Accès libre

Evaluation of the visibility of peri-implant bone defects using ultrasonography with two types of probes

À propos de cet article

Citez

Introduction

For the success of dental implants, the implant surface must be covered with bone. There is a positive correlation between alveolar bone thickness and primary osteointegration(1,2). Peri-implant defects are supportive bone tissue loss with a prevalence ranging from 28% to 56%(3). Routine radiographic evaluations are a technique used to evaluate whether peri-implantitis develops. When peri-implantitis goes unnoticed early on, marginal bone loss progresses and leaves the clinician with increasingly narrowing treatment options. From the patient’s point of view, there will be a decrease in the quality of final oral rehabilitation unless the condition is promptly addressed(4).

Intraoral and panoramic radiography techniques show the mesial and distal areas of the bone(5); but there can be geometric distortions and anatomical superimpositions(6). If 3-dimensional observation of the bone is necessary, conventional and cone-beam computed tomography (CBCT) can provide alternative options(7). There are certain major limitations of these techniques, though, including their high cost, increased radiation exposure, and formation of metal artifacts(8).

In view of such disadvantages of CBCT, the effectiveness of ultrasonography (US) on bone surface evaluation, bone thickness measurement, and peri-implant defects visibility has started to be evaluated in the literature in recent years(912). It is reported that US is the preferred modality due to its advantages such as non-invasive nature, use of non-ionizing radiation, good tolerability by patients, and low cost(13,14).

US is based on the principle of measuring the energy loss caused by the emission, reflection, and scattering of acoustic waves lower than 20 kHz in different tissues. The energy loss of the wave propagating throughout the tissue is associated with the acoustic properties of the waves(15). Due to the high-frequency mode on US, the depth of signal penetration into the tissue decreases, but the image quality is improved. This means that there is an inverse correlation between the image resolution and the measured depth. It has been reported in the literature that high-frequency ultrasounds can be used to scan the bone surface(16). In another study, it was found that the combined use of high- and low-frequency ultrasound may be a new approach in cortical bone evaluation(9).

The present study aimed to evaluate the efficacy of intraoral and extraoral US evaluations performed with two different types of probes (linear and hockey stick) for the visibility of peri-implant bone defects.

Material and methods

This study was carried out using three sheep heads including soft tissues with 14 implants [zirconium (n = 2) titanium implants (n = 12)]. Before starting the implant surgery, CBCT scanning was done (Planmeca 3D max, Helsinki, Finland) with the following exposure parameters: 96 kVp, 12 mA, and 18 s from each sheep head to plan and identify the implant sites properly. A radiologist with 10 years’ experience in oral and maxillofacial radiology performed the implant surgery as well as defect creations. The implants were inserted with a sub-crestal incision to reflect the mucoperiosteal flap. The osteotomy was performed using a pilot drill, and after that sequential drilling was done to prepare the area, taking into account the size of the implant. Copious irrigation with saline was done during the surgical procedure. The implant was inserted with the help of an insertion tool and a torque wrench. Following dental implant placement, the same radiologist created standardized defects around the dental implants. These simulated defects were created only in the buccal surfaces of the implants, app. 3 mm in diameter and semi-elliptical in form, with high-speed equipment, using copious air/water spray and rounded diamond burs (KG Sorensen, Zenith Dental ApS, Agerskov, Denmark). A total of 9 fenestrations (all buccal surfaces) were created, while 5 surfaces were free of fenestrations and used as a control group. The bone defects were created without knowing the depth. After creating the defects the around dental implants, the mucoperiosteal flap was closed. Care was taken not to damage the gingiva around the implants. The defects noted by the same radiologist served as the gold standard for US imaging.

