An Experimental Setup for Power Loss Measurement up to 1 kHz using an Epstein Frame at CMI
17 nov 2023
INFORMAZIONI SU QUESTO ARTICOLO
Pubblicato online: 17 nov 2023
Pagine: 275 - 280
Ricevuto: 13 lug 2023
Accettato: 27 ott 2023
DOI: https://doi.org/10.2478/msr-2023-0035
Parole chiave
© 2023 Michal Ulvr, published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
This paper describes an experimental setup used at the Czech Metrology Institute (CMI) to measure the specific power loss of oriented and non-oriented electrical steel sheets up to 1 kHz using an Epstein frame. Special attention is given to a) a description of the hardware that is used, b) a description of the feedback control and measurement software, and c) an analysis of the sources of uncertainty and validation. Calibration expanded uncertainty of (0.5 up to 1.6)% for k = 2 can be achieved with this setup.