Introduction

Nowadays, measuring the Electrical Bioimpedance (EBI) in humans is a common practice in several clinical applications e.g. electronic biopsy for skin cancer screening [1], Body Composition Analysis (BCA) for assessment on body fluids distribution [2], impedance cardiography for non-invasive hemodynamic monitoring [3]. Several EBI applications make use of EBI spectroscopy (EBIS) measurements analyzing the impedance spectrum and, in most cases, a Cole-based model analysis is performed.

ZCole(ω)=R+R0R1+(jωτ)α$${{Z}_{Cole}}\left( \omega \right)={{R}_{\infty }}+\frac{{{R}_{0}}-{{R}_{\infty }}}{1+{{\left( j\omega \tau \right)}^{\alpha }}}$$

To perform a Cole model analysis, EBIS data must be fitted to the Cole Function, see eq. (1), [4]. The Cole function experimentally resembles EBIS data from a single dispersion in the given frequency range and it is defined by four parameters R0, R, α and τ. The Cole parameters can be used for visualization producing the Cole plot and as data features to characterize an EBI system. Through further processing they can be also used for tissue constitution assessment like it is the case of BCA applications [5].

In BCA, using the values obtained from EBIS data for the Cole parameters and applying Hanai mixture theory [6], [7] and [2] or empirically derived prediction equations [8], it is possible to estimate the value of the BCA parameters: TBW, ECF, ICF and FM.

The estimation of the values of the Cole parameters from the EBIS data is usually obtained through iterative Curve fitting [9], [10], [11]. Since the impedance is a complex function of frequency, curve fitting can be done on the spectral domain or in the impedance plane [9] [12] and [13].

The impedance–only approach presented in [13] estimates the value of the Cole parameters from the modulus of the EBI, allowing the use of a non-phase sensitive spectrometer. This reduction of hardware requirements is obtained at the expenses of producing the parameter estimation on the impedance plane disregarding the frequency information.

Recently, the Non-linear Least Squares (NLLS) approach to fit EBIS measurements into the Cole function on the frequency domain was theoretically introduced by Ayllon et al in [14] and empirically proven by Buendia in [15]. Although in both works the NLLS approach was applied on the immitance spectral components of the complex EBI data, which require EBI measurements taken with a phase sensitive EBI spectrometer, the NLLS approach allows performing Cole parameter estimation on the modulus of the impedance as well.

Combining both the impedance-only estimation approach suggested by Ward et al in [13] and the NLLS method, it would be possible to implement a Cole function fitting method that would estimate the Cole parameter in the frequency domain and would not require phase-sensitive EBI measurements. A recent algorithm evaluation performed by Nordbotten et al. in [16] confirms the validity of such approach.

To validate the use of the NLLS impedance-only approach for a well known and spread application of EBIS, in this work, the BCA parameters have been estimated through Cole function fitting from total right side wrist-to-ankle EBI measurements. The BCA parameters have been calculated with the Bioimp software for assessment of body composition while, for comparison purposes, the Cole function fitting has been performed both, applying the NLLS method on the spectrum of the impedance modulus and using the Cole fitting tool available on Bioimp software.

A positive validation of the NLLS impedance-only approach would enable the design of impedance spectrometer for body composition analysis and assessment on nutritional status without the need of phase detection capabilities, which would reduce considerably the complexity of the hardware requirements.

Materials and methods
Cole Based analysis and BCA parameters estimation

In 1940 Kenneth S. Cole introduced the Cole equation, eq (1), an empirical complex nonlinear function of frequency that accurately fitted experimental EBI measurements. Such function is built by 4 parameters R0, R, α and τ, but only 2 of them, the resistance at DC frequency R0 and the resistance at infinite frequency R, are used to estimate the BCA parameters through Hanai’s mixture theory [6]. Using the Cole parameters, together with morphological data from a human subject and certain constants [17], it is possible to predict the volume of the extra- and intracellular fluid and consequently the total content of body water, ECF, ICF and TBW correspondently, which are known as BCA parameters. Such parameters are estimated according to equations 2 and 3, [17] & [18].

