Bioelectrical impedance changes of the trunk are opposite the limbs following acute hydration change

Pubblicato online: 25 Jun 2022
Volume & Edizione: Volume 13 (2022) - Edizione 1 (January 2022)
Pagine: 25 - 30
Ricevuto: 26 Apr 2022
Dettagli della rivista
License
Formato
Rivista
eISSN
1891-5469
Prima pubblicazione
01 Jan 2010
Frequenza di pubblicazione
1 volta all'anno
Lingue
Inglese
Introduction

The use of bioelectrical impedance analysis (BIA) to estimate human body composition parameters such as total body water (TBW), fat-free mass (FFM), and body fat percentage (%BF) dates back to the mid-1980s [1, 2, 3, 4]. The principles, assumptions, and clinical applications of the BIA method were thoroughly reviewed by Kushner [5]. These early studies used a single-frequency (50 kHz) BIA device with gel electrodes placed on the hands and feet in a tetrapolar configuration with the study participant lying in a supine position. Soon after, researchers began experimenting with segmental BIA, evaluating the limbs independently of the trunk [6], and multi-frequency BIA to distinguish between intracellular and extracellular water [7]. By the end of the 20th century, user-friendly hand-to-hand (hand-held) [8] and foot-to-foot (scale) [9] BIA devices with plate or contact electrodes were available for home use. Today, these BIA variations have merged such that user-friendly segmental multi-frequency BIA (SMFBIA) devices are commonplace in research laboratories, clinical settings, and fitness centers.

SMFBIA was validated against isotope dilution for TBW measurement [10,11], and against DXA [12,13] and a 4-compartment model [10] for FFM. The SMFBIA method is quick, easy to administer, non-invasive, and less expensive than many other body composition methods. For these reasons, SMFBIA is an appealing assessment method. Despite these advantages, numerous factors can affect the validity of this method. For example, accurate body positioning and electrode placement are critical for valid measurements [14,15]. Another variable known to affect the impedance of electrical current, and consequently the body composition measures derived from SMFBIA, is hydration status [16].

A substantial volume of research exists on the effect hydration status and hydration change has on bioimpedance measures; however, some questions remain. For example, O’Brien et al. [16] reviewed a number of hydration change studies that used single-frequency BIA and SMFBIA, and they found equivocal results with resistance to the electrical current both increasing and decreasing in response to acute hydration change. More importantly, all of the previous SMFBIA research on this topic has been limited to total body results; segmental impedance values following acute hydration change have not been reported. This segmental data could help elucidate the confounding question of what happens to impedance values following an acute change in hydration status. Thus, the purpose of this study was to evaluate the change in SMFBIA-estimated body composition variables (TBW, FFM, and %BF) as well as segmental impedance values following both acute dehydration and acute hyperhydration.

Materials and methods

The SMFBIA data collection was a sub-investigation of the acute effects of hyperhydration and dehydration on A-mode ultrasound measures of subcutaneous fat thickness [17]; thus, the methods presented here are similar to those presented previously [17] with the exception that SMFBIA information is included, and ultrasound information is excluded.

Sample

The study was open to adults, aged 18-65 y, and recruitment was via word-of-mouth within a university exercise science department and an advertisement posted at the university’s fitness center. Exclusion criteria included pregnancy, a self-reported inability to be confined in a hot environment or to drink a large volume of lightly-salted water, current use of an NSAID, or prior history of hyponatremia.

Procedures

Participants came to the laboratory on two consecutive days. They followed a pre-session euhydration protocol of drinking 10 ml∙kg-1 water approximately 2 h before arriving at the lab, maintained their typical daily routine, and refrained from exhaustive exercise prior to each trial. Upon arrival at the lab, participants emptied their bladder and provided a urine sample. Urine specific gravity (USG) was measured with a refractometer (Uricon-N, Atago, Tokyo, Japan) to verify euhydration status. A USG of < 1.020 was the criterion for euhydration [18]. Height was measured to the nearest 0.1 cm with a custom-made, wall-mounted stadiometer. Body mass was measured to the nearest 0.02 kg with a digital scale (DI/10, Wedderburn Scales Ltd., Dunedin, New Zealand). Male participants wore shorts only, while female participants wore shorts and a sports bra for all measurements.

SMFBIA measures were taken with an InBody 230 analyzer (Seoul, South Korea) following the manufacturer’s instructions for body position [19]. This included standing erect with the feet positioned on the foot electrodes and the arms slightly abducted with the hands grasping the electrodes on the handrails. This SMFBIA model provided impedance for five segments (right and left arms, right and left legs, and trunk) at two frequencies (20 kHz and 100 kHz). TBW, FFM, and %BF were estimated from the manufacturer’s proprietary formulas.

Participants were randomly assigned to dehydration or hyperhydration, with the opposite treatment applied during the subsequent trial. Dehydration took place in an environmental chamber with the temperature and relative humidity held constant at 40 °C and 60%, respectively. Participants had access to a cycle ergometer and treadmill, and exercise was ad libitum. A scale was available for participants to self-monitor changes in mass. Participants remained in the chamber until approximately 2% of body mass was lost or 2 h elapsed, whichever occurred first. Post-dehydration measurements occurred 30 min after exiting the chamber.

