Rivista e Edizione

Volume 29 (2023): Edizione 1 (March 2023)

Volume 28 (2022): Edizione 4 (December 2022)

Volume 28 (2022): Edizione 3 (September 2022)

Volume 28 (2022): Edizione 2 (June 2022)

Volume 28 (2022): Edizione 1 (March 2022)

Volume 27 (2021): Edizione 4 (December 2021)

Volume 27 (2021): Edizione 3 (September 2021)

Volume 27 (2021): Edizione 2 (June 2021)

Volume 27 (2021): Edizione 1 (March 2021)

Volume 26 (2020): Edizione 4 (December 2020)

Volume 26 (2020): Edizione 3 (September 2020)

Volume 26 (2020): Edizione 2 (June 2020)

Volume 26 (2020): Edizione 1 (March 2020)

Volume 25 (2019): Edizione 4 (December 2019)

Volume 25 (2019): Edizione 3 (September 2019)

Volume 25 (2019): Edizione 2 (June 2019)

Volume 25 (2019): Edizione 1 (March 2019)

Volume 24 (2018): Edizione 4 (December 2018)

Volume 24 (2018): Edizione 3 (September 2018)

Volume 24 (2018): Edizione 2 (June 2018)

Volume 24 (2018): Edizione 1 (March 2018)

Volume 23 (2017): Edizione 4 (December 2017)

Volume 23 (2017): Edizione 3 (September 2017)

Volume 23 (2017): Edizione 2 (June 2017)

Volume 23 (2017): Edizione 1 (March 2017)

Volume 22 (2016): Edizione 4 (December 2016)

Volume 22 (2016): Edizione 3 (September 2016)

Volume 22 (2016): Edizione 2 (June 2016)

Volume 22 (2016): Edizione 1 (March 2016)

Volume 21 (2015): Edizione 1 (December 2015)

Volume 20 (2014): Edizione 1 (March 2014)

Volume 19 (2013): Edizione 2 (December 2013)

Volume 19 (2013): Edizione 1 (March 2013)

Volume 18 (2012): Edizione 2 (June 2012)

Volume 18 (2012): Edizione 1 (March 2012)

Volume 17 (2011): Edizione 4 (December 2011)

Volume 17 (2011): Edizione 3 (September 2011)

Volume 17 (2011): Edizione 2 (June 2011)

Volume 17 (2011): Edizione 1 (March 2011)

Volume 16 (2010): Edizione 2 (June 2010)

Volume 16 (2010): Edizione 1 (March 2010)

Volume 15 (2009): Edizione 4 (December 2009)

Volume 15 (2009): Edizione 3 (September 2009)

Volume 15 (2009): Edizione 2 (June 2009)

Volume 15 (2009): Edizione 1 (March 2009)

Volume 14 (2008): Edizione 4 (December 2008)

Volume 14 (2008): Edizione 3 (September 2008)

Volume 14 (2008): Edizione 2 (June 2008)

Volume 14 (2008): Edizione 1 (March 2008)

Volume 13 (2007): Edizione 4 (December 2007)

Volume 13 (2007): Edizione 3 (September 2007)

Volume 13 (2007): Edizione 2 (June 2007)

Volume 13 (2007): Edizione 1 (March 2007)

Dettagli della rivista
Formato
Rivista
eISSN
1898-0309
Pubblicato per la prima volta
30 Dec 2008
Periodo di pubblicazione
4 volte all'anno
Lingue
Inglese

Cerca

Volume 26 (2020): Edizione 1 (March 2020)

Dettagli della rivista
Formato
Rivista
eISSN
1898-0309
Pubblicato per la prima volta
30 Dec 2008
Periodo di pubblicazione
4 volte all'anno
Lingue
Inglese

Cerca

6 Articoli
Accesso libero

Multi spectral classification and recognition of breast cancer and pneumonia

Pubblicato online: 03 Apr 2020
Pagine: 1 - 9

Astratto

Abstract

According to the Google I/O 2018 key notes, in future artificial intelligence, which also includes machine learning and deep learning, will mostly evolve in healthcare domain. As there are lots of subdomains which come under the category of healthcare domain, the proposed paper concentrates on one such domain, that is breast cancer and pneumonia. Today, just classifying the diseases is not enough. The system should also be able to classify a particular patient’s disease. Thus, this paper shines the light on the importance of multi spectral classification which means the collection of several monochrome images of the same scene. It can be proved to be an important process in the healthcare areas to know if a patient is suffering from a specific disease or not. The convolutional layer followed by the pooling layer is used for the feature extraction process and for the classification process; fully connected layers followed by the regression layer are used.

