Twinkle artifact in renal ultrasound, is it a solid point for the diagnosis of renal stone in children?

À propos de cet article

Citez

Introduction

Renal stone is one of the common findings in daily medical practice, and it is usually seen in the adult population rather than the pediatric population(1). However, it has been noticed that the incidence of renal stones has increased 5 times in comparison to the previous decade in the pediatric age group(2).

Renal stones are considered less common in children in comparison to the adult population, and result from a multifactorial process that involves both the patient’s underlying metabolic background and environmental conditions promoting nephrolithiasis, such as volume depletion, infection, or the intake of foods high in lithogenic solutes(3,4). In view of the fact that the vast majority of pediatric stones are calcium based(5), investigating causes should be focused on factors that contribute to an increased calcium excretion(4). There are some known underlying metabolic disturbances which increase the risk of stone development, such as Dent disease, primary hyperoxaluria, or Lesch-Nyhan syndrome(4).

There are no typical presenting symptoms of renal stones in the pediatric age group, and they usually present with nonspecific poorly localized pain in the abdomen or pelvis(6). Gross hematuria can also be seen in adolescents(4).

A nonenhanced computed tomography (CT) scan is the modality of choice to evaluate renal stones, being characterized by both high sensitivity and specificity. However, the evaluation of renal stones in children should be performed while limiting the ionizing radiation as far as possible. Therefore, ultrasound (US), being a widely available technique, is considered the modality of choice to obtain non-ionizing higher resolution images. According to the recently published guideline, US should be the initial imaging modality for use in children with nonspecific abdominal symptoms and those suspected of having renal stones(7).

Several ultrasonographic features are known to be associated with renal stones, which have been described in the literature, such as the presence of the twinkle artifact. One can identify the twinkle artifact as an intrinsic machine noise which presents as a focus of alternating colors on the Doppler mode signals, related to a strongly reflecting interface of an object and giving the appearance of turbulent blood flow.

A few studies have confirmed the usefulness of the twinkle artifact in detecting renal stones, but with variable degrees of sensitivity. For example, a study performed in an adult age group found that the twinkle artifact was associated with a high sensitivity for detecting renal stones but also with a high false positive rate(8). Another study compared the presence of the renal twinkling artifact in association with nephrolithiasis, revealing that it was relatively insensitive in routine clinical practice and had a high false positive rate(9). On the other hand, some studies mention that the use of color-Doppler US is a preferable option for the sensitive detection of tiny nephrolithiasis(10). One recent study conducted in a pediatric age group showed that the twinkle artifact was characterized by sensitivity reaching up to 88% (95% CI 72–96%)(11).

The purpose of our study was to assess the sensitivity of the twinkle artifact as an aiding tool to diagnose kidney stones in the pediatric age group by ultrasound.

Material and methods

The study had a retrospective design. Local institutional approval was acquired. The study population consisted of pediatric patients (<14 years) who had undergone abdomen/renal ultrasound in a single healthcare facility between February, 2013, and August, 2019. All finalized abdomen/renal ultrasound reports containing the words twinkle, twinkling, and kidney stones (n = 287) were identified by query of the institutional picture archiving and communication system (PACS).

Our inclusion criteria were: patient age <14 years, availability of images from a reference standard nonenhanced thin-section (≤2.5 mm), and abdominal CT examination performed within 90 days before or after the abdomen/renal ultrasound examination (n = 33).

The exclusion criteria were: any patient who received treatment or intervention for kidney stones such as nephrostomy tubes in between the ultrasound and CT, and any patient with nephrocalcinosis.

Anonymized US and CT reports and images were randomized and retrospectively reviewed by three radiologists (pediatric radiology consultant, pediatric radiology fellow, and senior radiology resident). The following parameters were recorded: gender of patient (male/female), age of patient (months), stone with the twinkle artifact (yes/no), number of stones, location of stones (right/left, kidney), size of stone (mm), is the stone seen in CT? (yes/no), and number of days between the US and CT assessments.

Abdominal CT was used as the reference standard. Subgroup analysis was based on the discrepancies regarding the presence of the twinkle artifact stone between the US and CT. The first group was true positive (US showed the twinkle artifact and a stone was also seen in CT), the second group was false positive (US showed the twinkle artifact but a stone was not seen in CT), and the third group was false negative (US did not show the twinkle artifact but a stone was seen in CT). SPSS V25 was used for data analysis.

Results

Thirty-three abdomen/renal ultrasound scans of 33 patients (21 males,12 females) fulfilled the entry criteria. The interval between the US and CT was <3 months for all patients. The median overall age was 4 years (IQR: 3.125, range: 1–165 months), and the median number of days between the US and CT was 13 (IQR: 26, range: 0–81 days) (Tab. 1). US detected 33 hyperechoic foci suspected for stones; 26 were confirmed true positive (showed the twinkle artifact and stones were seen in CT) (Fig. 1), 4 were false positive (showed the twinkle artifact but no stones were seen in CT) (Fig. 2), and 3 were false negative (did not show the twinkle artifact but stones were seen in CT). The overall median stone size was 2 mm in the right kidney, and 5 mm in the left kidney (IQR: 6,11 mm), respectively.