Imaging modalities

Two US consoles (ProSound Alpha 6, Hitachi Aloka Medical Ltd., Tokyo, Japan) with a hockey stick intra-operative probe, 13 MHz (UST-536, Hitachi Aloka Medical Ltd., Tokyo, Japan), and a linear probe, 5–13.3 MHz (UST-5413, Hitachi Aloka Medical Ltd., Tokyo, Japan), and also a high-resolution ACUSONS 2000 ultrasound unit (Siemens, Munich, Germany) with a 4–9 MHz linear probe (9L4 Transducer) and a hockey stick intra-operative probe 14 MHz (14L5 SP Transducer) were used in the study. All fenestrations and dental implants were scanned both intraorally and extraorally (Fig. 1, Fig. 2, Fig. 3, Fig. 4, Fig. 5). Extraoral scanning was performed with a linear probe, while intraoral scanning was done only with hockey stick probes with a different frequency bandwidth. After adjusting the probe to the desired trajectory, US images were obtained using both probes. Two observers (H.K., K.O) carried out all US scanning and evaluations. The observers were dentomaxillofacial radiologists with 10 years’ and 18 years’ of experience with US, respectively.

Fig. 1.

A. Image using both zirconium and titanium implants with fenestrations, B. Fenestration defects around titanium implants, C. Preparation of scanning site; note that the flap was again placed on the scanned region, D. Intra-oral scan, E. Extra-oral scanning of the dental implant sites

Fig. 2.

A. US image showing fenestration defects around titanium implants (arrows), B. US showing a fenestration defect around zirconium implant (arrow) scanned with a 13 MHz intra-operative probe that was used intraorally

Fig. 3.

A. US image showing a fenestration defect around titanium implant (arrowhead), while the fenestration defects were non-visible around A. titanium and B. zirconium implant scanned with a 5–13.3 MHz linear probe that was used extraorally (arrows)

Fig. 4.

A. US image showing fenestrations around titanium (left arrow) and zirconium implant (right arrow) scanned with a 14 MHz intra-operative probe, B. Same implants scanned with a 4–9 MHz linear probe that was used extraorally without any fenestration defect visible

Fig. 5.

A. US image showing fenestrations around titanium implants scanned with a 14 MHz intra-operative probe, B. Same implants scanned with a 4–9 MHz linear probe that was used extraorally with the defect

The two observers conducted two separate US sessions independently. The study was performed twice, with an interval of 2 weeks after the initial US scanning. The same sheep heads were used for both US scanning procedures. The observers were free to position the probe when taking images.

Image analysis

Before the US evaluations, both observers were trained to appropriately use the US software in a special session. However, no calibration was made for the US evaluations since the scanning itself was in real-time. The fenestration defects were detected simultaneously in each US session during scanning, so the detection of defects was done in real-time while scanning. US system proprietary software was used (Hitachi Aloka Medical Ltd., Tokyo, Japan, Siemens S2000, Munich, Germany). The observers were free to use any enhancement procedure that was available in the US unit. Moreover, the observers were aware of the existence of the defects, however they did not know the dental implant type (titanium or zirconium).

For all imaging methods, a five-point scale was used to assess the visibility of each fenestration: (1) definitely absent; (2) probably absent; (3) unsure; (4) probably present; (5) definitely present.

Examiner reliability and statistical analysis

Intraclass Correlation (ICC) analysis was used for the assessment of intraobserver and intraobserver reliability. Kappa coefficients were calculated to evaluate the gold standard and the observers’ agreements for each image set. Kappa values were interpreted according to the guidelines proposed by Landis and Koch(17), and adapted by Altman(18) κ ≤0.20, poor; κ = 0.21–0.40, fair; κ = 0.41–0.60, moderate; κ = 0.61–0.80, good; and κ = 0.81–1.00, very good. Scores obtained from the (1) 1st linear probe extraorally; (2) 2nd linear probe extraorally; (3) 1st hockey stick probe intraorally; and (4) 2nd hockey stick probe intraorally were compared with the gold standard. A probability level of less than 5% (p <0.05) was accepted as statistically significant. Statistical analyses were performed using SPSS 21.0 software (SPSS Inc., Chicago, IL, USA).

