A Comparison of ETV and LA for the Determination of Trace Elements in Solid Samples by MIP OES
Pubblicato online: 11 ott 2019
Pagine: 429 - 441
DOI: https://doi.org/10.1515/eces-2019-0032
Parole chiave
© 2019 Henryk Matusiewicz et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
The performance of electrothermal vaporization (ETV) and laser ablation (LA) of dry aerosols as sample introduction systems for microwave induced plasma optical emission spectrometry (MIP OES) are compared and evaluated in terms of detection limits, precision and accuracy for the determination of trace elements (Ca, Cd, Cu, Fe, Mg, Mn, Sr, Zn) in the same solid micro samples. In MIP OES both radiation sources can be independently adjusted to optimize the sampling process and then its subsequent excitation. A univariate approach and simplex optimization procedure were used to obtain the best signal/noise (S/N) ratio and derive analytical figures of merit. A comparison using a Student’s