Frequency distribution of demographic variables

Gender male 21 (63.6%)
female 12 (36.4%)
Age median 4.00
IQR 3.125
Number of stones Median 1.00
IQR 1.00

Fig. 1.

A 10-year-old female: a known case of thoracolumbar scoliosis. A. Gray-scale ultrasound of the left kidney showed left renal pelvis stone with posterior shadowing, measuring 13.6×12.2×4.3 mm. B. Color-Doppler ultrasound of the left kidney showed the twinkle artifact within the stone. Nonenhanced CT scan of the abdomen and pelvis was done 1 month later, confirming the presence of the stone

Fig. 2.

A 6-year-old male: a known case of acute lymphoblastic leukemia with hematuria. A. Grey-scale ultrasound of the left kidney showed mild left hydronephrosis with non-shadowing echogenic foci in the lower pole. B. Color-Doppler ultrasound of the left kidney showed the twinkle artifact within the left lower pole echogenic focus. Enhanced CT scan of the abdomen and pelvis showed bilateral mild hydronephrosis with no stones in the left kidney

Twinkle artifact sensitivity was found to be 89.7% (95% CI 39.574%–90%), and specificity could not be calculated due to a lack of true negatives (Tab. 2).

Subgroups analysis

US findings
True positive False positive False negative
Number of cases 26 4 3
Days between US andCT (9,23) (32.5, 57.25) (26, --)
% of twinkle artifact positive cases 86.7% 13.3% N/A
Size in mm (median, IQR) (10,11) (8,10) (15, --)

The true positive subgroup consisted of 26/30 (kidney) stones. The median stone size was 10 mm (IQR: 11 mm, range: 2–25 mm), and the median number of days between the US and CT in this subgroup was 9 (IQR: 23, range: 0–66 days). The twinkle artifact was assessed in all true positive stones, revealing a relatively high PPV of 26/29 (86.7%) for the twinkle artifact. The twinkle artifact was not dependent on stone size (Tab. 2).

The false positive subgroup consisted of 4/30 (kidney) foci (13.3%). The median stone size was 8 mm (IQR: 10 mm, range: 3–15 mm), and the median number of days between the US and CT in this subgroup was 32.5 (IQR: 57.25, range: 9 –81 days) (Tab. 2).

The false negative subgroup comprised 3 stones (100%). The median size was 15 mm (IQR: --, range: 2–15 mm), and the median number of days between the US and CT was 26 (IQR: --, range: 11–28 days) (Tab. 2).

There was no difference between the subgroups regarding stone size, age or number of days between the US and CT (p = 0.767, p = 0.736, and p = 0.194, respectively).

Discussion

This was a retrospective observational study using abdominal CT scans as a reference standard to test the hypothesis that the twinkle artifact occurs in most pediatric kidney stones, and with high sensitivity.

The twinkle artifact is a complex phenomenon. The most accepted theory nowadays is that the presence of a narrowband noise due to fluctuations in the circuits of the Doppler US equipment is the underlying cause of this sign. It is mainly observed on rough, hyperechoic, irregular surfaces with multiple cracks which cause a strong reflection of the incident US waves and multiple internal reflections which widen the spectrum. The appearance of the twinkle artifact is correlated with the roughness of the stones – the greater the surface roughness, the greater the artifact(12). This can explain why in 3 cases we did not find the twinkle artifact due to the smooth, regular interface of the stone surface. False positive twinkle artifact is thought to be a result of various causes such as parenchyma-vascular interface, nephrocalcinosis and (to a lesser extent in children) vascular calcification(13).

Our data show high sensitivity of the twinkle artifact in pediatric patients, reaching up to 89.7% (95% CI 39.574%– 90%), with a relatively high positive predictive value reaching up to 86.7%. However, one needs to take in consideration the fact that there is a small risk of false positive results. Our results are comparable to those obtained in a recent study, showing the sensitivity of 88% (95% CI 72–96%)(11). We found that the twinkle artifact can be detected already in small stones (as small as 2 mm).

Although not significant, the median stone size in the true positive subgroup (median: 10 mm) was larger than in the false positive subgroup but smaller than in the false negative subgroup median (median values: 8 and 15 mm, respectively).

The limitations of our study include small population size, since we use ultrasound in our center as a primary tool to detect kidney stones in pediatric patients and reserve CT for complex or questionable cases, if clinically needed; and the lack of knowledge about the biochemical composition of stones which may influence the twinkle artifact(14).

Conclusion

The twinkle artifact is a sensitive US tool for detecting pediatric kidney and ureter stones, but with a small risk of false positive findings.