Parole chiave

  • convolutional layer
  • pooling layer
  • multi spectral classification
  • breast cancer
  • pneumonia
Accesso libero

Design of a Monte Carlo model based on dual-source computed tomography (DSCT) scanners for dose and image quality assessment using the Monte Carlo N-Particle (MCNP5) code

Pubblicato online: 03 Apr 2020
Pagine: 11 - 20

Astratto

Abstract

The purpose of this work was to develop and validate a Monte Carlo model for a Dual Source Computed Tomography (DSCT) scanner based on the Monte Carlo N-particle radiation transport computer code (MCNP5). The geometry of the Siemens Somatom Definition CT scanner was modeled, taking into consideration the x-ray spectrum, bowtie filter, collimator, and detector system. The accuracy of the simulation from the dosimetry point of view was tested by calculating the Computed Tomography Dose Index (CTDI) values. Furthermore, typical quality assurance phantoms were modeled in order to assess the imaging aspects of the simulation. Simulated projection data were processed, using the MATLAB software, in order to reconstruct slices, using a Filtered Back Projection algorithm. CTDI, image noise, CT-number linearity, spatial and low contrast resolution were calculated using the simulated test phantoms. The results were compared using several published values including IMPACT, NIST and actual measurements. Bowtie filter shapes are in agreement with those theoretically expected. Results show that low contrast and spatial resolution are comparable with expected ones, taking into consideration the relatively limited number of events used for the simulation. The differences between simulated and nominal CT-number values were small. The present attempt to simulate a DSCT scanner could provide a powerful tool for dose assessment and support the training of clinical scientists in the imaging performance characteristics of Computed Tomography scanners.

Parole chiave

  • Monte Carlo simulation
  • dual-source CT
  • dual-energy CT
  • CTDI
  • image quality
Accesso libero

Investigation of imaging properties of novel contrast agents based on gold, silver and bismuth nanoparticles in spectral computed tomography using Monte Carlo simulation

Pubblicato online: 03 Apr 2020
Pagine: 21 - 29

Astratto

Abstract

In the present paper, some imaging properties of nanoparticles-based contrast agents including gold, bismuth, and silver were assessed and compared with conventional (iodinated) contrast agent in spectral computed tomography (CT). A spectral CT scanner with photon-counting detectors (PCD) and 6 energy bins was simulated using the Monte Carlo (MC) simulation method. The nanoparticles were designed with a diameter of 50 nm at concentrations of 2, 4, and 8 mg/ml. Water-filled cylindrical phantom was modeled with a diameter of 10 cm containing a hole with a diameter of 5 cm in its center, where was filled with contrast agents. The MC results were used to reconstruct images. Image reconstruction was accomplished with the filtered back-projection (FBP) method with hamming filter and linear interpolation method. CT number and contrast-to-noise ratio (CNR) of all studied contrast materials were calculated in spectral images. The simulations indicated that nanoparticle-based contrast agents have a higher CT number and CNR than the iodinated contrast agent at the same concentration and for all energy bins. In general, gold nanoparticles produced the highest CT number and CNR compared to silver and bismuth nanoparticles at the same concentration. However, at low energies (below 80 keV), silver nanoparticles performed similarly to gold nanoparticles and at high energies (120 keV), bismuth nanoparticles can be a good substitute for gold nanoparticles.

