Accesso libero

An Experimental Setup for Power Loss Measurement up to 1 kHz using an Epstein Frame at CMI

  
17 nov 2023
INFORMAZIONI SU QUESTO ARTICOLO

Cita
Scarica la copertina

This paper describes an experimental setup used at the Czech Metrology Institute (CMI) to measure the specific power loss of oriented and non-oriented electrical steel sheets up to 1 kHz using an Epstein frame. Special attention is given to a) a description of the hardware that is used, b) a description of the feedback control and measurement software, and c) an analysis of the sources of uncertainty and validation. Calibration expanded uncertainty of (0.5 up to 1.6)% for k = 2 can be achieved with this setup.

Lingua:
Inglese
Frequenza di pubblicazione:
6 volte all'anno
Argomenti della rivista:
Ingegneria, Elettrotecnica, Ingegneria dell'automazione, metrologia e collaudo