The hyperhydration protocol required participants to drink 2% body mass of lightly-salted water (30 mmol∙L-1 NaCl; 1.76 g NaCl∙L-1). They were encouraged to finish this volume within 30 min. Once consumed, they waited an additional 30 min before beginning the post-treatment measurements.

Participants emptied their bladder, and body mass was measured. The change in mass (Δmass) between the baseline and this post-treatment measurement confirmed and quantified dehydration and hyperhydration. SMFBIA measurements were repeated following the same standards described for baseline testing.

Statistical analyses

Data were checked for normality by visual inspection of plots and the Shapiro-Wilk test. Mean differences in body mass, TBW, FFM, %BF, and impedance for each body segment at both 20 kHz and 100 kHz between the two baseline measurements and the dehydration and hyperhydration post-treatments were evaluated using repeated-measures ANOVA. If the assumption of sphericity was violated, the Greenhouse-Geisser correction was used. When the F-score was significant, pairwise comparisons were made using the least significant difference. Statistical significance was accepted as p < 0.05. Effect sizes of the pairwise differences were reported as Cohen’s d with the benchmarks of 0.2, 0.5, and 0.8 defining small, medium, and large effects, respectively [20]. All statistical analyses were performed with SPSS version 25 (IBM, Armonk, NY).

Informed consent

Informed consent has been obtained from all individuals included in this study.

Ethical approval

The research related to human use has been complied with all relevant national regulations, institutional policies and in accordance with the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.

Results

Eleven recreationally-active adults (8 male, 3 female), ranging in age (19 to 54 y; 27.1 ± 10.5 y), height (156.5 to 188.8 cm; 176.2 ± 11.5 cm), mass (49.2 to 120.3 kg; 78.0 ± 20.6 kg), and body mass index (20.3 to 33.5 kg∙m-2; 24.7 ± 3.9 kg∙m-2) completed the study. Measured data from the dehydration trial are in Table 1, and data from the hyperhydration trial are in Table 2. All participants began each trial with USG < 1.020. The difference in USG was not significant between trials (Δ = 0.0035, 95%CI = -0.0074 to 0.0005, p = 0.081), suggesting similar euhydration at the start of each session. Body mass was similar at the start of each trial (Δ = 0.03 kg, 95%CI = -0.60 to 0.66 kg, p = 1.000) but decreased significantly post-dehydration (Δ = -2.0% or -1.56 kg, 95%CI = -2.13 to -0.99 kg, p < 0.001) and increased significantly post-hyperhydration (Δ = 1.2% or 0.90 kg, 95%CI = 0.68 to 1.11 kg, p < 0.001). Six of the 11 subjects stayed in the heat chamber for the full 2 h, while the others achieved the desired mass loss in less time. The average time to dehydrate 2% was 99 ± 28 min. The average amount of water consumed during the hyperhydration trial was 1,560 ± 413 ml.

Mean ± SD of the sample (N = 11) for the dehydration trial.

Variable Baseline Post-Dehydration Change from baseline (%) Pairwise p-value Cohen's d effect size
Body mass (kg) 78.0 ± 20.4 765 ± 20.2 -2.0±0.6 <0.001 2,71
TBW (kg) 48.9 ± 12.8 49.2 ± 12.5 0.9 ± 2.5 0.374 0.28
FFM (kg) 66.9 ± 17.7 67.1 ± 17.1 0.6 ± 2.5 0.617 0.16
Body fat (%) 14.1 ± 7.3 11.6 ± 8.1 -24.6 ±23.6 0.004 1.11
Z RA 20 kHz (Ω) 306.2 ± 60.4 281.4 ± 48.4 -7.7 ± 4.6 0.001 1.37
Z LA 20 kHz (Ω) 309.5 ±62.6 282.9 ± 50.1 -8.2 ± 4.3 <0.001 1.54
Z RL 20 kHz (Ω) 251.1 ± 33.6 232.1 ± 31.0 -7.5 ± 4.2 <0.001 1.82
Z LL 20 kHz (Ω) 250.6 ±33.8 230.4 ±29.1 -7.9 ± 4.4 < 0.001 1.70
ZTR 20 kHz (Ω) 21.1 ± 3.0 22.5 ± 3.1 6.9 ± 8.8 0.029 0.77
Z RA 100 kHz (Ω) 270.3 ± 57.3 246.8 ±46.1 -8.2 ± 4.9 0.001 1.37
Z LA 10D kHz (Ω) 273.8 ± 59.7 249.4 ± 47.5 -8.4 ± 4.9 0.001 1.42
Z RL 100 kHz(Ω) 219.0 ±31.7 202.7 ±28.8 -7.3 ± 4.3 <0.001 1.75
Z LL 100 kHz (Ω) 218.1 ± 31.7 200.9 ± 27.5 -7.7 ± 4.3 <0.001 1,71
Z TR 100 kHz (Ω) 17.1 ± 2.9 18.2 ± 3.1 6.6 ± 8.1 0.023 0.81

TBW = total body water; FFM = fat-free mass; Z = impedance; RA = right arm; LA = left arm; RL = right leg; LL = left leg; TR = trunk

Mean ± SD of the sample (N = 11) for the hyperhydration trial.

