equipped with a six degree of freedom (DOF) couch, which allows for more variable beam arrangement. Final dose distribution was calculated with the anisotropic analytic algorithm with dose grid of 2 mm.
For the MRIdian plans, CT images and contours that were used for VMAT plans were imported into MRIdian TPS. On average, 10 to 15 beams spaced 20° to 28° apart (110° to 221°) were used to generate a step-and-shoot IMRT plan. Beams entering through the corners of the couch were removed to avoid dosimetric uncertainty. The isocenter was placed in the PTV. Final dose calculation was done using grid size of 2 mm with Monte Carlo. Final dose distribution from MRIdian and Pinnacle were exported to MIM to tabulate and compare clinically relevant DVH parameters.
For dosimetric analysis, VMAT and MRIdian Linac plans were compared using plan metrics such as near minimum dose (D98%-Dose to 98% of PTV), near maximum dose (D2%), median dose (D50%), conformity index (CI) and dose homogeneity index (HI) for PTV. CI and HI were calculated as shown below18:
To evaluate the impact of intermediate dose on the normal tissue R50, the ratio of volume inside 50% isodose line to the PTV volume was calculated.
Dosimetric data for cord was compared between the two plans. Wilcoxon matched pairs signed-rank test, a non-parametric equivalent of paired t-test, was used to compare dosimetric parameters between VMAT and MRIdian Linac plans. For all statistical analysis SPPS version 25 was used.
Ten metastatic lesions from nine patients were included in this study for plan comparison. Lesions were spread across thoracic spine (T6-T12) and also lumbosacral spine (L2-S1). Dose-volume histogram (DVH) parameters for both plans are shown in Table 1. Detailed dosimetric results for all cases are summarized in Table 2. All plans were able to meet the planning parameters. R50 was lower for MRIdian Linac plans when compared to VMAT plans, indicating a lower spread of intermediate doses with MRIdian (5.16
Dose volume parameters for PTV, GTV, Spinal Cord and Cauda Equina for MRIdian Linac IMRT plans Vs TrueBeamTM STx VMAT plans
Structures | TrueBeamTM STx VMAT | MRIdian Linac IMRT | p-value |
---|---|---|---|
D98% (Gy) | 25.7 (15.6–29.5) | 26.5 (17.7–29.7) | 0.20 |
D50% (Gy) | 29.0 (27.9–39.1) | 30.0 (26.4–33.2) | 0.77 |
Conformity Index | 0.97 (0.93–1.0) | 0.97 (0.90–1.0) | 0.13 |
Dose Homogeneity Index | 0.22 (0.05–0.6) | 0.19 (0.1–0.57) | 0.49 |
R50 | 6.1 (2.9–16.7) | 5.2 (2.8–11.9) | 0.05 |
D50% (Gy) | 31.6 (30.4–34.2) | 31.7 (31.0–34.6) | 0.36 |
D98% (Gy) | 30.12 (25–33.63) | 30.35 (29.17–32.89) | 0.23 |
Conformity Index | 0.99 (0.95–1.0) | 0.99 (0.96–1.0) | 0.58 |
Dose Homogeneity Index | 0.09 (0.04–0.33) | 0.08 (0.01–0.18) | 0.30 |
R50 | 32 (6.08–69.0) | 29 (5.8–66.0) | 0.01 |
Max dose (Gy) | 12.6 (9.1–15.6) | 13.3 (10.35–17.3) | 0.13 |
D0.03 (Gy) | 12 (9.0–15.3) | 12.8 (9.5–16.5) | 0.25 |
Max dose (Gy) | 16 (0.22–28.0) | 18 (0.3–32) | 0.38 |
D0.03 (Gy) | 13 (0.2–20.0) | 16 (0.25–31) | 0.07 |
Detailed Dose volume histogram parameters for PTV, GTV, Spinal Cord and Cauda Equina for MRIdian Linac IMRT plans Vs TrueBeamTM STx VMAT plans
PTV | GTV | Spinal Cord | Cauda Equina | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Case # | Volume | Maximum | Dose (Gy) | D95 ( | Gy) | Volume | Maximum | Dose (Gy) | D95 ( | Gy) | Maximum | Dose (Gy) | D0.03 cc | (Gy) | Maximum | Dose (Gy) | D0.03 | cc (Gy) |
(cc) | MRIdian Linac IMRT | True Beam VMAT | MRIdian Linac IMRT | True Beam VMAT | (cc) | MRIdian Linac IMRT | True Beam VMAT | MRIdian Linac IMRT | True Beam VMAT | MRIdian Linac IMRT | True Beam VMAT | MRIdian Linac IMRT | True Beam VMAT | MRIdian Linac IMRT | True Beam VMAT | MRIdian Linac IMRT | True Beam VMAT | |
1 | 60.37 | 34.69 | 35.44 | 27.43 | 27.50 | 7.02 | 34.69 | 35.44 | 32.46 | 33.76 | 14.82 | 14.77 | 14.67 | 14.47 | 19.70 | 21.72 | 19.67 | 20.19 |
2 | 50.63 | 32.45 | 32.27 | 27.00 | 27.31 | 6.82 | 32.94 | 32.41 | 30.11 | 30.05 | 13.33 | 13.45 | 13.02 | 13.10 | 20.00 | 21.15 | 19.92 | 18.95 |
3 | 66.3 | 36.41 | 37.00 | 27.75 | 27.12 | 27.67 | 36.41 | 37.00 | 32.58 | 29.61 | 17.01 | 11.69 | 16.56 | 10.90 | 21.07 | 14.12 | 20.23 | 13.58 |
