Calculated neutron energy dependence of the dose-response of large recombination chamber
, and
Dec 11, 2019
About this article
Published Online: Dec 11, 2019
Page range: 117 - 121
Received: Apr 13, 2015
Accepted: Sep 16, 2019
DOI: https://doi.org/10.2478/nuka-2019-0015
Keywords
© 2019 Katarzyna Tymińska et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.
A model of REM-2-type chamber was modeled with MCNPX code to study the dose-response to monoenergetic neutrons in wide energy range from thermal to 20 MeV for various compositions of gas in the chamber. The energy dependence of the total dose absorbed in the filling gas was compared with the energy dependence of ambient absorbed dose D*(10) and with experimental data. The results of the studies will be useful for designing new, improved generation of recombination chambers.