Effect of adding a capsule with activated charcoal to abdominal ultrasound preparation on image quality

INFORMAZIONI SU QUESTO ARTICOLO

Cita

Introduction

Abdominal ultrasound plays a major role in the diagnosis of patients referred from outpatient clinics, receiving inpatient care and treated at emergency departments. In recent decades, ultrasound has become a cornerstone in diagnosis of many acute and chronic abdominal diseases, such as infections, inflammations, stones, masses, fluid collections and obstructions. The use of ultrasonography as the first-choice examination has become a standard procedure for doctors all over the world since it is available, safe and non-invasive. In routine abdominal ultrasound, patient preparation involving fasting for 8 hours and having a light meal the night before the examination is a common practice to avoid digestive motility and gases that may cause artifacts and degrade image quality. Even though patient preparation before abdominal ultrasound plays a major role in ensuring better visualization of internal organs and pathologies by minimizing artifacts caused by abdominal gases, which makes the scanning process much easier, abdominal gases still exist in many patients and in large amounts, especially in the pancreatic region, causing artifacts. For this reason, it is important to evaluate new, safe, efficient, reliable and cost-effective methods to improve patient preparation in order to eliminate excessive abdominal gases.

Literature review

Ultrasound is a form of mechanical energy used in medical imaging. Ultrasound waves are produced by the transducer and can travel within the body. They are reflected back from different structures with different attenuation values, thereby creating an image displayed on the monitor after being processed and reconstructed by the computer. Ultrasound is cheap and safe compared to other modalities. It plays a significant role in ruling out many pathologies. All sonographers must manage challenging image artifacts caused by gases present in the gastrointestinal tract. Usually, these are dirty shadowing and reverberation artifacts distal to gases, and they obscure underlying structures and degrade image quality(1). Factors causing excessive intestinal gas include: intestinal obstruction, motility disorders, infections, irritable bowel syndrome, malabsorption as well as psychological and dietary issues. Different drugs and agents are used for gas reduction; they include: beano, simethicone and activated charcoal. Some studies support the use of activated charcoal in the treatment of intestinal gas, while others do not(2). “Activated charcoal is a highly adsorbent powder made from superheated, high surface area, porous particles produced by pyrolysis of organic material. Its extensive surface area is covered with a carbon-based network that also includes functional groups (e.g., carbonyl, hydroxyl) that adsorb chemicals within minutes of contact, preventing gastrointestinal absorption and subsequent toxicity”(3). Many diseases have been treated with activated charcoal, including epilepsy, vertigo, and anthrax. Activated charcoal has recently been used as an antidote for different poisons and as an anti-gas agent. Large doses of activated charcoal administered over a long time have been demonstrated to be very safe and to produce no side effects(4). Higher BMI, abdominal wall thickness and waist circumference are the main factors that can predict the presence of excessive intestinal gas, which impairs the visualization of the upper abdominal structures and pathologies by ultrasound(5). In abdominal ultrasound, the gall bladder must be full and there should be as little gas in the gastrointestinal tract as possible. The intestinal contents (gas and food particles) have a direct impact on the diagnostic value of abdominal ultrasound since they can induce confusing shadows and inconclusive images. All patients undergoing ultrasound examination are always instructed to fast for several hours prior to the scan, mostly for 8 hours, sometimes for as long as 12 hours, to get the full, extended gall bladder. Otherwise the gall bladder will be contracted and difficult to assess. In addition, fasting patients will have less gas in the duodenum and colon. Patient preparation in the form of a two-day low-calorie diet, administration of laxatives and fasting have reported favorable results(6).

Oral simethicone is used as a pre-medication to reduce the amount of gas in the gastrointestinal tract. In the comparison of two protocols of simethicone pre-medication for ultrasound examination, where patients in one group were given a single dose for three days and patients in the other group were administered one single dose of simethicone 1 hour before ultrasound examination, it was found that the single dose administered 1 hour before scanning was superior in terms of the cost and benefit ratio(5). Significant improvement in the visualization of the pancreas and other retroperitoneal structures was reported after the use of simethicone-coated cellulose as an oral contrast agent. This combined agent was used in order to decrease gas-related artifacts from the bowel. Also, it can serve as an effective oral contrast agent for non-fating patients and in emergency cases(7). In Italy, activated charcoal is widely used to reduce intestinal gas and to prepare patients for abdominal US examinations. The results of using activated charcoal and alpha-galactosidase indicate a significant improvement in abdominal ultrasound in the study settings, but more evidence is needed regarding the time and cost of this form of preparation(8).

Our research was conducted to confirm or disprove the efficiency of using activated charcoal to improve the visualization of upper abdominal organs and achieve better image quality in ultrasound of the abdomen.