Parole chiave

  • spectral CT
  • photon-counting detector
  • contrast agent
  • Monte Carlo simulation
  • nanoparticles
Accesso libero

Conversion coefficients from fluence and air kerma to personal dose equivalent for monoenergetic photons using analytical fit and Monte Carlo simulation

Pubblicato online: 03 Apr 2020
Pagine: 31 - 44

Astratto

Abstract

The present study aims to calculate a new database of conversion coefficients from fluence and air Kerma to personal dose equivalent in two terms: absorbed dose and Kerma-approximations. In this work, we propose a new equation to perform an analytical fit of our Monte Carlo (MC) calculated conversion coefficients for photons for different angles. Also, we have calculated the conversion coefficients using the EGSnrc code. The conversion coefficients have been calculated for beams of monoenergetic photons from 0.015 to 10 MeV, incident on phantom ICRU for angles of incidence from 0° to a 75° in steps of 15°. Our computed values agree well when compared with those published for the ICRU 57 in Kerma-approximations with statistical uncertainties in the calculation around 2%. We can conclude from this work that the analytical approach is successful and powerful such as Monte Carlo methods to calculate the operational quantities.

Parole chiave

  • operational quantity
  • personal dose equivalent
  • conversion coefficients
  • fluence
  • air kerma
  • MC
Accesso libero

Occupational radiation exposures for medical workers in Pakistan – An overview

Pubblicato online: 03 Apr 2020
Pagine: 45 - 53

Astratto

Abstract

The imperative use of ionizing radiation in medicine causes the inevitable occupational exposure of the medical workers during the course of routine duties. The magnitude of health risk due to such radiation exposures has been described in terms of occupational radiation doses. In this context, it is obligatory to monitor, measure and document the radiation dose of occupationally exposed medical workers. This study aims to review the whole-body occupational radiation exposures of medical workers in Pakistan. Specifically, online literature published during 2000-2018 was reviewed for the occupational radiation exposures of Pakistani medical workers. Analysis of the extracted personal dosimetry data revealed that the total number of monitored medical occupational workers was 26046. The range of total cumulative and annual average effective doses was 94-15785 Person-mSv and 0.66-7.37 mSv, respectively. A significant number of the workers (25477; ~98%) received an annual dose below 5 mSv, while only 18 workers received an occupational exposure exceeding the annual dose limit of 20 mSv. It is expected that this study will provide a useful reference for evaluating and improving radiation protection and safety policies in the country.

Parole chiave

  • occupational exposure
  • radiation protection
  • effective dose
  • dose limits
Accesso libero

Evaluation of the effect of random setup errors on dose delivery in Intensity Modulated Radiotherapy

Pubblicato online: 03 Apr 2020
Pagine: 55 - 60

Astratto

Abstract

Aim: To conduct a study on the effect of random setup errors inpatient for dose delivery in Intensity Modulated Radiotherapy plans using Octavius 4D phantom.

Materials and methods: 11 patients with cancer of H&N were selected for this study. An IMRT plan was created for each patient. The IMRT quality assurance plans were transferred to Mosaiq workstation in a linear accelerator. These plans were delivered at the reference treatment position. Subsequently, the QA plans were delivered on the Octavius 4D phantom after introducing errors in various translational and rotational directions. The setup inaccuracies introduced varied from 1 mm to 5 mm along X, Y. These setup uncertainties were then introduced along X and Y direction simultaneously in equal measures. Similarly, IMRT plans were delivered also after introducing roll and yaw rotation of 1, 2 and 3 degrees in phantom. The deviation of gamma indices at all these positions was analyzed with respect to the reference setup position.

Results: The percentage of points passing the gamma acceptance criterion decrease as we increase the setup error. The change is found to be very insignificant with setup error up to 2 mm along X, Y or XY direction. Similarly, the rotational error of up to 3 degrees is found to be acceptable.

Conclusions: Small setup (< 2 mm) correction in patients may not adversely affect the dose delivery. But an error of similar magnitude in 2 directions simultaneously has a much greater impact on IMRT dose delivery.

