Voltammetric determination of Nordiazepam at meniscus modified silver solid amalgam electrode
and
Dec 20, 2014
About this article
Published Online: Dec 20, 2014
Page range: 105 - 108
DOI: https://doi.org/10.2478/acs-2014-0018
Keywords
© Faculty of Chemical and Food Technology, Slovak University of Technology in Bratislava
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
New electrochemical method for determination of Nordiazepam (NDZ) at meniscus modified silver solid amalgam electrode (m-AgSAE) was developed utilizing differential pulse voltammetry (DPV). The pH effect on the current response of NDZ was studied in the mixture of Britton-Robinson buffer (BR) and methanol (9:1) with optimum pH value of 10. The calibration dependences were examined under the optimal conditions and linearity in the range from 2 × 10−6 to 1 × 10−4 mol L−1 with limit of quantification of 1.7 × 10−6 mol L−1 were accomplished.