Methodology

A prospective research involved a population sample of 52 adult patients and volunteers, including both genders, persons at different age groups, with different body mass indices (BMI) (Tab. 1) and with no history of previous major abdominal surgery or acute illness. Pediatric patients, pregnant women, persons with a history of abdominal surgery, bedridden patients, and patients with bowel obstruction where excluded. The Ethics Committee of King Abdulaziz University approved the study protocol. Informed consent was obtained from all patients and volunteers. The routine upper abdominal ultrasound was preceded with usual preparation involving fasting for 8 hours prior to the examination (only water was allowed). Each patient was scanned twice; the first scan was conducted after the routine preparation, while the second scan was conducted after modified preparation where the same patient was instructed to keep fasting and additionally take one charcoal capsule orally 8 hours prior to the examination. Subsequently, we compared image quality using the pancreas as the reference. Ultrasound images obtained in both examinations from each patient were evaluated by a qualified radiologist and a senior sonographer using a specially designed evaluation form.

Body mass index

Body Mass Index
Underweight <18.5
Normal weight 18.5-24.9
Overweight 25-29.9
Obesity 30-34.9
Extreme obesity >35
Results

A sample size of 52 patients and volunteers included both genders. Twelve of them were excluded and only 40 candidates were included: 23 men and 17 women at the age range from 18 to 61 years, and with different BMI values (Tab. 1, Tab. 2, Tab. 3).

Gender and age group classification

Sex Age Group
16-25 26-35 36-45 46-55 55-65 Total
Males 3 7 7 3 3 23
Females 6 3 2 4 2 17
Total 9 10 9 7 5 40

Body mass index classification

Sex Body Mass Index
Underweight Normal weight Overweight Obesity Extreme obesity
Male 0 8 4 8 3
Female 2 5 4 3 3
Total 2 13 8 11 6

The images obtained before and after administration of an activated charcoal capsule were evaluated by two physicians experienced in radiology at King Abdulaziz University Hospital (KAUH): body-imaging radiologist and a senior sonographer with a PHD degree in ultrasound. They evaluated and classified the images into three categories: very good, good and poor. As for images of the head of the pancreas obtained before charcoal administration, 10% were classified as very good, 47.5% as good and 42.5% as poor. The results of the neck of the pancreas were as follows: 33.75% of very good images, 42.5% of good images and 28.75% of poor images. The images of the body of the pancreas were very good for 15% of images, good for 33.75% of images and poor for 51.2% of images. As for the tail of the pancreas, there were 0 very good images, 21.2% of good images and 78.75% of poor images, which was the highest result recoded (Fig. 1). The mean values regarding the visualization of the head of the pancreas after charcoal administration were: 35% in the very good group, 45% in the good group and 20% in the poor group. The results for the neck of the pancreas were as follows: 45% of very good images, 45% of good images and 10% of poor images. The images of the body of the pancreas were very good in 17.5% of cases, good in 60% of cases and poor in 22.5% of cases. The respective values for the tail of the pancreas were: 7.5%, 30% and 62.5% (Fig. 2).

Fig. 1.

Average values from two evaluations before activated charcoal administration

Fig. 2.

Average values from two evaluations after activated charcoal administration

Head of the pancreas

The results regarding the visualization of the head of the pancreas in 40 examined patients changed from 10% before charcoal administration to 35% after charcoal administration in the very good group, from 47.5% to 45% in the good group, and from 42.5% to 20% in the poor group, respectively; therefore the overall improvement was 50%.

Neck of the pancreas

As for the neck of the pancreas, the results were as follows: 33.75% and 45% of very good images, 42.5% and 45% of good images, and 28.75% and 10% of poor images, respectively for images obtained before and after charcoal administration. The overall improvement was 32.5%.

Body of the pancreas

For the body of the pancreas, the results were as follows: 15% and 17.5% in the very good group, 33.75% and 60% for the good group, and 51.2% and 22.5% for the poor group, respectively for images obtained before and after charcoal administration. The overall improvement was therefore 57.45%.

Tail of the pancreas

There were 0 and 7.5% of very good images, 21.2% and 30% of good images, and 78.75% and 62.5% of poor images, respectively for images obtained before and after charcoal administration. The overall improvement was therefore 32.55%.

Overall evaluation of image quality after administration of activated charcoal

The mean result obtained by both evaluators in terms of the overall criteria (much better, better, no difference and worse) was 21% for “much better,” 42% for “better,” 33% for “no difference” and only 4% for “worse.” The overall improvement in pancreas visualization was therefore 63% in all 40 cases (Fig. 3).

Fig. 3.

Average values for individual categories

Improvement in relation to body mass index

The overall improvement of 57.2% (much better and better) was noted for overweight, obese and extremely obese patients, whereas the value of only 39.05% was noted for underweight patients and for patients with normal body weight. In overweight, obese and extremely obese males the overall improvement reached 66.9% and the respective value for females was 49.9% (Tab. 4).

Improvement in two evaluations in relation to BMI and gender

Result Body radiologist Senior-sonographer Mean value
Improvement (overweight, obese, extremely obese) 53.3% 61.1% 57.2%
Improvement (underweight and normal weight) 40% 38.09% 39.045%
Improvement in males 58.8% 75% 66.9%
Improvement in females 46% 53.8% 49.9%
No improvement

The overall value for no improvement (no difference) was 62.4% in overweight, obese and extremely obese patients and only 37.2% for underweight patients and for patients with normal weight. In overweight, obese and extremely obese males the “no difference” value reached 36.9%, and the respective value for females was 62.5% (Fig. 4).