Parole chiave

  • gamma indices
  • random setup error
  • IMRT
  • quality assurance
  • dose delivery
6 Articoli
Accesso libero

Multi spectral classification and recognition of breast cancer and pneumonia

Pubblicato online: 03 Apr 2020
Pagine: 1 - 9

Astratto

Abstract

According to the Google I/O 2018 key notes, in future artificial intelligence, which also includes machine learning and deep learning, will mostly evolve in healthcare domain. As there are lots of subdomains which come under the category of healthcare domain, the proposed paper concentrates on one such domain, that is breast cancer and pneumonia. Today, just classifying the diseases is not enough. The system should also be able to classify a particular patient’s disease. Thus, this paper shines the light on the importance of multi spectral classification which means the collection of several monochrome images of the same scene. It can be proved to be an important process in the healthcare areas to know if a patient is suffering from a specific disease or not. The convolutional layer followed by the pooling layer is used for the feature extraction process and for the classification process; fully connected layers followed by the regression layer are used.

Parole chiave

  • convolutional layer
  • pooling layer
  • multi spectral classification
  • breast cancer
  • pneumonia
Accesso libero

Design of a Monte Carlo model based on dual-source computed tomography (DSCT) scanners for dose and image quality assessment using the Monte Carlo N-Particle (MCNP5) code

Pubblicato online: 03 Apr 2020
Pagine: 11 - 20

Astratto

Abstract

The purpose of this work was to develop and validate a Monte Carlo model for a Dual Source Computed Tomography (DSCT) scanner based on the Monte Carlo N-particle radiation transport computer code (MCNP5). The geometry of the Siemens Somatom Definition CT scanner was modeled, taking into consideration the x-ray spectrum, bowtie filter, collimator, and detector system. The accuracy of the simulation from the dosimetry point of view was tested by calculating the Computed Tomography Dose Index (CTDI) values. Furthermore, typical quality assurance phantoms were modeled in order to assess the imaging aspects of the simulation. Simulated projection data were processed, using the MATLAB software, in order to reconstruct slices, using a Filtered Back Projection algorithm. CTDI, image noise, CT-number linearity, spatial and low contrast resolution were calculated using the simulated test phantoms. The results were compared using several published values including IMPACT, NIST and actual measurements. Bowtie filter shapes are in agreement with those theoretically expected. Results show that low contrast and spatial resolution are comparable with expected ones, taking into consideration the relatively limited number of events used for the simulation. The differences between simulated and nominal CT-number values were small. The present attempt to simulate a DSCT scanner could provide a powerful tool for dose assessment and support the training of clinical scientists in the imaging performance characteristics of Computed Tomography scanners.

Parole chiave

  • Monte Carlo simulation
  • dual-source CT
  • dual-energy CT
  • CTDI
  • image quality
Accesso libero

Investigation of imaging properties of novel contrast agents based on gold, silver and bismuth nanoparticles in spectral computed tomography using Monte Carlo simulation

Pubblicato online: 03 Apr 2020
Pagine: 21 - 29

Astratto

Abstract

In the present paper, some imaging properties of nanoparticles-based contrast agents including gold, bismuth, and silver were assessed and compared with conventional (iodinated) contrast agent in spectral computed tomography (CT). A spectral CT scanner with photon-counting detectors (PCD) and 6 energy bins was simulated using the Monte Carlo (MC) simulation method. The nanoparticles were designed with a diameter of 50 nm at concentrations of 2, 4, and 8 mg/ml. Water-filled cylindrical phantom was modeled with a diameter of 10 cm containing a hole with a diameter of 5 cm in its center, where was filled with contrast agents. The MC results were used to reconstruct images. Image reconstruction was accomplished with the filtered back-projection (FBP) method with hamming filter and linear interpolation method. CT number and contrast-to-noise ratio (CNR) of all studied contrast materials were calculated in spectral images. The simulations indicated that nanoparticle-based contrast agents have a higher CT number and CNR than the iodinated contrast agent at the same concentration and for all energy bins. In general, gold nanoparticles produced the highest CT number and CNR compared to silver and bismuth nanoparticles at the same concentration. However, at low energies (below 80 keV), silver nanoparticles performed similarly to gold nanoparticles and at high energies (120 keV), bismuth nanoparticles can be a good substitute for gold nanoparticles.

