About this article
Published Online: Aug 06, 2016
Page range: 227 - 232
Received: Dec 11, 2015
Accepted: Jul 22, 2016
DOI: https://doi.org/10.1515/ama-2016-0035
Keywords
© 2016 Bialystok University of Technology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
The paper presents the results of investigations on the air gages dynamic characteristics in the measurement of the round profiles of motor cylinders. The principle of the measuring device is explained, and the analysis of the air gages dynamics is described. The results of dynamic calibration enabled to eliminate those configurations of air gages that may not meet the requirements of the measurement they were designed for. After the proper air gages were chosen, the entire system underwent the accuracy test and passed it successfully revealing the method accuracy better than 10% compared to the reference measurement.