1. bookVolume 13 (2019): Issue 6 (December 2019)
Journal Details
License
Format
Journal
First Published
01 Jun 2007
Publication timeframe
6 times per year
Languages
English
access type Open Access

The design and manufacture of atmospheric plasma jet surgical handpiece

Published Online: 25 Jun 2020
Page range: 235 - 240
Journal Details
License
Format
Journal
First Published
01 Jun 2007
Publication timeframe
6 times per year
Languages
English

Nowadays, with the appearance of new methods in curative medicine, increasing the efficiency and reducing the cost of energy and technology, both medical and nonmedical, has become the focus of interest. Plasma science is receiving increasing interest from the researchers in these areas. The unique nature of plasma has made it widely used in industry and medicine [1]. Plasma is one of the four main phases of a matter; in other words, quasi-neutral gas of charged particles provides collective behavior. The other three phases are solid, liquid, and gas. The term “plasma” refers to ionized gas, all or a significant portion of its atoms lost one or more electrons and turned into positive ions [2].

The passage of electric current from inside an ionized gas leads to a phenomenon called gaseous discharges. This phenomenon is the most common method of plasma production. At present, the focus has been given on atmospheric pressure plasma due to its simplicity and low cost than low-pressure plasma [3].

This type of plasma is non-equilibrium; in other words, there is no local thermodynamic equilibrium and content of hot electrons and cold ions and neutral species in the room temperature range [4]. From a macroscopic point of view, this is cold plasma. The importance of cold atmospheric pressure air plasma makes it possible to use in the processing of materials susceptible to heat and living tissue in the industry and medicine [5]. One type of cold atmospheric pressure plasma is plasma jet. In the contact of plasma with the surface of the material, energy is transferred from the plasma and secondary reactions occur at the surface of the material. Plasma surgery techniques, such as coagulation or withdrawal, are primarily based on the deadly effects of plasma on living systems. To achieve the selective effects and avoid the specific hazards, plasma torches should include characteristic components, and it is necessary to know how to control them by external work parameters [6].

Plasma jets include a nozzle with one or two electrodes. The plasma is generated inside the nozzle and transmitted by a gas stream to the bodies. Principally, the differences between all these plasma jets are in the configuration of the electrode, electrode material, the type of gas, frequency, and the amount of applied voltage. The plasma handpiece simply includes a plasma jet that can be taken by the hand. The gas is converted to plasma by passing through the electrodes and is used for various applications [7].

The advantage of plasma with low temperature is the high UV intensity. This radiation has a high microbicidal effect. Many reactions occur in the plasma due to different mechanisms of sterilization. These reactions include ultraviolet irradiation, reactive species, and the eruption of charged particles and spot heat distribution. The degree of penetration of these rays is limited [8]. The plasma technology is a desirable method for modifying surface properties. About the plasma advantage compared to ultraviolet ray generators, gas can penetrate the gaps. The potential benefits of non-heat plasma technology are that ultraviolet radiation, radicals, and other reactive species can affect microbes from anywhere in the plasma. Cold plasma produces high microbial effects by producing ultraviolet rays, causing small chemical changes and can be used in the food industry [9].

Atmospheric plasma jet has different medical applications due to its low temperature at room temperature. The plasma jet interaction with cells and living tissues of the body can be utilized in procedures such as sterilization, wound healing, and cancer treatment. In recent years, the effect of nonthermal plasmas on cancer cells has been studied, and it has been shown that this type of plasma has anti-proliferative effects on cancer cells [10].

Materials and methods

Designing is known for obtaining optimal solutions for the invention of industrial machines and machines. Design methods are done in a systematic method, and finding the solutions does not occur accidentally.

Despite the widespread use of torches, the process and arc limitations are considered major issues for its application. One of the main factors for this limitation is the lack of recognition and control of the arc dynamics inside the torch and the anode corrosion effect on plasma jet forces. The presence of geometry and symmetric boundary conditions, the gas flow is an intrinsic three-dimensional. Moreover, any arc movement will have a significant effect on the torch and jet output flow.