Variable Baseline Post-Hyperhydration Change from baseline (%) Pairwise p-value Cohen's d effect size
Body mass (kg) 78.0 ± 20.6 78.9 ± 20.8 1.2 ± 0.2 < 0.001 4.18
TBW (kg) 48,8 ± 12.7 49.0 ± 13.0 0.3 ± 09 0.240 0.38
FFM (kg) 66.7 ± 17.4 67.0 ± 17.9 0.3 ± 0.9 0.242 0.38
Body fat (%) 13.8 ± 8.4 14.7 ± 8.1 10.8 ± 15.6 0.024 0.80
Z RA 20 kHz (Ω) 309.6 ± 58.3 314.9 ± 64.7 1.4 ± 2.7 0.079 0.59
Z LA 20 kHz (Ω) 314,0 ± 60.1 318.3 ± 65.2 1.2 ± 2.7 0.149 0.47
Z RL 20 kHz (Ω) 252.8 ± 35.1 257.4 ± 43.3 1.5 ± 3.8 0.168 0.45
Z LL 20 kHz (Ω) 251.4 ± 35.5 255.8 ± 44.0 1.4 ± 4.0 0.195 0.42
Z TR 20kHz(Ω) 21.2 ± 3.3 20.9 ± 3.1 -1.3 ± 2.7 0.099 0.55
Z RA 100 kHz (Ω) 272.2 ± 55.6 276.5 ± 60.8 1.3 ± 2.8 0.100 0.55
Z LA 100 kHz (Ω) 276.8 ± 57.3 280.3 ± 61.5 1.0 ± 2.7 0.179 0.44
Z RL 100 kHz (Ω) 220.5 ± 32.8 224.5 ± 39.2 1.5 ± 3.6 0.141 0.48
Z LL 100 kHz (Ω) 218.5 ± 32.9 222.5 ± 39.5 1.5 ± 3.5 0.131 0.50
Z TR 100 kHz (Ω) 16.9 ± 3.1 16.7 ± 3.1 -1.5 ± 3.2 0.154 0.47

TBW = total body water; FFM = fat-free mass; Z = Impedance; RA = right arm; LA = left arm; RL = right leg; LL = left leg; TR = trunk

SMFBIA estimates of TBW did not change significantly across trials (p = 0.507) despite changes of approximately ±2% of body mass following dehydration and hyper-hydration. Correspondingly, estimates of FFM did not change significantly across trials (p = 0.796). SMFBIA failed to detect any loss in TBW in 8 of the 11 subjects following the dehydration protocol; in fact, the TBW estimate increased in these subjects (Figure 1a). Similarly, SMFBIA TBW estimate failed to increase more than 0.1 kg in 8 of the 11 subjects following the hyperhydration trial (Figure 1b). Consequently, the estimates of %BF from SMFBIA differed significantly across the 4 measurements (p < 0.001). The %BF post-dehydration was significantly lower than the other measurements, which did not differ significantly among each other.

Fig.1

Changes from baseline in body mass (solid bars) and SMFBIA-estimated total body water (hashed bars) following A) dehydration and B) hyperhydration.

Consistent with the laws of electrical conduction for a capacitive impedance system (negative phase angle) [5], the higher frequency (100 kHz) resulted in less impedance than the lower frequency (20 kHz) for a given body segment, and the arms, with their smaller cross-sectional area, created the most impedance followed by the legs and trunk (Tables 1 and 2). Baseline impedance values were consistent from day-today (p > 0.05). Dehydration resulted in a significant decrease (p < 0.01) of about 8% in limb impedance at both 20 kHz and 100 kHz.

Although there was a tendency for impedance to increase in the limbs following hyperhydration, this did not reach statistical significance (p > 0.05). Impedance changes in the trunk followed an opposite pattern of the limbs with a significant increase from dehydration (p < 0.05) and a nonsignificant (p > 0.05) trend to decrease following hyperhydration.

Discussion

The primary finding from this investigation was that the change in TBW estimated from SMFBIA did not match, or even approach, the change in body mass following dehydration or hyperhydration. This finding was expected as others reported that BIA and SMFBIA were not accurate at tracking acute hydration changes [21, 22, 23, 24, 25]. Further, changes in impedance were more pronounced following dehydration than hyperhydration in the present study. One possible reason for this was the greater change in mass following the dehydration trial compared to the hyperhydration trial. Participants changed by approximately 2% of their mass in both trials; however, there was a 30-minute equilibration period following hyperhydration after which participants urinated, thereby reducing the total mass change.

Theoretically, impedance should increase when one is in a state of hypohydration relative to euhydration because there is less fluid to conduct the current. The opposite is true of hyperhydration. However, a review of the literature indicates that the impedance change is often in the direction opposite of expectations following acute dehydration or hyperhydration [16]. This is because the water-electrolyte balance typically is not maintained following acute hydration change, and the electrolyte concentration might have a greater effect on BIA than the change in TBW [16]. Further, physiological changes that typically coincide with exercise-induced dehydration, such as sweating, increased skin temperature, and blood flow redistribution, may transiently alter impedance independent of TBW changes [16].

The most unexpected and novel finding from this study was truncal impedance changes were similar in magnitude (percent change) yet opposite in direction from limb impedance changes in response to hydration changes. Following dehydration, truncal impedance increased, matching the theoretical expectation that a decrease in fluid increases resistance to electrical current flow. However, limb impedance, at both low and high frequencies, consistently and significantly decreased in response to dehydration, suggestive of influence by a high electrolyte concentration. Although the magnitude of change was not great (and not statistically significant) following the hyperhydration trial, truncal impedance decreased consistent with expected fluid change while limb impedance increased consistent with electrolyte change. Thus, from this observation, it appears that fluid or TBW changes influence the trunk while ion concentration influences the limbs following acute hydration change.