Results

The lowest ICC value in the intraobserver reliability assessment was obtained with a linear probe (UST-5413) that was used extraorally. The highest ICC value in the intraobserver reliability assessment was obtained with a 14 MHz (14L5 SP Transducer) hockey stick probe (0.966) for Observer 1, and with a 13 MHz (UST-536 Transducer) hockey stick probe that was used intraorally (0.952) for Observer 2 (Tab. 1).

Intraobserver agreement for Observer 1 and Observer 2. p value less than 0.05 considered as statically significant (95% CI)

Observer 1 Observer 2
ICC p ICC p
1st hockey stick probe intraorally 0.878(0.632-0.960) 0.0001 0.952(0.817-0.986) 0.001
2nd hockey stick probe intraorally 0.966(0.885-0.989) 0.0001 0.921(0.752-0.975) 0.001
1st linear probe extraorally 0.696(0.095-0.901) 0.019 0.727(0.120-0.913) 0.016
2nd linear probe extraorally 0.849(0.524-0.952) 0.001 0.875(0.609-0.960) 0.001

The interobserver ICC coefficients were presented in Tab. 2. Good interobserver reliability was achieved in all probes with ICC values between 0.762 and 0.914. The linear probe (UST-5413) that was used extraorally had the lowest interobserver reliability (Tab. 2).

Interobserver agreement between observers. p value less than 0.05 considered as statically significant (95% CI)

ICC p
1st hockey stick probe intraorally 0.784 (0.349-0.930) 0.002
2nd hockey stick probe intraorally 0.909 (0.721-0.971) 0.001
1st linear probe extraorally 0.762 (0.300-0.922) 0.006
2nd linear probe extraorally 0.914 (0.728-0.972) 0.001

The 13 MHz hockey stick intraoral probe (UST-536 Transducer) had a high level of agreement with the gold standard (p <0.05; κ = 0.837), while the other hockey stick intraoral probe showed a moderate to high agreement with the gold standard for Observer 1 (p <0.05; κ = 0.689). No significant differences were found in the agreement between Observer 1 and the gold standard for the linear probes that were used extraorally (p >0.05) (Tab. 3).

Comparison of observers with the gold standard. p value less than 0.05 considered as statically significant

Sensitivity Specificity PPV NPV κ p
1st hockey stick probe intraorally Observer 1-Gold Standard 80% 100% 100% 90% 0.837 0.001
2nd hockey stick probe intraorally Observer 1-Gold Standard 80% 89% 80% 89%hz 0.689 0.011
1st linear probe extraorally Observer 1-Gold Standard 100% 33% 46% 100% 0.263 0.145
2nd linear probe extraorally Observer 1-Gold Standard 100% 33% 46% 100% 0.263 0.145
1st hockey stick probe intraorally Observer 2-Gold Standard 100% 89% 83% 100% 0.851 0.001
2nd hockey stick probe intraorally Observer 2-Gold Standard 100% 78% 71% 100% 0.714 0.005
1st linear probe extraorally Observer 2-Gold Standard 60% 22% 30% 50% –0.145 0.481
2nd linear probe extraorally Observer 2-Gold Standard 80% 11% 33% 50% –0.068 0.649

PPV – positive predictive value, NPV – negative predictive value

The agreement between Observer 2 with the gold standard was high (p <0.05; κ = 0.851 and 0.714, respectively) for both hockey stick probes used intraorally, while again no significant differences were noted in terms of agreement between Observer 2 and the gold standard for the linear probes that were used extraorally (p >0.05) (Tab. 3).

Discussion

Early diagnostic criteria of peri-implantitis include radiographic bone loss greater than one-third of implant height(19). For this reason, it is important to monitor bone loss in the follow-up period after the operation(20). In dentistry, radiographic examination is the most common choice to evaluate peri-implantitis(5). In the meta-analysis of Bohner et al. (21), the majority of the reviewed studies used CBCT or intraoral radiography to diagnose peri-implant bone defects. However, Bornstein et al. (22) reported that the evaluation of bone around dental implants was limited due to the formation of beam-hardening artifacts. In recent years, the question of whether bone defects can be examined without using ionizing radiation has started to be investigated, with the conclusion that US is a promising technique for the evaluation of bone defects and soft tissue pathologies caused by surgical complications(23).