The arc dynamics resulting from the balancing of the drag force generated from the interactions of the incoming gas flow along the arc and the Lorentz electromagnetic force generated from the local curvature of the arc is continuously displaced in the anode wall. The arc length is proportional to the number of voltage fluctuations. F=ρE+J×B{\rm{\vec F}} = {\rm{\rho \vec E}} + {\rm{\vec J}} \times {\rm{\vec B}}

It can be said that there is a direct relationship between the flow density and the Lorentz force. The generated arc is constantly inclined to reshape and reconnect, and one of its causes is the presence of Lorentz force. This force in the case of local arc curvature plays an important role. Because of the diverse shape of the arc, the Lorentz force creates an angular momentum that agrees or disagrees with arc coupling. Paschen's law about the electrical discharge of gases, which states the necessary voltage for the electrical breakdown of a gas placed in a uniform field, is a nonlinear function of the gas pressure and the distance between the electrodes. Vb=f(p.d){{\rm{V}}_{\rm{b}}} = {\rm{f}}({\rm{p}}.{\rm{d}})

Paschen's law curve determines that the voltage breakdown of a given gas is placed at a distance between the electrodes of a uniform field with a specific material, which is a function of the gas pressure and the distance between the electrodes [11]. The validity of Paschen's law is shown in a wide range of PD values. Nevertheless, in relatively large values of PD, it is seen, in some cases, that the voltage breakdown is lower in the same values of PD with larger distance. On the other hand, at very low pressures, the invalidity of Paschen's law has been observed. In these pressures, the voltage breakdown does not depend on the characteristics of the gas but depends on the purity and properties of the electrodes. The plasma torch is the most critical element of plasma-processing equipment. The design of electrodes is the main factor in the performance of most torches, measuring current and operating power over its lifetime. It is difficult to describe the arc and electrode interactions due to different physical phenomena within a range of parameters, such as in various pressures, currents, and densities. Different structures of the electrode are, therefore, used for various applications [12]. The thermionic cathodes include a wider range of arc flows, and it preferred the cold cathode in designs due to the dominance of field emission. As can be seen in Figure 1, the cathode is placed in the plasma jet as a copper rod at an adjustable distance from the annular anode.

Figure 1

The manufactured handpiece (left) and its cathode and anode of plasma core (right)

From the material behavior's point of view, the anode should be selected from the kind of electron absorber to allow for continuity of flow and cathode generates the additional electrons for the arc. The thermionic cathodes provide high-temperature electrons for producing a large number of electron emissions. Usually, strengthen materials are required in the form of bars, and materials with a lower functionality are added to increase the electron emission in lower temperatures of cathodes. Experimental results indicate that the geometry of electrodes can significantly affect the erosion rate and arc stability [13]. Also, electrodes with an acute head shape can suitably affect the stable discharge. The nozzle is essentially a part of a plasma torch with a high heat load. The nozzle is exposed to the hot plasma, radiation, Joule heating, ion bombardment, and also under electric arc occurrence. The electrode is cooled by the convection, while the nozzle is heated by the convection. When the cold nonplasma gas flows are passed through the cathode and entered into the ionized electrical arc, it is heated intensely, and this hot plasma gas flow with a convective phenomenon exited from the crater the nozzle [14].

The high-temperature plasma flow heats the anode by the radiation. However, this heating effect is usually negligible than other factors. Also, anodic hot spots can radiate the nozzle's heat to a colder gas around the nozzle. Joule heating is an important source of anode heating, but not as much as the heating effect in the cathode. The most important sources of heat transfer to the anode are ion bombardment and electric arc impact. Similar to the cathode that absorbs positive ions and converts their kinetic energy to heat energy, the heat is generated in the nozzle that absorbs the negative ions. The nozzle is cooled by the convection phenomenon, thermal conductivity, and radiation [15].

Results

By finalizing the design, all parts were made. The argon as suitable gas for the plasma process is selected due to ineffectualness, tastelessness, and inodorous. The argon is very neutral and not combined with any other gas and dissolved in the water rarely. In addition, argon is much more suitable than other gases because it greatly reduces the oxidation process of the cathode and the anode.

The dependence of jet length on applied voltage

The cold atmospheric plasma jet length depends on factors such as power supply, applied voltage, gas flow rate, and the distance between the electrodes. In this test, all parameters keep constant and only voltage value changes and are addressed its effect on the jet length. By assembling the system and testing, the breakdown occurs in a voltage of 2.5 kV (Figure 2). From observations, it is found that by increasing the voltage, the jet length increases and, in an extreme state, reaches to threshold value due to plasma density stabilization in the beam electric discharge mode (Figure 3). In the continual increasing of voltage, the plasma state turns from the beam to the arc, and the plasma converts from the cold plasma to hot plasma, which is not desirable. Gas flow is constant at 8 l/min.

