The ratio of weight loss to planning target volume significantly impacts setup errors in nasopharyngeal cancer patients undergoing helical tomotherapy with daily megavoltage computed tomography
Categoria dell'articolo: Research Article
Pubblicato online: 08 set 2016
Pagine: 427 - 432
Ricevuto: 15 mar 2016
Accettato: 19 lug 2016
DOI: https://doi.org/10.1515/raon-2016-0047
Parole chiave
© 2016 Radiol Oncol
This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Background
Changes in head and neck anatomy during radiation therapy (RT) produce setup uncertainties of nasopharyngeal cancer (NPC) irradiation. We retrospectively analyzed image guidance data to identify clinical predictors of setup errors.
Patients and methods
The data of 217 NPC patients undergoing definitive RT on a helical tomotherapy (HT) unit were analyzed. Factors including tumor stage, body mass index, weight loss, and planning target volume (PTV) were assessed as predictors of daily megavoltage computed tomography (MVCT) setup displacements, which were automatically registered using software.
Results
Mean daily setup displacements (in mm) were 1.2 ± 0.6, 1.8 ± 0.8, 3.4 ± 1.4 in the medial-lateral (ML), superior-inferior (SI), and anterior-posterior (AP) directions, respectively. Mean weight loss was 4.6 ± 3.3 kg (6.8 ± 4.9%). Patients with weight loss > 5% had significantly larger setup displacements in the AP (3.6 ± 1.5
Conclusions
Patients with weight loss > 5% and smaller PTVs, possibly because of small body frame or neck girth, were more likely to have increased setup errors in the AP direction.