The author can only speculate on the reason for the divergence of impedance change between the trunk and limbs in response to acute hydration change. First, it is important to recognize that acute TBW changes are not consistent throughout the body [26,27]. For example, heat exposure shifts water to the extracellular component to facilitate sweating [21,28]. Second, sweat gland density varies widely from one area of the body to another [29], and not all sweat glands are recruited simultaneously [30]. Third, heat and exercise increase skin temperature, which consistently decreases impedance [15,16,21,31]. Fourth, it might take more than 30 min for skin temperature to return to neutral following heat exposure [32]. Taken together, these factors may help explain why truncal and limb impedance values move in opposite directions following acute hydration change. Post-dehydration measurements were made at 30 min of exiting the environmental chamber, and skin temperature at the SMFBIA contact points (hands and feet) could have still been elevated following the heat/dehydration trial, leading to decreased limb impedance. Conversely, following the hyperhydration trial, a majority of the water would have emptied from the gut within 30 min [33], but much of it would have still been in the intestines, thereby increasing the fluid volume of the trunk and decreasing truncal impedance.

Results from the present study suggest that hyperhydration may have less of an influence on SMFBIA impedance values than dehydration. The magnitude of impedance change post-dehydration was statistically significant and greater than post-hyperhydration (not statistically significant). Similarly, other investigators found that excess hydration had either no significant influence on impedance or increased impedance only slightly from euhydration [14,23,34]. Fluid volume is likely a critical factor for inducing an impedance change [23,24]. Fluid intake of ≤ 1,000 mL appears to have negligible impact on impedance [14,34]. Others reported statistically significant changes in BIA-measured body composition variables following acute hydration of 500-750 mL, but they determined that these changes were not clinically meaningful [35,36]. The fluid intake in the present study was substantial, averaging > 1,500 mL; nevertheless, the impedance change remained negligible at ≤ 1.5% at all body segments.

This study has limitations and strengths. Additional measures, such as plasma osmolality and skin temperature, potentially could have helped explain the mechanism underlying why the trunk and limbs responded differently to changes in hydration status. Also, the post-treatment SMFBIA measurement was taken at a single time point (approximately 30 min after either dehydration or hyperhydration). Tracking impedance change over time (e.g., immediate post-treatment, 30 min, 60 min, 90 min, and 120 min) could have aided in determining the optimal time to conduct SMFBIA measures following a rapid alteration in hydration status. Despite these limitations, this is the first paper to identify a contrary impedance response between the limbs and trunk following acute hydration change. Another strength of the study was the inclusion of both a dehydration and hyperhydration trial, as previous investigators of BIA hydration status studies evaluated one or the other, or rehydrated the participants immediately after dehydration. Further, the methods to induce hyperhydration (e.g., fluid volume relative to mass) and dehydration (e.g., temperature- and humidity-controlled chamber) were tightly controlled.

Conclusion

In conclusion, TBW estimates from SMFBIA failed to match the body mass changes that accompanied acute dehydration and hyperhydration. Further, impedance values at the limbs moved in opposite direction to impedance values at the trunk at both 20 kHz and 100 kHz following acute dehydration. Impedance at the trunk appears to be influenced by fluid volume, while impedance at the limbs appears to be influenced by ion concentration; however, this is speculation, and more research is needed to determine the mechanistic reason for the divergence in SMFBIA impedance values between the trunk and limbs following acute hydration change.

Fig.1

Changes from baseline in body mass (solid bars) and SMFBIA-estimated total body water (hashed bars) following A) dehydration and B) hyperhydration.
Changes from baseline in body mass (solid bars) and SMFBIA-estimated total body water (hashed bars) following A) dehydration and B) hyperhydration.

Mean ± SD of the sample (N = 11) for the dehydration trial.

Variable Baseline Post-Dehydration Change from baseline (%) Pairwise p-value Cohen's d effect size
Body mass (kg) 78.0 ± 20.4 765 ± 20.2 -2.0±0.6 <0.001 2,71
TBW (kg) 48.9 ± 12.8 49.2 ± 12.5 0.9 ± 2.5 0.374 0.28
FFM (kg) 66.9 ± 17.7 67.1 ± 17.1 0.6 ± 2.5 0.617 0.16
Body fat (%) 14.1 ± 7.3 11.6 ± 8.1 -24.6 ±23.6 0.004 1.11
Z RA 20 kHz (Ω) 306.2 ± 60.4 281.4 ± 48.4 -7.7 ± 4.6 0.001 1.37
Z LA 20 kHz (Ω) 309.5 ±62.6 282.9 ± 50.1 -8.2 ± 4.3 <0.001 1.54
Z RL 20 kHz (Ω) 251.1 ± 33.6 232.1 ± 31.0 -7.5 ± 4.2 <0.001 1.82
Z LL 20 kHz (Ω) 250.6 ±33.8 230.4 ±29.1 -7.9 ± 4.4 < 0.001 1.70
ZTR 20 kHz (Ω) 21.1 ± 3.0 22.5 ± 3.1 6.9 ± 8.8 0.029 0.77
Z RA 100 kHz (Ω) 270.3 ± 57.3 246.8 ±46.1 -8.2 ± 4.9 0.001 1.37
Z LA 10D kHz (Ω) 273.8 ± 59.7 249.4 ± 47.5 -8.4 ± 4.9 0.001 1.42
Z RL 100 kHz(Ω) 219.0 ±31.7 202.7 ±28.8 -7.3 ± 4.3 <0.001 1.75
Z LL 100 kHz (Ω) 218.1 ± 31.7 200.9 ± 27.5 -7.7 ± 4.3 <0.001 1,71
Z TR 100 kHz (Ω) 17.1 ± 2.9 18.2 ± 3.1 6.6 ± 8.1 0.023 0.81