Figure 2

The gas breakdown inside the nozzle

Figure 3

Graph of plasma jet length vs. voltage (A) and gas flux (B)

The dependence of the jet length on the gas flux

The amount of gas flux is an affecting factor on the plasma jet. By voltage 4.5 kV, experiments are performed to measure the effect of the gas flux on the plasma jet. Starting by the zero for gas flow, with the gradual increase in the velocity of gas flow, it is shown the jet length reaches to the threshold state. This behavior can be attributed to two main factors. First, at higher velocities, excited atoms, ions, and molecules emitting photons make a long distance before relaxation due to the return of the disturbed system to balance state. Second, at higher velocities, fluid pressure also decreases; as a result, the free mean distance increases slightly and leads to an increase in the energy relaxation length. Both processes lead to an increase in the plasma jet length (Figure 3). On the other hand, as the velocity increases, flame and jet length decreases sharply due to high losses of plasma, including ions and electrons. The results of the tests showed that by increasing the rate of argon gas flux, its concentration decreases. By a very high increase of gas flux, the jet length reduces because the plasma will not have the opportunity to recover between the electrodes and the gas flow will be turbulent.

Conclusion

In this work, the characteristics of an atmospheric plasma jet were first introduced. Then, the handpiece that was fitting this plasma was designed. The lightweight and ergonomic design, insulating against the high-voltage current, durable and beautiful body, was analyzed. After construction, it has been explained that jet length is one of the important factors in the processing of surfaces. The atmospheric plasma jet length depends on factors such as the electrodes voltage and the exhaust gas intensity. In this study, failure occurred at a voltage of 2.5 kV. Further, the flame length increased with increasing voltage, and due to stabilizing, the plasma density in a beam discharge mode, the jet length threshold occurred in particular voltage value. As more applied voltage, the plasma mode changed and turned from beam to arc. It was observed that with the increase in the voltage, jet length increased. It was also observed that with the increase in the velocity of the gas flow, the jet length reached the threshold state. This behavior can be attributed to two main factors. First, at a higher velocity, excited atoms, ions, and molecules emitting photons, travel longer distances before relaxation. At higher velocities, the fluid pressure also decreased, and as a result, the free distance increased slightly and led to increasing the energy's relaxation length. Both of these processes led to an increase in the plasma jet length. On the other hand, with increasing velocity, the flame, and jet lengths decreased greatly due to high losses of plasma, including ions and electrons. The results showed that with increasing the velocity of argon gas, its concentration decreased. This study concluded based on the result achieved, the performance of the proposed design is successful. The advantages include low-cost manufacturing, highly stable performance, and low erosion; they can make considerations for future development.

Figure 1

The manufactured handpiece (left) and its cathode and anode of plasma core (right)
The manufactured handpiece (left) and its cathode and anode of plasma core (right)

Figure 2

The gas breakdown inside the nozzle
The gas breakdown inside the nozzle

Figure 3

Graph of plasma jet length vs. voltage (A) and gas flux (B)
Graph of plasma jet length vs. voltage (A) and gas flux (B)

Reiazi R, Akbari ME, Norozi A, Etedadialiabadi M. Application of cold atmospheric plasma (CAP) in cancer therapy: a review. Int J Cancer Manage. 2017; 10:e8728.ReiaziRAkbariMENoroziAEtedadialiabadiMApplication of cold atmospheric plasma (CAP) in cancer therapy: a reviewInt J Cancer Manage201710e8728Search in Google Scholar

Laroussi M. Low-temperature plasma jet for biomedical applications: a review. IEEE T Plasma Sci. 2015; 43:703–12.LaroussiMLow-temperature plasma jet for biomedical applications: a reviewIEEE T Plasma Sci20154370312Search in Google Scholar

Ulrich C, Kluschke F, Patzelt A, Vandersee S, Czaika VA, Richter H, et al. Clinical use of cold atmospheric pressure argon plasma in chronic leg ulcers: a pilot study. J Wound Care. 2015; 24:196, 198–200, 202–3.UlrichCKluschkeFPatzeltAVanderseeSCzaikaVARichterHClinical use of cold atmospheric pressure argon plasma in chronic leg ulcers: a pilot studyJ Wound Care2015241961982002023Search in Google Scholar

Wende K, Bekeschus S, Schmidt A, Jatsch L, Hasse S, Weltmann KD, et al. Risk assessment of a cold argon plasma jet with respect to its mutagenicity. Mutat Res Genet Toxicol Environ Mutagen. 2016; 798–99:48–54.WendeKBekeschusSSchmidtAJatschLHasseSWeltmannKDRisk assessment of a cold argon plasma jet with respect to its mutagenicityMutat Res Genet Toxicol Environ Mutagen2016798–994854Search in Google Scholar

Isbary G, Shimizu T, Li YF, Stolz W, Thomas HM, Morfill GE, Zimmermann JL. Cold atmospheric plasma devices for medical issues. Expert Rev Med Devices. 2013; 10:367–77.IsbaryGShimizuTLiYFStolzWThomasHMMorfillGEZimmermannJLCold atmospheric plasma devices for medical issuesExpert Rev Med Devices20131036777Search in Google Scholar