4 | 47.63 | 31.68 | 30.85 | 27.19 | 27.94 | 18.8 | 34.02 | 33.45 | 30.06 | 30.52 | 13.25 | 15.65 | 12.62 | 15.34 | 24.08 | 21.19 | 22.48 | 19.79 |
5 | 2.6 | 33.87 | 30.71 | 27.73 | 27.26 | 0.9 | 33.87 | 32.02 | 30.24 | 30.21 | 12.58 | 13.81 | 11.92 | 12.90 | 9.31 | 12.37 | 7.73 | 7.56 |
6 | 3.93 | 33.52 | 31.40 | 27.67 | 27.89 | 1.65 | 33.52 | 34.96 | 32.17 | 27.39 | 10.35 | 9.10 | 10.32 | 9.09 | 16.44 | 18.01 | 12.58 | 12.32 |
7 | 3.05 | 32.64 | 30.88 | 27.50 | 27.82 | 1.07 | 32.65 | 33.70 | 30.44 | 29.02 | 11.17 | 10.10 | 11.15 | 10.09 | 32.13 | 28.01 | 31.16 | 19.32 |
8 | 25.43 | 32.73 | 34.07 | 27.92 | 27.99 | 2.31 | 32.73 | 34.07 | 30.31 | 31.43 | 13.72 | 10.98 | 13.22 | 10.36 | 13.72 | 10.6 | 13.19 | 10.25 |
9 | 29.02 | 33.95 | 34.14 | 27.50 | 27.60 | 1.89 | 35.70 | 34.14 | 30.05 | 30.52 | 11.04 | 9.25 | 9.53 | 8.96 | 0.32 | 0.22 | 0.25 | 0.21 |
10 | 38.29 | 33.02 | 33.95 | 27.88 | 27.02 | 2.51 | 33.02 | 33.95 | 30.45 | 30.51 | 17.30 | 14.11 | 14.51 | 13.75 | 3.54 | 3.67 | 3.20 | 3.40 |
The average beam on time for the VMAT plans was 7 minutes compared to 12 minutes for the MRIdian Linac plans. Dose to the spinal cord and Cauda equina was also calculated and shown to be comparable between MRIdian Linac and VMAT plans (Table 1). Isodose distribution and DVH for MRIdian Linac
Increased global incidence of cancer in combination with improved systemic therapies has led to an increase in the prevalence of oligometastatic disease involving bone.19 Synchronous or metachronous bony vertebral/spinal metastases are diagnosed in approximately 40–70% of patients with cancer, mainly secondary to breast, lung, or prostate adenocarcinoma.20, 21 Aggressive treatment of isolated metastases in select patients may lead to improved outcomes.22, 23
Multiple studies have compared treatment planning quality among dynamic conformal arcs, static IMRT, VMAT and tomotherapy for spine.24, 25, 26 Matuszak
In our experience, the ability to accurately visualize the spinal cord is a significant advantage with the ViewRay MRgRT for several reasons. MRI-guided radiotherapy allows physician contouring to rely on the optimally visualized spinal cord on MR simulation images and not on CT-MRI fusion with accompanying errors associated with registration.10 Figure 2 shows a representative image of a T1 weighted image obtained on a 1.5T diagnostic MRI compared to TRUFI sequence on 0.345T ViewRay MRIdian system demonstrating excellent tumor demarcation on both images sets. At several clinics, including our institution spine SBRT treatments are delivered not only on TrueBeam but also Tomotherapy. Figures 3 and 4 show examples of 0.345T MRI images compared to kV CBCT from TrueBeamTM STx and to MVCT acquired on TomoTherapy®. As can be seen, MRI provides images with superior soft tissue contrast, allowing better visualization of the spinal cord and spinal canal on the volumetric image acquired for setup verification. This enables accurate patient setup for treatment delivery and could potentially minimize set up errors.
Spine SBRT plans usually have a steep dose gradient at the spinal cord/PTV interface. In a systematic review by Chang
MRI-guided therapy with MRIdian does have a few limitations. The MRIdian couch does not allow for six degrees of freedom or non-coplanar beam angles. It also does not permit modification of collimator angle or allow dynamic treatment delivery. Beam entry from couch edges for treatment plans are restricted due to high couch attenuation. Also, the number of beams used for planning are restricted to keep the total treatment time reasonable. Another limitation of the system is the lowest monitor units (MU) that can be delivered with the MRIdian Linac system is 1. Additionally, MRgRT is contraindicated in the post-operative setting if patients have magnetic resonance imaging (MRI)-incompatible metal implants. In many cases, however, the benefits of substantially improving soft tissue contrast, ability of real-time tracking and online adaptation may outweigh any planning and delivery difficulties encountered with MRgRT.
Here, we have shown that 3-fraction spine SBRT plans are dosimetrically comparable between MRIdian and TrueBeamTM STx VMAT plans and MRgRT can be successfully used for SBRT spine with reasonable delivery time.