ECF=1100Kb2ρecf2Db*H4WR023$$ECF=\frac{1}{100}\sqrt[3]{\frac{K_{b}^{2}\rho _{ecf}^{2}}{{{D}_{b}}}*\frac{{{H}^{4}}W}{R_{0}^{2}}}$$

The ECF parameter in liters is obtained with eq. (2), Where W is body weight in kg, H is height in cm, R0 is the value of the Cole parameter in Ω, Kb is the body proportion, typically 4.3 for wrist to ankle measurements, ρecf is the resistivity of the extracellular fluid and Db is the body density in kg/l with a estimated value of 1.05.

Once the value ECF is estimated introducing the value for R0, the ICF parameter is calculated following eq. (4). Note that equations (2), (3) and (4) have been adapted to the Cole parameters and the nomenclature of body fluid and not body water.

(1+ICFECF)52=(R0R)(1+ρiρeICFECF)$${{\left( 1+\frac{ICF}{ECF} \right)}^{{}^{5}/{}_{2}}}=\left( \frac{{{R}_{0}}}{{{R}_{\infty }}} \right)\left( 1+\frac{{{\rho }_{i}}}{{{\rho }_{e}}}\frac{ICF}{ECF} \right)$$

Where ICF and ECF are the volumes of intra and extra cellular fluid in liters respectively and ρi and ρe are the apparent intra- and extracellular resistivities respectively [17].

Eq. (3) can be solved by expanding it into the form of eq. (4) where x=ICFECF.$x={}^{ICF}/{}_{ECF}.$

x5+5x4+(10(R0R)2+(ρiρe)2)x2+(52(R0R)2+(ρiρe))x+1(R0R)2=0$$\begin{align}& {{x}^{5}}+5{{x}^{4}}+\left( 10-{{\left( \frac{{{R}_{0}}}{{{R}_{\infty }}} \right)}^{2}}+{{\left( \frac{{{\rho }_{i}}}{{{\rho }_{e}}} \right)}^{2}} \right){{x}^{2}} \\ & \,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,\,+\left( 5-2{{\left( \frac{{{R}_{0}}}{{{R}_{\infty }}} \right)}^{2}}+\left( \frac{{{\rho }_{i}}}{{{\rho }_{e}}} \right) \right)x+1-{{\left( \frac{{{R}_{0}}}{{{R}_{\infty }}} \right)}^{2}}=0 \\ \end{align}$$

The expression in eq. (4) can be solved iteratively by using various values of x between 0 and 5, until the result is approximately zero (within 0.00001).

Then ICF may be calculated from x and ECF (obtained earlier) as in eq. (5) [17].

ICF=x*ECF$$ICF=x*ECF$$

Once the values of ECF and ICF are obtained, the value of the TBW is consequently obtained just by addition of ICW and ECW like in eq. (6).

TBW=ECW+ICW$$TBW=ECW+ICW$$

The amount of Fat Free Mass (FFM) can be derived directly from the TBW value applying the hydration constant, Kh, as in eq. (7). The typical value for Kh is 0.732 [19].

FFM=TBWKh$$FFM=\frac{TBW}{{{K}_{h}}}$$

Therefore to work out the value of the fat mass (FM) that is the parameter used for analisys eq. (8) is used.

FM=WFFM$$FM=W-FFM$$

In this work the tool used to obtain the BCA parameters from the Cole parameters has been the Bioimp software analysis tool for Body composition assessment (v5.3.1.1, Impedimed Ltd, Brisbane).