Guerrero-Preston R, Ogawa T, Uemura M, Shumulinsky G, Valle BL, Pirini F, et al. Cold atmospheric plasma treatment selectively targets head and neck squamous cell carcinoma cells. Int J Mol Med. 2014; 34:941–6.Guerrero-PrestonROgawaTUemuraMShumulinskyGValleBLPiriniFCold atmospheric plasma treatment selectively targets head and neck squamous cell carcinoma cellsInt J Mol Med2014349416Search in Google Scholar

Walsh JL, Kong MG. Room-temperature atmospheric argon plasma jet sustained with sub-microsecond high-voltage pulses. Appl Phys Lett. 2007; 91:221502.WalshJLKongMGRoom-temperature atmospheric argon plasma jet sustained with sub-microsecond high-voltage pulsesAppl Phys Lett200791221502Search in Google Scholar

Robert E, Sarron V, Darny T, Riès D, Dozias S, Pouvesle JM, et al. Rare gas flow structuration in plasma jet experiments. Plasma Sources Sci Technol. 2014; 23:5.RobertESarronVDarnyTRièsDDoziasSPouvesleJMRare gas flow structuration in plasma jet experimentsPlasma Sources Sci Technol2014235Search in Google Scholar

Hosseinzadeh Colagar A, Memariani H, Sohbatzadeh F, Valinataj Omran A. Nonthermal atmospheric argon plasma jet effects on Escherichia coli biomacromolecules. Appl Biochem Biotechnol. 2013; 171:1617–29.Hosseinzadeh ColagarAMemarianiHSohbatzadehFValinataj OmranANonthermal atmospheric argon plasma jet effects on Escherichia coli biomacromoleculesAppl Biochem Biotechnol2013171161729Search in Google Scholar

Ladislav B, Baránková H. Cold atmospheric plasma: sources, processes, and applications. Thin Solid Films. 2010; 518:6705–13.LadislavBBaránkováHCold atmospheric plasma: sources, processes, and applicationsThin Solid Films2010518670513Search in Google Scholar

Keidar M, Shashurin A, Volotskova O, Ann Stepp M, Srinivasan P, Sandler A, Trink B. Cold atmospheric plasma in cancer therapy. Phys Plasmas. 2013; 20:1–8.KeidarMShashurinAVolotskovaOAnn SteppMSrinivasanPSandlerATrinkBCold atmospheric plasma in cancer therapyPhys Plasmas20132018Search in Google Scholar

Case A, Messer S, Brockington S, Wu L, Witherspoon FD, Elton R, Merging of high-speed argon plasma jets. Phys Plasmas. 2013; 20:012704.CaseAMesserSBrockingtonSWuLWitherspoonFDEltonRMerging of high-speed argon plasma jetsPhys Plasmas201320012704Search in Google Scholar

Sohbatzadeh F, Hossienzadeh Colagar A, Mirzanejhad S, Motallebi S, Farhadi M, Bagheri M, et al. Design and construction of triplet atmospheric cold plasma jet for sterilization. Int J Psychol Res. 2014; 13:363–74.SohbatzadehFHossienzadeh ColagarAMirzanejhadSMotallebiSFarhadiMBagheriMDesign and construction of triplet atmospheric cold plasma jet for sterilizationInt J Psychol Res20141336374Search in Google Scholar

Mirpour S, Ghomi H, Piroozmand S, Nikkhah M, Tavassoli SH, Azad SZ, The selective characterization of a non-thermal atmospheric pressure plasma jet on the treatment of human breast cancer and normal cells. IEEE T Plasma Sci. 2014; 2:315–22.MirpourSGhomiHPiroozmandSNikkhahMTavassoliSHAzadSZThe selective characterization of a non-thermal atmospheric pressure plasma jet on the treatment of human breast cancer and normal cellsIEEE T Plasma Sci.2014231522Search in Google Scholar

Sohbatzadeh F, Hosseinzadeh Colagar A, Mirzanejhad S. Hajiahmadi S, Talebzadeh M, Ghasemi M. Investigation of remote atmospheric pressure plasma effect on Candida albicans fungus. In: Proceedings of 30th ICPIG, 2011 August 28th – September 2nd, Belfast, Northern Ireland; 2011.SohbatzadehFHosseinzadeh ColagarAMirzanejhadSHajiahmadiSTalebzadehMGhasemiMInvestigation of remote atmospheric pressure plasma effect on Candida albicans fungusIn:Proceedings of 30th ICPIG2011 August 28th – September 2ndBelfast, Northern Ireland2011Search in Google Scholar