Aortic patient-prosthesis mismatch - does it matter? A review for cardiologists and cardiac surgeons
, , et
29 sept. 2024
À propos de cet article
Catégorie d'article: Review
Publié en ligne: 29 sept. 2024
Pages: 107 - 121
DOI: https://doi.org/10.2478/rjc-2024-0001
Mots clés
© 2024 Alexandru C Visan et al., published by Sciendo
This work is licensed under the Creative Commons Attribution 4.0 International License.
Figure 1

Figure 2

Figure 3

Figure 4

Figure 5

Figure 6

Figure 7

Figure 8

Figure 9

Classification of patient-prosthesis mismatch based on severity_ Values above represent the effective orifice of area indexed to the BSA (EOAi)
Severity of patient prosthesis-mismatch (based on iEOA) | |||
---|---|---|---|
Mild | Moderate | Severe | |
BMI < 30 kg/m2 | > 0.85 cm2/m2 | 0.66 – 0.85 cm2/m2 | ≤ 0.65 cm2/m2 |
BMI ≥ 30 kg/m2 | > 0.70 cm2/m2 | 0.56 – 0.70 cm2/m2 | ≤ 0.55 cm2/m2 |
Proportion of patients with severe PPM after SAVR or TAVR, according to stroke volume index_ From Abbas AE, Ternacle J, Pibarot P, et al_ Impact of flow on prosthesis-patient mismatch following transcatheter and surgical aortic valve replacement_ Circ Cardiovasc Imaging_ 2021;14(8):E012364_ doi:10_1161/CIRCIMAGING_120_012364
SAVR | Severe PPM (%) |
---|---|
normal flow | 8 |
low flow | 42 |
normal flow | 5 |
low flow | 20 |
Impact of using manufacturer-predicted AVA, rather than continuity-measured, AVA on the prevalence of PPM, in SAVR and TAVR valves_ From Ternacle J, Pibarot P, Herrmann HC, et al_ Prosthesis-Patient Mismatch After aortic valve replacement in the PARTNER 2 trial and registry_ Cardiovascular Interventions_ 2021;14(13):1466–1477_ doi:10_1016/J_JCIN_2021_03_069
PPM prevalence (%) | ||
---|---|---|
SAVR | Measured AVA | Predicted AVA |
Moderate PPM | 31 | 28.4 |
Severe PPMT | 23.6 | 1.2 |
Moderate PPM | 17 | 21 |
Severe PPM | 0 | 0.1 |
For any given size label, the actual external diameter of the valve varies according to the model_ This is the reason why labelled size cannot be used as a surrogate for LVOT diameter in the continuity equation for calculating the aortic valve area (AVA)_ From Doenst T, Amorim PA, Al-Alam N, Lehmann S, Mukherjee C, Faerber G_ Where is the common sense in aortic valve replacement? A review of hemodynamics and sizing of stented tissue valves_ Journal of Thoracic and Cardiovascular Surgery_ 2011;142(5):1180—1187_ doi:10_1016/j_jtcvs_2011_05_007
Size label Valves | External diameter (mm) | |||||
---|---|---|---|---|---|---|
Epic | Supra | Mosaic | Perimount | Magna | Mitroflow | |
19 | – | 25 | 25 | 26 | 24 | 21 |
21 | 25 | 27 | 27 | 29 | 26 | 24 |
23 | 27 | 29 | 30 | 31 | 28 | 26 |
25 | 29 | 31 | 33 | 33 | 30 | 28 |
27 | 31 | 33 | 36 | 35 | 32 | 32 |
19 | – | 25 | 17 | 19 | 23 | 21 |
21 | 21 | 27 | 18.5 | 21 | 25 | 23 |
23 | 23 | 29 | 20.5 | 23 | 27 | 26 |
25 | 25 | 31 | 22.5 | 25 | 28.7 | 28 |
27 | 27 | 33 | 24 | 27 | 31 | 31 |