Mean ± SD of the sample (N = 11) for the hyperhydration trial.

Variable Baseline Post-Hyperhydration Change from baseline (%) Pairwise p-value Cohen's d effect size
Body mass (kg) 78.0 ± 20.6 78.9 ± 20.8 1.2 ± 0.2 < 0.001 4.18
TBW (kg) 48,8 ± 12.7 49.0 ± 13.0 0.3 ± 09 0.240 0.38
FFM (kg) 66.7 ± 17.4 67.0 ± 17.9 0.3 ± 0.9 0.242 0.38
Body fat (%) 13.8 ± 8.4 14.7 ± 8.1 10.8 ± 15.6 0.024 0.80
Z RA 20 kHz (Ω) 309.6 ± 58.3 314.9 ± 64.7 1.4 ± 2.7 0.079 0.59
Z LA 20 kHz (Ω) 314,0 ± 60.1 318.3 ± 65.2 1.2 ± 2.7 0.149 0.47
Z RL 20 kHz (Ω) 252.8 ± 35.1 257.4 ± 43.3 1.5 ± 3.8 0.168 0.45
Z LL 20 kHz (Ω) 251.4 ± 35.5 255.8 ± 44.0 1.4 ± 4.0 0.195 0.42
Z TR 20kHz(Ω) 21.2 ± 3.3 20.9 ± 3.1 -1.3 ± 2.7 0.099 0.55
Z RA 100 kHz (Ω) 272.2 ± 55.6 276.5 ± 60.8 1.3 ± 2.8 0.100 0.55
Z LA 100 kHz (Ω) 276.8 ± 57.3 280.3 ± 61.5 1.0 ± 2.7 0.179 0.44
Z RL 100 kHz (Ω) 220.5 ± 32.8 224.5 ± 39.2 1.5 ± 3.6 0.141 0.48
Z LL 100 kHz (Ω) 218.5 ± 32.9 222.5 ± 39.5 1.5 ± 3.5 0.131 0.50
Z TR 100 kHz (Ω) 16.9 ± 3.1 16.7 ± 3.1 -1.5 ± 3.2 0.154 0.47

Kushner RF, Schoeller DA. Estimation of total body water by bioelectrical impedance analysis. Am J Clin Nutr 1986;44:41724. https://doi.org/10.1093/ajcn/44.3.417Kushner RF Schoeller DA Estimation of total body water by bioelectrical impedance analysis Am J Clin Nutr 19864441724 https://doi.org/10.1093/ajcn/44.3.4173529918Apri DOISearch in Google Scholar

Lukaski HC, Bolonchuk WW, Hall CB, Siders WA. Validation of tetrapolar bioelectrical impedance method to assess human body composition. J Appl Physiol 1986;60:1327-32. https://doi.org/10.1152/jappl.1986.60.4.1327Lukaski HC Bolonchuk WW Hall CB Siders WA Validation of tetrapolar bioelectrical impedance method to assess human body composition J Appl Physiol 1986601327 32 https://doi.org/10.1152/jappl.1986.60.4.13273700310Apri DOISearch in Google Scholar

Lukaski HC, Johnson PE, Bolonchuk WW, Lykken GI. Assessment of fat-free mass using bioelectrical impedance measurements of the human body. Am J Clin Nutr 1985;41:810-7. https://doi.org/10.1093/ajcn/41.4.810Lukaski HC Johnson PE Bolonchuk WW Lykken GI Assessment of fat-free mass using bioelectrical impedance measurements of the human body Am J Clin Nutr 198541810 7 https://doi.org/10.1093/ajcn/41.4.8103984933Apri DOISearch in Google Scholar

Segal KR, Gutin B, Presta E, Wang J, Van Itallie TB. Estimation of human body composition by bioelectrical impedance methods: a comparative study. J Appl Physiol 1985;58:156571. https://doi.org/10.1152/jappl.1985.58.5.1565Segal KR Gutin B Presta E Wang J Van Itallie TB Estimation of human body composition by bioelectrical impedance methods: a comparative study J Appl Physiol 198558156571 https://doi.org/10.1152/jappl.1985.58.5.15653997721Apri DOISearch in Google Scholar

Kushner RF. Bioelectrical impedance analysis: a review of principles and applications. J Am Coll Nutr 1992;11:199-209. https://doi.org/10.1080/07315724.1992.12098245Kushner RF Bioelectrical impedance analysis: a review of principles and applications J Am Coll Nutr 199211199 209 https://doi.org/10.1080/07315724.1992.12098245Apri DOISearch in Google Scholar