Fig. 4.

Images that were not improved in relation to BMI and gender

Relationship of BMI with the individual categories

The mean results from both analyses in relation to the BMI were as follows: the “much better” category: 12% in underweight patients, 29% in patients with normal weight, 29% in overweight patients, 18% in obese patients and 12% in extremely obese patients; the “better” category: 3% in underweight patients, 35% in patients with normal weight, 29% in overweight patients, 21% in obese patients and 12% in extremely obese patients; the “no difference” category: 4% in underweight patients, 35% in patients with normal weight, 4% in overweight patients, 38% in obese patients and 19% in extremely obese patients; the “worse” category: 67% in obese patients and 33% in extremely obese patients (Fig. 5).

Fig. 5.

Average values from two evaluations for individual categories in relation to BMI

The mean results from both analyses in relation to the gender were as follows: the “much better” category: 11% in males and 35% in females; the “better” category: 43% in males and 41% in females; the “no difference” category: 39% in males and 24% in females; the “worse” category 7% in males and 0 in females (Fig. 6).

Fig. 6.

Average values from two evaluations for individual categories in relation to gender

Relationship with age group

The results for the “much better” category were as follows: 41% in age group 16–25 years (the highest value), 23% in age group 36–45 years and 18% in two age groups: 26–35 years and 46–55 years. For the “better” category, the following values were noted: 29% (the highest value) in age group 36–45 years, 26% in age group 26–35 years, 21% in age group 16–25 years and 12% in two age groups 26–35 years and 46–55 years. The highest result in the “no difference” category was 31% in age group 26–35 years, followed by 23% in age group 56–65 years, 19% in age group 46–55 years, 15% in age group 16–25 years and only 12% in age group 36–45 years The “worse” category was selected only in two age groups: 33% in age group 36–45 years and 66% in age group 46–55 years.

Discussion

Activated charcoal is a highly adsorbent powder made from superheated, high surface area, porous particles produced from organic material(3). It is an excellent adsorbent of many chemical substances, including gases(9). The unsatisfactory ultrasound scans of the pancreas are frequently caused by an excessive amount of abdominal gases. This study showed that activated charcoal can improve pancreas visualization in ultrasound. The results mentioned above indicate that considerable improvement concerned the head of the pancreas and the body of the pancreas, where the highest improvement rates were obtained. The total improvement in the visualization of the pancreatic head was 50%, which was expected since the head of the pancreas is easy to access compared to other parts of the pancreas. Administration of activated charcoal to patients undergoing abdominal ultrasound contributed to a noticeable reduction of gases, which led to an increase in the quality of the visualization of the pancreatic head. As for the neck of the pancreas, the improvement in visualization was also observed (32.5%) for images obtained after charcoal administration compared to pre-administration scans.

Moreover, the study showed that the greatest improvement concerned the visualization of the body of the pancreas compared to the rest of the organ, where the overall improvement was 57.45%. The tail of the pancreas is usually very difficult to visualize with routine preparation for abdominal ultrasound because of its location between the stomach and the spleen. In this case, image quality improved by 32.5% after activated charcoal administration.

We found that significant overall improvement was related to BMI; in overweight, obese and extremely obese patients, improvement of image quality reached 57.2%, which is higher than the result for underweight patients and in patients with normal weight (only 39.05%). The improvement of image quality after charcoal administration in males was 66.9% while in females 49.9%, as shown in Tab. 4. On the other hand, the “not improved” images in males were found for 36.9% of cases. The respective value for females was 62.5%, and this had a negative effect on the overall improvement in females, which was 49.9%, as shown in Fig. 4. A low level of physical activity in females as seen in our society may be the major factor that affects the efficiency of activated charcoal in women. From our study, we concluded that adding one capsule of activated charcoal to the patient preparation routine in a sample of 40 patients directly improved the visualization of the pancreas by 63%. Image quality was not improved in 37% of the cases, as shown in Fig. 3.

Conclusion

Our preliminary study concluded that activated charcoal is useful, safe and cost-effective, and can be used to reduce gastrointestinal gases and improve the visualization of the pancreas and other structures in routine abdominal ultrasound in adults. Patient preparation for a routine abdominal ultrasound examination can be updated by adding an activated charcoal capsule to eliminate gases that usually degrade image quality in abdominal ultrasonography.

Limitations

Limited time to complete the research

The sample size was not large as patients refused participation in this research as additional preparation and a re-scan were needed.

Recommendations

It is recommended to conduct further investigation with an increased dose to two capsules of activated charcoal instead of one.

The sample size should be increased in future research for both genders.

It is recommended to fine-tune patient preparation by instructing patients to restrain from having fizzy drinks and avoid heavy meals for two days before the examination in addition to administration of activated charcoal.

Patient instructions should be revised and the statement “avoid milk, coffee and smoking on the day of the examination” should be added.