An optimized compact-based method is introduced that verifies the distance measuring precision of surveying instruments. The paper describes the improvement of the compact based method not only in establishing a compact facility by using shorter lengths and folded path but also in the assessment process by implementing the official ISO standard for electronic distance measuring equipment (ISO 17123-4). The method produces statistically comparable results to ISO standards and is suited for regular instrument checks in institutions with limited space available.