Fuller NJ, Elia M. Potential use of bioelectrical impedance of the ‘whole body’ and of body segments for the assessment of body composition: comparison with densitometry and anthropometry. Eur J Clin Nutr 1989;43:779-91.Fuller NJ Elia M Potential use of bioelectrical impedance of the ‘whole body’ and of body segments for the assessment of body composition: comparison with densitometry and anthropometry Eur J Clin Nutr 198943779 91Search in Google Scholar

Thomas BJ, Cornish BH, Ward LC. Bioelectrical impedance analysis for measurement of body fluid volumes: a review. J Clin Eng 1992;17:505. https://doi.org/10.1097/00004669-199211000-00016Thomas BJ Cornish BH Ward LC Bioelectrical impedance analysis for measurement of body fluid volumes: a review J Clin Eng 199217505 https://doi.org/10.1097/00004669-199211000-0001610124463Apri DOISearch in Google Scholar

Gibson AL, Heyward VH, Mermier CM. Predictive accuracy of Omron body logic analyzer in estimating relative body fat of adults. Int J Sport Nutr Exerc Metab 2000;10:216-27. https://doi.org/10.1123/ijsnem.10.2.216Gibson AL Heyward VH Mermier CM Predictive accuracy of Omron body logic analyzer in estimating relative body fat of adults Int J Sport Nutr Exerc Metab 200010216 27 https://doi.org/10.1123/ijsnem.10.2.21610928828Apri DOISearch in Google Scholar

Nunez C, Gallagher D, Visser M, Pi-Sunyer FX, Wang Z, Heymsfield SB. Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact foot-pad electrodes. Med Sci Sports Exerc 1997;29:524-31. https://doi.org/10.1097/00005768-199704000-00015Nunez C Gallagher D Visser M Pi-Sunyer FX Wang Z Heymsfield SB Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact foot-pad electrodes Med Sci Sports Exerc 199729524 31 https://doi.org/10.1097/00005768-199704000-000159107636Apri DOISearch in Google Scholar

Bosy-Westphal A, Schautz B, Later W, Kehayias JJ, Gallagher D, Müller MJ. What makes a BIA equation unique? Validity of eight-electrode multifrequency BIA to estimate body composition in a healthy adult population. Eur J Clin Nutr 2013;67:S14-S21. https://doi.org/10.1038/ejcn.2012.160Bosy-Westphal A Schautz B Later W Kehayias JJ Gallagher D Müller MJ What makes a BIA equation unique? Validity of eight-electrode multifrequency BIA to estimate body composition in a healthy adult population Eur J Clin Nutr 201367S14 S21 https://doi.org/10.1038/ejcn.2012.16023299866Apri DOISearch in Google Scholar

Matias CN, Judice PB, Santos DA, Magalhaes JP, Minderico CS, Fields DA, Sardinha LB, Silva AM. Suitability of bioelectrical based methods to assess water compartments in recreational and elite athletes. J Am Coll Nutr 2016;35:413-21. https://doi.org/10.1080/07315724.2015.1058198Matias CN Judice PB Santos DA Magalhaes JP Minderico CS Fields DA Sardinha LB Silva AM Suitability of bioelectrical based methods to assess water compartments in recreational and elite athletes J Am Coll Nutr 201635413 21 https://doi.org/10.1080/07315724.2015.105819826934568Apri DOISearch in Google Scholar

Bracco D, Thiebaud D, Chiolero RL, Landry M, Burckhardt P, Schutz Y. Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans. J Appl Physiol 1996;81:2580-7. https://doi.org/10.1152/jappl.1996.81.6.2580Bracco D Thiebaud D Chiolero RL Landry M Burckhardt P Schutz Y Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans J Appl Physiol 1996812580 7 https://doi.org/10.1152/jappl.1996.81.6.25809018509Apri DOISearch in Google Scholar

Ling CHY, de Craen AJM, Slagboom PE, Gunn DA, Stokkel MPM, Westendorp RGJ, Maier AB. Accuracy of direct segmental multi-frequency bioimpedance analysis in the assessment of total body and segmental body composition in middle-aged adult population. Clin Nutr 2011;30:610-5. https://doi.org/10.1016/j.clnu.2011.04.001Ling CHY de Craen AJM Slagboom PE Gunn DA Stokkel MPM Westendorp RGJ Maier AB Accuracy of direct segmental multi-frequency bioimpedance analysis in the assessment of total body and segmental body composition in middle-aged adult population Clin Nutr 201130610 5 https://doi.org/10.1016/j.clnu.2011.04.00121555168Apri DOISearch in Google Scholar

Evans WD, McClagish H, Trudgett C. Factors affecting the in vivo precision of bioelectrical impedance analysis. Appl Radiat Isot 1998;49: 485-7. https://doi.org/10.1016/S0969-8043(97)00061-4Evans WD McClagish H Trudgett C Factors affecting the in vivo precision of bioelectrical impedance analysis Appl Radiat Isot 199849 485 7 https://doi.org/10.1016/S0969-8043(97)00061-4Apri DOISearch in Google Scholar

