Comparing intrapartum ultrasound and clinical examination in the assessment of fetal head position in African women


Cite

Introduction

The fetal head position plays an important role in the mode of delivery and is one of the known predictors of labor outcome. Particularly, the persistent occiput posterior position might be associated with prolonged labor and operative delivery(13). Undiagnosed occiput posterior positions increase the risk of perineal lacerations and neonatal injuries in the process of delivery(4). An accurate method for determining the occiput position is therefore useful in assessing the progress of labor.

Studies conducted in different populations have shown that ultrasonography is more accurate than digital vaginal examination (VE) in assessing the occiput position in labor(5). A systematic review of these various studies found no such study conducted in a black African population(5). Due to variations in the pelvic architecture of women from different geographical locations(68), and the relatively higher incidence of malposition and malpresentation in black women compared to whites(9), a context-specific study on ultrasound versus digital VE was needed on the assessment of the occiput posterior position.

Again, existing studies mainly assessed the overall agreement between ultrasound and digital VE which considered all fetal head positions, without being specific about statistical agreement on the occiput posterior position, which is the most important reason for the examination.

In addition, while existing studies have shown that there is significant agreement between ultrasound and digital VE in the second stage of labor(5), which coincides with when membranes have mostly ruptured, the influence of ruptured membranes on the agreement between ultrasound and digital VE was not very clear.

Consequently, the primary objective of this prospective study included assessing the overall agreement between ultrasound and digital VE in a black African population, and to analyze the agreement on the occiput posterior position. The secondary objective was to assess the influence of ruptured membranes on this agreement.

Material and methods

This cross-sectional study was carried out at the labor and delivery ward of Komfo Anokye Teaching Hospital (KATH) in Kumasi, Ghana. Women who were eligible for inclusion were those who had received antenatal care at KATH, had singleton pregnancy, cephalic presentation, were in spontaneous labor and had given consent to participate in the study. Women with multiple pregnancy, breech presentation, macrosomia, intrauterine growth restriction, known fetal abnormality, previous cesarean section, and those with obstetrical comorbidities, such as hypertensive disorders in pregnancy and diabetes in pregnancy were all excluded from the study.

Recruitment of potential study participants began during the antenatal period. Women receiving antenatal care at KATH who met the inclusion criteria, had no contraindications to vaginal delivery and were in their third trimester, were approached individually by a member of the research team. After explaining to them the purposes and benefits of the study, they were invited to participate and written informed consent was obtained. The signed informed consent form was then attached to the antenatal record booklet of all consenting potential study participants. The women were then approached again during labor at KATH.

The criteria for labor at KATH are painful uterine contractions accompanied by cervical effacement and dilatation. Women with cervical dilatation of at least 4 cm are usually admitted to the labor ward. Digital VE is performed for every woman during admission. Depending on the cervical dilatation, subsequent digital VE is performed every 4 hours, or earlier if clinically indicated. During labor, all potential study participants were again approached individually by a member of the research team, who asked them about their interest in participating in the study. Women who reaffirmed their consent to participate were then recruited. Women were recruited consecutively until the desired sample size was attained. Except for the intrapartum ultrasound scan, which was done for the study participants, all other processes of labor were assessed and monitored as is usually done. Ultrasound examination to determine the fetal head position was arranged to coincide with the next digital VE, in accordance with the protocols of the labor ward at KATH.

Digital VE was performed by the attending obstetrician, who determined the occiput position along with other parameters for assessing the progress of labor. Only obstetricians with at least 5 years of experience in intrapartum care were allowed to determine the occiput position.

Immediately after performing digital VE, a transabdominal ultrasound scan was performed by an experienced sonographer, who had over 10 years of experience in obstetric ultrasound. The sonographer was blinded from the digital VE findings of the attending obstetrician. The ultrasound scans were performed with a mobile ultrasound unit (P 300, Siemens-Acuson, Italy), using a 2–5 MHz curvilinear transducer. Each participant had one ultrasound examination performed for the study.

The transabdominal ultrasound approach was used in assessing the fetal head position by placing the covered transducer in transverse probe orientation at the suprapubic region of the maternal abdomen. Confirmation of cephalic presentation was made before proceeding with the study. Scans were performed in the absence of uterine contractions by obtaining an axial view of the fetal head. In the axial view, midline intracranial structures, including the thalami, falx cerebri, cavum septum pellucidi and the cerebellar hemisphere, served as anatomical landmarks for determining the occipital region of the fetal head. Anterior and posterior cranial structures such as the orbits, nasal bridge, and cervical spine were also used for the identification of direct posterior position. The classification of the fetal head position was therefore determined by following guidelines of the conventional probe orientation for transverse sonograms. Fig. 1, Fig. 2, and Fig. 3 show the orientations for occiput posterior positions.

Fig. 1.