Non-Linear Least Squares for Cole Function Fitting

This method aims to obtain the best coefficients for a given model that fits the curve, the method given by eq. (9) aims to minimize the summed squared of the error between the measured data value and the fitted value |Zi|¯,$\overline{\left| {{Z}_{i}} \right|},$ which is obtained from the modulus of the Cole function shown in eq. (10).

mini=1Nei2=mini=1N(|Zi||Zi|¯)2$$\min \sum\nolimits_{i=1}^{N}{e_{i}^{2}=\min {{\sum\nolimits_{i=1}^{N}{\left( \left| {{Z}_{i}} \right|-\overline{\left| {{Z}_{i}} \right|} \right)}}^{2}}}$$Zi¯=R+R0R1+ωταcosαπ/21+2ωταcosαπ/2+ωτ2α2+R0Rωταsinαπ/21+2ωταcosαπ/2+ωτ2α2$$ \overline{\left| {{Z}_{i}} \right|}=\sqrt{{{\left( {{R}_{\infty }}+\frac{\left( {{R}_{0}}-{{R}_{\infty }} \right)\left( 1+{{\left( \omega \tau \right)}^{\alpha }}\cos \left( {}^{\alpha \pi }/{}_{2} \right) \right)}{1+2{{\left( \omega \tau \right)}^{\alpha }}\cos \left( {}^{\alpha \pi }/{}_{2} \right)+{{\left( \omega \tau \right)}^{2\alpha }}} \right)}^{2}}+{{\left( \frac{\left( {{R}_{0}}-{{R}_{\infty }} \right){{\left( \omega \tau \right)}^{\alpha }}\sin \left( {}^{\alpha \pi }/{}_{2} \right)}{1+2{{\left( \omega \tau \right)}^{\alpha }}\cos \left( {}^{\alpha \pi }/{}_{2} \right)+{{\left( \omega \tau \right)}^{2\alpha }}} \right)}^{2}}}$$

Where N is the number of frequency data points included in the fitting. This approach was validated in [14] as working approach to estimate the Cole parameters from the resistance spectrum and the reactance spectrum, as well as from the complex impedance spectrum. In this case, the minimization cost function has been built with the modulus of the complex EBI. Thus the term |Zi|$\left| {{Z}_{i}} \right|$ is the modulus of the measured impedance and |Zi|¯$\overline{\left| {{Z}_{i}} \right|}$ is the absolute value of the Cole function in eq. (1) as shown in (10).

This method has been implemented in Matlab, fitting the generated data to a non-linear real parametric model with coefficients, using the natural frequency ω as an independent variable. Performing the curve fitting using a Cole-based function like in eq. (1) allows the estimation of the values for the four Cole parameters.

EBI Measurements and Descriptive Statistics

Right side 4-electrode wrist-to-ankle EBI spectroscopy measurements have been taken in five healthy volunteers. The EBIS measurements were performed with the SFB7 bioimpedance spectrometer manufactured by Impedimed ltd. using repositionable Red Dot Ag/AgCl electrodes manufactured by 3M. The frequency range of performed EBI measurements was 3.096 to 1000 kHz and 100 complex EBI spectroscopy measurements were obtained for each of the volunteers.

The body parameters of the volunteers can be observed in Table I. Subject 5 is female and all the others are male.

Subjects body features

Subjects FeaturesAge (years)Height (cm)Weight (kg)
Subject 13217699.0
Subject 23016560.0
Subject 32617494.4
Subject 42618283.8
Subject 52417572.5
EBI Data Analysis and Comparison

As the work flow on Fig. 1 indicates, the Cole curve fitting and BCA parameters estimation were performed on a total of 500 measurements. The mean, minimum and maximum values of the BCA parameters estimated from the corresponding Cole parameters were calculated for each of the subjects. The values obtained for both Cole curve fittings were compared.

Fig. 1

Work Flow

The performed EBI measurements were fitted to the Cole function with both, the curve fitting implemented on the Bioimp software and the NLLS approach on the modulus of the impedance implemented with MATLAB.

The Cole fitting with the Bioimp software was produced with the following curve fitting setup, Td compensation off, rejection threshold of 1% and frequency limits from 3.096 kHz to 1000 kHz, i.e. full spectral range.

Two sets of BCA parameters were obtained per subject, each of them corresponding to the Cole parameters estimated with each of the approaches. The BCA parameters were calculated using the BCA tool of Bioimp with the morphological subject information listed in Table I and the proportionality and body resistivity constants indicated in Table II.