Kushner RF, Gudivaka R, Schoeller DA. Clinical characteristics influencing bioelectrical impedance analysis measurements. Am J Clin Nutr 1996;64:423S-427S. https://doi.org/10.1093/ajcn/64.3.423SKushner RF Gudivaka R Schoeller DA Clinical characteristics influencing bioelectrical impedance analysis measurements Am J Clin Nutr 199664423S 427 https://doi.org/10.1093/ajcn/64.3.423S8780358Apri DOISearch in Google Scholar

O’Brien C, Young AJ, Sawka MN. Bioelectrical impedance to estimate changes in hydration status. Int J Sports Med 2002;23:361-6. https://doi.org/10.1055/s-2002-33145O’Brien C Young AJ Sawka MN Bioelectrical impedance to estimate changes in hydration status Int J Sports Med 200223361 6 https://doi.org/10.1055/s-2002-3314512165888Apri DOISearch in Google Scholar

Wagner DR, Cotter JD. Ultrasound measurements of subcutaneous fat thickness are robust against hydration changes. Int J Sport Nutr Exerc Metab 2021;31:244-9. https://doi.org/10.1123/ijsnem.2020-0240Wagner DR Cotter JD Ultrasound measurements of subcutaneous fat thickness are robust against hydration changes Int J Sport Nutr Exerc Metab 202131244 9 https://doi.org/10.1123/ijsnem.2020-024033508779Apri DOISearch in Google Scholar

American College of Sports Medicine (ACSM). Nutrition and athletic performance. Med Sci Sports Exerc 2016;48:543-68. https://doi.org/10.1249/MSS.0000000000000852American College of Sports Medicine (ACSM) Nutrition and athletic performance Med Sci Sports Exerc 201648543 68 https://doi.org/10.1249/MSS.000000000000085226891166Apri DOISearch in Google Scholar

InBody. InBody230 user’s manual. Seoul: InBody; 1996. https://www.manualslib.com/manual/1206816/Inbody-230.html. Accessed 2 February 2022.InBody InBody230 user’s manual. Seoul: InBody; 1996 https://www.manualslib.com/manual/1206816/Inbody-230.html Accessed 2 February 2022Search in Google Scholar

Cohen J. Statistical Power Analysis for the Behavioral Sciences (2nd ed). Hillsdale, NJ: Lawrence Earlbaum. 1988.Cohen J Statistical Power Analysis for the Behavioral Sciences (2nd ed) Hillsdale, NJ Lawrence Earlbaum 1988Search in Google Scholar

Asselin M-C, Kriemler S, Chettle DR, Webber CE, Bar-Or O, McNeill FE. Hydration status assessed by multi-frequency bioimpedance analysis. Appl Radiat Isot 1998;49:495-7. https://doi.org/10.1016/S0969-8043(97)00179-6Asselin M-C Kriemler S Chettle DR Webber CE Bar-Or O McNeill FE Hydration status assessed by multi-frequency bioimpedance analysis Appl Radiat Isot 199849495 7 https://doi.org/10.1016/S0969-8043(97)00179-6Apri DOISearch in Google Scholar

Cutrufello PT, Dixon CB, Zavorsky GS. Hydration assessment among marathoners using urine specific gravity and bioelectrical impedance analysis. Res Sports Med 2016;24:219-27. https://doi.org/10.1080/15438627.2016.1202831Cutrufello PT Dixon CB Zavorsky GS Hydration assessment among marathoners using urine specific gravity and bioelectrical impedance analysis Res Sports Med 201624219 27 https://doi.org/10.1080/15438627.2016.120283127373703Apri DOISearch in Google Scholar

Dixon CB, Ramos L, Fitzgerald E, Reppert D, Andreacci JL. The effect of acute fluid consumption on measures of impedance and percent body fat estimated using segmental bioelectrical impedance analysis. Eur J Clin Nutr 2009;63:1115-22. https://doi.org/10.1038/ejcn.2009.42Dixon CB Ramos L Fitzgerald E Reppert D Andreacci JL The effect of acute fluid consumption on measures of impedance and percent body fat estimated using segmental bioelectrical impedance analysis Eur J Clin Nutr 2009631115 22 https://doi.org/10.1038/ejcn.2009.4219536161Apri DOISearch in Google Scholar

Ugras S. Evaluating of altered hydration status on effectiveness of body composition analysis using bioelectrical impedance analysis. Libyan J Med 2020;15:1741904. https://doi.org/10.1080/19932820.2020.1741904Ugras S Evaluating of altered hydration status on effectiveness of body composition analysis using bioelectrical impedance analysis Libyan J Med 2020151741904 https://doi.org/10.1080/19932820.2020.1741904714421232182203Apri DOISearch in Google Scholar

Utter AC, McAnulty SR, Riha BF, Pratt BA, Grose JM. The validity of multifrequency bioelectrical impedance measures to detect changes in the hydration status of wrestlers during acute dehydration and rehydration. J Strength Cond Res 2012;26:9-15. https://doi.org/10.1519/JSC.0b013e318238ea51Utter AC McAnulty SR Riha BF Pratt BA Grose JM The validity of multifrequency bioelectrical impedance measures to detect changes in the hydration status of wrestlers during acute dehydration and rehydration J Strength Cond Res 2012269 15 https://doi.org/10.1519/JSC.0b013e318238ea5121964427Apri DOISearch in Google Scholar