Parole chiave

  • spectral CT
  • photon-counting detector
  • contrast agent
  • Monte Carlo simulation
  • nanoparticles
Accesso libero

Conversion coefficients from fluence and air kerma to personal dose equivalent for monoenergetic photons using analytical fit and Monte Carlo simulation

Pubblicato online: 03 Apr 2020
Pagine: 31 - 44

Astratto

Abstract

The present study aims to calculate a new database of conversion coefficients from fluence and air Kerma to personal dose equivalent in two terms: absorbed dose and Kerma-approximations. In this work, we propose a new equation to perform an analytical fit of our Monte Carlo (MC) calculated conversion coefficients for photons for different angles. Also, we have calculated the conversion coefficients using the EGSnrc code. The conversion coefficients have been calculated for beams of monoenergetic photons from 0.015 to 10 MeV, incident on phantom ICRU for angles of incidence from 0° to a 75° in steps of 15°. Our computed values agree well when compared with those published for the ICRU 57 in Kerma-approximations with statistical uncertainties in the calculation around 2%. We can conclude from this work that the analytical approach is successful and powerful such as Monte Carlo methods to calculate the operational quantities.

Parole chiave

  • operational quantity
  • personal dose equivalent
  • conversion coefficients
  • fluence
  • air kerma
  • MC
Accesso libero

Occupational radiation exposures for medical workers in Pakistan – An overview

Pubblicato online: 03 Apr 2020
Pagine: 45 - 53

Astratto

Abstract

The imperative use of ionizing radiation in medicine causes the inevitable occupational exposure of the medical workers during the course of routine duties. The magnitude of health risk due to such radiation exposures has been described in terms of occupational radiation doses. In this context, it is obligatory to monitor, measure and document the radiation dose of occupationally exposed medical workers. This study aims to review the whole-body occupational radiation exposures of medical workers in Pakistan. Specifically, online literature published during 2000-2018 was reviewed for the occupational radiation exposures of Pakistani medical workers. Analysis of the extracted personal dosimetry data revealed that the total number of monitored medical occupational workers was 26046. The range of total cumulative and annual average effective doses was 94-15785 Person-mSv and 0.66-7.37 mSv, respectively. A significant number of the workers (25477; ~98%) received an annual dose below 5 mSv, while only 18 workers received an occupational exposure exceeding the annual dose limit of 20 mSv. It is expected that this study will provide a useful reference for evaluating and improving radiation protection and safety policies in the country.

Parole chiave

  • occupational exposure
  • radiation protection
  • effective dose
  • dose limits
Accesso libero

Evaluation of the effect of random setup errors on dose delivery in Intensity Modulated Radiotherapy

Pubblicato online: 03 Apr 2020
Pagine: 55 - 60

Astratto

Abstract

Aim: To conduct a study on the effect of random setup errors inpatient for dose delivery in Intensity Modulated Radiotherapy plans using Octavius 4D phantom.

Materials and methods: 11 patients with cancer of H&N were selected for this study. An IMRT plan was created for each patient. The IMRT quality assurance plans were transferred to Mosaiq workstation in a linear accelerator. These plans were delivered at the reference treatment position. Subsequently, the QA plans were delivered on the Octavius 4D phantom after introducing errors in various translational and rotational directions. The setup inaccuracies introduced varied from 1 mm to 5 mm along X, Y. These setup uncertainties were then introduced along X and Y direction simultaneously in equal measures. Similarly, IMRT plans were delivered also after introducing roll and yaw rotation of 1, 2 and 3 degrees in phantom. The deviation of gamma indices at all these positions was analyzed with respect to the reference setup position.

Results: The percentage of points passing the gamma acceptance criterion decrease as we increase the setup error. The change is found to be very insignificant with setup error up to 2 mm along X, Y or XY direction. Similarly, the rotational error of up to 3 degrees is found to be acceptable.

Conclusions: Small setup (< 2 mm) correction in patients may not adversely affect the dose delivery. But an error of similar magnitude in 2 directions simultaneously has a much greater impact on IMRT dose delivery.

Parole chiave

  • gamma indices
  • random setup error
  • IMRT
  • quality assurance
  • dose delivery

Pianifica la tua conferenza remota con Sciendo