Proportionality and body resistivity constants used with Bioimp

MaleFemale
ρe340322
ρi859784
Body density (Db)1.05
Body proportion (Kb)4.30
Hydration constant (Kh)0.732
Results

The following figures and tables present the values of the BCA parameters obtained with both fitting processes from the EBIS measurement for all 5 subjects.

Table III reports the mean values for the BCA parameters in liters and kilograms correspondingly to the fluid body contents and the fat mass.

Mean values for the BCA parameters of each subject

Subject 1Subject 2Subject 3Subject 4Subject 5
NLLS ZBioimpNLLS ZBioimpNLLS ZBioimpNLLS ZBioimpNLLS ZBioimp
TBW (l)56.356.432.633.049.649.750.150.433.233.4
ECF (l)28.428.417.817.726.026.026.726.620.120.1
ICF (l)27.928.014.815.323.623.723.423.813.113.4
FM (kg)22.122.015.414.826.726.515.414.927.126.8

Figures 2, 3, 4, 5 present the mean, minimum and maximum values, in %, obtained for the BCA parameters, i.e. TBW, ECF, ICF or FM, per subject for both approaches, while Tables IV-VII contain the mean values and the standard deviation (SD).

Fig. 2

Mean, minimum and maximum values for TBW, in %, estimated for all the subjects with both methods

Fig. 3

Mean, minimum and maximum values for ECF, in %, estimated for all the subjects with both methods

Fig. 4

Mean, minimum and maximum values for ICF, in %, estimated for all the subjects with both methods

Fig. 5

Mean, minimum and maximum values for FM, in %, estimated for all the subjects with both methods

Mean and Standard Deviation of the estimated values for TBW in %

TBW (%)NLLS |Z|$\left| Z \right|$-OnlyBioimp
MeanSDMeanSD
Subject 156.880.0356.930.04
Subject 254.400.1255.110.16
Subject 352.530.2952.660.31
Subject 459.780.0860.190.08
Subject 545.860.0546.110.07

Fig. 2 and Table IV present the amount of TBW in the body given as a percentage of the total body mass. A high level of agreement between the values obtained with both methods, with a correlation coefficient of 0.9987, is obtained, as could be expected. It is also possible to observe that the mean values obtained with the NLLS impedance-only fitting are marginally smaller, as well as the SD values. Fig. 3 and Table V present the amount of fluid in the extracellular space given as a percentage of the amount of total body water. In this case the mean values estimated with the NLLS impedance-only fitting are slightly larger and the SD values are again slightly smaller, being nevertheless the result given by the two methods very close with a correlation coefficient of 0.9956.

Mean and Standard Deviation of the estimated values for ECF in %

ECF (%)NLLS |Z|$\left| Z \right|$-OnlyBioimp
MeanSDMeanSD
Subject 150.450.0750.330.06
Subject 254.620.0853.750.12
Subject 352.480.1452.290.15
Subject 453.310.1052.720.08
Subject 560.480.1159.990.12

Fig. 4 and Table VI present the amount of fluid in the intracellular space given as a percentage of the amount of total body water. Since the values of ICF and EFC are complementary, in this case the values estimated with the NLLS impedance-only fitting are slightly smaller and the SD values reported in tables IV & V are identical.

Mean and Standard Deviation of the estimated values for ICF in %

ICF (%)NLLS |Z|$\left| Z \right|$-OnlyBioimp
MeanSDMeanSD
Subject 149.550.0749.670.06
Subject 245.380.0846.250.12
Subject 347.520.1447.710.15
Subject 446.690.1047.280.08
Subject 539.520.1140.010.12

Fig. 5 and Table VII present the amount of FM in the body given as a percentage of total body mass, the FM results are a linear function of the TBW results as can be deduced from equations (7) and (8).