Armstrong LE. Assessing hydration status: the elusive gold standard. J Am Coll Nutr 2007;26:575S-584S. https://doi.org/10.1080/07315724.2007.10719661Armstrong LE Assessing hydration status: the elusive gold standard J Am Coll Nutr 200726575S 584 https://doi.org/10.1080/07315724.2007.1071966117921468Apri DOISearch in Google Scholar

Barley OR, Chapman DW, Abbiss CR. Reviewing the current methods of assessing hydration in athletes. J Int Soc Sports Nutr 2020;17:52. https://doi.org/10.1186/s12970-020-00381-6Barley OR Chapman DW Abbiss CR Reviewing the current methods of assessing hydration in athletes J Int Soc Sports Nutr 20201752 https://doi.org/10.1186/s12970-020-00381-6760233833126891Apri DOISearch in Google Scholar

Sawka MN, Coyle EF. Influence of body water and blood volume on thermoregulation and exercise performance in the heat. Exerc Sports Sci Rev 1999;27:167-218. https://doi.org/10.1249/00003677-199900270-00008Sawka MN Coyle EF Influence of body water and blood volume on thermoregulation and exercise performance in the heat Exerc Sports Sci Rev 199927167 218 https://doi.org/10.1249/00003677-199900270-00008Apri DOISearch in Google Scholar

Taylor NAS, Machado-Moreira CA. Regional variations in transepidermal water loss, eccrine sweat gland density, sweat secretion rates and electrolyte composition in resting and exercising humans. Extrem Physiol Med 2013;2:4. https://doi.org/10.1186/2046-7648-2-4Taylor NAS Machado-Moreira CA Regional variations in transepidermal water loss, eccrine sweat gland density, sweat secretion rates and electrolyte composition in resting and exercising humans Extrem Physiol Med 201324 https://doi.org/10.1186/2046-7648-2-4371019623849497Apri DOISearch in Google Scholar

Buono MJ. Limb vs trunk sweat gland recruitment patterns during exercise in humans. J Therm Biol 2000;25:263-6. https://doi.org/10.1016/S0306-4565(99)00095-9Buono MJ Limb vs trunk sweat gland recruitment patterns during exercise in humans J Therm Biol 200025263 6 https://doi.org/10.1016/S0306-4565(99)00095-910745121Apri DOISearch in Google Scholar

Cornish BH, Thomas BJ, Ward LC. Effect of temperature and sweating on bioimpedance measurements. Appl Radiat Isot 1998;49:475-6. https://doi.org/10.1016/S0969-8043(97)00057-2Cornish BH Thomas BJ Ward LC Effect of temperature and sweating on bioimpedance measurements Appl Radiat Isot 199849475 6 https://doi.org/10.1016/S0969-8043(97)00057-2Apri DOISearch in Google Scholar

Kenny GP, McGinn R. Restoration of thermoregulation after exercise. J Appl Physiol 2017;122:933-44. https://doi.org/10.1152/japplphysiol.00517.2016Kenny GP McGinn R Restoration of thermoregulation after exercise J Appl Physiol 2017122933 44 https://doi.org/10.1152/japplphysiol.00517.201627881668Apri DOISearch in Google Scholar

Mudie DM, Murray K, Hoad CL, Pritchard SE, Garnett MC, Amidon GL, Gowland PA, Spiller RC, Amidon GE, Marciani L. Quantification of gastrointestinal liquid volumes and distribution following a 240 mL dose of water in the fasted state. Mol Pharm 2014;11:3039-47. https://doi.org/10.1021/mp500210cMudie DM Murray K Hoad CL Pritchard SE Garnett MC Amidon GL Gowland PA Spiller RC Amidon GE Marciani L Quantification of gastrointestinal liquid volumes and distribution following a 240 mL dose of water in the fasted state Mol Pharm 2014113039 47 https://doi.org/10.1021/mp500210c25115349Apri DOISearch in Google Scholar

Matthews EL, Hosick PA. Bioelectrical impedance analysis does not detect an increase in total body water following isotonic fluid consumption. Appl Physiol Nutr Metab 2019;44:1116-20. https://doi.org/10.1139/apnm-2019-0106Matthews EL Hosick PA Bioelectrical impedance analysis does not detect an increase in total body water following isotonic fluid consumption Appl Physiol Nutr Metab 2019441116 20 https://doi.org/10.1139/apnm-2019-010630998857Apri DOISearch in Google Scholar

Androutsos O, Gerasimidis K, Karanikolou A, Reilly JJ, Edwards CA. Impact of eating and drinking on body composition measurements by bioelectrical impedance. J Hum Nutr Diet 2015;28:165-71. https://doi.org/10.1111/jhn.12259Androutsos O Gerasimidis K Karanikolou A Reilly JJ Edwards CA Impact of eating and drinking on body composition measurements by bioelectrical impedance J Hum Nutr Diet 201528165 71 https://doi.org/10.1111/jhn.1225925158295Apri DOISearch in Google Scholar

Kutac P. The effect of intake of water on the final values of body composition parameters in active athletes using two different bioimpedance analyzers. Acta Gymnica 2014;44:10716. https://doi.org/10.5507/ag.2014.011Kutac P The effect of intake of water on the final values of body composition parameters in active athletes using two different bioimpedance analyzers Acta Gymnica 20144410716 https://doi.org/10.5507/ag.2014.011Apri DOISearch in Google Scholar