Left occiput posterior (LOP) position

Fig. 2.

Right occiput posterior (ROP) position

Fig. 3.

Direct occiput posterior (DOP) position

Statistical analysis

The data was double-entered into Microsoft Excel Spread Sheet and analyzed with XLSTAT version 2015.6 (Addinsoft, Brooklyn, NY, USA).

The main outcome variable, i.e. fetal head position, was categorized into eight possible options using the 360o of a circle with a 45o interval as follows: left occiput anterior (LOA, 45o), left occiput transverse (LOT, 90o), left occiput posterior (LOP, 135o), direct occiput posterior (DOP, 180o), right occiput posterior (ROP, 225o), right occiput transverse (ROT, 270o), right occiput anterior (ROA, 315o), and direct occiput anterior (DOA, 360o).

The level of agreement between ultrasound and digital VE was analyzed by Cohen’s kappa statistics.

The study was approved by the Committee on Human Research, Publications and Ethics of Kwame Nkrumah University of Science and Technology and Komfo Anokye Teaching Hospital, and informed consent was obtained from all study participants.

Results

In total, results were obtained from 196 participants as presented in Tab. 1. The ages ranged from 18 to 39 years with a mean of 26.8 years (standard deviation 4.9 years). Nearly half (47%) of the women were primiparas. The average gestational age of the participants was 39 weeks 4 days. At the time of the ultrasound examination, 36% of the women had ruptured membranes, and labor was being augmented in about 14% of the participants.

Reproductive characteristics of the women

Variables Number a Percentage (%)
Age group (years)
<20 9 4.5
20–29 134 68.0
30–39 55 27.5
Mean ± SD (years) 26.8 ± 4.9
Body mass index (kg/m2) 27.8 ± 3.5
Parity
Nulliparous 92 47.0
Primiparous 43 22.0
Multiparous 61 31.0
Mean gestational age ± SD (weeks) 39.6 ± 1.07
Ruptured membranes
Yes 69 36.0
No 127 64.5
Fetal head positions (ultrasound)
Left occiput anterior 21 10.7
Left occiput transverse 60 30.6
Left occiput posterior 35 17.9
Direct occiput posterior 5 2.6
Right occiput posterior 30 15.3
Right occiput transverse 33 16.9
Right occiput anterior 12 6.1

Number unless otherwise indicated.

SD – standard deviation

The determination of the fetal head position was obtainable with ultrasound in all participants, but only in 134 participants with digital VE. In the remaining 62 cases (32%), the clinicians could not determine the fetal head position during digital VE. Of this group, 37 cases were in transverse positions (60%) and 16 in posterior positions (26%). In the 134 fetal head positions determined by digital VE, the agreement with ultrasound was moderate (kappa = 0.405). However, the overall agreement for the 196 participants was low (kappa = 0.21) as shown in Tab. 2.

Kappa table of agreement between Ultrasound and Digital VE on fetal head position

Overall agreement in the range of 45o
Ultrasound (+)/ (–)
Digital VE (+)/ (–) + Total:
+ (n: 68) (n: 128) (n: 196)
(n: 0) (n: 66) (n: 66)
Total: 68 194 262
Observed agreements % 134 (51.15%) Kappa = 0.211; 95% CI: From 0.151 to 0.271
Agreements expected by chance %: 99.7 (38.07)
Agreement on occiput posterior positions
Ultrasound (+)/ (–)
Digital VE (+)/ (–) + Total:
+ (n: 14) (n: 56) (n: 70)
(n: 14) (n: 112) (n: 126)
Total: (n: 28) (n: 168) (n: 196)
Observed agreements % 126 (64.29%) Kappa = 0.103; 95% CI: From –0.022 to 0.227
Agreements expected by chance %: 118.0 (60.209%)
Agreement on transverse positions
Ultrasound (+)/ (–)
Digital VE (+)/ (–) + Total:
+ (n: 23) (n: 70) (n: 93)
(n: 23) (n: 80) (n: 103)
Total: (n: 46) (n: 150) (n: 196)
Observed agreements % 103 (52.55%) Kappa = 0.025 95% CI: From –0.182 to 0.057
Agreements expected by chance %: 100.7 (51.35%)

Out of the 196 participants, ultrasound obtained 70 occiput posterior positions, including ROP, LOP and DOP positions. Ultrasound and digital VE agreed on the occiput posterior position in 14 out of the 70 (20%), and disagreed in 56 out of the 70 (80%) cases. However, they both agreed that 112 fetuses were not in occiput posterior positions. This yielded a between-method percentage agreement of 64% on occiput posterior positions, and a kappa value of 0.1 (Tab. 2). Agreement on the transverse occiput position is also presented in Tab. 2.

