Figure 1

Impact of deep inspiration on cardiac dose during postoperative radiotherapy for left-sided breast cancer. Doses from 0% to three-dimensional dose maximum are depicted by color wash. Upper two slices are taken at mid-breast level and the lower pair 6 slices below. At both levels, the proportion of cardiac volume is reduced by voluntary deep inspiration breath-hold.
Impact of deep inspiration on cardiac dose during postoperative radiotherapy for left-sided breast cancer. Doses from 0% to three-dimensional dose maximum are depicted by color wash. Upper two slices are taken at mid-breast level and the lower pair 6 slices below. At both levels, the proportion of cardiac volume is reduced by voluntary deep inspiration breath-hold.

Mean values +/– standard deviations of dose volume histogram (DVH) parameters for heart, left anterior descending coronary artery (LAD), lung and optimized planning target volume (OPTV) achieved by the free breathing and breath-hold technique.

DVH parameterFree breathingBreath-holdP-value
Heart
Dmean [Gy]6.1 +/– 2.53.2 +/– 1.4
Dmax [Gy]51.1 +/– 1.448.5 +/– 6.8
V50Gy [%]0.7 +/– 1.10.2 +/– 0.6
V40Gy [%]6.2 +/– 3.52.1 +/– 1.9
V30Gy [%]7.8 +/– 42.8 +/– 2.3< 0.05
V25Gy [%]8.5 +/– 4.23.2 +/– 2.5
V20Gy [%]9.2 +/– 4.43.6 +/– 2.7
V15Gy [%]10 +/– 4.64.1 +/– 2.9
V10Gy [%]11.2 +/– 54.9 +/– 3.2
Left lung
Dmean [Gy]11.3 +/– 3.210.6 +/– 2.6
Dmax [Gy]51.6 +/– 1.850.6 +/– 1.1
V30Gy [%]17.9 +/– 6.317 +/– 4.6NS
V20Gy [%]20.5 +/– 719.5 +/– 5.1
V10Gy [%]25.9 +/– 8.125.4 +/– 5.8
LAD
Dmean [Gy]23.0 +/– 6.714.8 +/– 7.6
Dmax [Gy]49.7 +/– 3.444.3 +/– 12.2< 0.05
D50% [Gy]20.4 +/– 17.68.4 +/– 11.1
OPTV
V95% [%]95.6 +/– 4.195.2 +/– 6.3NS

Vertical, longitudinal and lateral couch shifts and corresponding systematic and random errors for the initial 18 patients treated with voluntary deep inspiration breath-hold (V-DIBH) and a historic control treated with free-breathing (FB) technique. Differences between the two approaches were non-significant.

Vertical [mm]Longitudinal [mm]Lateral [mm]
Setup parameterFBV-DIBHFBV-DIBHFBV-DIBH
Mean shift11.22.82.71.92.1
Systematic error0.71.51.62.10.81.3
Random error0.60.62.10.71.10.4

Impact of supraclavicular fossa (SCF) radiotherapy on heart and lung dose volume histogram (DVH) parameters

DVH parameterFree breathingp-valueBreath holdp-value
No SCF (n = 23)SCF (n = 31)No SCF (n = 23)SCF (n = 31)p-value
HeartHeart
Dmean [Gy]5.5 +/– 2.26.6 +/– 2.72.9 +/– 1.13.5 +/– 1.6
Dmax [Gy]51.1 +/– 1.551.1 +/– 1.348.2 +/– 7.848.8 +/– 5.9
V50Gy [%]1 +/– 1.30.5 +/– 0.90.3 +/– 0.70.2 +/– 0.3
V40Gy [%]5.9 +/– 3.66.5 +/– 3.4NS1.6 +/– 1.42.4 +/– 2.2NS
V30Gy [%]7.2 +/– 4.18.3 +/– 3.92.3 +/– 1.83.3 +/– 2.6
V25Gy [%]7.8 +/– 4.29.1 +/– 4.12.6 +/– 1.93.7 +/– 2.8
V20Gy [%]8.4 +/– 4.49.8 +/– 4.42.9 +/– 2.14.2 +/– 3
V15Gy [%]9.2 +/– 4.710.6 +/– 4.53.4 +/– 2.34.7 +/– 3.2
V10Gy [%]10.4 +/– 5.112 +/– 4.84.2 +/– 2.65.6 +/– 3.5
LungLung
Dmean [Gy]9 +/– 2.613.2 +/– 2.2< 0.059.3 +/– 3.211.8 +/– 2.4< 0.05
Dmax [Gy]51.1 +/– 1.452 +/– 2.1NS50.4 +/– 1.150.8 +/– 1.1NS
V30Gy [%]13.4 +/– 4.921.7 +/– 4.714 +/– 419.6 +/– 3.3
V20Gy [%]15.5 +/– 5.324.9 +/– 5.1< 0.0516.3 +/– 4.422.3 +/– 3.8< 0.05
V10Gy [%]19.8 +/– 6.131.2 +/– 5.521.5 +/– 5.528.8 +/– 3.6