Mean and Standard Deviation of the estimated values for FM in %

FM (%)NLLS |Z|$\left| Z \right|$-OnlyBioimp
MeanSDMeanSD
Subject 122.300.0422.230.05
Subject 225.690.1724.710.22
Subject 328.240.4028.060.43
Subject 418.330.1117.770.11
Subject 537.350.0737.010.09

Table VIII presents the difference between the values obtained with the NLLS impedance-only method and the values produced from the Bioimp. fitting. From the values in Table VIII it is possible to observe that the difference on percentage of the estimated BCA parameter is below 1 %.

Differences obtained on the estimation of the BCA parameters

Δ (%) / (l)(NLLS |Z|-Only) - Bioimp
TBWECFICFFM
Subject 1-0.04 / -0.040.12 / 0.04-0.12 / -0.090.06 / 0.06
Subject 2-0.71 / -0.420.87 / 0.05-0.87 / -0.480.97 / 0.59
Subject 3-0.13 / -0.120.19 / 0.03-0.19 / -0.150.18 / 0.17
Subject 4-0.41 / -0.350.59 / 0.12-0.59 / -0.460.57 / 0.47
Subject 5-0.25 / -0.180.49 / 0.05-0.49/ -0.240.34 / 0.25
Average-0.31 / -0.220.45 / 0.06-0.45 / -0.280.42 / 0.31

Note: The percentage of ECF and ICF are referred to the TBW and the FM is expressed in percentage of the total mass and it is given in kg

Discussion

The volume for the ECF obtained with both approaches is practically the same according to the obtained correlation coefficient. The volume differences observed on the estimation of the body fluid distribution occur mainly in the amount of ICF. The amount of volume estimated with the NLLS impedance–only fitting is slightly smaller than the amount of volume estimated by Bioimp. In any case a correlation coefficient of 0.9956, indicates that the difference obtained between both approaches is very small.

Since the only difference in the applied estimation process is the curve fitting method applied to obtain the Cole function and the Cole parameters, any difference in the obtained value for the BCA parameters comes from the curve fitting process. The negligible difference in the estimation of ECF indicates that the value estimated for R0 is very similar for both methods. This leaves the source for the difference obtained between both approaches in the estimation of the value R∞.

When performing EBI measurements any capacitive parasitic effect associated to the measurement load or leads influence largely on the estimation of the reactance and the phase spectra, while the influence on the resistance or the modulus spectra is remarkable smaller [15]. If the estimation of the Cole parameters is done in the impedance plane [9], reactance vs. resistance, fitting the measured data to a semicircle, the reactance data containing corrupted values produced by capacitive leakage would lead to underestimation of the value of R∞, which consequently influences on the calculation of the value of ICF.

To compensate for such parasitic effect the Bioimp software has a built-in option known as Td compensation [17, 20]. In this study such option has been left unused to allow for a fair comparison between fitting approaches, this way being able to show that the NLLS impedance-only fitting approach produces a slightly larger value for ICF distribution than the one obtained with the Bioimp fitting. In any case the measurements used in this study presented a reactance deviation at high frequencies that was completely un-noticeable to the naked eye, which indicates that the capacitive leakage present on the measurement was almost negligible [15].

The NLLS impedance-only not only presents the advantage of exhibiting a specific robustness to the effects of capacitive leakage due to the fact the modulus of the impedance spectrum is not affected by capacitive leakage as much as the phase or the reactance, but brings an additional and significant advantage also for the electronics of measurement acquisition systems. The measurement of the modulus of the impedance does not require phase detection, which decreases the complexity required to build the spectrometer device.

Conclusion

The performance exhibited by NLLS impedance-only approach for estimating the BCA parameters indicates that such fitting approach is a strong alternative to the currently in use Cole curve fitting-based on the impedance plane. In addition to the advantages related to the curve fitting and Cole parameters estimation processes, the use of the NLLS impedance–only allows the simplification of the measurement instrumentation eliminating the need for a phase detector. Consequently, since BCA and fluid distribution assessment is currently the application of electrical bioimpedance most spread worldwide, this novel Cole parameter estimation approach might have significant repercussions in the design of future electrical bioimpedance spectrometers.