Out of the 196 participants, 71 had ruptured membranes. The number of occiput posterior positions obtained by ultrasound in these 71 women was 30, but digital VE agreed with this in only 7 participants. On the other hand, digital VE obtained occiput posterior positions in 7 out of the 71 women with ruptured membranes, on which ultrasound disagreed. However, both examinations agreed that 34 participants in the ruptured membrane group were not in occiput posterior positions. Tab. 3 shows details on agreement between ultrasound and digital VE in the population with ruptured membranes versus intact membranes.

Kappa table of agreement on OP positions in the sub-group with ruptured and intact membranes

Occiput posterior position with ruptured membranes
Ultrasound (+)/ (–)
Digital VE (+)/ (–) + Total:
+ (n: 7) (n: 23) (n: 30)
(n: 7) (n: 34) (n: 41)
Total: (n: 14) (n: 57) (n: 71)
Observed agreements % 41 (57.75%) Kappa = 0.067; 95% CI: From –0.137 to 0.272
Agreements expected by chance %: 38.8 (54.69%)
Occiput posterior position with intact membranes
Ultrasound (+)/(–)
Digital VE (+)/ (–) + Total:
+ (n: 7) (n: 33) (n: 40)
(n: 9) (n: 76) (n: 85)
Total: (n: 16) (n: 109) (n: 125)
Observed agreements % 83 (66.40%) Kappa = 0.082; 95% CI: From –0.077 to 0.241
Agreements expected by chance %: 79.2 (63.39%)
Discussion

The purpose of the study was to compare intrapartum ultrasound and digital VE in the assessment of the fetal head position in black African women in the first stage of labor.

The occiput transverse position was the commonest fetal head position in this study population (47%), followed by the occiput posterior position (36%) and then the occiput anterior position (17%). The high rate of the occiput transverse position agrees with Ahmad et al.(10) whose study comprised 5% of blacks, 23% of Asians and over 50% of whites. However, our population had the highest percentage of occiput transverse positions in the first stage of labor. Again, the 36% of the occiput posterior positions is within the 30% reported by studies conducted in other populations(11).

In a third of the study population, digital VE could not determine the fetal head position. This is comparable to similar studies conducted in other geographical locations, which reported up to 60% difficulty in determining the fetal head position with digital VE in the first stage of labor. While Souka et al.(12) attributed this difficulty to occiput posterior positions, the dominant position in such cases in our study was occiput transverse.

The overall agreement between ultrasound and digital VE in our study was low (kappa = 0.2) and comparable to studies by Sherer et al.(13), who obtained a kappa value of 0.12 in their American population, and Shetty et al. 14, who obtained a value of 0.15 in an Indian population. However, a moderate kappa concordance of 0.4 was obtained after removing cases that could not be determined by VE, which is also comparable to the moderate agreement obtained by Souka et al.12 after they had removed the same variable.

Earlier studies by Souka et al.(12) and Shetty et al.(14) had observed that the detection of occiput posterior position by digital VE was more difficult than the occiput anterior position, but they provided no statistical analysis for the agreement on the occiput posterior position. Hence, as further investigation, our study analyzed the between-method agreement on the occiput posterior position, which was very low (kappa = 0.1). This confirms the extreme difficulty associated with detecting the occiput posterior position with digital VE. We further identified that the transverse position was much more difficult to determine with digital VE (kappa = 0.02) than the posterior position in our study population.

In addition, the study also found that about 85% of fetal head positions that could not be determined by digital VE were transverse and posterior positions, which confirms that digital VE has difficulty in detecting malpositions. The implications of the difficulty in detecting occiput posterior and transverse positions in the first stage of labor were beyond the scope of our study design. However, Choi et al.(15) reported that occiput posterior positions in the first stage of labor had higher percentage of cesarean sections than first stage occiput anterior positions. Further research is therefore needed in the black African population to establish the relationship between mode of delivery and first stage occiput positions. In a related subject, the administration of oxytocin has been reported as an influence on the rotation of occiput posterior position(16), which would make the use of ultrasound in confirming fetal head position a useful technology to employ when considering labor augmentation.

Regarding our hypothesized view on the possible influence of ruptured membranes on agreement, no statistically significant influence was obtained in our study. It therefore suggests that ruptured membranes do not improve the accuracy of digital VE.

The major limitation of our study was that it had not been designed to examine participants from the onset of labor to delivery, which could have identified the relationship between mode of delivery and the first stage occiput position.

Conclusions

In conclusion, this study suggests poor agreement between ultrasound and digital VE on the occiput posterior position in black African women who were in the first stage of labor. Again, over 85% of fetal head positions that could not be determined by digital VE were occiput transverse and posterior positions. This shows that Digital VE has difficulty in detecting malpositions, with no significant influence of intact or ruptured membranes. Ultrasound is therefore more useful than digital VE whenever malposition is suspected in the first stage of labor.