A new quartz for 𝛾-transfer calibration of radiation sources

INFORMAZIONI SU QUESTO ARTICOLO

Cita

Quartz grains extracted from aeolian sand (LexCal2014) are tested for suitability as transfer material for radiation source calibration. After thermal pre-treatment, this quartz exhibits luminescence and dose-response properties which appear appropriate for transfer calibration purposes with little dependence of recovered β-dose on preheat temperature in the single aliquot (SAR) procedure and satisfying performance in β-dose recovery (0.98−1.00 given/measured β-dose for various experiments). Additional support is obtained by SARA and interpolation procedures, where OSL-data from γ-dosed LexCal2014 is interpolated on data obtained for zeroed LexCal2014 quartz, which is β-irradiated by the source to be calibrated. Initial results on fine-grain material agree with the coarse grain results presented.

The γ-dose of 3.00 ± 0.07 Gy is administered in a scatter-free geometry at the IAEA/WHO Secondary Standard Dosimetry Laboratory (SSDL) of the Helmholtz Zentrum München, with absorbed dose calculations obtained by Monte Carlo simulations.