To the best of our knowledge, there are only a few studies on bone surfaces with evaluations performed using US(9,13,16,24). In their study, Degen et al. (9) compared low- and high-frequency US to CBCT in measuring cortical bone thickness, concluding that US can assist CBCT in measuring bone thickness. Bohner et al. (10) used CBCT, Micro-CT, and US to evaluate peri-implant bone defects, and reported that US underestimated the measurements for the supra-alveolar and intra-bony surfaces, compared with CBCT and Micro-CT. Also, it was found that US was accurate in measurements of the width of peri-implant defects, although vertical measurements were underestimated by about 1 mm compared to those performed with CBCT and Micro-CT(10). In the study of Choi et al. (25), US was found to be capable of showing representative features for implant planning in a porcine model; these included implants that were placed in edentulous ridges; implants for single missing teeth; implants and teeth with simulated dehiscences; and mental foramina.

Several studies have explored the effect of probes of different frequencies on the quality of imaging periodontal defects. Mahmoud et al. (24) investigated the effectiveness of high- and low-frequency probes in imaging periodontal diseases. The authors stated that high-frequency probes were more effective in detecting periodontal diseases early. Similarly, Chifor et al. (26) reported that using high-frequency probes was more effective in imaging periodontal bone defects. However, no study comparing the effectiveness of the intraoral probe and linear probe in the visibility of defects has been found. Therefore, the aim of this study was to investigate the effectiveness of different probes in the visibility of peri-implant defects.

CBCT can also be used routinely for detecting these kinds of defects. However, the occurrence of metal artifacts around dental implants, as scattering or complete absorption of the beam can exist and be concluded with image degradation. This situation can prevent the observation of the implant-bone interface, and make it difficult to evaluate peri-implant bone defects(27,28). In the head and neck area, high-resolution images in multiple planes are obtained with modern US units with high-frequency linear probes (7.5–12 MHz)(29). In dental practice, US is mainly used in cases of maxillofacial fractures(30), cervical lymphadenopathy(31),soft tissue masses(32), masticatory and neck muscles(33,34), temporomandibular joint(35), periapical(36,37) salivary gland diseases(38), intraosseous jaw pathologies(39), and carotid paragangliomas(40).

There are several limitations of this study. Firstly, two different ultrasound systems with probes slightly differing in frequency were used. Although the frequencies of probes are similar to each other, they may still influence the results obtained. This can be due to the value of the signal to noise ratio (SNR) which increases as the frequency of the ultrasound signal rises. It was stated that less speckle noise was produced as higher frequency signal was applied(41). In a recent paper, it was also found that lower frequency may achieve a better depth penetration, while higher frequencies are associated with better resolution. While the output power may improve image quality by increasing the intensity of transmitted sound energy, the impact is usually insignificant(42). This issue can be addressed in more depth in future studies.

Another limitation of this study is that even though two observers performed two separate US sessions independently, observational differences in US can affect the results. Also, no comparison was performed for implant types (zirconia and titanium implants). Since the number of implants of both types in this study is not sufficient, the comparison between them has not been studied statistically. However, further studies with more implant types and numbers should be done to elucidate the differences between implant types in US images.

Conclusion

Hockey stick probes used intraorally can be an effective option for the evaluation of the visibility of peri-implant bone defects. No report was found in the literature regarding a comparison of peri-implant bone defect visibility with different US probes. To the best of our knowledge, this study is the first one addressing this subject. Thus, the conclusion is that US can be an alternative method of evaluating defects. However, further studies are needed to determine the effectiveness of US in the visualization of peri-implant bone defects.

eISSN:
2451-070X
Langue:
Anglais
Périodicité:
4 fois par an
Sujets de la revue:
Medicine, Basic Medical Science, other